← Product Code [DIO](/productcode/DIO) · K031512

# EMIT II PLUS COCAINE METABOLITE ASSAY (K031512)

_Dade Behring, Inc. · DIO · Jul 23, 2003 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K031512

## Device Facts

- **Applicant:** Dade Behring, Inc.
- **Product Code:** [DIO](/productcode/DIO.md)
- **Decision Date:** Jul 23, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3250
- **Device Class:** Class 2
- **Review Panel:** Clinical Toxicology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K031512 · Jul 23, 2003 | EMIT II PLUS COCAINE METABOLITE ASSAY | Dade Behring, Inc. | Retrospective human urine samples | Retrospective clinical urine samples were used to compare the performance of the EMIT II Plus Cocaine Metabolite Assay against GC/MS reference results to establish substantial equivalence. | Method comparison; Retrospective samples; GC/MS comparison |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Retrospective method comparison | Human urine samples; Sample Size: 125 (120 native retrospective samples and 5 diluted samples); Number of Sites: 1 (manufacturer's site) | GC/MS | Qualitative agreement at 150 ng/ml and 300 ng/ml cutoffs |

## Indications for Use

The Emit® II Plus Cocaine Metabolite Assay is a homogeneous enzyme immunoassay with a 150 ng/mL or 300 ng/mL cutoff (SAMSHA initial test cutoff level). The assay is intended for use in the qualitative and semi-quantitative analyses of benzoylecgonine (cocaine metabolite) in human urine. Emit® II Plus assays are designed for use with a number of chemistry analyzers. The Emit® II Plus Monoclonal Cocaine Metabolite Assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.

## Device Story

The EMIT II Plus Cocaine Metabolite Assay is a homogeneous enzyme immunoassay used in clinical laboratories on automated chemistry analyzers. It detects benzylecgonine in human urine samples. The assay utilizes antibodies, substrates, and enzyme-labeled benzylecgonine; the presence of the drug in the sample competes with the enzyme-labeled drug for antibody binding sites, modulating enzyme activity. The analyzer measures the resulting signal to provide qualitative or semi-quantitative results. Results are preliminary and require confirmation via gas chromatography/mass spectrometry (GC/MS). The assay assists clinicians in identifying cocaine use; it is intended for professional use in clinical settings.

## Clinical Evidence

Method comparison study performed against GC/MS reference method. At 150 ng/mL cutoff, 93% agreement observed. At 300 ng/mL cutoff, 94% agreement observed. Data demonstrates performance consistency with predicate.

## Technological Characteristics

Homogeneous enzyme immunoassay; utilizes chemistry analyzers; qualitative and semi-quantitative detection of benzoylecgonine; reagent-based system.

## Regulatory Identification

A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.

## Special Controls

*Classification.* Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

## Predicate Devices

- Syva EMIT II Plus Cocaine Metabolite Assay ([K993988](/device/K993988.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE

A. 510(k) Number: K031512
B. Analyte: cocaine metabolite
C. Type of Test: qualitative or quantitative homogeneous enzyme immunoassay
D. Applicant: Syva Company, Dade Behring Inc.
E. Proprietary and Established Names: EMIT II Plus Cocaine Metabolite Assay

F. Regulatory Information:
1. Regulation section: 21CFR862.3250, Cocaine and Cocaine Metabolite Test System.
2. Classification: Class II
3. Product Code: DIO
4. Panel: Toxicology (91)

G. Intended Use:
1. Indication(s) for use: The EMIT II Plus Cocaine Metabolite Assay is a homogeneous enzyme immunoassay with a 150 ng/ml or 300 ng/ml cutoff. The assay is intended for use in the qualitative and semi-quantitative analyses of benzylecgonine (cocaine metabolite) in human urine. EMIT II Plus Assays are designed for use with a number of chemistry analyzers.
2. Special condition for use statement(s): The EMIT II Plus Cocaine metabolite Assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/Mass spectrometry is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgement should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

For prescription use.

3. Special instrument Requirements: The device is for use on automated clinical chemistry analyzers.

{1}

Page 2 of 5

H. Device Description:
The device consists of two reagent sets (Reagents 1 and 2) containing all components of the immunoassay (antibodies, substrates, enzyme-labeled benzylecgonine and inactive components).

I. Substantial Equivalence Information:
1. Predicate device name(s): Syva EMIT II Plus Cocaine Metabolite Assay
2. Predicate K number(s): K993988
3. Comparison with predicate: Indications for use, operating principle and reagent composition are similar to the predicate device. The change to the device includes updated (more stringent) performance claims for correct identification of near-cutoff samples.

J. Standard/Guidance Document Referenced (if applicable):

K. Test Principle: The test is a homogeneous enzyme immunoassay for use on automated clinical chemistry analyzers.

L. Performance Characteristics (if/when applicable):
Performance was evaluated on the Syva 30R Biochemical System.

1. Analytical performance:
a. Precision/Reproducibility: Precision was determined by assaying calibrators and in-house control material for 20 days, 2 runs per day in triplicate. Calculations were according to NCCLS EP5-A.

