K031540 · Lares Research, Inc. · EKX · Jun 30, 2003 · Dental
Device Facts
Record ID
K031540
Device Name
APEX ELECTRIC HANDPIECE SYSTEM
Applicant
Lares Research, Inc.
Product Code
EKX · Dental
Decision Date
Jun 30, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4200
Device Class
Class 1
Attributes
3rd-Party Reviewed
Indications for Use
The Lares Research Apex™ Electric Handpiece System ("System" defined as and limited to: A. Contra Angle Dental Instrument B. Electric Motor and C. Control System) device consists of a dental handpiece and accessories as listed in 21 CFR 872.4200, product code EKX that is a Class I, non-exempt hand-held device for dental drilling applications. This device is indicated for use where high speeds are required in general dentistry with or without the use of coolant where precision control of the cutting bur is desired such as cutting a tooth for caries, and/or crown preparation and finishing, cutting and/or finishing for dentures, denture bases, crowns, inlays and metal plates and/or root canal preparation for restoration. The device is also indicated for use in difficult to reach areas in general dentistry with or without the use of coolant where precision control of the cutting bur is desired such as the preparing dental caries for restorations, and for use at low speeds in teeth cleaning applications and dental surface polishing. The Apex™ Electric Handpiece System device is driven by an electronically controlled DC electric motor with a pneumatic/electronic actuated control system that is connected to the Contra Angle dental instrument via an ISO-3964 coupling. The Apex™ Electric Handpiece System device is indicated for use by dental professionals only.
Device Story
Apex™ Electric Handpiece System is a dental drilling system comprising a desktop control console, DC electric motor, and contra-angle handpiece. The system connects to existing dental unit air supply via four-hole tubing and 110V/220V electrical power. The dentist operates the system using the existing air-delivery foot pedal to actuate the motor. The control console allows the operator to select motor speed and rotation direction (forward/reverse). The contra-angle instrument attaches via ISO-3964 coupling and accepts various burs. The system provides precise torque, speed, and illumination at the procedure site. Used by dental professionals in clinical settings to facilitate restorative and cleaning procedures. Benefits include enhanced precision and control during dental drilling and polishing tasks.
Clinical Evidence
Bench testing only. Biocompatibility/cytotoxicity evaluation performed by Nelson Laboratories using MEM Elution method on Mouse Heteroploid Connective Tissue L-929 cells. Results showed grade zero (no cellular destruction) for all samples tested after treatment with handpiece conditioner.
Technological Characteristics
Desktop console with electronic microprocessor; DC controlled micromotor; ISO-3964 coupling; autoclavable motor cover (135°C); pneumatic-to-electrical power translation; integrated lighting; triple-port water spray. Standards: IEC 601-1, IEC 601-1-2, ISO 3964, ISO 7785-2, ISO 9687, ISO 11498.1.
Indications for Use
Indicated for dental professionals performing general dentistry procedures including caries removal, crown preparation/finishing, denture/inlay/metal plate finishing, and root canal preparation. Also indicated for teeth cleaning and surface polishing. Applicable for high and low speed use, with or without coolant, in difficult-to-reach areas.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
## June 23, 2003 Revised -Safety and Effectiveness and of Summary of Performance Data and Cytotoxicity Data Review for the Apex™ Electric Handpiece System - Original 510(k)
# K031540
Premarket Notification, Lares® Research, Inc.
This 510(k) summary of safety and effectiveness information was submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92
| I. Submitted By: | Lares Research, Inc. |
|------------------|-----------------------------------|
| | 295 Lockheed Avenue |
| | Chico, CA 95973 USA |
| | Contact: John Dahlgren |
| | Telephone: 530-345-1767 x262 |
| | Fax: 530-345-1870 |
| | E-Mail: jdahlgren@laresdental.com |
II. Date: 510(k) prepared April 6, 2003
III. 510(k) Reason: Marketing and commercial distribution of this device for the first time.
IV. Trade Name: Apex™ Electric Handpiece System
V. Classification Name: Dental Handpiece and Accessories (21 CFR 872.4200)
VI. Classification: The device is a general control of Class I according to 21CFR 872.4200
#### VII. The Apex™ Device Description and List of Predicate Devices
| Company | Device | 510(k) No. | Date Cleared |
|--------------------|--------------------------------|------------|--------------|
| Bien Air USA, Inc. | ORL-E-92 Surgical Drill System | K984244 | 02/23/1999 |
| W & H | Synea WA-99LT | K993526 | 01/11/2000 |
| NSK | Speed Increaser Contra Angle | K972569 | 10/08/1997 |
The Apex™ Electric Handpiece System is a pneumatic/electronic controlled drive for dental drilling applications. The system consists of (3) components: A commercially-available electronic control system and electric motor and, Contra Angle dental instrument.
