The CVIEW™ is a high performance, digital, cardiac image viewer. The System supports DICOM 3.x and higher cardiac and angio images and displays them in their original dynamic mode. The CVIEW accepts images from DICOM 3.x CDs or can be configured to read DICOM 3 images from a DICOM Archive Server. The CVIEW allows immediate playback of images from any patient stored on the system and burning CDs in the DICOM part 10 - file format. It represents the next generation in network cardiac review systems. Features include 1024 viewing, security, report generation, multi-modality viewing, databasing as well as instant access to hundreds of cases on-line or thousands through the network.
Device Story
CVIEW is a digital cardiac image viewing workstation for cath, echo, and angio studies. Input: DICOM 3.x images from CDs or network archive servers. Operation: System provides 1024-resolution viewing, multi-modality display, and database management. Output: Visual display of dynamic cardiac images and DICOM Part 10 CD burning. Used in clinical settings by healthcare professionals for case review and report generation. Benefits: Provides immediate access to patient studies, facilitating efficient clinical decision-making and workflow in cardiac review environments.
Clinical Evidence
No clinical data provided; not required for this device type.
Technological Characteristics
Digital workstation for DICOM 3.x image viewing and management. Supports 1024-resolution display, multi-modality integration, and network connectivity. Features include report generation, database management, and DICOM Part 10 CD burning capability.
Indications for Use
Indicated for use as a digital cardiac image viewer for clinicians to review, manage, and report on cardiac and angio case studies. Supports DICOM 3.x images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo with the letter 'C' in a bold, curved font, with a circle above it. Below the 'C' is the text 'Electromed' in a bold font, with the word 'imaging' underneath in italics. The logo appears to be for a company called Electromed Imaging.
JUL 2 1 2003
# 510(k) SUMMARY
K 831638
### DATE PREPARED 1
May 21st, 2003
### CONTACT PERSON 2
Warren Baker President, Chief Executive Officer (Official Correspondent)
### DEVICE NAME 3
| Proprietary Name: | CARDIO VASCULAR INFORMATION ENHANCED WORKSTATION™<br>(CVIEW)™ |
|---------------------|---------------------------------------------------------------|
| Common Name: | DICOM Workstation |
| Radiology Panel: | 90LLZ - System, Image Processing |
| Classification Name | Picture Archiving and Communications Systems (PACS) |
| | 21CFR892.2050 |
| | (Federal Register / Vol. 63, No. 82 / April 29, 1998) |
### DEVICE DESCRIPTION AND INTENDED USE 4
Cardio Vascular Information Enhanced Workstation™ (CVIEW)™ is a viewing stations for Cath, Echo and Angio case studies, represent the next generation in network cardiac review systems. Features include 1024 viewing, security, report generation, multi-modality viewing, databasing as well as instant access to hundreds of cases on-line or thousands through the network.
C-510(k) Summary
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo for "Electromed imaging". The logo consists of a stylized letter "C" inside of a circle. Above the letter "C" is a small dot. Below the circle are the words "Electromed" and "imaging" stacked on top of each other.
### SUBSTANTIAL EQUIVALENCE TO PREDICATE DEVICE(S) 5
## Substantially Equivalent (Predicate) Devices
| Product Tradename | Manufacturer | 510(k) Number |
|------------------------------------------------|----------------------------|----------------------|
| ViewNT | Electromed International | K000474 |
| Centricity PACS Plus | GE Medical Systems | K023557 |
| Video Plus System &<br>Echocardiography System | Camtronics Medical Systems | K992666 &<br>K992259 |
| Impax Workstation | AGFA Corporation | K022292 |
### SAFETY AND EFFECTIVENESS 6
The intended use and technological characteristic of the Cardio Vascular Information Enhanced Workstation (CVIEW)™ are similar or equivalent to the Predicate Device(s). Any differences between the CVIEW™and the Predicate Device(s) have no significant influence on the safety and/or effectiveness of the Device.
### CLINICAL PERFORMANCE DATA 7
Not required for determination of substantial equivalence for this type and class of device.
### CONCLUSION DRAWN FROM CLINICAL AND NONCLINICAL TEST DATA 8
Not required for determination of substantial equivalence for this type and class of device.
C-510(k) Summary
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo. The logo consists of a circular border with text around the perimeter. Inside the circle is a stylized image of three birds in flight, overlapping each other. The text around the perimeter of the circle reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 2 1 2003
Mr. Warren Baker President, Chief Executive Officer Electromed Imaging 440 boul. Armand-Frappier, bureau 250, Laval (Québec) H7V 4B4 CANADA
Re: K031638
Trade/Device Name: Cardio Vascular Information Enhanced Workstation™ (CVIEWTM) Regulation Number: 21 CFR 892.1000 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: May 22, 2003 Received: June 4, 2003
Dear Mr. Baker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. Ixxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo with a stylized letter 'C' inside a black circle. Above the 'C' is a small, white, circular dot, possibly representing a light or highlight. Below the circle, the words 'Electromed' are printed in a bold, sans-serif font, with the word 'imaging' directly underneath in a smaller, italicized font. The overall design is simple and professional, likely representing a company in the medical imaging field.
## Indication for Use Statement
510(k) Number:(if known):
No 3 1638
Device Name:
Cardio Vascular Information Enhanced Workstation™ (CVIEW™M)
Indications for Use:
The CVIEW™ is a high performance, digital, cardiac image viewer. The System supports DICOM 3.x and higher cardiac and angio images and displays them in their original dynamic mode. The CVIEW accepts images from DICOM 3.x CDs or can be configured to read DICOM 3 images from a DICOM Archive Server. The CVIEW allows immediate playback of images from any patient stored on the system and burning CDs in the DICOM part 10 - file format.
It represents the next generation in network cardiac review systems. Features include 1024 viewing, security, report generation, multi-modality viewing, databasing as well as instant access to hundreds of cases on-line or thousands through the network.
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Symm
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number *K031638*
Prescription Use
OR
Over-the-Counter Use
(per 21CFR801.109)
CVIEW-USE-STAT
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.