3D INVIVO BY CSIST

K031567 · Chung-Shan Institute of Science and Technology · LLZ · Oct 2, 2003 · Radiology

Device Facts

Record IDK031567
Device Name3D INVIVO BY CSIST
ApplicantChung-Shan Institute of Science and Technology
Product CodeLLZ · Radiology
Decision DateOct 2, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

3D inVIVO™ is a PACS System used to acquire, transmit, view, and store image or patient data. This data can be transmitted, stored, and viewed over a computer network or off-site using an Internet connection. The typical users of this system are trained professionals including radiologist, physicians, technicians, and nurses. The 3D inVIVO™ is indicated for capture and storage of 2D images from ultrasound systems and reconstructing them into 3D ultrasound images. These images provide an approximate representation of the 3D volume and are not intended for use in diagnosis or quantitative measurements.

Device Story

3D inVIVO™ is a PC-based PACS software system operating on Windows NT/2000/XP. It acquires successive 2D ultrasound images in real-time from ultrasound systems; performs 3D reconstruction; stores images in a DICOM server; and supports transmission over intranet/internet. Features include freehand scanning, DICOM/non-DICOM format conversion, and GUI tools (zoom, pan, pseudo-color, threshold, Gamma correction). Rendering modes include MIP, X-ray, translucent, surface, slicing, and 3D cine. Supports distance/area/angle measurements and telemedicine (audio/video/text). Used by clinicians to view and store patient data; provides approximate 3D volume representations for visualization. Not intended for primary diagnosis or quantitative measurement.

Clinical Evidence

Bench testing only. No clinical data provided. Hazard analysis performed with potential hazards classified as Minor.

Technological Characteristics

PC-based software application; Windows NT/2000/XP environment. Supports DICOM 3.x and HL7 v2.4 standards. Connectivity via intranet/internet. No patient-contacting components. Does not control life-sustaining equipment.

