← Product Code [GCJ](/productcode/GCJ) · K031882

# ENDOSCOPIC APPLICATOR (K031882)

_Baxter Healthcare Corp · GCJ · Sep 5, 2003 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K031882

## Device Facts

- **Applicant:** Baxter Healthcare Corp
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Sep 5, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

This Endoscopic Applicator is intended for use in delivering hemostatic agents to bleeding surgical sites through a 5mm or larger trocar.

## Device Story

The Endoscopic Applicator is a reusable surgical instrument designed to deliver hemostatic agents to bleeding sites during endoscopic or laparoscopic procedures. The device consists of a non-reflective stainless steel cannula and a stainless steel stylet (obturator). It is operated by a surgeon or trained clinician in a surgical setting. The device is inserted through a 5mm or larger trocar to reach the surgical site. The clinician manually uses the stylet to push the hemostatic agent through the cannula to the target area. The device is designed to be cleaned and sterilized by the user between uses, with a service life of up to 20 uses. By facilitating the precise application of hemostatic agents, the device assists in managing surgical bleeding.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Materials: Stainless steel cannula and stylet with ME-92 coating. Dimensions: 5mm outer diameter, 292mm length. Operation: Manual delivery of hemostatic agents. Sterilization: Non-sterile; user-validated cleaning and sterilization required between uses. Reusability: Up to 20 cycles.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- MedChem Surgical Endoscopic Accessory/ Delivery System ([K913172](/device/K913172.md))
- MedChem Surgical Delivery System ([K932466](/device/K932466.md))
- Davol Surgical Mesh Delivery System ([K930147](/device/K930147.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Baxter Healthcare Corporation 34175 Ardenwood Blv ont. California 94

SEP - 5 2003

K 031882

Image /page/0/Picture/3 description: The image shows the logo for Baxter BioScience. The word "Baxter" is in a bold, italicized font. Below the word "Baxter" is the word "BioScience" in a smaller, non-italicized font. The logo is simple and professional.

## 510(k) SUMMARY

This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92.

#### Name. Address, Phone and Fax Number of the Applicant A.

Baxter Healthcare Corporation 34175 Ardenwood Blvd. Fremont, CA 94536

(510) 818-4600 Telephone: Fax: (510) 818-4700

#### B. Contact Person

Lori DonDiego Senior Manager, Regulatory Affairs

#### C. Date Prepared

June 16, 2003

#### D. Device Name

Trade Name: Endoscopic Applicator Common Name: Endoscopic Applicator Classification Name: Endoscopes and Accessories

#### E. Device Description

The Endoscopic Applicator is a re-usable applicator used to deliver hemostatic agents to bleeding surgical sites. The Endoscopic Applicator consists of two components; (1) a non-reflective stainless steel cannula, and (2) a stainless steel stylet (obturator). The Endoscopic Applicator is to be thoroughly cleaned and sterilized before each use, and can be used up to 20 times.

#### F. Intended Use

This Endoscopic Applicator is intended for use in delivering hemostatic agents to bleeding surgical sites through a 5mm or larger trocar.

{1}------------------------------------------------

### Substantial Equivalence G.

The Endoscopic Applicator is substantially equivalent to the following commercially available laparoscopic instruments:

| MedChem Surgical Endoscopic<br>Accessory/ Delivery System | MedChem Products, Inc. | K913172 |
|-----------------------------------------------------------|------------------------|---------|
| MedChem Surgical Delivery System                          | MedChem Products, Inc. | K932466 |
| Davol Surgical Mesh Delivery System                       | Davol, Inc.            | K930147 |

The Endoscopic Applicator and the predicate devices have similar intended uses. They are all intended to deliver surgical products into the body. The Endoscopic Applicator is substantially equivalent to the predicate devices in intended use, design and components, materials, and performance characteristics. See Table 1.