VIII. Biocompatibility/Cytotoxicity Data Evaluation: Nelson Laboratories (Salt Lake City, UT) was contracted to carry out a biocompatibility-cytotoxicty evaluation of this Lares Research Apex™ Electric Handpiece System after treatment with the Lares Handpiece Conditioner (with lubricant and solvent). Nelson Labs performed a Minimum Essential Medium (MEM) Elution on three (3) separate Lares Handpiece samples [See Appendix #3: Nelson Laboratories MEM Elution [and Cytotoxicty Evaluation] Method and Appendix #4: Nelson Laboratories MEM Elution ~ Final Report for the Lares Research Apex 200LS Contra-Angle Electric Handpiece Instrument] beginning June 13 and ending June 19, 2003.
{1}------------------------------------------------
This type of testing was designed to evaluate the cytotoxicty of any extractable substances or residues associated with the 3 Lares Handpiece samples immediate after the described treatment protocol with the separately sold Lares Research Conditioner [with lubricant and solvent]. The extract of the samples was added to cell line monolayers (Mouse Heteroploid Connective Tissue L-929), and the extract and confluent monolayers incubated at 37°C with 5 +/-1% CO2 for 48 +/- 3 hrs. Cell monolavers were examined microscopically, and scored on the basis of the degree of cellular death and/or damage. Pages 5 - 8 of the attached Nelson Laboratories MEM Elution -Final Report [dated 6-19-03] for the Lares Research Apex 200LS Contra-Angle Electric Handpiece Instrument has provided results demonstration grade zero (0) = No Cellular Destruction was indicated. The 3 Lares Handpiece samples tested, even after the described treatment protocol with the separately sold Lares Research Device Conditioner [with Jubricant and solvent], qave No indication that any cytotoxicity substances were extractable [page 8 of the Nelson report of 6/19/03). Thus the Lares Handpiece Conditioner was deemed to be biocompatible for use with these Handpieces.
IX. Indications for Use: The Lares Research Apex™ Electric Handpiece System ("System" defined as and limited to: A. Contra Angle Dental Instrument B. Electric Motor and C. Control System) device consists of a dental handpiece and accessories as listed in 21 CFR 872.4200, product code EKX that is a Class I, non-exempt hand-held device for dental drilling applications. This device is indicated for use where high speeds are required in general dentistry with or without the use of coolant where precision control of the cutting bur is desired such as cutting a tooth for caries, and/or crown preparation and finishing, cutting for dentures, denture bases, crowns, inlays and metal plates and/or root canal preparation for restoration. The Apex™ Electric Handpiece System device is also indicated for use in difficult to reach areas in general dentistry with or without the use of coolant where precision control of the cutting bur is desired such as the preparing dental caries for restorations, and for use at low speeds in teeth cleaning applications and dental surface polishing. The Apex™ Electric Handpiece System device is driven by an electronically controlled DC electric motor with a pneumaticlelectronic actuated control system that is connected to the Contra Angle dental instrument via an ISO-3964 coupling. The Apex™ Electric Handpiece System device is indicated for use by dental professionals only.
X. Technical Characteristics: The Apex™ Electric Handpiece System consists of a console type desktop housing with the electronic microprocessor with integrated electric and pneumatic-electric controls, a high-torque motor sealed against lubrication from handpieces with removable motor sleeve and, Contra Angle dental instrument.
The Apex™ Electric Handpiece System uses a DC controlled micromotor as the driving element of the system, controlled by an electronic microprocessor that is actuated by the practicing dentist's existing air delivery system via foot pedal. Rotation of the dental instrument is controlled by the translation of air to electrical power wherein a wide range of speeds can be achieved and maintained resulting in a precise control of torque and light at the procedure site.
XI. Principle of Operation: The three components of the Apex™ Electric Handpiece System work in unison to deliver precise torque, speed and light at the procedure site. The principle of operation is as follows:
Apex™ Control System: The Apex™ Control System connects to an existing air unit allowing the use of the Apex™ Electric Motor. Connection of the Apex™ Control System is accomplished via existing four-hole tubing from air supply. The air pressure is reduced to accommodate proper connection of the control. The Control System is then plugged into an existing 110V electrical outlet.
