K031487 · Nonin Medical, Inc. · DXN · Mar 22, 2004 · Cardiovascular
Device Facts
Record ID
K031487
Device Name
MODEL AVANT 2120 PULSE OXIMETER AND NIBP
Applicant
Nonin Medical, Inc.
Product Code
DXN · Cardiovascular
Decision Date
Mar 22, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Nonin® Model Avant™ 2120 Pulse Oximeter and NIBP Monitor is a portable device indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, and blood pressure of adult and pediatric patients in hospitals, medical facilities, and subacute environments. The Model 2120 is intended for spot-checking and / or continuous monitoring of patients during both no motion and motion conditions, for patients who are well or poorly perfused. Its functions may be used separately or simultaneously.
Device Story
Portable monitor combining pulse oximetry and NIBP; used in hospitals, medical facilities, and subacute environments. Pulse oximetry: passes red/infrared light through tissue; detects fluctuating signals from arterial pulses; measures ratio of absorbed light to determine functional oxygen saturation. NIBP: uses oscillometric step-deflate technique; internal pump inflates cuff; two valves control deflation; microprocessor analyzes cuff pressure changes during deflation to determine blood pressure. Device supports spot-checking and continuous monitoring; functions used separately or simultaneously. Memory playback feature allows data transfer to computer via acquisition software for analysis. Healthcare providers use displayed values for clinical assessment.
Clinical Evidence
No clinical data provided. Substantial equivalence based on technological characteristics and performance comparisons to predicate devices.
Technological Characteristics
Portable monitor; oscillometric NIBP measurement; pulse oximetry via red/infrared light absorption. Microprocessor-controlled signal analysis. Connectivity via data acquisition software for computer transfer. Powered by internal pump and valves for NIBP. No specific material standards or software architecture details provided.
Indications for Use
Indicated for adult and pediatric patients (arm circumference 18-42 cm) in hospitals, medical facilities, and subacute environments for monitoring SpO2, pulse rate, and blood pressure. Pulse oximetry also indicated for infants and neonates. Not indicated for NIBP use in neonates.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Model Avant™ 2120 Pulse Oximeter and NIBP (K013319)
{0}------------------------------------------------
# Special 510(k): Device Modification
MAR 2 2 2004
# Attachment 4
# 510(k) Summary
Nonin Medical, Inc.
Model 2120 Avant™ Pulse Oximeter and Noninvasive Blood Pressure (NIBP) Monitor
# 510(k) Summary
9 May 2003
#### (1) Submitter information
| Name: | Nonin Medical, Inc. |
|----------------------|-----------------------------------------------------------------------------------------------------------------------|
| Address: | 2605 Fernbrook Lane North<br>Plymouth, MN 55447-4755 |
| Telephone: | 763-553-9968 |
| Registration Number: | 2183646 |
| Contact person: | Richard P. Bennett (Official Correspondent) |
| | Nonin Medical, Inc.<br>2605 Fernbrook Lane North<br>Plymouth, MN 55447-4755<br>Tel: 763-577-3166<br>Fax: 763-553-7807 |
| Date prepared: | 9 May 2003 |
## (2) Name of Device
| Device Trade Name: | NONIN® Model Avant™ 2120 Pulse Oximeter and<br>Noninvasive Blood Pressure (NIBP) Monitor |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Common /<br>Classification Name: | Pulse Oximeter with Noninvasive Blood Pressure |
| Device Class: | Class II; 74 DQA, 870.2700; 74 DXN, 870.1130.<br>No performance standards have been established<br>under Section 514 of the Food, Drug and Cosmetic<br>Act for Pulse Oximeters. |
### (3) Legally-marketed predicate devices
Model Avant™ 2120 Pulse Oximeter and NIBP, K013319, clearance date January 3, 2002, by Nonin Medical, Inc.
{1}------------------------------------------------
# Special 510(k): Device Modification
Model N-395 Pulse Oximeter with Extension of Motion Performance Claims to cover three additional oximetry sensors, K993637, clearance date November 24, 1999 made by Nellcor Puritan Bennett.
Masimo SET® Radical Pulse Oximeter, K013792, clearance date December 11, 2001, made by Masimo Corporation.
