K031646 · Bausch & Lomb, Inc. · LPN · Nov 14, 2003 · Ophthalmic
Device Facts
Record ID
K031646
Device Name
BAUSCH & LOMB MULTI-PURPOSE SOLUTION NRC03
Applicant
Bausch & Lomb, Inc.
Product Code
LPN · Ophthalmic
Decision Date
Nov 14, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5928
Device Class
Class 2
Indications for Use
Bausch & Lomb Multi-Purpose Solution NRC03 is indicated for use in the daily cleaning, removal of protein deposits, rinsing, chemical (not heat) disinfection and storage of soft (hydrophilic) contact lenses as recommended by your eye care practitioner.
Device Story
Bausch & Lomb Multi-Purpose Solution NRC03 is a sterile, isotonic chemical solution for soft contact lens maintenance. It contains boric acid, sodium chloride, sodium phosphate, Hydranate (hydroxyalkylphosphonate), poloxamer 407, polyquaternium-10, poloxamine 1107, and polyaminopropyl biguanide (0.0001%) as a preservative. The solution is used by patients at home for daily cleaning, rinsing, chemical disinfection, and storage of lenses. It functions by surfactant-based cleaning and chemical disinfection to remove protein deposits and maintain lens hygiene. The solution is packaged in a plastic bottle with a tamper-evident seal. It benefits patients by providing a convenient, non-heat disinfection method that maintains lens wettability and prevents protein buildup, thereby supporting comfortable lens wear.
Clinical Evidence
A 3-month controlled, randomized clinical study evaluated safety and efficacy. Preclinical testing included cytotoxicity and eye irritation studies (no toxicity found), lens compatibility studies, and microbiology disinfection efficacy testing per FDA guidance. Cleaning efficacy was validated via Critical Micelle Concentration determination and protein deposition analysis using image analysis. Results demonstrated statistical equivalence to control regimens in protein removal and lens cleanliness, with statistically significant improvements in lens wettability.
Technological Characteristics
Sterile isotonic solution. Ingredients: boric acid, sodium chloride, sodium phosphate, Hydranate (hydroxyalkylphosphonate), poloxamer 407, polyquaternium-10, poloxamine 1107, purified water, and polyaminopropyl biguanide (0.0001%). Form factor: plastic bottle with tamper-evident seal. Function: chemical disinfection and surfactant-based cleaning. No software or electronic components.
Indications for Use
Indicated for daily cleaning, protein deposit removal, rinsing, chemical disinfection, and storage of soft (hydrophilic) contact lenses for patients as recommended by an eye care practitioner.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
{0}------------------------------------------------
510(k) Premarket Notification Bausch & Lomb® Multi-Purpose Solution NRC03
# K031646 510 (k) SUMMARY SUMMARY OF SAFETY AND EFFECTIVENESS
NOV 1 4 2003
### FOR
### BAUSCH & LOMB® MULTI-PURPOSE SOLUTION NRC03
### (TRADENAME TO BE DETERMINED)
#### 1. Submitter Information
Bausch & Lomb Incorporated 1 400 North Goodman Street Rochester, New York 14609
Contact Person: Paul G. Stapleton Director, Regulatory Affairs
585-338-8172 Telephone Number:
#### 2. Device Name
Classification Name: Soft (hydrophilic) Contact Lens Care Solution
Proprietary Name: Bausch & Lomb Multi-Purpose Solution NRC03
#### Predicate Devices 3.
Bausch & Lomb ReNu Multi-Purpose Solution, Alcon Opti-Free Express Multi-Purpose Disinfecting Solution and AMO Complete brand Multi-Purpose Solution have been selected as the predicate devices for Bausch & Lomb Multi-Purpose Solution NRC03.