{2}------------------------------------------------
Apex™ Electric Motor: The output tubing from the Apex™ Control System is located and connected to the Apex™ Electric Motor. When connected, the existing air supply foot pedal is depressed and the air pressure is increased to operating levels while the electric motor is running.
Apex™ Contra Angle Dental Instrument: The Apex™ Contra Angle Dental Instrument is connected to the Apex™ Electric Motor via an ISO 3964 coupler. The Apex™ Electric Handpiece System is now ready for use. The dental professional then mounts the bur of choice for the procedure.
At this time, the operator then selects an appropriate speed via the Apex™ Control System control panel labeled "Motor Speed". Rotation direction can be set at the Apex™ Control System control panel to either "forward" or "reverse". Once desired speed and proper rotation initiated, the operator can proceed with the dental procedure accordingly.
| | Bien Air ORL-E-92 | Lares Apex™ | Substantial<br>Equivalence |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Indications for<br>Use | Indicated for the<br>preparation of intra-oral<br>bone for microsurgery and<br>implantology and for<br>apicoectomies. The<br>device is driven by an<br>electric micromotor or an<br>air motor handpiece that<br>has the ISO-3964<br>Coupling. | Indicated for use where high<br>speeds are required in general<br>dentistry with or without the use of<br>coolant where precision control of<br>the cutting bur is desired such as<br>cutting a tooth for caries, and/or<br>crown preparation and finishing,<br>cutting and/or finishing for<br>dentures, denture bases, crowns,<br>inlays and metal plates and/or root<br>canal preparation for restoration.<br>The device is driven by an<br>electronically controlled DC electric<br>motor with a pneumatic/electronic<br>actuated control system that is<br>connected to the Contra Angle<br>dental instrument via an ISO-3964<br>coupling. | Yes |
| Input Voltage | 115V/230V AC 50/60 Hz | 110V/220V AC 50/60 Hz | Yes |
| Materials | Electronic and mechanical<br>parts and components | Electronic and mechanical parts<br>and components | Yes |
| Drive Delivery | DC Micromotor with<br>Controller | DC Micromotor with Controller | Yes |
| Sterilization | Removable Motor Cover<br>is Autoclavable at 135°C | Removable Motor Cover is<br>Autoclavable at 135°C | Yes |
| | Bien Air ORL-E-92 | Lares Apex™ | Substantial<br>Equivalence |
| Indications for<br>Use | When used in conjunction<br>with the micromotor and<br>control indicated above,<br>the device is intended for<br>use where high speed is<br>required in general<br>dentistry with or without<br>use of coolant; such as<br>cutting a tooth for cavity<br>and/or crown preparation<br>and finishing, cutting<br>and/or finishing of<br>dentures, denture bases,<br>crowns, inlays and metal<br>plates. | When used in conjunction with the<br>micromotor and control indicated<br>above, the device is indicated for<br>use where high speeds are<br>required in general dentistry with<br>or without the use of coolant where<br>precision control of the cutting bur<br>is desired such as cutting a tooth<br>for caries, and/or crown<br>preparation and finishing, cutting<br>and/or finishing for dentures,<br>denture bases, crowns, inlays and<br>metal plates and/or root canal<br>preparation for restoration. The<br>device is also indicated for use in<br>difficult to reach areas in general<br>dentistry with or without the use of<br>coolant where precision control of<br>the cutting bur is desired such as<br>the preparing dental caries for<br>restorations, and for use at low<br>speeds in teeth cleaning<br>applications and dental surface<br>polishing. | Yes |
| Speeds | Bien Air CA 1541 1:5 0-<br>200K | Apex™ 200LS 1:5 0-200K | Yes |
| | Bien Air CA 1142 1:1 0-<br>40K | Apex™ 40LS 1:1 0-40K | Yes |
| | Bien Air CA 10141 10:1 0-<br>4K | Apex™ 10LS 5:1 0-7K | Yes |
| Lighting | Yes | Yes | Yes |
| Water Spray | Yes-Triple Port | Yes-Triple Port | Yes |
| Bur Release | Push Button | Push Button | Yes |
| Coupler | ISO 3964 Compatible | ISO 3964 Compatible | Yes |
| Sterilization | 135°C Autoclavable | 135°C Autoclavable | Yes |
#### XII. Substantial Equivalence Comparison - Micromotor and Control:
{3}------------------------------------------------
#### XII. Substantial Equivalence Comparison ~ Contra Angle:
The Lares Research Apex™ Electric Handpiece System device is substantially equivalent to the Bien Air ORL-E-92 Surgical Drill System in commercial distribution by Bien Air USA, Inc. The fundamental technical characteristics of the Apex™ Electric Handpiece System are similar to those of the predicate devices, principally the Bien Air ORL-E-92 Surgical Drill System. The Apex™ Electric Handpiece System is equivalent to the Bien Air ORL-E-92 Surgical Drill System in design, speed, rotation, irrigation and autoclavability. Both Apex™ Electric Handpiece System and Bien Air ORL-E-92 Surgical Drill System use an electric micromotor, parts of which are autoclavable and an electronic control for speed regulation and direction. The range of speed of this device is equivalent to predicate devices and both systems use contra angled handpieces.
