← Product Code [NHA](/productcode/NHA) · K031719

# ESTHETIC ZIRCONIA ABUTMENT (K031719)

_Nobel Biocare AB · NHA · Jul 2, 2003 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K031719

## Device Facts

- **Applicant:** Nobel Biocare AB
- **Product Code:** [NHA](/productcode/NHA.md)
- **Decision Date:** Jul 2, 2003
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 872.3630
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

Nobel Biocare's Esthetic Zirconia Abutment is intended for use as an anchor to which prosthetic devices, such as artificial teeth, may be attached using dental cement to restore a patient's chewing function.

## Device Story

Esthetic Zirconia Abutment serves as anchor for prosthetic devices (e.g., artificial teeth) to restore chewing function in partially edentulous patients. Device is manufactured from Zirconium Oxide (Y-TZP) to provide improved esthetics compared to traditional Aluminum Oxide abutments. Used in dental clinical settings by dental professionals; attached to endosseous implants via dental cement. Provides structural support for dental prosthetics, benefiting patients through improved aesthetic appearance and restored masticatory function.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on material characterization and comparison to the predicate device.

## Technological Characteristics

Material: Zirconium Oxide (Y-TZP). Form factor: Dental abutment. Function: Endosseous dental implant component. Sterilization: Not specified. Connectivity: None. Software: None.

## Regulatory Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

## Predicate Devices

- Ceramic Abutment ([K913255](/device/K913255.md))

## Reference Devices

- Y-TZP Powder ([K010630](/device/K010630.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 510(k) Summary of Safety and Effectiveness 1.4

| Submitted by:                          | Nobel Biocare AB                                      |
|----------------------------------------|-------------------------------------------------------|
| Address:                               | 15 Bohusgatan<br>Göteborg, SE-402 26<br>Sweden        |
| Name of Contact Person:                | Elizabeth J. Mason, Sr. Regulatory Affairs Specialist |
| Telephone:                             | (714) 282-4800, ext. 7830                             |
| Facsimile:                             | (714) 998-9348                                        |
| Date of Submission:                    | May 30, 2003                                          |
| Classification Name:                   | Endosseous Dental Implant (21 CFR 872.3640)           |
| Trade or Proprietary<br>or Model Name: | Esthetic Zirconia Abutment                            |
| Legally Marketed Device:               | Ceramic Abutment (K913255)                            |

## Device Description:

Nobel Biocare's Esthetic Zirconia Abutment incorporates a material change to the predicate device, Ceramic Abutment (K913255). The intended use of the modified Esthetic Zirconia Abutment, as described in its labeling, has not changed from the unmodified device, Ceramic Abutment (K913255), as a result of this modification. Like the original device, Ceramic Abutment (K913255), the Esthetic Zirconia Abutment is intended for use as an anchor to which prosthetic devices, such as artificial teeth, may be attached using dental cement to restore a patient's chewing function.

The material change consists of using Zirconium Oxide, rather than Aluminum Oxide, in the manufacturing of the device. In order to improve esthetics, a whiter prosthetic device is desired within the dental industry. To achieve this, Nobel Biocare's Esthetic Zirconia Abutment will be manufactured using Zirconium Oxide powder, rather than Aluminum Oxide powder utilized in the predicate device, Ceramic Abutment (K913255).

The FDA previously deared the Zirconium Oxide powder (trade name: Y-TZP Powder) under 510(k) number K010630.

## Indications for Use:

Nobel Biocare's Esthetic Zirconia Abutment is indicated for the treatment of partially edentulous patients requiring prosthetic devices and/or endosseous implants to restore chewing function.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

2 2003 JUL

2003

Nobel Biocare AB C/O Ms. Elizabeth J. Mason Nobel Biocare USA, Incorporated 22715 Savi Ranch Parkway Yorba Linda, California 92887

Re: K031719

Trade/Device Name: Esthetic Zirconia Abutment Regulation Number: 872.3640 Regulation Name: Endosseous Implant Regulatory Class: III Product Code: NHA Dated: May 30, 2003 Received: June 3, 2003

Dear Ms. Mason:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Page 2 - Ms. Mason

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

Suser Russo

Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Statement of Indications for Use 1.3

Page 1 of 1

K031719 510(k) number (if known):

-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Device Name: Esthetic Zirconia Abutment

Indications for Use:

---------

Nobel Biocare's Esthetic Zirconia Abutment is indicated for the treatment of partially edentulous patients requiring prosthetic devices and/or endosseous implants to restore chewing function.

(Please do not write below this line - Continue on another page if needed.)

------------------Concurrence of CDRH, Office of Device Evaluation (ODE)

(Optional Format 3-10-98)

Keen Muluy for MSR
(Division Sign-Off)

Division of Anesthesiology, General Hospital, Infection Control. Dent

510(k) Number: K031719

Nobel Biocare USA, Inc. Special 510(k) Notification Esthetic Zirconia Abutment May 2003

000008

---

**Source:** [https://fda-staging.innolitics.com/device/K031719](https://fda-staging.innolitics.com/device/K031719)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