Qualitative Analysis
|  Concentration (ng/ml) | Mean (mAU/min) | Within-run precision |   | Total precision  |   |
| --- | --- | --- | --- | --- | --- |
|   |   |  standard deviation | %CV | standard deviation | %CV  |
|  0 | 251.4 | 1.5 | 0.6 | 1.7 | 0.7  |
|  113 | 286.1 | 1.8 | 0.6 | 2.0 | 0.7  |
|  150 | 305.0 | 2.0 | 0.7 | 2.3 | 0.8  |
|  188 | 322.6 | 2.3 | 0.7 | 2.5 | 0.8  |
|  375 | 377.6 | 2.1 | 0.6 | 2.6 | 0.7  |

{2}

Page 3 of 5

## Semi-Quantitative Analysis

|  Concentration (ng/ml) | Mean (ng/mL) | Within-run precision |   | Total precision  |   |
| --- | --- | --- | --- | --- | --- |
|   |   |  standard deviation | %CV | standard deviation | %CV  |
|  113 | 98.4 | 4.1 | 4.2 | 4.4 | 4.5  |
|  150 | 141.1 | 4.6 | 3.3 | 5.4 | 3.8  |
|  188 | 183.7 | 5.8 | 3.2 | 6.3 | 3.4  |
|  225 | 217.5 | 6.2 | 2.9 | 6.8 | 3.1  |
|  300 | 302.4 | 8.3 | 2.7 | 9.6 | 3.2  |
|  375 | 372.6 | 10.2 | 2.7 | 12.5 | 3.4  |

## b. Linearity/assay reportable range:

Recovery was determined by spiking negative human urine with known concentrations of benzylecgonine throughout the range 45 ng/ml to 900 ng/ml. Samples at each concentration level were analyzed in replicates of 20. The averaged recoveries of these replicate determinations ranged from 95-112%.

Semiquantitation of positive results enables the laboratory to determine an appropriate dilution of the specimen for confirmation by GC/MS. The semiquantitative mode also permits the laboratory to establish quality control procedures and assess control performance.

## c. Traceability (controls, calibrators, or method):

EMIT calibrators and commercial control materials are required but not supplied.

## d. Detection limit:

Sensitivity was determined by assaying the negative calibrator in duplicate for ten runs and determining a standard curve. The analytical sensitivity (mean plus two standard deviations) is less than 35 ng/ml.

## e. Analytical specificity:

Cross-reactivity was tested by spiking four levels of cocaine and ecgonine. The best line of the nominal value versus the semi-quantitative response was determined. The process was performed across multiple calibrators, lots and instruments. The following concentrations produce responses approximately equivalent to benzylecgonine at the cutoff concentrations.

|  Compound | Concentration (ug/ml) at the 150 ng/ml cutoff | Concentration (ug/ml) at the 300 ng/ml cutoff  |
| --- | --- | --- |
|  Cocaine | 18-53 | 40-119  |
|  ecgonine | 2-6 | 7-20  |

{3}

Page 4 of 5

A variety of over-the-counter and prescription drugs were tested for interference by spiking into negative urine. The drugs and levels in the samples tested are listed in the package insert. No unusual interference was observed from these tests.

f. Assay cut-off: Assay cutoffs are 150 or 300 ng/ml. (See precision/reproducibility section above).

2. Comparison studies:

a. Method comparison with predicate device:

Comparison to GCMS was evaluated at the manufacturer's site for 125 urine samples (120 native retrospective samples and 5 diluted samples). Sample concentrations ranged from 60 to greater than 1000 ng/ml and included at least 11 samples within +25% and 11 samples within -25% of the two cutoff concentrations. Results of the 150 ng/ml and 300 ng/ml cutoffs for the EMIT qualitative assay and GCMS are shown:

Method Comparison Tables

150 ng/ml cutoff, qualitative:

GCMS

|  + EMIT | + 62 | - 2*  |
| --- | --- | --- |
|   | 7* | 54  |

* 8 of these 9 results were near-cutoff samples (within +/-25% of the cutoff concentration).

300 ng/ml cutoff, qualitative:

GCMS

|  + EMIT | + 34 | - 6*  |
| --- | --- | --- |
|   | 1* | 84  |

* 6 of these 7 results were near-cutoff samples (within +/-25% of the cutoff concentration).

b. Matrix comparison: Not applicable. The device is indicated only for urine specimens.

3. Clinical studies:

a. Clinical sensitivity: Not applicable. Clinical studies are not typically submitted for this device type.

{4}

Page 5 of 5

b. Clinical specificity: Not applicable. Clinical studies are not typically submitted for this device type.

c. Other clinical supportive data (when a and b are not applicable):

4. Clinical cut-off: Not applicable

5. Expected values/Reference range: Not applicable

**M. Conclusion:**

I recommend that the EMIT II Plus Cocaine Metabolite Assay is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K031512](https://fda-staging.innolitics.com/device/K031512)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).