K031634 · Wellhofer Dosimetrie · IYE · Aug 29, 2003 · Radiology
Device Facts
Record ID
K031634
Device Name
OMNIPRO I'MRT
Applicant
Wellhofer Dosimetrie
Product Code
IYE · Radiology
Decision Date
Aug 29, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Indications for Use
The intended use of the OmniPro I'mRT system is to: • verify the treatment plan and delivered dose of intensity modulated or static beams prior to treatment • verify the intensity maps during IMRT delivery prior to treatment • verify the absolute dose in given points for IMRT fields.
Device Story
OmniPro I'mRT is a radiotherapy quality assurance (QA) system used by medical physicists and dosimetry experts in clinical settings. It functions as a software-hardware suite, supporting film scanners, detector arrays, and single-point dosimeters (e.g., Dose 1). The system imports planned dose/intensity data from treatment planning systems (DICOM/RTOG formats) and measured data from irradiated phantoms. It performs signal-to-dose calibration, visualizes dose distributions, and compares measured data against planned distributions using tools like isodose comparison, subtraction, and the Gamma method. The system aids clinical decision-making by verifying the accuracy of IMRT delivery plans prior to patient treatment, ensuring the delivered dose matches the intended plan, thereby enhancing treatment safety and precision.
Clinical Evidence
No clinical testing was performed as the device is a quality assurance tool not involved in direct patient treatment. Performance was evaluated in a clinical environment by medical physicists to confirm operational parameters and intended use.
Technological Characteristics
System consists of software and hardware (film scanners, detector arrays, phantoms, single detectors). Connectivity via standard data formats (DICOM, RTOG). Electrical safety/EMC compliance per IEC 601-1, EN 60 601-1-2, IEC 61010, and CSA 22.2-601.1. Calibration involves two-step conversion: signal to optical density, then optical density to dose. Supports gantry or couch mounting.
Indications for Use
Indicated for use by physicists and dosimetry experts in radiotherapy departments to verify treatment plans, delivered dose, intensity maps, and absolute dose for intensity modulated radiotherapy (IMRT) or static beam treatments prior to patient delivery.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
K∅3/63Y
Section 4 018/1
## AUG 2 9 2003
(a)
510(k) Summary
This summery is submitted in compliance with 21 CFR 807.92
| (1) | Submitted by:<br>Trade name of the company: | Scanditronix Wellhöfer GmbH<br>Bahnhofsstraße 5,<br>D-90592 Schwarzenbruck<br>Germany<br>Scanditronix Wellhöfer |
|-----|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Contact persons: | Thomas Matzen<br>+46 18 18 07 00 |
| | Date of preparation: | 23 May -2003 |
| (2) | Trade name of device: | OmniPro I'mRT |
| | Common name: | Dosimetry System for Quality Assurance |
| | Classification name: | (Accessory to) Radionuclide radiation therapy<br>system, §892,5750; X-ray radiation therapy<br>system, §892,5900; and Medical charged-particle<br>radiation therapy system, §892.5050. |
| (3) | Identification of predicate<br>marketed device: | RIT113 Film Analysis System FDA K935928<br>12/13/1993 |
### Description of the device: (4)
The Scanditronix & Wellhöfer OmniPro I'mRT system is a system similar to Radiological Imaging Technology: RIT113 Film Analysis System FDA K935928 12/13/1993
The OmniPro I'mRT system and the marketed predicated product are radiotherapy quality assurance systems designed to measure dose, doseor intensity-distributions, to analyse these data and to compare the measurement data with calculated dose- or intensity-distributions, especially in the field of Intensity Modulated Radiotherapy.
The OmniPro I'mRT system can consist of
- software (herafter called OmniPro I'mRT) .
- hardware supported by OmniPro I'mRT (like detector arrays, . filmscanning devices, single detectors, bodyphantoms for positioning and pacing of film and/or detectors or detector arrays).
A typical workflow could looks like this:
{1}------------------------------------------------
K\$\phi\$3/63Y
- Section 4
- 1 A treatment planning system (not part of the OmniPro I'mRT system) calculates the dose distribution inside the I'mRT Phantom for a specific IMRT plan.
- 2- Films are placed in the I'mRT Phantom.
- 3- The I'mRT Phantom (and the films) is irradiated according to the above IMRT plan. (Radiation device not part of the OmniPro I'mRT svstem)
- 4- The films are developped (development machine not part of the OmniPro I'mRT system) and scanned using a film scanner supported by OmniPro I'mRT (e.g. Vidar VXR-12).
- 5- Exposure is translated into dose, data are visualised in OmniPro I'mRT where data can be analysed
- 6- The dose plan (point 1-) is imported to OmniPro I'mRT
- 7- Measured dose and planned dose are compared.
- 8- The data and/or the result of the comparison are saved and exported (e.g. printed out).
Note that this is just an example to illustrate the general workflow: Steps in the workflow may be left out, others may be altered or some may be added for example if using other detectors than film. In addition to film detector arrays or single detectors are supported.
