K031561 · Ge Medical Systems Information Technologies · DQK · Aug 1, 2003 · Cardiovascular
Device Facts
Record ID
K031561
Device Name
CARDIOSOFT/CASE CARDIAC TESTING SYSTEM
Applicant
Ge Medical Systems Information Technologies
Product Code
DQK · Cardiovascular
Decision Date
Aug 1, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1425
Device Class
Class 2
Attributes
Pediatric
Indications for Use
CardioSoft/CASE Cardiac Testing System is intended to be used by trained operators under direct supervision of a licensed health care practitioner on adult and pediatric patients. The CardioSoft/CASE Cardiac Testing System is designed to acquire process, record, archive, analyze and output (12 and 15 lead) ECG data during a period of physiologic stress or during a resting ECG test, acquire data from ancillary devices (such as Spirometry and Ambulatory Blood Pressure), provide median morphology recordings and record ECG in real-time with and without arrhythmia detection. The arrhythmia detection portion of CardioSoft/CASE Cardiac Testing System is provided to the user for the convenience of automatic detection of arrhythmias but does not provide alarms. CardioSoft/CASE Cardiac Testing System provides the control of external device (typically a treadmill or Ergometer) and communicates with centralized electronic/digital storage system via network. CardioSoft/CASE Cardiac Testing System provides a user selectable option for printouts of prognostic scores on select reports. Vector loops are also available. CardioSoft/CASE Cardiac Testing System can be configured in a network environment for multiple CASE and CardioSoft stations allowing the user to create a central database of patient demographics and collected patient physiological data.
Device Story
System acquires, processes, records, archives, and outputs 12/15-lead ECG data during resting or stress testing; integrates data from ancillary devices (spirometry, ambulatory blood pressure). Operates via software-only package or turnkey system; controls external devices (treadmills/ergometers). Features include median morphology recording, real-time ECG monitoring, and automatic arrhythmia detection (no alarms). Used in hospitals, clinics, and physician offices by trained operators. Provides interpretive statements on morphology, rhythm, and conduction; physician renders final medical opinion. Supports networked central database for patient demographics and physiological data. Benefits include streamlined cardiac testing workflow and centralized data management.
Clinical Evidence
No clinical data. Bench testing only; system validated through requirements specification review, code inspections, software/hardware testing, safety testing, environmental testing, and final validation.
Technological Characteristics
Programmable diagnostic computer for ECG analysis. Software-based data acquisition and processing. Connectivity via network for centralized storage. Controls external exercise equipment (treadmills/ergometers). Supports 12/15-lead ECG, spirometry, and ambulatory blood pressure inputs.
Indications for Use
Indicated for adult and pediatric patients undergoing resting ECG, stress testing, spirometry, or ambulatory blood pressure monitoring. Used by trained operators under licensed healthcare practitioner supervision. Not for transport, home use, vital signs monitoring, or intracardiac use.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
GE Marquette ECG Analysis Program (12SL) (K002209)
Submission Summary (Full Text)
{0}------------------------------------------------
K031561
p. 1/2
# 510(k) Summary of Safety and Effectiveness
| Date: | May 15, 2003 | |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|
| Submitter: | GE Medical Systems Information Technologies | |
| | 8200 West Tower Avenue | |
| | Milwaukee, WI 53223 USA | |
| Contact Person: | David Wahlig | |
| | Sr. Regulatory Affairs Specialist | |
| | GE Medical Systems Information Technologies | |
| | Phone: (262) 293-1705 | |
| | Fax: (414) 918-8112 | |
| Device: | Trade Name: | CardioSoft/CASE Cardiac Testing System |
| Common/Usual Name: | ECG Analysis Computer | |
| Classification Names: | 21 CFR 870.1425 Programmable diagnostic computer | |
| | 21 CFR 870.2920 Transmitters and Receivers, Electrocardiograph, Telephone | |
| | 21 CFR 870.2340 Electrocardiograph | |
| | 21 CFR 870.2340 System, ECG Analysis | |
| | 21 CFR 870.1025 Detector and Alarm, Arrhythmia | |
| Predicate Devices: | CASE 8000 exercise testing system | K991014 |
| | CardioSoft CardioSys | K951130 |
| | Twave Alternans | K023380 |
| | GE Marquette ECG Analysis Program (12SL) | K002209 |
| Device Description: | The CardioSoft/CASE Cardiac Testing System is designed to be used for<br>resting ECG, stress test ECG, Spirometery, Ambulatory Blood Pressure and<br>for recording ECG in real-time with and without arrhythmia detection.<br>The CardioSoft will be offered as a software only package including a front<br>end for data acquisition. | |
| Intended Use: | The CASE is a turnkey product utilizing the CardioSoft software. The CASE<br>device is designed for the US markets. A version designed for the European<br>market, the CardioSys, will not be offered for sale in the US.<br>CardioSoft/CASE Cardiac Testing System is intended to be used by trained<br>operators under direct supervision of a licensed health care practitioner on<br>adult and pediatric patients. The CardioSoft/CASE Cardiac Testing System is<br>designed to acquire process, record, archive, analyze and output (12 and 15<br>lead) ECG data during a period of physiologic stress or during a resting ECG<br>test, acquire data from ancillary devices (such as Spirometry and Ambulatory<br>Blood Pressure), provide median morphology recordings and record ECG in<br>real-time with and without arrhythmia detection. The arrhythmia detection<br>portion of CardioSoft/CASE Cardiac Testing System is provided to the user<br>for the convenience of automatic detection of arrhythmias but does not provide<br>alarms.<br>CardioSoft/CASE Cardiac Testing System provides the control of external<br>device (typically a treadmill or Ergometer) and communicates with centralized | |
| | electronic/digital storage system via network. CardioSoft/CASE Cardiac<br>Testing System provides a user selectable option for printouts of prognostic<br>scores on select reports. Vector loops are also available. | |
| | CardioSoft/CASE Cardiac Testing System can be configured in a network<br>environment for multiple CASE and CardioSoft stations allowing the user to<br>create a central database of patient demographics and collected patient<br>physiological data. | |
{1}------------------------------------------------
CardioSoft/CASE Cardiac Testing System is intended to be used primarily in the hospital but can be used in clinics, physician offices, outreach centers or wherever exercise, stress testing, ECG, Spirometry or ambulatory blood pressure testing is performed.
