K031266 · Terumo Corp. · FPA · May 30, 2003 · General Hospital
Device Facts
Record ID
K031266
Device Name
SURSHIELD SAFETY WINGED INFUSION SET
Applicant
Terumo Corp.
Product Code
FPA · General Hospital
Decision Date
May 30, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Surshield Safety Winged Infusion Set is intended for intravenous administration of fluids and/or withdrawal of blood specimens using a syringe, luer adapter, or other compatible/appropriate devices. The winged infusion set may be used for any patient population with consideration given to patient size. Additionally, after withdrawal of the needle from the patient's vein, the shield cover can be manually activated to cover the needle to minimize risk of accidental needlestick.
Device Story
Sterile, single-use winged infusion set; consists of needle, winged hub, tubing, connector, and hinged shield cover. Used for IV fluid administration or blood withdrawal. Manual operation: user inserts needle into vessel; upon withdrawal, user manually pivots shield cover 180 degrees over needle until it latches with audible click. Transparent shield allows visual confirmation of needle containment; design keeps user fingers behind needle point to reduce needlestick injury risk. Used in clinical settings by healthcare professionals. Benefits include safe needle disposal and reduced risk of accidental injury.
Clinical Evidence
Bench testing only. Performance testing included shield cover/needle locking strength, break strength, activation force, puncture resistance, flow rate, priming volume, connection strengths (wing-to-tubing, tubing-to-connector, needle-to-wing), needle penetration resistance, and leak/blockage tests. Biocompatibility testing performed per ISO-10993.
Technological Characteristics
Manual winged infusion set; 19G-25G needles; 90mm or 300mm tubing. Materials are biocompatible per ISO-10993. Sterilization via Ethylene Oxide (EtO) to SAL 10^-6, meeting residual limits for EtO and Ethylene Chlorohydrin. No software or electronic components.
Indications for Use
Indicated for intravenous fluid administration or blood specimen withdrawal in any patient population, with selection based on patient size.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
KO31266
TERUMO MEDICAL PRODUCTS - HANGZHOU CHINA SURSHIELD™ SAFETY WINGED INFUSION SET Section II - Summary and Certification
# SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION PERTAINING TO SUBSTANTIAL EQUIVALENCE
#### DEVICE NAME
#### Proprietary Name
SURSHIELD™ Safety Winged Infusion Set
#### Classification Name
Intravascular Administration Set (80FPA) 21CFR, Section 880.5440 Classification: Class II
#### Common Name
Intravascular Administration Set
#### INTENDED USE
The Surshield Safety Winged Infusion Set is intended for intravenous administration of fluids and/or withdrawal of blood specimens using a syringe, luer adapter, or other compatible/appropriate devices. The winged infusion set may be used for any patient population with consideration given to patient size. Additionally, after withdrawal of the needle from the patient's vein, the shield cover can be manually activated to cover the needle to minimize risk of accidental needlestick.
#### DESCRIPTION
The Surshield Safety Winged Infusion Set is a sterile, single-use device consisting of a needle attached to a winged type hub, tubing, connector and connector cap, and a hinged shield cover that attaches to the wing just below the needle-to-wing junction. The Surshield Safety Winged Infusion Set manufactured by Terumo Medical Products in Hangzhou, China is the same as the Surshield Winged Infusion Set manufactured by Terumo Corporation in Kofu, Japan and cleared under K010103.
{1}------------------------------------------------
The shield cover can be turned 180 degrees on the hinge. As the needle is removed from the patient's vessel, the user's finger actively pushes the shield cover until it latches onto needle using a one- or two- handed technique. An audible click is noted upon activation. The shield cover is designed to allow the user's finger to remain behind the needle point so that the risk of needle stick injury is minimized. The shield cover is transparent for easy confirmation of the needle held in it.
The device possesses both 90 and 300mm length tubing.
### SUBSTANTIAL EQUIVALENCE
The Surshield Safety Winged Infusion Set manufactured by Terumo Medical Products in Hangzhou. China is substantially equivalent to the Surshield Winged Infusion set manufactured by Terumo Corporation in Kofu, Japan and cleared under K010103.
## PRINCIPLE OF OPERATION/TECHNOLOGY
Both devices are operated manually.
## MATERIALS
The materials used in the Surshield Safety Winged Infusion Set manufactured by Terumo Medical Products in Hangzhou, China are the same as the predicate devices, which do not raise any new issues of safety or effectiveness.