Indications for Use

Indicated for capture and storage of 2D ultrasound images and reconstruction into 3D volumes for trained professionals (radiologists, physicians, technicians, nurses). Not for diagnostic or quantitative use.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Ko3 1567 OCT = 2 2003 Image /page/0/Picture/2 description: The image shows a logo with a trident inside of a black circle on the left and the text "CSIST" on the right. The trident is white and has three prongs. The text "CSIST" is in a simple sans-serif font. ## 510(k) Summary of Safety and Effectiveness This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990. Date Prepared: July 25, 2003 Submitter's Information: 21 CFR 807.92(a)(1) CHUNG-SHAN INSTITUTE OF SCIENCE AND TECHNOLOGY 481 CHIA-AN SEC. CHUNG-CHENG RD. LUNG-TAN, TAO-YUAN, CHINA (TAIWAN) 32500 Trade Name, Common Name and Classification: 21 CFR 807.92(a)(2) | Trade Name: | 3D inVIVO™ by CSIST | |------------------------|-----------------------------------------| | Common Name: | Picture Archiving Communications System | | Device Classification: | 892.2050 | | Name: | System, Image Processing | Predicate Device: 21 CFR 807. 92(a)(3) | Device Classification<br>Name | SYSTEM, IMAGE PROCESSING, RADIOLOGICAL | |-------------------------------|------------------------------------------------------------------------------| | Regulation Number | 892.2050 | | 510(k) Number | K030635 | | Device Name | THE OBSIDIAN PACS SYSTEM | | Applicant | OBSIDIAN MEDICAL TECHNOLOGY, INC.<br>5108 CORONA CT.<br>PLEASANTON, CA 94588 | | Contact | BIANCA GREEN | | Product Code | LLZ | | Decision Date | 03/28/2003 | Device Description: 21 CFR 807 92(a)(4) The CSIST 3D inVIVO™ interactive 3D diagnostic ultrasound system (3DUS) acquires a set of successive 2D ultrasound images from Ultrasound system and coverts them into a 3D ultrasound image, stores the 3DUS images in the Image Archive of DICOM Image Server, views patient data and medical images via the Intranet or Internet, supplies image availability information to the Department System Scheduler/Order Filler with DICOM 3.x standard, and accept schedule information and updated procedure information from the Department System Scheduler/Order Filler with HL7 version 2.4 standard. {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows a logo with a trident inside of a black circle on the left and the text "CSIST" on the right. The trident is white and has three points. The text "CSIST" is in a simple sans-serif font and is also white. The logo appears to be for an organization or company called CSIST. CSIST Interactive 3DUS system is a PC-based Modality with operating system of MS Windows NT/2000/XP. 3D inVIVO™ provides the advanced process functions, such as : - Supports the freehand scanning, 3D image acquisition, and 3D reconstruction. । - Acquires images from ultrasound system in real time. -- - Supports the ultrasound image storage and ultrasound multi-frame image storage. | - -Converts non-DICOM to DICOM format. - Supports the GUI functions: zoom, pan, pseudo-color setting, threshold setting, | Gamma correction, and thumbnail. - Provides the DICOM query and retrieve services (images and presentation states). - - -Displays the cut planes of 3D image along the long axis or the short axis. - Provides 3D rendering modes: MIP, X-ray, translucent, surface, slicing, and 3D ! cine. - Supports distance/area/angle measurements. - - Supports Tele-medicine (audio, video, and text) between regional medicine center -----------------------------------------------------------------------------------------------------------------------------------------------------------------------------and remote clients. ## Indications for Use: 21 CFR 807 92(a)(5) 3D inVIVO™ is a PACS System used to acquire, transmit, view, and store image or patient data. This data can be transmitted, stored, and viewed over a computer network or off-site using an Internet connection. The typical users of this system are trained professionals including radiologist, physicians, technicians, and nurses. The 3D inVIVO™ is indicated for capture and storage of 2D images from ultrasound svstems and reconstructing them into 3D ultrasound images. These images provide an approximate representation of the 3D volume and are not intended for use in diagnosis or quantitative measurements. ## Technological Characteristics: 21 CFR 807 92(a)(6) The device is a software application and does not contact the patient, nor does it control any life sustaining devices. A physician, providing ample opportunity for competent human intervention interprets images and information being displayed and printed. Conclusion: 21 CFR 807 92(b)(1) The 510(k) Pre-Market Notification for 3D inVIVO™ contains adequate information and data to enable FDA - CDRH to determine substantial equivalence to the predicate device. 3D inVIVO™ will be manufactured in accordance with the voluntary standards listed in the enclosed voluntary standard survey. The submission contains the results of a hazard analysis and the potential hazards have been classified as Minor. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol featuring a staff with two snakes entwined around it. The caduceus is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the left side of the caduceus. The text is written in a sans-serif font and is in all capital letters. Public Health Service OCT - 2 2003 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Chung-Shan Institute of Science and Technology c/o Mr. Carl Alletto 1100 Lakeview Blvd. DENTON TX 76208 Re: K031567 Trade/Device Name: 3D inVIVO™ by CSIST Regulation Number: 21 CFR §892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: July 26, 2003 Received: July 28, 2003 Dear Mr. Aletto: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter: | 8xx. lxxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours. Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure l'age 2 {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows a logo with a trident symbol inside a black circle on the left and the text "CSIST" on the right. The trident is white and has three prongs pointing upwards. The text "CSIST" is in a simple, sans-serif font and is also white. The logo appears to be for an organization or company named CSIST. Page 1 of -1 ## (Indications for Use Form) 510(k) Number: K831567 Device Name: CSIST Interactive 3D Diagnostic Ultrasound System called "3D inVIVO™ Indications for Use: 3D inVIVO™ is a PACS System used to acquire, transmit, view, and store image or patient data. This data can be transmitted, stored, and viewed over a computer network or off-site using an Internet connection. The typical users of this system are trained professionals including radiologist, physicians, technicians, and nurses. The 3D inVIVO™ is indicated for capture and storage of 2D images from ultrasound systems and reconstructing them into 3D ultrasound images. These images provide an approximate representation of the 3D volume and are not intended for use in diagnosis or quantitative measurements. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) f Prescription Use OR Over-The-Counter Use (Per 21 CFR 801.109) (Optional Format 1-2-96) David Ch. Lyman (Division Sign-Off) Division of Reproductive, and Radiological Devices 510(k) Number
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...