SOLUTION FACE MASKS

K031353 · A Plus International, Inc. · FXX · Jun 19, 2003 · General, Plastic Surgery

Device Facts

Record IDK031353
Device NameSOLUTION FACE MASKS
ApplicantA Plus International, Inc.
Product CodeFXX · General, Plastic Surgery
Decision DateJun 19, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

The A Plus International, Inc. Solution Face Masks are designed as surgical apparel medical devices that are intended to fully cover the nose and mouth for infection control, and to be worn (for example) by operating room personnel during surgical procedures, so as to protect both the surgical patient as well as operating room personnel from transfer of microorganisms, body fluids, and particulate materials. Solution Face Masks may be used as isolation masks, procedure masks, or dental face masks, but not as combination surgical and respirator masks. Solution Face Masks are intended as single use disposable devices and not intended for reuse of any kind.

Device Story

Solution Face Masks are single-use, disposable surgical apparel. Device covers nose and mouth to provide barrier protection against microorganisms, body fluids, and particulates. Used by operating room personnel, dental staff, and in isolation settings. Protects both patient and wearer. No electronic or mechanical components; purely physical barrier device.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Disposable face mask; physical barrier material; non-sterile or sterile (implied by surgical apparel classification); single-use; no electronic components or software.

Indications for Use

Indicated for use by operating room personnel, dental staff, or in isolation settings to cover nose and mouth. Protects patient and personnel from transfer of microorganisms, body fluids, and particulates. Not for use as combination surgical/respirator masks. Single-use only.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight, composed of three curved lines. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 1 9 2003 A Plus International, Incorporated C/O Mr. Alfredo J. Quattone California Department of Health Services Food & Drug Bra nch P.O. Box 942732 (MS-357) Sacramento, California 94234-7320 Re: K031353 Trade/Device Name: SOLUTION Face Masks Regulation Number: 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: FXX Dated: June 5, 2003 Received: June 9, 2003 Dear Mr. Quattrone: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Page 2 - Mr. Quattrone Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Susan Peons Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # INDICATIONS FOR USE # 510(k) Number (if known): ### Device Name: # SOLUTION Face Masks K031353 # Indications for Use: The A Plus International, Inc. Solution Face Masks are designed as surgical apparel medical devices that are intended to fully cover the nose and mouth for infection control, and to be worn (for example) by operating room personnel during surgical procedures, so as to protect both the surgical patient as well as operating room personnel from transfer of microorganisms, body fluids, and particulate materials. Solution Face Masks may be used as isolation masks, procedure masks, or dental face masks, but not as combination surgical and respirator masks. Solution Face Masks are intended as single use disposable devices and not intended for reuse of any kind. ### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) St. Barns for Olin iology, General Hospital, 510(k) Number: K031353 Prescription Use (Per 21 CFR 801.109) OR #### Over-The-Counter Use (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...