← Product Code [CHH](/productcode/CHH) · K031824

# COBAS INTEGRA CHOLESTEROL GEN. 2 (K031824)

_Roche Diagnostics Corp. · CHH · Jul 9, 2003 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K031824

## Device Facts

- **Applicant:** Roche Diagnostics Corp.
- **Product Code:** [CHH](/productcode/CHH.md)
- **Decision Date:** Jul 9, 2003
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 862.1175
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

The cassette COBAS Integra Cholesterol Gen.2 (CHOL2) contains an in vitro diagnostic reagent system intended for use on COBAS Integra systems for the quantitative determination of total cholesterol in serum and plasma.

## Device Story

Enzymatic colorimetric assay for total cholesterol quantification; utilizes cholesterol esterase, cholesterol oxidase, and peroxidase to generate red quinone-imine dye; color intensity proportional to cholesterol concentration. Used on COBAS Integra automated clinical chemistry analyzers in clinical laboratory settings. Operated by laboratory personnel. Provides quantitative cholesterol results to clinicians for diagnosis and management of lipid metabolism disorders and hypercholesterolemia.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on analytical performance comparisons between the subject device and the predicate.

## Technological Characteristics

Enzymatic colorimetric reagent system. Components: cholesterol esterase, cholesterol oxidase, peroxidase. Form factor: cassette-based reagent for automated clinical chemistry analyzers. Measuring range: 0.1-800 mg/dL.

## Regulatory Identification

A cholesterol (total) test system is a device intended to measure cholesterol in plasma and serum. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.

## Predicate Devices

- Roche/Hitachi Cholesterol Assay ([K952127](/device/K952127.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

031824

## 510(k) Summary - COBAS Integra Cholesterol Gen.2

| Introduction                     | According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence                                                                                       |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter name, address, contact | Roche Diagnostics Corporation<br>9115 Hague Rd<br>Indianapolis IN 46250<br>(317) 521-3831<br><br>Contact person: Sherri L. Coenen<br><br>Date prepared: June 11, 2003                                                                                             |
| Device Name                      | Proprietary name: Roche Diagnostics COBAS Integra Cholesterol Gen.2<br><br>Common name: Total Cholesterol Assay<br><br>Classification name: Cholesterol (Total) test system                                                                                       |
| Device description               | The COBAS Integra Cholesterol Gen.2 is an enzymatic colorimetric assay using cholesterol esterase, cholesterol oxidase, and peroxidase to form a red quinone-imine dye. The color intensity of the dye is directly proportional to the cholesterol concentration. |
| Intended use                     | The cassette COBAS Integra Cholesterol Gen.2 (CHOL2) contains an in vitro diagnostic reagent system intended for use on COBAS Integra systems for the quantitative determination of total cholesterol in serum and plasma.                                        |
| Predicate Device                 | We claim substantial equivalence to the currently marketed Roche/Hitachi Cholesterol Assay. (K952127).                                                                                                                                                            |

{1}------------------------------------------------

# 510(k) Summary - COBAS Integra Creatinine plus ver.2,

continued

The following table describes the similarities and differences between the Reagent Summary COBAS Integra Cholesterol Gen.2 and the predicate device.

| Topic              | Roche Hitachi Cholesterol (K952127)                                                                                                                               | COBAS Integra Cholesterol Gen.2 (Modified Device)                                                                                                                                                                                        |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use       | Enzymatic in vitro test for the direct<br>quantitative determination of<br>cholesterol in human serum and<br>plasma on automated clinical<br>chemistry analyzers. | The cassette COBAS Integra<br>Cholesterol Gen.2 (CHOL2) contains<br>an in vitro diagnostic reagent system<br>intended for use on COBAS Integrasystems for the quantitative<br>determination of total cholesterol in<br>serum and plasma. |
| Method             | Enzymatic, colorimetric test                                                                                                                                      | Same                                                                                                                                                                                                                                     |
| Sample type        | Human Serum and Plasma                                                                                                                                            | Same                                                                                                                                                                                                                                     |
| Measuring<br>range | 3 - 800 mg/dl                                                                                                                                                     | 0.1 - 800 mg/dl                                                                                                                                                                                                                          |
| Expected<br>values | According to the recommendations of<br>the European Atherosclerosis Society<br>and the recommendations of the<br>NCEP Adult Treatment Panel.                      | Same                                                                                                                                                                                                                                     |

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with three lines representing its wings and three curved lines representing its body. The seal is black and white.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

#### 9 2003 JUL

Ms. Sherri L. Coenen Regulatory Affairs Consultant Regulatory Submissions, Centralized Diagnostics Roche Diagnostics Corporation 9115 Hague Road P.O. Box 50457 Indianapolis, IN 46250-0457

Re: k031824

Trade/Device Name: COBAS Integra Cholesterol Gen 2 Regulation Number: 21 CFR 862.1175 Regulation Name: Cholesterol (total) test system Regulatory Class: Class I Product Code: CHH Dated: June 11, 2003 Received: June 13, 2003

Dear Ms Coenen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{3}------------------------------------------------

### Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

### Indications for Use Statement

510(k) Number (if known): - MA

Device Name: COBAS Integra Cholesterol Gen. 2

Indications For Use:

The cassette COBAS Integra Cholesterol Gen.2 (CHOL2) contains an in vitro diagnostic reagent system intended for use on COBAS Integra systems for the quantitative determination of total cholesterol in serum and plasma,

Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in blood and lipoprotein metabolism disorders,

Jean Copy
Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K031824

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation (ODE) |                          |
|--------------------------------------------------------|--------------------------|
| Prescription Use (Per 21 CFR 801.109)                  |                          |
| OR                                                     |                          |
| Over-The-Counter Use                                   |                          |
|                                                        | (Optional Format 1-2-96) |

---

**Source:** [https://fda-staging.innolitics.com/device/K031824](https://fda-staging.innolitics.com/device/K031824)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).