← Product Code [LTK](/productcode/LTK) · K031297

# ACCESS OV MONITOR ASSAY (K031297)

_Beckman Coulter, Inc. · LTK · May 2, 2003 · Immunology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K031297

## Device Facts

- **Applicant:** Beckman Coulter, Inc.
- **Product Code:** [LTK](/productcode/LTK.md)
- **Decision Date:** May 2, 2003
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 866.6010
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Indications for Use

The Access OV Monitor assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of CA 125 antigen levels in human serum and plasma, using the Access Immunoassay Systems. This device is indicated for use in the measurement of CA 125 antigen to aid in the management of ovarian cancer patients. Serial testing for patient CA 125 antigen concentrations should be used in conjunction with other clinical methods used for monitoring ovarian cancer.

## Device Story

Access OV Monitor assay is a paramagnetic particle, chemiluminescent immunoassay; measures CA 125 antigen levels in human serum and plasma. Used on Beckman Coulter Access Immunoassay Systems (including UniCel Dxl 800). Operated by laboratory personnel in clinical settings. Input: patient serum/plasma samples. Process: automated immunoassay system performs quantitative analysis. Output: CA 125 antigen concentration values. Healthcare providers use serial test results alongside other clinical methods to monitor ovarian cancer management; aids in tracking disease progression or response to therapy.

## Clinical Evidence

Bench testing only. Method comparison, precision, and analytical sensitivity studies were conducted to compare performance on the UniCel Dxl 800 system versus the Access 2 system. Results met all established acceptance criteria.

## Technological Characteristics

Paramagnetic particle, chemiluminescent immunoassay. Reagents and calibrators are identical to those used in the predicate Access 2 system. System is an automated immunoassay analyzer. No changes to assay formulations, substrate, or wash buffer.

## Regulatory Identification

A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.

## Special Controls

*Classification.* Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.

## Predicate Devices

- Access OV Monitor Assay ([K023597](/device/K023597.md))

## Reference Devices

- Beckman Coulter UniCel Dxl 800 Access Immunoassay System ([K023764](/device/K023764.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# MAY - 2 2003

510(K) Summary 1.4

(Page 1 of 2)

#### 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

510(k) Number K031297

Date Prepared: April 22,2003

| Submitter                 | Contact Person                      |
|---------------------------|-------------------------------------|
| Beckman Coulter, Inc      | Lynn Weist                          |
| 1000 Lake Hazeltine Drive | Staff Regulatory Affairs Specialist |
| Chaska, MN 55318          | Phone: 952-368-1271                 |
|                           | Fax: 952-368-7710                   |

#### General Information

| Proprietary Name                        | Access® OV Monitor Assay                            |
|-----------------------------------------|-----------------------------------------------------|
| Classification Name                     | Tumor Associated Antigen Immunological Test System  |
| Device Class                            | Class II                                            |
| Legally Marketed<br>(Unmodified) Device | Access OV Monitor Assay (K023597, cleared 12/11/02) |

#### Device Description

The Access® OV Monitor reagents, calibrators and the Access Immunoassay Systems family comprise the Access Immunoassay System for the quantitative determination of CA 125 antigen in human serum and plasma.

### Intended Use

The Access OV Monitor assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of CA 125 antigen levels in human serum and plasma, using the Access Immunoassay Systems. This device is indicated for use in the measurement of CA 125 antigen to aid in the management of ovarian cancer patients. Serial testing for patient CA 125 antigen concentrations should be used in conjunction with other clinical methods used for monitoring ovarian cancer.

{1}------------------------------------------------

# 510(k) Summary

## (Page 2 of 2)

# Description of the Modification to the Legally Marketed Device

The modification to the Access OV Monitor assay is to add a new instrument platform, the Beckman Coulter UniCel™ Dxl 800 Access® Immunoassay System, to the family of Access Immunoassay Systems. The Dxl System is a new model within the same model series of Access Immunoassay Systems manufactured and distributed by Beckman Coulter, Inc. The DxI System was cleared for marketing by FDA on January 28, 2003, (K023764).

The DxI uses the same Access OV Monitor reagents and calibrators, packaged the same as for the Access 2 System. The formulations of the substrate and wash buffer used with the Access OV Monitor assay are unchanged. There are no changes to the intended uses, technical specifications or final performance specifications and claims for the assay.

# Supporting Data

In order to demonstrate that the Access OV Monitor assay on the Dxl system is substantially equivalent to the Access OV Monitor assay on the Access 2 system. method comparison, precision and analytical sensitivity studies were conducted. The Access OV Monitor assay met the established acceptance criteria for method comparison, precision and analytical sensitivity.

## Conclusion

The information provided in this submission supports a substantial equivalence determination, and therefore, 510(k) premarket notification clearance of the Access OV Monitor assay on the UniCel Dxl 800 Access Immunoassay System.

Section of the consisted

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are black and have a flowing, abstract design. The image is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in black, uppercase letters.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Lynn Weist Staff Regulatory Affairs Specialist Beckman Coulter, Inc. 1000 Lake Hazeltine Drive Chaska, Minnesota 55318-1084

MAY - 2 2003

Re: k031297 Trade/Device Name: Access® OV Monitor Assay Regulation Number: 21 CFR § 866.6010 Regulation Name: Tumor associated antigen immunological test system Regulatory Class: II Product Code: LTK Dated: April 23, 2003 Received: April 24, 2003

Dear Ms. Weist:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title. 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{3}------------------------------------------------

Page 2 -

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known): K031297

PAGE 1 OF 1

Device Name: Access® OV Monitor Assay

Indications for Use:

The Access OV Monitor assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of CA 125 antigen levels in human serum and plasma, using the Access Immunoassay Systems. This device is indicated for use in the measurement of CA 125 antigen to aid in the management of ovarian cancer patients. Serial testing for patient CA 125 antigen concentrations should be used in conjunction with other clinical methods used for monitoring ovarian cancer.

# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use OR Over-the-Counter Use (per 21 CFR 801.109) (Optional Format 1-2-96) al Laboratory Devices 12031297 510(k) Number.

---

**Source:** [https://fda-staging.innolitics.com/device/K031297](https://fda-staging.innolitics.com/device/K031297)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
