K031715 · Applied Imaging Corp. · NOT · Jan 8, 2004 · Hematology
Device Facts
Record ID
K031715
Device Name
ARIOL HER-2/NEU IHC
Applicant
Applied Imaging Corp.
Product Code
NOT · Hematology
Decision Date
Jan 8, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.1860
Device Class
Class 2
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
HER2/neu protein expression scoring
—
—
100% concordance across replicates, runs, and instruments.
—
—
Comparison study: 125 clinical slides of known scoring intensity.
3 (pathologists)
Indications for Use
The Applied Imaging Ariol™ is an automated scanning microscope and image analysis system. It is intended for in vitro diagnostic use as an aid to the pathologist in the detection, classification, and counting of cells of interest based on particular color, intensity, size, pattern, and shape. This Hersight application is intended for use as an accessory to the HercepTest™ (DAKO USA, Carpinteria, CA) and is intended to provide semi-quantitative immunohistochemical (IHC) results to aid in the determination of HER-2/neu (HER2) over-expression in breast cancer tissues routinely processed for histological evaluation. When used with DAKO HercepTest, it is indicated for use as an aid in the assessment of breast cancer patients for whom Herceptin® (Trastuzumab) treatment is being considered. Note: The actual correlation of the DAKO HercepTest to Herceptin clinical outcome has not been cstablished. The Ariol system is an adjunctive computer-assisted methodology to assist the reproducibility of a qualified pathologist in the acquisition and measurement of images from microscope slides of breast cancer specimens stained for the presence of HER2 receptor protein. The accuracy of the test result depends upon the quality of the immunohistochemical staining. It is the responsibility of a qualified pathologist to employ appropriate morphological studies and controls as specified in the instructions for the DakoCytomation HercepTest to assure the validity of the Ariol-assisted HER2 score.
Device Story
Ariol™ is an automated scanning microscope and image analysis system. It captures digital images of histological breast cancer slides stained with DAKO HercepTest. The system performs automated detection, classification, and counting of cells based on color, intensity, size, pattern, and shape. It provides semi-quantitative IHC results to assist pathologists in determining HER2 over-expression. Used in clinical laboratory settings by pathologists to improve reproducibility of HER2 scoring. The output serves as an adjunctive tool; the pathologist remains responsible for final interpretation and validation using morphological studies and controls. The system benefits patients by providing standardized, computer-assisted measurements to support treatment decisions regarding Herceptin therapy.
Clinical Evidence
No clinical data provided in the document.
Technological Characteristics
Automated scanning microscope with motorized stage, focus, and filter wheels. Uses colorimetric pattern recognition for image analysis. Calibration requires laboratory-stained 1+ and 3+ training slides per staining run. Software-based classification of membrane and nuclei. Standalone workstation including computer, monitor, and printer. Bar-code sample identification.
Indications for Use
Indicated for use as an aid to pathologists in the detection, classification, and counting of cells in breast cancer tissue specimens stained for HER2 receptor protein to assess patients being considered for Herceptin (Trastuzumab) treatment.
Regulatory Classification
Identification
Immunohistochemistry test systems (IHC's) are in vitro diagnostic devices consisting of polyclonal or monoclonal antibodies labeled with directions for use and performance claims, which may be packaged with ancillary reagents in kits. Their intended use is to identify, by immunological techniques, antigens in tissues or cytologic specimens. Similar devices intended for use with flow cytometry devices are not considered IHC's.
Special Controls
(2) Class II (special control, guidance document: “FDA Guidance for Submission of Immunohistochemistry Applications to the FDA,” Center for Devices and Radiologic Health, 1998). These IHC's are intended for the detection and/or measurement of certain target analytes in order to provide prognostic or predictive data that are not directly confirmed by routine histopathologic internal and external control specimens. These IHC's provide the pathologist with information that is ordinarily reported as independent diagnostic information to the ordering clinician, and the claims associated with these data are widely accepted and supported by valid scientific evidence. Examples of class II IHC's are those intended for semiquantitative measurement of an analyte, such as hormone receptors in breast cancer.
Submission Summary (Full Text)
{0}
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE AND INSTRUMENT TEMPLATE
A. 510(k) Number:
k031715
B. Analyte:
Her2/neu protein on formalin-fixed paraffin-embedded breast cancer specimens
C. Type of Test:
Computer-assisted image analyzer for immunohistochemistry (immunocytochemistry)
D. Applicant:
Applied Imaging Corporation
E. Proprietary and Established Names:
Applied Imaging Ariol™ with HER2 Application
F. Regulatory Information:
1. Regulation section:
21 CFR §864.1860 Immunohistochemistry reagents and kits
2. Classification:
Class II
3. Product Code:
NOT (microscope, automated, image analysis, operator intervention)
4. Panel:
Pathology 88
G. Intended Use:
The Applied Imaging Ariol™ is an automated scanning microscope and image analysis system. It is intended for in vitro diagnostic use as an aid to the pathologist in the detection, classification, and counting of cells of interest based on particular color, intensity, size, pattern, and shape.
