The Dynatron Solaris Series of products, including Solaris 709, 708, 706, and 705, provide electrical stimulation, ultrasound and/or combination electrical stimulation and ultrasound, as well as infrared therapy. 1) Electrical Muscle stimulation therapy for: Relaxation of muscle spasms Prevention or retardation of disuse atrophy Increasing local blood circulation Muscle re-education Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis Maintaining or increasing range of motion 2) Transcutaneous electrical nerve stimulation and Interferential Current Therapy for: Symptomatic relief of chronic intractable pain and/or management of post-traumatic or post-surgical pain. 3) Ultrasound therapy for: Applying therapeutic deep heat within body tissues for the treatment of selected medical conditions such as: Relief of pain Muscle spasms Joint contractures 4) Infrared therapy to provide topical heating for: Temporary increase in local blood circulation Temporary relief of minor muscle and joint aches, pains and stiffness Relaxation of muscles Muscle spasms Minor pain and stiffness associated with arthritis
Device Story
Dynatron Solaris Series (models 705, 706, 708, 709) provides multi-modal physical therapy; includes electrical stimulation (Hi Volt, Russian, Biphasic, Direct Current, IFC, Premodulated, Microcurrent, Diadynamic), ultrasound, and optional infrared therapy (D880 probe). System comprises console with alpha-numeric display, software, control electronics, and various transducers/electrodes. Used in clinical settings by healthcare professionals to deliver therapeutic deep heat, muscle stimulation, or nerve stimulation. Output parameters controlled via console; clinician monitors patient response to adjust therapy. Benefits include pain relief, muscle spasm reduction, and improved circulation/range of motion.
Clinical Evidence
No clinical data. Bench testing only; device verified and validated under design/change control per 21 CFR Part 820 and applicable performance standards (21 CFR 1010, 1050).
Technological Characteristics
System console with alpha-numeric display, software, and control electronics. Modalities: electrical stimulation (various waveforms), ultrasound, infrared. Conforms to 21 CFR 1010 and 1050. Components: ultrasound transducer, infrared probe, electrotherapy multi-probe, patient cables, electrodes, gel, straps.
Indications for Use
Indicated for patients requiring electrical muscle stimulation (spasm relaxation, atrophy prevention, blood circulation, re-education, post-surgical calf stimulation, range of motion), TENS/IFC (chronic intractable pain, post-traumatic/post-surgical pain), ultrasound (deep heat for pain, spasms, contractures), or infrared therapy (topical heating for circulation, aches, stiffness, arthritis).
Regulatory Classification
Identification
An ultrasonic diathermy for use in applying therapeutic deep heat for selected medical conditions is a device that applies to specific areas of the body ultrasonic energy at a frequency beyond 20 kilohertz and that is intended to generate deep heat within body tissues for the treatment of selected medical conditions such as relief of pain, muscle spasms, and joint contractures, but not for the treatment of malignancies.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a sequence of handwritten alphanumeric characters. The characters are 'K031329'. The characters are written in a dark ink on a white background.
## OCT 2 2 2003
## SUMMARY OF SAFETY AND EFFECTIVENESS
DYNATRONICS CORPORATION SUBMITTED BY: 7030 Park Centre Drive Salt Lake City UT 84121 Phone: (800) 874-6251; (801) 568-7000; Fax: (801) 568-7711
- 1. DEVICE NAME (Trade/common, and classification): Dynatron® Solaris™ Series (model numbers 709, 708, 706, and 705); Therapeutic Ultrasound, Interferential Current Therapy, Electrical Muscle Stimulator, Electrotherapy, Non-Invasive Electrical Nerve Stimulator, Ultrasound and Muscle Stimulator (and Accessories), and Infrared therapy.
| Classification: | Class II |
|-------------------|-----------------------------------------------------|
| Regulation Nos .: | 882.5890, 890.5300, 890.5850, 890.5860 and 890.5500 |
| Product Codes: | GZJ, IMI, IPF, IMG, LIH and ILY |
## 2. PREDICATE DEVICES:
- Solaris 709 Previously Dynatron 950, cleared under K950349 (Sep 13, 1995); Solaris 708 - Previously Dynatron 850, cleared under K941461 (Sep 01, 1995); Solaris 706 - Previously Dynatron 650, cleared under K950348 (Sep 13, 1995); Solaris 705 - Previously Dynatron 550, cleared under K941577 (Sep 01, 1995); Solaris D880 Infrared Probe -- SE to MedX 1000 Series, cleared under K020017 (July 12, 2002)
- PERFORMANCE STANDARDS: The Dynatron Solaris Series of devices conform to 3. the applicable requirements of 21 CFR sections 1010 (Performance Standards for Electronic Products: General) and 1050 (Performance Standards for Sonic, Infrasonic, and Ultrasonic Radiation-Emitting Products).
- 4. DESCRIPTION: The Dynatron® Solaris™ Series (Model numbers 709, 708, 706, and 705) provide electrical stimulation, ultrasound, and/or electrical stimulation and ultrasound, as well as optional infrared therapy.
Components:
System console, containing software and control electronics with alpha-numeric displays
Ultrasound transducer, infrared probe, and multi-probe for administering electrotherapy
Accessories such as patient cables, electrodes, transducer gel and elastic straps.
{1}------------------------------------------------
- 5. INTENDED USE/INDICATIONS FOR USE: The Dynatron Solaris Series of products, including Solaris 709, 708, 706, and 705, provide electrical stimulation, ultrasound and/or combination electrical stimulation and ultrasound, as well as infrared therapy.
