← Product Code [GEH](/productcode/GEH) · K031697

# DR. SCHOLL'S FREEZE AWAY WART REMOVER, MODEL 40226 (K031697)

_Schering-Plough Healthcare Products, Inc. · GEH · Aug 29, 2003 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K031697

## Device Facts

- **Applicant:** Schering-Plough Healthcare Products, Inc.
- **Product Code:** [GEH](/productcode/GEH.md)
- **Decision Date:** Aug 29, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4350
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Dr. Scholl's Freeze Away Wart Remover is indicated for the over-the-counter removal of common warts and plantar warts on the bottom of the foot.

## Device Story

Dr. Scholl's Freeze Away Wart Remover is an OTC cryosurgical system for wart removal. Device components include a pressurized canister containing a dimethyl ether and propane cryogen mixture, a reusable activator, and disposable applicators. Operation involves attaching an applicator to the canister, pressing it onto the activator to saturate the applicator with cryogen, and applying the cold tip directly to the wart. The applicator features a luer lock to prevent accidental detachment during use. The device maintains a surface temperature of approximately -42°C to freeze the wart tissue. It is intended for patient self-use at home. The system provides a non-invasive method for wart removal, with labeling providing instructions for self-diagnosis and guidance on when to consult a physician.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing included temperature measurements of the applicator surface (-42°C) and biocompatibility studies conducted per ISO 10993 for limited surface contact with breached/compromised skin.

## Technological Characteristics

Portable cryosurgical system. Cryogen: dimethyl ether and propane mixture. Applicator: disposable, features luer lock for secure attachment. Energy source: pressurized canister. Temperature: -42°C at applicator tip. Biocompatibility: ISO 10993 compliant. Operation: manual activation via reusable activator.

## Regulatory Identification

(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.

## Predicate Devices

- Wartner Wart Removal System ([K011708](/device/K011708.md))
- Histofreezer Wart Removal System ([K023487](/device/K023487.md), [K924114](/device/K924114.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# K031697

113

# PREMARKET NOTIFICATION 510(k) SUMMARY

- 1. Submitter's Name: Schering-Plough HealthCare Products, Inc. Address: 3030 Jackson Avenue, Memphis, TN 38151 Telephone Number: (901) 320-2917 Contact Person: John M. Clayton, PhD Date Prepared: August 8, 2003
- 2. Device Name:

Proprietary Name: Dr. Scholl's® Freeze Away™ Wart Remover Usual Name: OTC Wart Removal System Classification Name: Cryosurgical Unit, Accessories

- 3. Devices to Which Substantial Equivalence is Claimed: Modification to Wartner Wart Removal System, by Wartner Medical Products (Primary Predicate) - K011708 Histofreezer Wart Removal System, by OraSure Technologies, Inc. (Plantar Wart Claim) - K023487, K924114
- 4. Description of Device:

Dr. Scholl's Freeze Away Wart Remover is a cryosurgical system for the over-thecounter removal of common warts and plantar warts on the bottom of the foot. The device consists of the following:

- Pressurized canister with a cryogen mixture of liquid dimethyl ether and propane a.
- b. Reusable activator that releases the cryogen into the applicators
- Disposable applicators C.
- d. Information booklet
- 5. Intended Use Statement:

Dr. Scholl's Freeze Away Wart Remover is indicated for the over-the-counter removal of common warts and plantar warts on the bottom of the foot.

- 6. Comparison of Technological Characteristics:
Dr. Scholl's Freeze Away Wart Remover is substantially equivalent to the Modification to Wartner Wart Remover System by Wartner Medical Products. The only significant difference between the two products is that the Freeze Away Wart Remover product is additionally indicated for the removal of plantar warts. For the over-the-counter plantar wart indication, we are claiming substantial equivalence to the Histofreezer Wart Removal System.

All 3 devices (Freeze Away, Wartner, Histofreezer) are portable cryosurgical systems comprised of a canister containing cryogen and an applicator that is saturated with cryogen and then applied to the wart to be treated.

{1}------------------------------------------------

# K031697

#### a. Laboratory Testing

The average temperature of the Dr. Scholl's Freeze Away Wart Remover applicator surface after saturation is -42°C and is similar to that of the predicate devices (see below).