CardioSoft/CASE Cardiac Testing System offers no diagnostic opinion to the user. Instead it provides interpretive statements of morphology, rhythm, and conduction for which the physician renders his/her own medical opinion.
CardioSoft/CASE Cardiac Testing System is not intended to be used as a transport device or for home use. CardioSoft/CASE Cardiac Testing System is not intended for the use as a vital signs physiological monitor or for intracardiac use.
#### Technology:
The proposed CardioSoft/CASE Cardiac Testing System employs the same technology as the predicate device.
#### Test Summary:
The CardioSoft/CASE Cardiac Testing System complies with the voluntary standards as detailed in Section 9 of this submission. The following quality assurance measures were applied to the development of the system:
- Requirements specification review
- Code inspections .
- Software and hardware testing ●
- Safety testing ●
- Environmental testing .
- Final validation .
#### Conclusion:
The results of these measurements demonstrated that the CardioSoft/CASE Cardiac Testing System is as safe, as effective, and performs as well as the predicate device.
212
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized design featuring three abstract human profiles facing to the right, with flowing lines suggesting movement or energy.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG - 1 2003
GE Medical Systems Information Technologies c/o Mr. David Wahlig Senior Regulatory Affairs Specialist 8200 West Tower Avenue Milwaukee. WI 53223
Re: K031561
Trade Name: CardioSoft/CASE Cardiac Testing Systems Regulation Number: 21 CFR \$870.1025 Regulation Name: Patient Physiological Monitor with arrhythmia detection and alarms Regulatory Class: Class III (three) Product Code: MHX Dated: May 15, 2003 Received: May 19, 2003
Dear Mr. Wahlig:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
## Page 2 - Mr. David Wahlig
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Dele Tum
U-Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### 510(k) Number (if known):
CardioSoft/CASE Cardiac Testing System Device Name:
Indications For Use:
CardioSoft/CASE Cardiac Testing System is intended to be used by trained operators under of a licensed health care practitioner on adult and pediatic patients. The Cardiac Testing System is designed to acquire process, record, archive, and output (12 and 15 lead) ECG data during a period of physiologic stress or during a resting ECG test, acquire data from ancillary devices (such as Spirometry and Ambulatory Blood Pressure), provide median morphology record ECG in real-time with and without arrhythmia detection. The arrhythmia detection portion of Cardiac Testing System is provided to the user for the convenience of automatic detection of arrhythmias but does not provide alarms.
CardioSoft/CASE Cardiac Testing System provides the control of external device (typically a treadmill or Ergoneter) and communicates with centralized electronic/digital storage system via network. Cardiac Testing System provides a user selectable option for printouts of prognostic scores on select reports. Vector loops are also available.
CardioSoft/CASE Cardiac Testing System can be configured in a network environment for multiple CASE and CardioSoft stations allowing the user to create a central database of patient demographics and collected patient physiological data.
CardioSoft/CASE Cardiac Testing System is intended to be used primarily in the hospital but can be used in clinics, physician offices, outreach centers or wherever exercise, stress testing, ECG, Spirometry or ambulatory blood pressure testing is performed.
CardioSoft/CASE Cardiac Testing System offers no diagnostic opinion to the user. Instead it provides interpretive statements of morphology, rhythm, and conduction for which the physician renders hisher own medical opinion.
CardioSoft/CASE Cardiac Testing System is not intended to be used as a transport device or for home use, CardioSoft/CASE Cardiac Testing System is not intended for the use as a vital signs physiological monitor or for intracardiac use.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
N.Qetta
**510(k) Number** K03156
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.