- 1) Electrical Muscle stimulation therapy for: Relaxation of muscle spasms Prevention or retardation of disuse atrophy Increasing local blood circulation Muscle re-education Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis Maintaining or increasing range of motion
- 2) Transcutaneous electrical nerve stimulation and Interferential Current Therapy for:
Symptomatic relief of chronic intractable pain and/or management of post-traumatic or post-surgical pain.
- 3) Ultrasound therapy for:
Applying therapeutic deep heat within body tissues for the treatment of selected medical conditions such as:
- Relief of pain Muscle spasms Joint contractures
- 4) Infrared therapy to provide topical heating for:
Temporary increase in local blood circulation Temporary relief of minor muscle and joint aches, pains and stiffness Relaxation of muscles
Muscle spasms
Minor pain and stiffness associated with arthritis
The Intended Use/Indications For Use stated herein are consistent with the cleared indications for the predicate devices.
- 6. SUBSTANTIAL EQUIVALENCE (SE) RATIONALE: The Dynatron Solaris 709, 708, 706, and 705, offer electrical stimulation, ultrasonic therapy and/or a combination of the two, and share the same or similar basic characteristics and the same general intended use in physical medicine, general and plastic surgery and neurology as the predicate devices. Therefore, the proposed Dynatron Solaris products are substantially equivalent to the Dynatron 950, 850, 650 and 550. The infrared therapy probe provides topical heating for treatment of selected medical conditions and shares the same or similar basic characteristics, features and intended use as the predicate and, therefore, is substantially equivalent to the MedX 1000 Series infrared probe (applicable 'K' numbers listed above).
{2}------------------------------------------------
- 7. SAFETY AND EFFECTIVENESS: There are no substantive differences between the products defined in this 510(k) submission and the predicate devices. They are similar to the technologies that are currently used in other similar medical devices. They were developed and documented under Dynatronics' mature Quality Management System, under the Quality System Regulation, 21 CFR Part 820, under design/change control, and is verified/validated to applicable standards/guidance documents. The Solaris family of electrical stimulation, ultrasound, and/or combination products, and accessories, are safe and effective, when used as indicated in specific applications under a clinician's supervision/therapy program.
Signed: Ronald J. Hatch
Ronald J. Hatch, VP Operations/RA
DYNATRONICS CORPORATION
Dated: April 24, 2003
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three lines representing its body and head. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 2 2 2003
Ronald J. Hatch VP Operations/Regulatory Affairs Dynatronics, Corp. 7030 Park centre Dr. Salt Lake City, Utah 84121
Re: K031329
Trade/Device Name: Solaris Models 708 and 709 Regulation Number: 21 CFR 890.5300 Regulation Name: Ultrasonic Diathermy Regulatory Class: Class II Product Code: IMI
Trade/Device Name: Solaris Models 705, 706, 708 and 709 Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: IPF
Trade/Device Name: Solaris Models 705, 706, 708 and 709 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator v Regulatory Class: Class II Product Code: GZJ
Trade/Device Name: Solaris Models 705, 706, 708 and 709 Regulation Number: 21 CFR 882.5890 Regulation Name: Interferential Current Therapy Regulatory Class: Class II Product Code: LIH
Trade/Device Name: D880 Infrared Probe for use with Solaris Models 705, 706, 708 and 709 Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Heat Lamp Regulatory Class: Class II Product Code: ILY
{4}------------------------------------------------
Page 2 - Mr. Ronald J. Hatch
Dated: August 29, 2003 Received: September 4, 2003
## Dear Mr. Hatch:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll
{5}------------------------------------------------
Page 3 – Mr. Ronald J. Hatch
free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Sincerely yours,
Mark N. Milkerson
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number: K031329
Device Name: Dynatron® Solaris™ Series
Indications for Use:
Hi Volt, Russian, and Biphasic modes of stimulation are all forms of electrical muscle stimulation therapy available with the Solaris 709, 708, 706 & 705 and indicated for:
> Relaxation of muscle spasms Prevention or retardation of disuse atrophy Increasing local blood circulation Muscle re-education Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis Maintaining or increasing range of motion
The Direct Current mode of stimulation is also a form of electrical muscle stimulation therapy available with the Solaris 709, 708, 706 & 705 and indicated for:
Relaxation of muscle spasms
IFC, Premodulated, Microcurrent and Diadynamic modes of stimulation are all forms of transcutaneous electrical nerve stimulation (TENS) or Interferential Current Therapy (IFC) available with the Solaris 709, 708, 706 & 705 and indicated for:
> Symptomatic relief of chronic intractable pain and/or management of post traumatic or post-surgical pain
for Mark A. Milliner
[continued]
Division Sign-Off.
Division of General Restorative
and Neurological Devices
510(k) Number K031329
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
{7}------------------------------------------------
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Optional Format 3-10-98)
Ultrasound therapy is available with the Solaris 709 & 708 and indicated for:
Applying therapeutic deep heat within body tissues for the treatment of selected medical conditions such as:
> Relief of pain , Muscle spasms Joint contractures
Infrared therapy is available as an optional accessory (D880) probe with the Solaris 709, 708, 706 & 705 and indicated to provide topical heating for:
> Temporary increase in local blood circulation Temporary relief of minor muscle and joint aches, pains and stiffness Relaxation of muscles Muscle spasms Minor pain and stiffness associated with arthritis
for Mark N. Millerson
vision Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number K031329
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.