TI FENESTRATED PLATE

K031772 · Interpore Cross Intl. · KWP · Sep 5, 2003 · Orthopedic

Device Facts

Record IDK031772
Device NameTI FENESTRATED PLATE
ApplicantInterpore Cross Intl.
Product CodeKWP · Orthopedic
Decision DateSep 5, 2003
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 888.3050
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Ti Fenestrated Plate is intended to be used as a temporary construct that assists normal healing and is not intended to replace normal body structures. The Ti Fenestrated Plate is intended to stabilize the spinal operative site during fusion procedures and should be removed after fusion. The Ti Fenestrated Plate is attached to the spine posteriorly by means of Songer Cables. The levels of attachment are T1 to L5. The Ti Fenestrated Plate is indicated for the following: 1. Degenerative Disc Disease (as defined by chronic back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). 2. Idiopathic scoliosis. 3. Kyphotic deformities of the spine. 4. Lordotic deformities of the spine. 5. Vertebral fracture or dislocation. 6. Tumors. 7. Spondylolisthesis. 8. Stenosis. 9. Pseudarthrosis. 10. Unsuccessful previous attempts at spinal fusion.

Device Story

Titanium plate with multiple holes; used for posterior spinal stabilization (T1-L5). Attached to spine via Songer Cables passed under/around lamina; secured using tensioner/crimper instrument. Temporary construct; assists healing; removed post-fusion. Used in surgical setting by surgeons. Provides mechanical stabilization to spinal operative site; facilitates fusion; benefits patient by supporting structural integrity during healing process.

Clinical Evidence

No clinical data provided; substantial equivalence based on biomechanical performance and design similarity.

Technological Characteristics

Titanium plate with multiple fenestrations; posterior spinal fixation; secured via Songer Cables and tensioner/crimper instrument; T1-L5 levels.

Indications for Use

Indicated for patients requiring spinal stabilization during fusion procedures (T1-L5) for degenerative disc disease, idiopathic scoliosis, kyphotic/lordotic deformities, vertebral fracture/dislocation, tumors, spondylolisthesis, stenosis, pseudarthrosis, or failed prior fusion.

Regulatory Classification

Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(k) SUMMARY #### SUBMITTED BY Wendy Spielberger Lead, Regulatory and Clinical Affairs Staff Interpore Cross International 181 Technology Drive Irvine, California 92618 (949) 453-3200 June 6, 2003 This summary of 510(k) substantial equivalence information is being submitted in accordance with the requirements of 21 CFR §807.92. #### CLASSIFICATION, COMMON OR USUAL NAME, DEVICE NAME | Classification Name: | Appliance, Fixation, Spinal Interlaminal | |-------------------------|------------------------------------------| | Common/Usual Name: | Spinal Interlaminal Fixation | | Product Classification: | Class II | | Proprietary Name: | Ti Fenestrated Plate | #### PREDICATE DEVICE Predicate device information is provided in this premarket notification. #### INDICATIONS-FOR-USE The Ti Fenestrated Plate is intended to be used as a temporary construct that assists normal hing and is not intended to replace normal body structures. The Ti Fenestrated Plate is intended to stabilize the spinal operative site during fusion procedures and should be removed after fusion. The Ti Fenestrated Plate is attached to the spine posteriorly by means of Songer Cables. The levels of attachment are T1 to L5. The Ti Fenestrated Plate is indicated for the following: - 1. Degenerative Disc Disease (as defined by chronic back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). - 2. Idiopathic scoliosis. - Kyphotic deformities of the spine. 3. - Lordotic deformities of the spine. ব - 5. Vertebral fracture or dislocation. - Tumors. 6. - Spondylolisthesis. 7. - 8. Stenosis. - Pseudarthrosis. 9. - 10. Unsuccessful previous attempts at spinal fusion. ### DEVICE DESCRIPTION The Interpore Cross International Ti Fenestrated Plate is a titanium plate containing multiple holes along the length of the plate. The Ti Fenestrated Plate is intended for use in the thoracolumbar spine from T1-L5. The plate is secured to the posterior spine via Songer Cables. The Songer Cables are {1}------------------------------------------------ K031772-PIGe 2/2 placed through the holes in the Ti Fenestrated Plate and passed under and around the lamina. The plate and cable(s) are then secured to the spine using a tensioner/crimper instrument to obtain the appropriate tension level and to secure the cable crimp. ### COMPARISON TO THE PREDICATE DEVICE Based on the same indications for use, intended use, similarity in materials of construction and equivalent biomechanical performance, the Ti Fenestrated Plate is considered substantially equivalent to the legally marketed predicate device. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a bird-like figure, composed of three curved lines. Public Health Service SEP - 5 2003 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Wendy Spielberger, RAC Lead, Regulatory and Clinical Affairs Staff Interpore Cross International 181 Technology Drive Irvine, California 92618 Re: K031772 > Trade Name: Ti Fenestrated Plate Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: II Product Code: KWP Dated: June 6, 2003 Received: June 10, 2003 Dear Ms. Spielberger: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ ### Page 2 - Ms. Wendy Spielberger, RAC This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594- . Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely vours. b Mach A. Mellem Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # 510(k) Number (if known): #### Device Name: Ti Fenestrated Plate # Indications-For-Use: The Ti Fenestrated Plate is intended to be used as a temporary construct that assists normal healing and is not intended to replace normal body structures. The Ti Fenestrated Plate is intended to stabilize the spinal operative site during fusion procedures and should be removed after fusion. The Ti Fenestrated Plate is attached to the spine posteriorly by means of Songer Cables. The levels of attachment are T1 to L5. The Ti Fenestrated Plate is indicated for the following: - 1. Degenerative Disc Disease (as defined by chronic back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). - 2. Idiopathic scoliosis. - 3. Kyphotic deformities of the spine. - 4. Lordotic deformities of the spine. - 5. Vertebral fracture or dislocation. - 6. Tumors. - 7. Spondylolisthesis. - 8. Stenosis. - 9. Pseudarthrosis. - 10. Unsuccessful previous attempts at spinal fusion. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use<br>(PER 21 CFR 801.109) | | |------------------------------------------|--| |------------------------------------------|--| | | Mark M | |--|--------| |--|--------| | OR | | |----|--| |----|--| | | Mill | |--|------| |--|------| | Over-The-Counter Use | | |--------------------------|--| | (Optional Format 1-2-96) | | (Division Sign-Off) 1 Division of General. Restorative and Neurological Devices KO317772 510(k) Number _
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...