#### Description of the Device 4.
Bausch & Lomb Multi-Purpose Solution NRC03 is a sterile isotonic solution containing boric acid, sodium chloride, sodium phosphate, Hydranate® (hydroxyalkylphosphonate), poloxamer 407, polyquaternium-10, poloxamine 1107, purified water and preserved with polyaminopropyl biguanide (0.0001%). The product is indicated for use in the daily cleaning, removing protein deposits, rinsing, chemical (not heat) disinfection and storage of soft (hydrophilic) contact lenses as recommended by your eye care practitioner. The sterile solution is
{1}------------------------------------------------
contained in a plastic bottle with a tamper evident seal and is labeled with a lot number and expiration date.
#### Indications for Use ડ.
Bausch & Lomb Multi-Purpose Solution NRC03 is indicated for use in the daily cleaning, removal of protein deposits, rinsing, chemical (not heat) disinfection and storage of soft (hydrophilic) lenses as recommended by your eye care practitioner.
#### Description of Safety and Substantial Equivalence Q.
A series of preclinical and clinical studies were completed on this product. The following studies have been completed:
#### Toxicology
A series of cytotoxicity and eye irritation studies of the Bausch & Lomb Multi-Purpose Solution NRC03 and lenses cycled in the solution were undertaken. In these studies, there was no evidence of toxicity.
#### Lens Compatibility
Lens compatibility studies were undertaken after cycling lenses in the solution. The results indicated that lenses were compatible with the solution.
#### Microbiology
Disinfection efficacy was evaluated according to the FDA Guidance document as well as modifications to established procedures. The results indicate satisfactory levels of disinfection efficacy.
#### Cleaning Efficacy
The cleaning efficacy of the solution has been evaluated through the determination of the Critical Micelle Concentration. The surfactant concentrations are well above the CMC for the individual surfactants.
The ability of the solution to prevent the deposition of protein has been established by the evaluation of protein present on the lens following solution usage in a no rub with a rinse regimen with a four (4) hour disinfection time. Protein deposition was significantly prevented as measured by Image Analysis.
The efficacy of the product to remove protein deposits measured by chemical analysis was evaluated with protein deposited lenses. The results indicated statistical equivalence between Test and Control.
{2}------------------------------------------------
### Cleanliness Analysis of Clinical Lens Returns
The analysis of lens cleanliness from both Test and Control regimens indicates that these regimens are substantially equivalent when evaluated by Image Analysis.
### Lens Wettability
Lens wettability over time has been assessed by measurement of contact angle and continued presence of wetting agents over time. The results indicated statistical significance for longer wetting ability over the controls.
### Clinical Studies
A 3-month controlled, randomized study designed to evaluate the safety and efficacy of the Multi-Purpose Solution NRC03 was completed to demonstrate the safety and efficacy of the solution and substantial equivalence to the Control.
### Substantial Equivalence
Bausch & Lomb Multi-Purpose Solution NRC03 is substantially equivalent to Bausch & Lomb ReNu Multi-Purpose Solution, Opti-Free Express Multi-Purpose Disinfecting Solution and Complete brand Multi-Purpose Solution.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three wing-like shapes, symbolizing health and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Public Health Service
NOV 1 4 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Bausch & Lomb C/O Paul G. Stapleton Director. Global Regulatory Affairs 1400 N. Goodman St. Rochester, NY 14609
Re: K031646 Trade/Device Name: Bausch & Lomb Multi-Purpose Solution NRC03 Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (hydrophilic) contact lens care products Regulatory Class: Class II Product Code: LPN Dated: October 14, 2003 Received: October 15, 2003
Dear Mr. Stapleton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Paul G. Stapleton
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A. Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Bausch & Lomb Incorporated 1400 North Goodman Street Rochester, NY 14692-0450
# Indications for Use Statement
0311646 510(k) Number (if known):
Device Name: Bausch & Lomb Multi-Purpose Solution NRC03 (Tradename to be determined)
Indications for Use:
Bausch & Lomb Multi-Purpose Solution NRC03 is indicated for use in the daily cleaning, removal of protein deposits, rinsing, chemical (not heat) disinfection and storage of soft (hydrophilic) contact lenses as recommended by your eye care provider.
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use _____________________________________________________________________________________________________________________________________________________________
OR
Over-The-Counter-Use
-Use
^
Kane Wahabik
510(k) Number K031646
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.