This Hersight application is intended for use as an accessory to the HercepTest™ (DAKO USA, Carpinteria, CA) and is intended to provide semi-quantitative immunohistochemical (IHC) results to aid in the determination of HER-2/neu (HER2) over-expression in breast cancer tissues routinely processed for histological evaluation.
Score.
1. Indication(s) for use:
When used with DAKO HercepTest, it is indicated for use as an aid in the assessment of breast cancer patients for whom Herceptin® (Trastuzumab)
{1}
Page 2 of 6
treatment is being considered. Note: The actual correlation of the DAKO HercepTest to Herceptin clinical outcome has not been established.
2. Special condition for use statement(s):
The Ariol system is an adjunctive computer-assisted methodology to assist the reproducibility of a qualified pathologist in the acquisition and measurement of images from microscope slides of breast cancer specimens stained for the presence of HER2 receptor protein. The accuracy of the test result depends upon the quality of the immunohistochemical staining. It is the responsibility of a qualified pathologist to employ appropriate morphological studies and controls as specified in the instructions for the DakoCytomation HercepTest to assure the validity of the Ariol-assisted HER2.
3. Special instrument Requirements:
Applied Imaging Ariol™
H. Device Description:
The Ariol system is comprised of a computer, monitor, keyboard, mouse, printer, installed software, and a microscope with motorized stage, focus and filter wheels. The Ariol displays images of tissue areas on the monitor. It is the user's responsibility to review the Ariol-generated results and designate the final result on the report form. Automatic relocation, capture and archiving of the cell images are performed by the instrument based upon operator selection.
The Ariol™ system uses the Hersight (HER2 IHC) assay to analyze slides from tissue sections immunohistochemically stained for the presence of the HER-2/neu protein. The system detects HER-2/neu positive cells, but may also identify non-specific brown staining on debris, cytoplasm or edge effects. The instrument software performs two tasks; first, it automatically scans the slide for positively stained areas (based on the training classifier), and secondly it provides capabilities allowing a human operator to identify target regions and confirm the scores of the areas presented by the system.
First, the user trains the Hersight classifiers. Breast cancer tissue slides previously hand-scored by a pathologist as 1+ and 3+ are used for the training. The Hersight classifiers consist of color and shape classifiers for the membrane and nuclei of cells from the 1+ and 3+ cases provided. This training provides sufficient information to allow the system to score all possible cases (0, 1+, 2+, and 3+).
Once the classifiers have been determined, the system is ready to begin automatic scoring of test slides. For each case, the pathologist reviews the Ariol-suggested score in the data grid. The information contained in the data grid assists the pathologist in determining the final score. The pathologist must type the final score in the blank text box provided in the bottom left-hand corner of the Review application.
I. Substantial Equivalence Information:
1. Predicate device name(s)
ChromaVision Medical Systems, Inc. ACIS Her2 software application
{2}
Page 3 of 6
2. Predicate K number(s):
k032113
3. Comparison with predicate:
| DEVICE | PREDICATE |
| --- | --- |
| A. Similarities | |
| Examines formalin-fixed paraffin-embedded breast cancer specimens stained by DakoCytomation HercepTest™ for Her2/neu receptor protein. | Examines formalin-fixed paraffin-embedded breast cancer specimens stained by DakoCytomation HercepTest™ for Her2/neu receptor protein. |
| B. Differences | |
| Ariol™ instrument and software | ACIS instrument and software |
J. Standard/Guidance Document Referenced (if applicable):
None
K. Test Principle:
Method of cell detection is by colorimetric pattern recognition by microscopic examination of prepared cells by size, shape, hue, and intensity as observed by an automated computer controlled microscope and/or by visual observation by a health care professional.
L. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
The Ariol™ HER-2/neu application was evaluated for precision in simulated clinical settings. Precision was assessed via three precision studies, each study with an increasing level of variation in study design.
Precision Study #1- Within-run, within-instrument
The study evaluated a total of 18 slides. The slides were represented by serial sections of each of three clinical tissue slides of known IHC intensities of 1+, 2+ and 3+ (Samples A, B, and C, n = 3 x 3 = 9), and three sets of commercial controls, Levels 1-3, with staining intensities of 0, 1+, and 3+, respectively (n = 3 x 3 = 9). The 18 slides were randomized, masked, scanned, and interpreted with one Ariol system and one pathologist. The slides were also read manually by the same pathologist. There was a minimum of three days between the manual and Ariol readings.
Precision Study #2- Between-run, within-instrument
The 18 slides, as described in Precision Study #1, were re-randomized, re-scanned, and re-interpreted by the same pathologist on the same Ariol instrument over two more "runs." (total of three
{3}
Page 4 of 6
runs). Ariol scores were compared to the manual scores from Study #1. There was a minimum of three days between Ariol runs.
## Precision Study #3- Between-instrument
The 18 slides, as described in Precision Study #1, were re-randomized, re-scanned, and re-interpreted by the same pathologist on two additional Ariol instruments (total of three instruments). Ariol scores were compared to the manual scores from Study #1. There was a minimum of three days between Ariol runs.
