← Product Code [HWC](/productcode/HWC) · K031937

# ALANTECH GRAFT FIXATION SCREW (K031937)

_Arthrocare Corp. · HWC · Sep 4, 2003 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K031937

## Device Facts

- **Applicant:** Arthrocare Corp.
- **Product Code:** [HWC](/productcode/HWC.md)
- **Decision Date:** Sep 4, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Atlantech Graft Fixation Screw is an interference screw for use in the fixation of bone-tendon-bone grafts in cruciate ligament reconstructive surgery.

## Device Story

Atlantech Graft Fixation Screw is a metallic interference screw used in cruciate ligament reconstructive surgery. Device features tapered rear profile to minimize graft impingement post-implantation. Screw is cannulated to accept 1.2mm guidewire. Used by surgeons in clinical settings to secure bone-tendon-bone grafts. Provides mechanical fixation to facilitate ligament reconstruction.

## Clinical Evidence

No clinical data provided. Substantial equivalence based on technological characteristics and established clinical use of interference screws.

## Technological Characteristics

Metallic bone fixation fastener. Features tapered rear profile and cannulated design for 1.2mm guidewire. Class II device (21 CFR 888.3040).

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- RCI Fixation Screws ([K992945](/device/K992945.md))
- Arthrex Cannulated Interference Screw ([K915424](/device/K915424.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

SEP - 4 2003

31937

page 1 of 2

### 510(k) Summary

# ArthroCare Corporation Atlantech Graft Fixation Screw

### General Information

| Submitter Name/Address:               | ArthroCare Corporation<br>680 Vaqueros Avenue<br>Sunnyvale, CA 94085-2936                                                                                                                         |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Phone Number:                         | (408) 736-0224                                                                                                                                                                                    |
| Contact Person:                       | Valerie Defiesta-Ng<br>Director, Regulatory Affairs                                                                                                                                               |
| Manufacturers Name/Address:           | Atlantech Medical Devices, Ltd.<br>(Subsidiary of ArthroCare Corporation)<br>Atlantech House, 38 Freemans Way<br>Harrogate Business Park, Harrogate<br>North Yorkshire, HG3 IDH<br>United Kingdom |
| Date Prepared:                        | June 16, 2003                                                                                                                                                                                     |
| Device Description                    |                                                                                                                                                                                                   |
| Trade Name:                           | Atlantech Graft Fixation Screw                                                                                                                                                                    |
| Generic/Common Name:                  | Screws, Fixation, Bone                                                                                                                                                                            |
| Classification Name:                  | Smooth or threaded metallic bone fixation<br>fastener (21 CFR 888.3040)                                                                                                                           |
| Predicate Devices                     |                                                                                                                                                                                                   |
| RCI Fixation Screws                   | K992945, cleared November 18, 1999                                                                                                                                                                |
| Arthrex Cannulated Interference Screw | K915424, cleared February 9, 1993                                                                                                                                                                 |

### Product Description

The Atlantech Graft Fixation Screw is for fixation of grafts in cruciate ligament reconstructive surgery. Standard features include a tapered rear profile, which museuzes graft impingement post implantation. The screw is cannulated to accept 1.2mm A-Fech Controlled Flexion Guidewire.

{1}------------------------------------------------

# K031937 page 2 of 2

### Intended Uses

The Atlantech Graft Fixation Screw is an interference screw for use in the fixation of bone-tendon-bone grafts in cruciate ligament reconstructive surgery.

### Substantial Equivalence

Based on the indications for use statement and technological characteristics to the predicate devices, the Atlantech Graft Fixation Screw is substantially equivalent to the predicate devices under the Federal Food, Drug and Cosmetic Act.

### Summary of Safety and Effectiveness

The Atlantech Graft Fixation Screw, as described in this 510(k), is substantially equivalent to the predicate device with respect to the indications for use and technological characteristics. The use of interference screws for the fixation of grafts in cruciate ligament reconstructive surgery is well established. Atlantech Graft Fixation Screws have been commercially available in Europe since 1996 and do not differ markedly from others on the market.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three horizontal lines forming its wings and two wavy lines below representing water. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP = 4 2003

Ms. Valerie Defiesta-Ng Director, Regulatory Affairs ArthroCare Corporation 680 Vaqueros Avenue Sunnyvale, California 94085-2936

Re: K031937

Trade/Device Name: Atlantech Graft Fixation Screw Regulation Numbers: 21 CFR 888.3040 Regulation Names: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Codes: HWC Dated: June 16, 2003 Received: June 24, 2003

Dear Ms. Defiesta-Ng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Valerie Defiesta-Ng

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Mark n Mellers

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use Statement

Device Name

Atlantech Graft Fixation Screw

K_031937 510(k) Number:

Indications for Use:

The Atlantech Graft Fixation Screw is an interference screw for use in the fixation of bone-tendon-bone grafts in cruciate ligament reconstructive surgery.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use the text is the following: Prescription Use X OR Over-the-Counter Use (Per 21 CFR 801.109) for Mark N Millen (Division Sign-Off) Division of General, Restorative and Neurological Devices Koz 1937

510(k) Number .

ix

---

**Source:** [https://fda-staging.innolitics.com/device/K031937](https://fda-staging.innolitics.com/device/K031937)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).