SonoView Pro™ by Medison Co. Ltd. is a software application for the for the display and 3D visualization of medical image data derived from various sources (i.e. Ultrasound, CT scanners, MRI scanners). Images and data can be acquired, stored, communicated, processed, printed, rendered, and displayed within the system and or across computer networks at distributed locations. Typical users of this system are trained professionals, including physicians, nurses, and technicians.
Device Story
SonoView Pro is an internet-based PACS software for managing medical images (US, CT, MR, NM, CR, DR, DX, XA, PT). It operates on standard Intel Pentium-based PCs running Windows 2000/XP over TCP/IP networks. The system acquires images from DICOM-compliant modalities; it does not generate original images. Features include 3D reconstruction, real-time US image acquisition, multi-frame storage, and GUI tools (zoom, pan, pseudo-color, threshold, Gamma correction). It provides 3D rendering (MIP, X-ray, translucent, surface, slicing, 3D cine) and measurement tools (distance, area, angle). Tele-medicine capabilities (audio, video, text) support remote collaboration. Physicians use the output for clinical interpretation; the system facilitates image review, storage, and communication, potentially improving diagnostic workflow and patient care through enhanced visualization.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Software application; runs on Intel Pentium-based hardware; Windows 2000/XP OS; TCP/IP network connectivity. DICOM compliant. Features 3D reconstruction, rendering (MIP, surface, slicing), and measurement tools. Non-contact device; does not control life-sustaining equipment.
Indications for Use
Indicated for display and 3D visualization of medical image data (US, CT, MRI, etc.) for trained professionals including physicians, nurses, and technicians.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K 031886
## JUN 3 0 2003
# 510(k) Summary of Safety and Effectiveness
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.
Date Prepared:
May 5, 2003
Submitter's Information: 21 CFR 807.92(a)(1)
Medison Co. Ltd. Medison Venture Tower, 997-10 Daechi-dong, Kangnam-gu, Seoul 135-280, Korea
Trade Name, Common Name and Classification: 21 CFR 807.92(a)(2)
| Trade Name: | SonoView Pro™ by Medison Co. Ltd. |
|------------------------|-----------------------------------------|
| Common Name: | Picture Archiving Communications System |
| Device Classification: | 892.2050 |
| Name: | System, Image Processing |
Predicate Device: 21 CFR 807. 92(a)(3)
| Device Classification Name | SYSTEM, IMAGE PROCESSING, RADIOLOGICAL |
|----------------------------|------------------------------------------------------------------------------------------------------------|
| Regulation Number | 892.2050 |
| 510(k) Number | K022692 |
| Device Name | VoxelPlus™ PACS |
| Applicant | Mevisys Co. Ltd.<br>EE&CS Bldg. Room 4203, KAIST<br>373-1 Guseong-dong Yuseong-gu<br>Daejeon 305-701 Korea |
| Product Code | 90 LLZ |
| Date Received | 08/13/2002 |
| Decision Date | 10/11/2002 |
#### Device Description: 21 CFR 807 92(a)(4)
SonoView Pro™ is an Internet based software picture archiving and communications system that provides users with capabilities relating to the acceptance, transfer, display, storage, and digital processing of medical images. SonoView Pro'™ includes features to access and manage medical imaging studies from ultrasound (US), cat-scan (CT), magnetic radiography (MR), nuclear medicine (NM), computerized radiography (CR), digital radiography (DR), digital x-ray (DX), x-ray angiography (XA), PET scan (PT), and other imaging modalities.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "medison" in a bold, sans-serif font. Below the word, there are three curved lines that resemble brush strokes. The lines are arranged in a parallel fashion, with the top line being the shortest and the bottom line being the longest.
SonoView Pro™ is deployed over conventional TCP/IP networking infrastructure available in most healthcare organizations and utilizes commercially available computer platforms (Intel Pentium-based) and operating systems (Microsoft Windows 2000/XP). The system does not produce any original medical images. All images located on SonoView Pro™ have been received from DICOM compliant imaging modalities and/or PACS components or systems. SonoView Pro™ provides the
advanced process functions, such as :
- Supports the freehand scanning, 3D image acquisition, and 3D reconstruction. -
- Acquires images from ultrasound system in real time. -
- -Supports the ultrasound image storage and ultrasound multi-frame image storage.
- Supports the GUI functions: zoom, pan, pseudo-color setting, threshold setting, -Gamma correction, and thumbnail.
- -Provides the DICOM query and retrieve services (images and presentation states).
- -Displays the cut planes of 3D image along the long axis or the short axis.
- Provides 3D rendering modes: MIP, X-ray, translucent, surface, slicing, and 3D cine.
- -Supports distance/area/angle measurements.
- Supports Tele-medicine (audio, video, and text) between regional medicine center and remote clients.
#### Indications for Use: 21 CFR 807 92(a)(5)
SonoView Pro™ by Medison Co. Ltd. is a software application for the display and 3D visualization of medical image data derived from various sources (i.e. Ultrasound, CT scanners). Images and data can be acquired, stored, communicated, processed, printed, rendered, and displayed within the system and or across computer networks at distributed locations. Typical users of this system are trained professionals, including physicians, nurses, and technicians.
Technological Characteristics: 21 CFR 807 92(a)(6)
The device is a software application and does not contact the patient, nor does it control any life sustaining devices. A physician, providing ample opportunity for competent human intervention interprets images and information being displayed and printed.
#### Conclusion: 21 CFR 807 92(b)(1)
- The 510(k) Pre-Market Notification for SonoView Pro™ contains adequate . information and data to enable FDA - CDRH to determine substantial equivalence to the predicate device.
- 프 The device will be manufactured in accordance with the voluntary standards listed in the enclosed voluntary standard survey.
- 미 The submission contains the results of a hazard analysis and the potential hazards have been classified as Minor.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three curved lines that resemble a human figure.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 3 0 2003
Medison Co., Ltd. % Mr. Ned Devine Responsible Third Party Entela, Inc. 3033 Medison Ave. SE GRAND RAPIDS MI 49548 Re: K031886 Trade/Device Name: SonoView ProTM Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: June 13, 2003 Received: June 18, 2003
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Page 2
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "medison" in a stylized font. Below the word are three curved lines that appear to be a design element. The word is in lowercase letters and has a bold appearance. The overall design is simple and clean.
Page 1 of -1
#### (Indications for Use Form)
510(k) Number:
Device Name: SonoView Pro™ by Medison Co. Ltd.
Indications for Use:
SonoView Pro™ by Medison Co. Ltd. is a software application for the for the display and 3D visualization of medical image data derived from various sources (i.e. Ultrasound, CT scanners, MRI scanners). Images and data can be acquired, stored, communicated, processed, printed, rendered, and displayed within the system and or across computer networks at distributed locations.
Typical users of this system are trained professionals, including physicians, nurses, and technicians.
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF
NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
OR
Over-The-Counter Use
(Per 21 CFR 801.109)
Daniel R. Segerson
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K031886
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.