← Product Code [ODG](/productcode/ODG) · K031347

# ULTRASONIC GASTROVIDEOSCOPE, TYPE UC160P-AT8 AND UCT160-AT8 (K031347)

_Olympus Optical Co., Ltd. · ODG · May 9, 2003 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K031347

## Device Facts

- **Applicant:** Olympus Optical Co., Ltd.
- **Product Code:** [ODG](/productcode/ODG.md)
- **Decision Date:** May 9, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Trans-esophageal(non-cardiac) Other 1) Gastrointestinal tract and the surrounding organs

## Device Story

Ultrasonic gastrovideoscope for real-time endoscopic ultrasound (EUS) imaging, EUS-guided fine needle aspiration (FNA), and endoscopic surgery in the upper gastrointestinal tract and surrounding organs. Device inputs: ultrasonic signals reflected from internal tissues. Principle of operation: piezoelectric transducer transmits sound waves; reflected waves converted to electrical signals; processed for display as B-mode, M-mode, Color Doppler, Amplitude Doppler, and spectral analysis. Used in clinical settings by physicians. Output: real-time ultrasound images and flow data on a monitor. Facilitates visualization of GI tract structures and biopsy guidance. Benefits: enables minimally invasive diagnostic and therapeutic procedures in the upper GI tract.

## Clinical Evidence

Bench testing only. Compliance with IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-2-18, and CISPR11 standards. Biocompatibility testing performed for new distal tip materials.

## Technological Characteristics

Piezoelectric transducer for ultrasound transmission/reception. Modes: B, M, PWD, Color Doppler, Amplitude Doppler. Combined modes: B/M, B/PWD, B/Color Doppler, B/Amplitude Doppler, B/PWD/Color Doppler, B/Amplitude Doppler/PWD. Connectivity: compatible with Olympus video system center, light source, and HDI-5000 ultrasound system. Standards: IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-2-18, CISPR11.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- EUS EXERA Ultrasonic Gastrovideoscope OLYMPUS GF TYPE UC160P-OL5 / OLYMPUS GF TYPE UCT160-OL5 ([K010591](/device/K010591.md))
- Ultrasonic Gastrovideoscope OLYMPUS GF TYPE UC140P-AL5 ([K011314](/device/K011314.md))
- HDI-5000 Ultrasound system ([K961459](/device/K961459.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Olympus America, Inc. % Ms. Laura Danielson Responsible Third Party Official TUV Product Service 1775 Old Highway 8 NW, Suite 104 New Brighton, MN 55112-1891

JUL 2 7 2015

Re: K031347

> Trade/Device Name: Ultrasonic Gastrovideoscope OLYMPUS GF Type UC160P-AT8 and UCT160-AT8 Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: ODG, FET, FDS, ITX Dated (Date on orig SE Itr): April 28, 2003 Received (Date on orig SE Itr): April 29, 2003

Dear Ms. Danielson,

This letter corrects our substantially equivalent letter of May 9, 2003.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA) , You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be

{1}------------------------------------------------

Page 2 -

found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

KD31347

## 4.3.1 Indications for Use Form for Ultrasonic Gastrovideoscope OLYMPUS GF TYPE UC160P-AT8 Ultrasonic Gastrovideoscope OLYMPUS GF TYPE UCT160-AT8

### Diagnostic Ultrasound Indications for Use Form

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  |   | Mode of Operation |   |     |     |                  |                      |                              |                       |                    |  |
|----------------------------------|---|-------------------|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| Clinical<br>Application          | A | B                 | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |  |
| Ophthalmic                       |   |                   |   |     |     |                  |                      |                              |                       |                    |  |
| Fetal                            |   |                   |   |     |     |                  |                      |                              |                       |                    |  |
| Abdominal                        |   |                   |   |     |     |                  |                      |                              |                       |                    |  |
| Intraoperative (specify)         |   |                   |   |     |     |                  |                      |                              |                       |                    |  |
| Intraoperative                   |   |                   |   |     |     |                  |                      |                              |                       |                    |  |
| Neurological                     |   |                   |   |     |     |                  |                      |                              |                       |                    |  |
| Pediatric                        |   |                   |   |     |     |                  |                      |                              |                       |                    |  |
| Small Organ (specify)            |   |                   |   |     |     |                  |                      |                              |                       |                    |  |
| Neonatal Cephalic                |   |                   |   |     |     |                  |                      |                              |                       |                    |  |
| Adult Cephalic                   |   |                   |   |     |     |                  |                      |                              |                       |                    |  |
| Cardiac                          |   |                   |   |     |     |                  |                      |                              |                       |                    |  |
| Transesophageal                  |   | N                 | N | N   |     | N                | N                    |                              | Note1                 | Non-<br>Cardiac    |  |
| Transrectal                      |   |                   |   |     |     |                  |                      |                              |                       |                    |  |
| Transvaginal                     |   |                   |   |     |     |                  |                      |                              |                       |                    |  |
| Transurethral                    |   |                   |   |     |     |                  |                      |                              |                       |                    |  |
| Intravascular                    |   |                   |   |     |     |                  |                      |                              |                       |                    |  |
| Peripheral Vascular              |   |                   |   |     |     |                  |                      |                              |                       |                    |  |
| Laparoscopic                     |   |                   |   |     |     |                  |                      |                              |                       |                    |  |
| Musculo-skeletal<br>Conventional |   |                   |   |     |     |                  |                      |                              |                       |                    |  |
| Musculo-skeletal<br>Superficial  |   |                   |   |     |     |                  |                      |                              |                       |                    |  |
| Other (specify)<br>Note2         |   | N                 | N | N   |     | N                | N                    |                              | Note1                 |                    |  |