- (5) Intended use:
The intended use of the OmniPro I'mRT system is to:
- · verify the treatment plan and delivered dose of intensity modulated or static beams prior to treatment
- · verify the intensity maps during IMRT delivery prior to treatment
- · verify the absolute dose in given points for IMRT fields.
- (6) Technological comparison:
The Scanditronix Wellhöfer OmniPro I'mRT is a Dosimetry System similar to
- Radiological Imaging Technology: RIT113 Film Analysis System FDA K935928 12/13/1993.
Both systems consist of a software supporting hardware. Both allow 1 - import of planned data.
- 2- import of measured data,
- 3- comparison of measured or planned data with measured or planned data.
Overview over relevant parameters:
{2}------------------------------------------------
j3/63y
Section 4 p3 Jul
| Area | RIT113 | OmniPro I'mRT |
|---------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for use | Verifying delivered dose or intensity maps and comparing it with treatment plans especially in<br>intensity modulated radiotherapy (IMRT) treatments | |
| Used by | Physicists and dosimetry<br>experts in radiotherapy<br>departments | Physicists, dosimetry<br>experts in radiotherapy<br>departments |
| Design | Analysing- and<br>measuring-software<br>supporting devices like<br>film scanners | Analysing- and<br>measuring-software<br>supporting devices like<br>film scanners |
| Energy used | Vidar: 95-130 or 190-<br>260VAC, <75W | Vidar 95-130 or 190-<br>260VAC, <75W<br>Lumisys: 100-120V,<br>50/60Hz, 1.5Amps<br>Or 200-240V, 50/60Hz,<br>1.0Amps,<br>BIS: 100-240V, 50/60Hz.<br>0.75-0.35A<br>Dose 1: 100-240VAC,<br>max 40W, typical 15W |
| Energy delivered<br>Electrical safety | No energy delivered.<br>UL1950 (Vidar Dos. Pro) | No energy delivered<br>EN 60-601, IEC 61010<br>(Dose 1), CSA C22.2-601<br>(Dose 1), UL1950 (Vidar<br>Dos. Pro) |
| Performance | RIT113 | OmniPro I'mRT |
| Dose/Fluence Map<br>import format | Dicom<br>RTOG<br>BMP<br>TIFF<br>BINARY<br>Vendor specific:<br>- ADAC Pinnacle<br>- CadPlan<br>- additonally different<br>other vendors | Dicom<br>RTOG<br>+ vendor specific:<br>- ADAC Pinnacle<br>- CadPlan (Varian) |
| Film digitizers Support? | Yes.<br>Vidar (VXR-12, -12plus, -<br>16 Dosimetry Pro) | Yes.<br>Vidar (VXR-12, -16, 16<br>Dosimetry Pro)<br>Lumisys |
| Calibration | Signal to dose | Signal to OD, OD to Dose |
| Body Phantom support | Yes (in principal all<br>available phantoms) | Yes. All phantoms that<br>allow filmpositioning<br>perpendicular to room<br>coordinate system. |
| Support of data from<br>Detector Array<br>measurements | Yes.<br>Import of data from a<br>number of Digital Portal<br>imagers<br>- Varian a500<br>- Varian Portalvision<br>- Elekta<br>- Kodak CR | Yes.<br>Controlling the Beam<br>imaging system (I'mRT<br>QA). |
| Support of data from<br>single detector<br>measurements | ? | Yes |
| | | |
| Difference | Yes | Yes |
| Sum | Yes | Yes |
| Multiplication | No (?) | Yes |
| Profile Comparison | Yes | Yes |
| Isodose comparison | Yes | Yes |
| Gammamethod | No (?) | Yes |
| Plan-plan intercomparison | Available for most operations | All operations |
| Plan –film intercomparison | All operations | All operations |
| Film-film intercomparison | Available for most operations | All operations |
2003-05-23/TM
{3}------------------------------------------------
K\$\phi\$31634
Section 4
PY 11
Comment to the above comparison list:
There are no principal differences between the systems, neither regarding support of measured data, import of data or tools for comparing or analysing of data.
The differences regarding performance are the following
- 1- Import of data: The RIT113 system supports more vendor specific dataformats than OmniPro I'mRT. OmniPro I'mRT supports rather data formats according to standards (DICOM, RTOG), since more and more vendors support either of the standards.
- 2- Film scanners: RIT113 does not support the VXR-16 scanner and Lymisys scanners. The use and the specifications of those scanners however are very similar to the VXR-12 and VXR-16 Dosimetry Pro scanners, parameters like geometric resolution, data deepth (12 or 16 bits) may vary slightly.
- 3- Calibration: The RIT113 performs the calibration in one step from signal to dose, in OmniPro I'mRT the calibration is performed in two steps: First from signal to optical density and then from optical density to dose. The two step method has the adventage that the calibration curves for the scanner and the film can be checked independently from each other.
- 4- Bodyphantom support: Both systems support almost all phantoms on the market (mounting of film in planes parallel or perpendicular to the room coordinate system). This way of mounting films is seen as the standard and all parameters that can be checked with film can be checked using this standard set up.