The NONIN® Model Avant™ 2120 Pulse Oximeter and Noninvasive Blood Pressure (NIBP) Monitor is substantially equivalent to these devices.
### (4) Description
The Model Avant™ 2120 combines Pulse Oximetry and NIBP, utilizing the same fundamental scientific technology and intended use as the predicate devices.
#### Pulse Oximetry
The NONIN® Model Avant™ 2120 Finger Pulse Oximeter passes red and infrared light through perfused tissue and detects the fluctuating signals caused by arterial blood pressure pulses. Well-oxygenated blood is bright red, while poorly oxygenated blood is dark red. The pulse oximeter determines functional oxygen saturation of arterial hemoglobin from this color difference by mcasuring the ratio of absorbed red and infrared light as the blood volume fluctuates with each heart beat.
#### Blood Pressure Measurement
The NONIN® Model Avant™ 2120 NIBP uses an oscillometric stcp deflate technique to determine blood pressure. An internal electric pump is used to inflate the cuff, and deflation is controlled by two valves. During cuff deflation, small cuff pressure changes (resulting from arterial blood pressure pulses) are analyzed by the microprocessor, in order to determine the blood pressure. The Model Avant™ 2120 has the ability to make blood pressure measurcments at predetermined intervals or on demand. The Model 2120 has a Memory playback feature, allowing stored data to be transferred to a computer through data acquisition software for analysis.
### (5) Intended Use
#### Indications for Use:
The NONIN® Model Avant™ 2120 Pulse Oximeter and NIBP Monitor is a portable device indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, and blood pressure of adult and pediatric patients in hospitals, medical facilities, and subacute environments. The Model 2120 pulse oximeter is intended for spot-checking and / or continuous monitoring of patients during both no motion and motion conditions, for patients who are well or poorly perfused. Its functions may be used separately or simultaneously.
{2}------------------------------------------------
# Special 510(k): Device Modification
### Pulse Oximeter Intended Use:
The pulse oximeter is intended for noninvasively monitoring the oxygen saturation and pulse rate of adult, pcdiatric, infant, and neonatal patients in hospitals, medical facilities, and subacute environments. The Model 2120 is intended for spot-checking and / or continuous monitoring of patients during both no motion and motion conditions, for patients who are well or poorly perfused.
#### Blood Pressure Monitor Intended Use:
The blood pressure monitor is intended for noninvasively monitoring the blood pressure of adult and pediatric patients in hospitals, medical facilities, and subacute environments. The blood pressure monitor is not intended for use with neonates. It is intended for attended care and may be used for spot-checking.
The Model 2120 should be used for patients with arm circumferences of 18-42 cm.
#### (6) Conclusion
The NONIN® Model Avant™ 2120 Pulse Oximeter and Noninvasive Blood Pressure (NIBP) Monitor is equivalent in safety and efficacy to the legally marketed predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring a stylized caduceus, which is a symbol often associated with healthcare and medicine.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 2 2 2004
Nonin Medical, Inc. c/o Mr. John R. Dalpee Director of Regulatory Affairs 2605 Fernbrook Lane North Plymouth, MN 55447-4755
Re: K031487
> Trade Name: Model 2120 Avant™ Pulse Oximeter and Noninvasive Blood Pressure (NIBP) Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: II (two) Product Code: 74 DXN Dated: February 16, 2004 Received: February 19, 2004
Dear Mr. Dalpee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 -- Mr. John R. Dalpee
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4648. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Donna R. LaClair
GBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K03i487/S2
# Indications for Use
510(K) Number: K031487
Device Name: Model 2120 Avant™ Pulse Oximeter and Noninvasive Blood Pressure (NIBP) Monitor
## Indications for Use:
The Nonin® Model Avant™ 2120 Pulse Oximeter and NIBP Monitor is a portable device indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, and blood pressure of adult and pediatric patients in hospitals, medical facilities, and subacute environments. The Model 2120 is intended for spot-checking and / or continuous monitoring of patients during both no motion and motion conditions, for patients who are well or poorly perfused. Its functions may be used separately or simultaneously.
Prescription Use
(Per 21 CFR 801.109)
✓
Over-The-Counter Use
Concurrence of CDRH, Office of Device Evaluation (ODE)
or
Duna R. Vochner
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number Ko 31487
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.