{4}------------------------------------------------
XIII. Performance Data: No formal performance data was submitted for this Class I device. Lares Research, Inc. has taken all steps necessary to assure that the Apex™ Electric Handpiece System meets all applicable ISO, IEC and FDA Guidance Standards listed in Section XV of this document and all data supporting this statement is available at Lares Research for review by any authorized agents of the Food and Drug Administration (FDA).
XIV. 510(k) Checklist: This notification contains all information required by 21 CFR 807.87
#### XV. Applicable Standards:
#### Guidance Documents:
Guidance Document on Dental Handpieces, issued July 1995 via the Internet at: http://www.fda.gov/cdrh/ode/556.pdf
Dental Handpiece Sterilization as issued September, 1992 via the Internet at: http://www.fda.gov/cdrh/comp/589.pdf
#### Device Safety:
IEC 601-1, IEC 601-1-2, IEC-IA (as equivalent and fully corresponding to the European standards EN 60601-1, EN 60601-1-2, EN 60601-14)
#### EMC Compatibility:
IEC 601-1-2 with reference to EN 55011, EN 610004-2, EN 610004-3, EN 6100044, EN 610004-5: conformity has been certified during an independent examination by a competent bodv.
#### Dental Handpieces:
ISO 3964 Dental Handpieces - Coupling Dimensions
ISO 7785-2 Dental Handpieces - Part 2: Straight and geared angle handpieces
ISO 9687 Dental Equipment - Graphical Symbols
ISO 11498.1 Dental Handpieces - Dental Low Voltage Electrical Motors
#### XVI. 510(k) Contact Person:
For further information, please contact:
John Dahlgren, Director of Regulatory Affairs Lares Research, Inc., 295 Lockheed Avenue, Chico, CA 95973 USA Telephone: 530-345-1767 x262; Fax: 530-345-1870 E-Mail: jdahlgren@laresdental.com
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo is circular in shape, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a symbol consisting of three stylized human profiles facing to the right, with flowing lines beneath them that could represent water or movement. The overall design is simple and conveys a sense of unity and service.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 3 0 2003
Lares Research, Incorporated C/O Mr. Alfredo J. Quattrone Responsible Third Party Official California Department of Health Services Food & Drug Branch P.O. Box 942732 (MS-357) Sacramento, California 94234-7320
Re: K031540
Trade/Device Name: Apex™ Electric Handpiece System Regulation Number: 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: I Product Code: EKX Dated: June 23, 2003 Received: June 24, 2003
Dear Mr. Quattrone:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Mr. Quattrone
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Susan Turner
Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## INDICATIONS FOR USE
## 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
#### Device Name: Apex™ Electric Handpiece System
#### Indications for Use:
The Lares Research Apex™ Electric Handpiece System ("System" defined as and limited to: A. Contra Angle Dental Instrument B. Electric Motor and C. Control System) device consists of a dental handpiece and accessories as listed in 21 CFR 872.4200, product code EKX that is a Class I, non-exempt hand-held device for dental drilling applications. This device is indicated for use where high speeds are required in general dentistry with or without the use of coolant where precision control of the cutting bur is desired such as cutting a tooth for caries, and/or crown preparation and finishing, cutting and/or finishing for dentures, denture bases, crowns, inlays and metal plates and/or root canal preparation for restoration. The device is also indicated for use in difficult to reach areas in general dentistry with or without the use of coolant where precision control of the cutting bur is desired such as the preparing dental caries for restorations, and for use at low speeds in teeth cleaning applications and dental surface polishing. The Apex™ Electric Handbiece System device is driven by an electronically controlled DC electric motor with a pneumatic/electronic actuated control system that is connected to the Contra Angle dental instrument via an ISO-3964 coupling. The Apex™ Electric Handpiece System device is indicated for use by dental professionals only.
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Ken Mely for VSR
(Division Sian-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
510(k) Number: K031546
Prescription Use:
(Per 21 CFR 801.109)
OR
Over-The-Counter Use:
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.