{2}------------------------------------------------

|                         | NEW DEVICE                                                                                                                                                                                                          | PREDICATE DEVICES                                                                                                                                                                                                                                                                                       |                                                                                                                                  |                                                                                                                                                                  |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FEATURE                 | Endoscopic Applicator                                                                                                                                                                                               | MedChem Surgical<br>Endoscopic Accessory/<br>Delivery System<br>K913172                                                                                                                                                                                                                                 | MedChem Surgical<br>Delivery System<br>K932466                                                                                   | Davol Surgical Mesh<br>Delivery System<br>K930147                                                                                                                |
| Intended Use            | This Endoscopic Applicator<br>is intended for use in<br>delivering hemostatic<br>agents to bleeding surgical<br>sites through a 5mm or<br>larger trocar.                                                            | The MedChem<br>Endoscopic Delivery<br>System is used in<br>endoscopic surgical<br>procedures to maintain the<br>already established<br>pneumoperitoneum while<br>providing a means of<br>facilitating delivery of<br>adjunctive surgical<br>products (ligatures,<br>sutures, etc.) through a<br>trocar. | A syringe device used to<br>facilitate the delivery of<br>adjunctive surgical<br>products during various<br>surgical procedures. | Intended to facilitate the<br>delivery of surgical mesh to<br>the surgical site during<br>laparoscopic soft tissue<br>repair procedures (e.g.<br>hernia repair). |
| Anatomical Sites        | Endoscopic / Laparoscopic                                                                                                                                                                                           | Endoscopic                                                                                                                                                                                                                                                                                              | Various                                                                                                                          | Laparoscopic (abdominal<br>soft tissue)                                                                                                                          |
| Sheath Outer Diameter   | 5mm                                                                                                                                                                                                                 | 0.394"                                                                                                                                                                                                                                                                                                  | 0.197" - 0.750" (various)                                                                                                        | 9.7mm                                                                                                                                                            |
| Sheath Length           | 292mm                                                                                                                                                                                                               | 13.31"                                                                                                                                                                                                                                                                                                  | 8" - 13" (various)                                                                                                               | 8"                                                                                                                                                               |
| Size cannula used with  | 5 mm or larger                                                                                                                                                                                                      | Minimum 10 mm OD                                                                                                                                                                                                                                                                                        | Various, depending on<br>size of delivery system.                                                                                | 10 mm or larger                                                                                                                                                  |
| Main Components         | Cannula<br>Stylet                                                                                                                                                                                                   | Sheath<br>Plunger<br>Plunger O-Ring                                                                                                                                                                                                                                                                     | Sheath<br>Plunger<br>Plunger O-Ring                                                                                              | Introducer Sheath/Handle<br>Introducer Rod/Handle                                                                                                                |
|                         | NEW DEVICE                                                                                                                                                                                                          | PREDICATE DEVICES                                                                                                                                                                                                                                                                                       |                                                                                                                                  |                                                                                                                                                                  |
| FEATURE                 | Endoscopic Applicator                                                                                                                                                                                               | MedChem Surgical<br>Endoscopic Accessory/<br>Delivery System<br>K913172                                                                                                                                                                                                                                 | MedChem Surgical<br>Delivery System<br>K932466                                                                                   | Davol Surgical Mesh<br>Delivery System<br>K930147                                                                                                                |
| Main Materials          | Stainless Steel<br>Stainless Steel<br>ME-92Coating                                                                                                                                                                  | Polycarbonate<br>Polycarbonate<br>Silicone, Medical Grade                                                                                                                                                                                                                                               | Polycarbonate<br>Polycarbonate<br>Silicone, Medical Grade                                                                        | Stainless Steel/PVC<br>Stainless Steel/PVC                                                                                                                       |
| Sterilization Method    | Non-sterile. (Intended to<br>be cleaned and sterilized<br>by the user prior to the first<br>use and then after each<br>subsequent use.)                                                                             | Gamma or dry heat cycle,<br>SAL 10-6                                                                                                                                                                                                                                                                    | Gamma or dry heat cycle,<br>SAL 10-6                                                                                             | EtO, SAL 10-6                                                                                                                                                    |
| Expiration Date (years) | N/A - reusable up to 20<br>times                                                                                                                                                                                    | Unknown                                                                                                                                                                                                                                                                                                 | Unknown                                                                                                                          | Unknown                                                                                                                                                          |
| Single Use Only         | Reusable                                                                                                                                                                                                            | Single Use                                                                                                                                                                                                                                                                                              | Single Use                                                                                                                       | Single Use                                                                                                                                                       |
| Packaging               | Packaging does not<br>provide a sterile barrier.<br>Polyethylene terephthalate-<br>glycol (PETG) tube, Low<br>Density Polyethylene<br>(LDPE) poly tubing, Vinyl<br>End Cap, 200 lb. test #3<br>white cardboard box. | Polycarbonate plastic tray<br>sealed with a polyethylene<br>lid                                                                                                                                                                                                                                         | Polycarbonate plastic tray<br>sealed with a polyethylene<br>lid.                                                                 | Tyvek pouch in a plastic<br>blister tray                                                                                                                         |

# Table 1. Comparison Table

{3}------------------------------------------------

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is enclosed within a circle, and the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES (USA)" is arranged around the upper half of the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 3 0 2006

Baxter Healthcare Corporation % Ms. Lori DonDiego Senior Manager, Regulatory Affairs 34175 Ardenwood Boulevard Fremont, California 94555

Re: K031882

Trade/Device Name: Endoscopic Applicator Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscopic and accessories Regulatory Class: II Product Code: GCJ Dated: June 16, 2003 Received: June 18, 2003

Dear Ms. DonDiego:

This letter corrects our substantially equivalent letter of September 5, 2003.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class 111 (PMA), it mav be subject to additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning vour device in the Federal Register.

Please be advised that FDA s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other

{5}------------------------------------------------

Page 2 - Ms. Lori Dondiego

Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0515. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

el.rree

Ro

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

{6}------------------------------------------------

510(k) Number (if known): K 031862

Device Name:

Endoscopic Applicator

Indications For Use:

This Endoscopic Applicator is intended for use in delivering hemostatic agents to bleeding surgical sites through a 5mm or larger trocar.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use

(Per 2

OR (Per 21 CFR 801.109) Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

Muriam C. Provost

(Division Sign-Off) Division of General, Restorative and Neurological Devices

510(k) Number K031882

---

**Source:** [https://fda-staging.innolitics.com/device/K031882](https://fda-staging.innolitics.com/device/K031882)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).