## Precision study results
The data exhibited 100% concordance across replicates, runs, and instruments, and among all time factors over which the studies were conducted.
b. Linearity/assay reportable range:
Not applicable.
c. Traceability (controls, calibrators, or method):
The analytical traceability of the system depends on the DakoCytomation HercepTest™. The Ariol™ instrument employs laboratory stained 1+ and 3+ training slides for every different staining run to calibrate the computer-assisted detection system.
d. Detection limit (functional sensitivity):
Not applicable
e. Analytical specificity
The specificity of the test result is dependent on the analytical performance of the DakoCytomation HercepTest™.
f. Assay cut-off:
The assay cut-off of the test result is dependent on the analytical performance of the DakoCytomation HercepTest™. The pathologist must follow the recommendations of the DakoCytomation HercepTest™.
2. Comparison studies:
a. Method comparison with predicate device:
The substantial equivalence studies were based on comparison to conventional manual microscopy performed in accordance with DakoCytomation HercepTest™ instructions for use.
Concordance was evaluated as the agreement between manual HER2 scores and raw Ariol HER2 scores, and manual HER2 scores, and Ariol HER2 scores after they had been reviewed by a pathologist. Approximately 120 clinical slides of known scoring intensity were obtained from a commercial vendor. Requested were: approximately one-third negative slides (0 and 1+), one-third 2+, and one-third 3+ for HER-2/neu over-expression. All slides were randomized, blinded, and read manually by each of three pathologists.
{4}
Page 5 of 6
All slides were bar-coded, re-randomized to ensure blinding, and scanned by an Ariol system. The Ariol then presented scan modes and interpretations three times, once each to the same three pathologists who performed the manual readings. When reviewing the interpretations done on the Ariol system, the pathologists were asked to agree or disagree with the automated score. In the event of disagreement, the pathologist then over-rode the score with his/her own score.
The data were analyzed using 4 x 4 cross-tabulations (0, 1+, 2+, and 3+) that compared the raw Ariol score to the result obtained from the pathologist’s manual evaluation of the same slide, and also for the adjusted Ariol HER2 score after pathologist review. There were 125 slides available for the entire comparison, although Pathologist #3 did not score one of the slides, so this comparison total is 124 instead of 125, as is the case for the other two pathologists against Ariol. Data were also evaluated for inter-pathologist scores. The data demonstrated that agreement of the Ariol raw scores to manual scores equaled or exceeded the agreement of the pathologists among themselves. Agreement is improved even further when the Ariol images are re-reviewed by the pathologist. The Ariol™ also improved the reproducibility between the pathologists.
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical sensitivity:
The clinical sensitivity of the test system is dependent on the analytical performance of the DakoCytomation HercepTest™. The pathologist must follow the recommendations of the DakoCytomation HercepTest™.
b. Clinical specificity:
The clinical specificity of the test system is dependent on the analytical performance of the DakoCytomation HercepTest™. The pathologist must follow the recommendations of the DakoCytomation HercepTest™.
4. Clinical cut-off:
The clinical cut-offs of the test result is dependent on the analytical performance of the DakoCytomation HercepTest™. The pathologist must follow the recommendations of the DakoCytomation HercepTest™.
5. Expected values/Reference range:
DakoCytomation HercepTest™ HER2 scoring range is 0 to 3+.
M. Instrument Name:
Applied Imaging Ariol™
{5}
Page 6 of 6
# N. System Descriptions:
See (H) Device Description.
1. **Modes of Operation:**
Semi-automated computer-assisted interpretation.
2. **Software:**
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types: Yes
3. **Sample Identification:**
Bar-coding of the microscope slides is done before the slides are loaded into the instrument.
4. **Specimen Sampling and Handling:**
The microscope slides to be examined are loaded into the Ariol™ and are scanned automatically. The Ariol™ constructs video images of the scanned data for the pathologist to examine and interpret.
5. **Assay Types:**
Computer-assisted image analysis of formalin-fixed paraffin-embedded breast tissue stained by immunohistochemistry reaction for Her2/neu protein.
6. **Reaction Types:**
Light microscopy
7. **Calibration:**
The Ariol™ instrument employs laboratory-stained 1+ and 3+ training slides for every different staining run to calibrate the computer-assisted detection system.
8. **Quality Control:**
The accuracy of the system depends on the laboratory following the quality control instructions recommended in the labeling of the accessory immunohistochemistry (immunocytochemistry) kit associated with the Ariol™.
# O. Other Supportive Instrument Performance Characteristics Data Not Covered In The "L. Performance Characteristics" Section Of The SE Determination Decision Summary.
# P. Conclusion:
Based on the results of the clinical studies described in this 510(k) submission, it is concluded that the Ariol™ device is as safe and effective (therefore substantially equivalent) as the predicate devices as an aid in the assessment of specimens from breast cancer patients for whom HERCEPTIN® (Trastuzumab) treatment is being considered.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.