There is the Gammex 469 phantoms where the film is mounted in a a helix inside a cylinders. This phantom is not supported by OmniPro I'mRT. However this does not limitate the intended use in any way, since with films using the standard set up all parameters to be checked with film can be measured.
- 5- Support of data from Detector Array measurements: RIT113 supports the import of data measured with a detector array (portal imaging systems), OmniPro I'mRT supports the whole measurement process of the detector array (I'mRT QA). The detector arrays supported by RIT113 are optmized for optimal imaging quality. The I'mRT QA has been developed for dosimetric purpose.
The principal of use is very similar: Both systems give a signal that is proportional to the intensity map. The signal can be imported to the main application and than be compared with a planned intensity, fluence- or dose distribution.
2003-05-23/TM
{4}------------------------------------------------
Kd3/63Y
Section 4
ρJΛ6
Both systems are gantry mounted, the I'mRT QA can additionally be placed onto the treatment couch.
- 6- Support of data from single detector measurements: There are two important tasks in verifying IMRT plans: Verifying the relative dose or intensity map (done with film or detector array), verifying the absolute dose in one single point. Therefore OmniPro I'mRT supports measurements done with a single detector. The Dose 1 (510(k) K000209) or other absolute dosimetry
measurement systems can be used.
- Analysing tools: The tools for IMRT field verification are almost 7 identical: Three of the most important tools/operations Film alignment. Subtraction and isodose comparison are available in both systems. In OmniPro I'mRT even multiplication and the gammamethod (see Daniel A. Low et al: A technique for the quantitative evaluation of dose distribution, Med Phys 25 (5) May 1998) are available, allowing the comparison of two datadistributions in one single step instead of using two different methods (subtraction and isodose comparison) in two steps.
#### Non-Clinical tests: (6b) (1)
The OmniPro I'mRT system consists of software and hardware. The hardware supported is mainly the same as supported by the RIT113. Also the data formats (and thereby the planning systems) supported for import are mainly the same.
Since the accuracy of the measured och planned data is limited by the technical specification of the measurement device or the raw planning data rather by the software reading these data we can conclude that a non-clinical comparison tests in a laboratory between the RIT113 and the OmniPro I'mRT system is not necessary to show equivalence.
To minimize potential electrical hazards, Scanditronix Wellhöfer adheres to recognized and established industry practice, and all devices are subject to final performance testing.
All electrical devices of the OmniPro I'mRT System are designed for conformance with national and international standards Dose 1:
Mechanical stability: IEC 61010
CSA 22.2-601.1, EN 60-601, IEC 61010 Electrical safety: EMC: EN 55022. FCC Class B BIS20 / I'mRT QA: IEC 601-1 standards for electrical isolation and leakage current, EN 60 601-1-2, Electromagnetic compatibility.
The quality assurance system at Scanditronix Wellhöfer GmbH is certified since 1995 according to DIN EN ISO 9001 and DIN EN 46001.
- Clinical tests: (2)
{5}------------------------------------------------
K\$\phi\$3/63y
Section 4
P (
Due to the fact that the system is a quality assurance device in radiation treatment not directly involved in the delivery of the treatment radiation, no clinical testing was performed.
However, the device was tested in clinical environment by medical physicists to evaluate the overal performance (indication for use) of the system.
- (3) Conclusion:
Testing operational parameters indicates that the Scanditronix Wellhöfer OmniPro I'mRT System is save, that it fulfills the intended use.
The technological comparison with the predicate device RIT113 Film indicates that it is equivalent with the product RIT113.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows a circular seal with an eagle-like bird in the center. The bird is stylized with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the top half of the circle, following its curvature. The seal appears to be a logo or emblem, possibly representing a government agency related to health and human services in the United States.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 2 9 2003
Scanditronix Wellhöfer GmbH % Mr. Thomas Matzen Official Correspondent Scanditronix Wellhöfer AB Stålgatan 14, 75450 Uppsala SWEDEN
Re: K031634
Trade/Device Name: OmniPro I'mRT System Regulation Number; 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: 90 IYE Dated: May 5, 2003 Received: June 3, 2003
Dear Mr. Matzen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{7}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely vours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Section 3
Ver/ 3 - 4/24/96
| Applicant: | Scanditronix Wellhöfer GmbH |
|---------------------------|-----------------------------|
| 510(k) Number (if known): | K431634 |
| Device Name: | OmniPro I'mRT |
| Indications For Use: | |
Intended Use
The intended use of the OmniPro I'mRT system is to:
• verify the treatment plan and delivered dose of intensity modulated or static beams prior to treatment
- · verify the intensity maps during IMRT delivery prior to treatment
- · verify the absolute dose in given points for IMRT fields.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Per 21 CFR 801.109) (Optional Format 1-2-96)
David G. Seggern
(Division Sign Division of Re and Radiologic 510(k) Num
3-1 (1)
Indication for Use 2003-04-28/TM
*Prescription Use* ✓
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.