← Product Code [KWP](/productcode/KWP) · K031595

# U & I AMERICA SPINAL HOOK SYSTEM (K031595)

_U&I Corp., America · KWP · Jun 24, 2003 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K031595

## Device Facts

- **Applicant:** U&I Corp., America
- **Product Code:** [KWP](/productcode/KWP.md)
- **Decision Date:** Jun 24, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3050
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The U & I Spinal Hook and Rod System intended for use as a posterior, noncervical, nonpedicle system indicated for degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. The U & I Spinal Hook and Rod System can be used in conjunction with the OPTIMA Spinal System or the GLOBAL Spinal Fixation System.

## Device Story

Top-loading posterior spinal fixation system; components include hooks, rods, set screws, connectors, and transverse linking mechanism. Used in posterior plane for unilateral or bilateral fixation; allows adjustment in sagittal and coronal planes for optimal anatomic placement. Implanted by surgeons in clinical/OR settings to stabilize spine. Benefits patient by providing structural support for fusion and deformity correction. Specialized stainless steel instruments used for application and removal.

## Clinical Evidence

Bench testing only. Mechanical testing conducted in accordance with ASTM F1717 to demonstrate equivalence to predicate devices.

## Technological Characteristics

Titanium alloy (Ti-6AI-4V ELI) per ASTM F136. Top-loading posterior spinal fixation system. Includes hooks, rods, set screws, connectors, and transverse linking mechanism. Manual surgical application. Non-sterile (implants) and surgical grade stainless steel instruments.

## Regulatory Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

## Predicate Devices

- Rogozinski Spinal System ([K983804](/device/K983804.md))
- Cotrel-Dubousset (CD™) Spinal system ([K001319](/device/K001319.md))
- Cotrel-Dubousset (CD™) Spinal system ([K955807](/device/K955807.md))

## Reference Devices

- OPTIMA Spinal System
- GLOBAL Spinal Fixation System

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# JUN 2 4 2003

## K031595 ' of 3

### 510(k) SUMMARY

The following 510(k) summary is being submitted as required by 21 CFR 807.92(a):

#### Submission Information:

Classification Name:

|                        | Contact:                    | Patrick Moore<br>Manager, Quality Assurance                                                        |
|------------------------|-----------------------------|----------------------------------------------------------------------------------------------------|
|                        | Sponsor:                    | U&I America<br>6132 South 380 West<br>Murray, UT 84107<br>Phone: 801.262.3100<br>Fax: 801.262.3151 |
|                        | Date Prepared:              | June 11, 2003                                                                                      |
| Device Identification: |                             |                                                                                                    |
|                        | Trade Name:<br>Common Name: | U&I America, Spinal Hook System™<br>Spinal Hook System                                             |

Appiliance, Fixation, Spinal Interlaminal (KWP) per 21 CFR ട്ട് 888.3050

Spinal Hook System

#### Substantially Equivalent Predicate Legally Marketed Devices:

United States Surgical Corporation, Rogozinski Spinal System – KWP, MNH, MNI, KWQ -- (K983804) Cotrel-Dubousset (CD™) Spinal system -- MNH, MNI, KWQ, KWP -- (K001319, K955807)

The substantial equivalence of this device is based on equivalence in intended use, materials, designs and operational principles to the above listed predicate devices.

{1}------------------------------------------------

#### K031595 2 cm 3 Device Description:

The U&I America, Spinal Hook System ™ is a top-loading posterior spinal fixation system which consists of hooks, rods, set screws, connectors, and a transverse (cross) linking mechanism. The Spinal Hook System ™ implant components are fabricated from titanium alloy (Ti-6AI-4V ELI) that conforms to ASTM F 136. Various sizes of these implants are available.

The U&I America, Spinal Hook System ™ can be used in the posterior plane providing unilateral and bilateral modes of fixation.

The U&I America, Spinal Hook System ™ design allows adjustment in both the sagittal and coronal planes permitting hook placement according to the best possible anatomic (spinal) location and orientation.

Specialized instruments made from surgical instrument grade stainless steel are available for the application and removal of the Spinal Hook System ™ implants.

From the foregoing, we conclude that the subject U&I America, Spinal Hook System ™ device is as safe and effective as named predicates and currently marketed competitive devices for the stated indications.

#### Indications for Use:

The U & I Spinal Hook and Rod System intended for use as a posterior, noncervical, nonpedicle system indicated for degenerative disc disease (defined as back pain of discogenic origin with deqeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

The U & I Spinal Hook and Rod System can be used in coniunction with the OPTIMA Spinal System or the GLOBAL Spinal Fixation System.

#### Statement of Technological Comparison:

The subject spinal implant system is substantially equivalent to the above listed predicate devices in terms of materials, design, indications for use and operational principles.

{2}------------------------------------------------

Image /page/2/Figure/0 description: The image shows the alphanumeric string 'K031595' on the top line. Below that, there is a handwritten line that reads '3 c4 3'. The characters in the second line are less clear and appear to be handwritten digits with a possible letter in the middle.

# Performance Data:

Bench testing as listed in Section XII which was conducted in accordance with ASTM F1717 demonstrates equivalence to the above listed predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a human figure with three flowing lines representing the department's services.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 2 4 2003

Mr. Patrick Moore Manager, Quality Assurance U&I America 6132 South 380 West, Suite 200 Murray, Utah 84107

Re: K031595

Trade Name: U&I America, Spinal Hook System™ Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: II Product Code: KWP Dated: May 16 and June 10, 2003 Received: May 21 and June 12, 2003

Dear Mr. Moore:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{4}------------------------------------------------

#### Page 2 - Mr. Patrick Moore

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours.

sincerely yours,

L. Mark N. Millerm

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K031595, Spinal Hook System Additional Information:

### 510(k) Number (if known): K031595

Device Name: U&I America, Spinal Hook System ™

#### Indications for Use:

The U & I Spinal Hook and Rod System intended for use as a posterior, noncervical, nonpedicle system indicated for degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

The U & I Spinal Hook and Rod System can be used in conjunction with the OPTIMA Spinal System or the GLOBAL Spinal Fixation System.

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use OR Over-the-Counter Use_ _____________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109) (Optional Format 1-2-96)

Mark A. Mikkelsen

vision Sign-Off) sion of General, Restorative Neurological Devices

s. (k) Number -

---

**Source:** [https://fda-staging.innolitics.com/device/K031595](https://fda-staging.innolitics.com/device/K031595)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