N= new Indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments:

Note1: Combined mode operation: B/M.B/PWD,B/Color Doppler,B/Amplitude Doppler,B/PWD/Color Doppler, B/Amplitude Doppler/PWD

Note2: the gastrointestinal tract and the surrounding organs

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

David R. Ingram

(Division Sign-Off) Division of Reproductive, and Radiological Devic 510(k) Number

Prescription Use {Per 21 CFR 801.109}

Section 4.3 Page 2

{3}------------------------------------------------

K031341

## MAY - 9 2003

# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21CFR, Part 807, Subpart E, Section 807.92.

## A. Submitter's name, address, telephone number, initial importer, contact person

#### 1. Manufacturer of the subject device

Name & Address of Manufacturer, Olympus Optical Co., Ltd.

Registration Number : Address, Phone and Fax of R & D Department Endoscope Division

34-3 Hirai Hinode-machi, Nishitama-gun, Tokyo, 190-0182 Japan 30036370927 2951 Ishikawa-cho Hachioji-shi, Tokyo 192-8507 Japan TEL 81-426-42-2891 FAX 81-426-46-5613

2. Initial Importer

Name: Address:

Olympus America Inc. Two Corporate Center Drive Melville, NY 11747-3157 TEL 516-844-5688 FAX 516-844-5416

#### 3. Name of Contact Person

Name:

Ichiro Funabashi Supervisor Regulatory Affairs Quality and Engineering Department Medical System Group Olympus Optical Co., Ltd.

Address, Phone and Fax:

2951 Ishikawa-cho Hachioji-shi, Tokyo 192-8507 Japan TEL 81-426-42-2891 FAX 81-426-46-5613

{4}------------------------------------------------

## B. Device Name, Common Name

## 1. Common/Usual Name

Ultrasonic endoscope

#### 2. Device Name

- · Ultrasonic Gastrovideoscope OLYMPUS GF TYPE UC160P-AT8
- · Ultrasonic Gastrovideoscope OLYMPUS GF TYPE UCT160-AT8

### 3.Classification Name

|                                  | FR Number | Product<br>Code | Class |
|----------------------------------|-----------|-----------------|-------|
| Endoscope and accessories        | 876.1500  | KOG             | II    |
| Diagnostic Ultrasound Transducer | 892.1570  | ITX             | II    |

## C. Identification of the predicate or legally marketed device

The following devices information demonstrates that this device is substantially equivalent to a legally marketed, predicate medical device.

| Device Name                                                                                      | #K      |
|--------------------------------------------------------------------------------------------------|---------|
| EUS EXERA Ultrasonic Gastrovideoscope OLYMPUS GF<br>TYPE UC160P-OL5 / OLYMPUS GF TYPE UCT160-OL5 | K010591 |
| Ultrasonic Gastrovideoscope OLYMPUS GF TYPE UC140P-<br>AL5                                       | K011314 |
| HDI-5000 Ultrasound system                                                                       | K961459 |

## D. Device Description

### 1. Summary

These subject devices have been designed to be used with the HDI5000 Ultrasound system (Philips Ultrasound), Olympus video system center, light Source, documentation equipment, video monitor, endo-therapy accessories such as an aspiration biopsy needle and electrosurgical unit except for endoscopic ultrasound (EUS) guided electrosurgery.

These subject devices are designed for endoscopic real-time ultrasonic imaging, for performing endoscopic ultrasound (EUS) guided fine needle aspiration (FNA) and for endoscopic surgery within the upper gastrointestinal tract and surrounding orgains.

{5}------------------------------------------------

## 2. Design

These subject devices are designed to comply with the standards listed below.

| IEC 60601-1    |
|----------------|
| IEC 60601-1-1  |
| IEC 60601-1-2  |
| IEC 60601-2-18 |
| CISPR11        |

### 3. Materials

The material for Distal Tip of these subject devices has a new patient-contacting material. The biocompatibility test reports of the new material show that the new material is safe for its intended use.

### E. Intended Use:

The intended use of these subject devices, as defined by FDA guidance documents, is:

Trans-esophageal(non-cardiac) Other 1) Gastrointestinal tract and the surrounding organs

#### F. Technological Characteristics:

These devices operate identically to the predicate device in that piezoelectric material in the transducer is used as an ultrasound source to transmit sound waves into the body. Sound waves are reflected back to the transducer and converted to electrical signals that are processed and displayed as images. Doppler shift caused by blood flow is displayed as Color Flow, or as spectrum analysis.

---

**Source:** [https://fda-staging.innolitics.com/device/K031347](https://fda-staging.innolitics.com/device/K031347)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).