#### b. Biocompatibility

The cryogen used is a mixture of dimethyl ether and propane, which is the same as the cryogen used in the primary predicate device (Wartner).

The standard biocompatibility studies recommended under ISO 10993 for a medical device with limited surface contact duration to breached/compromised skin were conducted on the applicator material and are included in this 510(k).

### c. Comparison to Predicate Devices:

#### Application

Dr. Scholl's Freeze Away Wart Remover includes an applicator that is saturated with cryogen and then applied to the wart to be treated.

The primary predicate device (Wartner) uses a similar applicator constructed of a different material.

## Applicator Effectiveness Duration

Dr. Scholl's Freeze Away Wart Remover maintains a temperature of -42°C at the applicator tip for approximately 170 seconds when tested using a temperature probe resting on a room temperature Teflon surface. A temperature of -57°C [similar to that listed in 510(k)s K011708 and K924114] is measured if tested using a temperature probe held in mid-air. The method that employs a Teflon surface under the temperature probe is a more appropriate method for simulation of actual use conditions and was used for the substantial equivalence comparison.

The primary predicate device (Wartner) maintains a similar temperature (-36℃) for approximately 75 seconds, while the Histofreezer device maintains the same temperature as Freeze Away (-42°C) for approximately 230 seconds. All devices maintain the freezing temperature for a sufficient length of time to allow the product to be used according to the instructions for use.

#### Crvogen

Dr. Scholl's Freeze Away Wart Remover uses a cryogen composed of dimethyl ether and propane.

The primary predicate device (Wartner) uses the same cryogen.

{2}------------------------------------------------

#### Safety / Ease of Use

Dr. Scholl's Freeze Away Wart Remover includes a safety feature whereby the cryogen will be released into the applicator only after the applicator is attached to the pressurized can and pressed onto the activator. The applicator on the Freeze Away product also contains a luer lock which allows the applicator to be twisted onto the can, preventing accidental detachment during activation or treatment.

The primary predicate device (Wartner) utilizes the first safety feature listed above but does not contain the second safety feature.

#### Indications for Use

Dr. Scholl's Freeze Away Wart Remover is indicated for the over-the-counter removal of common warts and plantar warts.

The primary predicate device (Wartner) is indicated for the over-the-counter removal of common warts, while a second predicate device (Histofreezer) is indicated for the over-the-counter removal of common warts and plantar warts.

#### Labeling

The labeling of Dr. Scholl's Freeze Away Wart Remover has been developed to ensure the consumer has adequate directions for safe use. Freeze Away labeling has also been developed to provide adequate information for the consumer to make a self diagnosis and to ensure that they contact their doctor if in any doubt, if stinging or aching persists after treatment, or if the wart does not improve after four treatments.

The safety and warning statements for the OTC predicate device (Modification to Wartner Wart Removal System) are essentially similar.

#### 7. Conclusion

Based on the information presented above, it is concluded that the proposed Dr. Scholl's Freeze Away Wart Remover is safe and effective for its intended use when used in accordance with label directions and is substantially equivalent to the predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.

AUG 2 9 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Philip Johnson Manager, Regulatory Affairs Schering-Plough Healthcare Products 3 Connell Drive Berkeley Heights, New Jersey 07922

Re: K031697

Trade/Device Name: Dr. Scholl's Freeze Away™ Wart Remover Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical unit and accessories Regulatory Class: II Product Code: GEH Dated: May 30, 2003 Received: June 3, 2003

#### Dear Mr. Johnson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Mr. Philip Johnson

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Miriam C. Provost

(W Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

#### Statement of Indications for Use

510(k) Number (if known): KO31697

Device Name: Dr. Scholl's® Freeze Away™ Wart Remover

Indications for Use: For the over-the-counter removal of common warts and plantar warts on the bottom of the foot.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Over-The-Counter Use

Miriam C. Provost

(Division Sign-Off) Division of General, Restorative and Neurological Devices

510(k) Number K031697

---

**Source:** [https://fda-staging.innolitics.com/device/K031697](https://fda-staging.innolitics.com/device/K031697)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