{2}------------------------------------------------
## TERUMO MEDICAL PRODUCTS - HANGZHOU CHINA SURSHIELD™ SAFETY WINGED INFUSION SET Section II - Summary and Certification
### SPECIFICATIONS
| Cannula<br>gauge | Color code | Product code | Winged<br>type<br>hub | Cannula<br>length | Tube<br>length |
|------------------|------------|--------------|-----------------------|-------------------|----------------|
| 19G | Cream | SV*S19BL | C type | 3/4"(19mm) | 300mm |
| | | SV*S19DL | D type | 5/8"(16mm) | |
| | | SV*S19BLS | C type | 3/4"(19mm) | 90mm |
| 21G | Green | SV*S21BL | C type | 3/4"(19mm) | 300mm |
| | | SV*S21DL | D type | 5/8"(16mm) | |
| | | SV*S21BLS | C type | 3/4"(19mm) | 90mm |
| 23G | Light blue | SV*S23BL | C type | 3/4"(19mm) | 300mm |
| | | SV*S23DL | D type | 5/8"(16mm) | |
| | | SV*S23BLS | C type | 3/4"(19mm) | 90mm |
| 25G | Orange | SV*S25BL | C type | 3/4"(19mm) | 300mm |
| | | SV*S25DL | D type | 5/8"(16mm) | |
| | | SV*S25BLS | C type | 3/4"(19mm) | 90mm |
## PERFORMANCE
The following tests were performed on the Surshield Safety Winged Infusion Set manufactured by Terumo Medical Products in Hangzhou, China:
- Shield cover/Needle Locking Strength .
- Break strength of the shield cover joint
- Force to lock the needle in the shield cover (Force to activate safety feature) .
- Puncture Resistance of Shield Cover
- Flow Rate
- Priming Volume
- Wing to Tubing Connection Strength
- Tubing to Connector Connection Strength
- Needle to Wing Connection Strength
- . Needle Penetration Resistance
- Wing Needle Protector Fit
- . Leak Test
- . Blockage Test
{3}------------------------------------------------
### TERUMO MEDICAL PRODUCTS - HANGZHOU CHINA SURSHIELD™ SAFETY WINGED INFUSION SET Section II - Summary and Certification
Additionally, a risk analysis was conducted and no new issues were identified since the Surshield China device is identical to the cleared Surshield Kofu. None of the data raises any new issues of safety and effectiveness.
The Surshield Safety Winged Infusion Set manufactured by Terumo Medical Products in Hangzhou. China is substantially equivalent to the Surshield Winged Infusion set manufactured by Terumo Corporation in Kofu, Japan and cleared under K010103.
## ADDITIONAL SAFETY INFORMATION
The sterilization conditions are validated according to EN550 to provide a Sterility Assurance Level (SAL) of 104.
Ethylene Oxide residual levels resulting from EtO sterilization will not exceed the maximum residue levels proposed for Part 821 of Title 21 in the Federal Register Notice issued June 23, 1978, and indicated as follows:
| Ethylene Oxide | 25 ppm |
|-----------------------|--------|
| Ethylene Chlorohydrin | 25 ppm |
The device's blood contacting materials were tested in accordance with the tests recommended in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part-1: Evaluation and Testing. Results of the testing demonstrate that the blood contacting materials are biocompatible.
## CONCLUSION
The Surshield Safety Winged Infusion Set manufactured by Terumo Medical Products in Hangzhou. China is substantially equivalent to the Surshield Winged Infusion set manufactured by Terumo Corporation in Kofu, Japan and cleared under K010103 with respect to intended use, design, technology/principles of operation, materials and performance. Differences between the devices do not raise any new issues of safety or effectiveness.
Date Prepared: 04/17/03
- Prepared by: Barbara Smith Sr. Regulatory Affairs Specialist 125 Blue Ball Rd Elkton, MD 21921 Phone: 410-392-7241
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized symbol featuring three abstract human profiles facing to the right, with flowing lines suggesting movement or connection.
#### Public Health Service
MAY 3 0 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Barbara Smith Senior Regulatory Affairs Specialist Terumo Medical Corporation 125 Blue Ball Road Elkton, Maryland 21921
Re: K031266
Trade/Device Name: SURSHIELD™ Safety Winged Infusion Set Regulation Number: 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: April 18, 2003 Received: April 22, 2003
Dear Ms. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Smith
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Suvor Russer
Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K 03/2 66
#### 510(k) Number (if known):
SURSHIELD™ SAFETY WINGED INFUSION SET Device Name:
#### Indications For Use:
The Surshield Safety Winged Infusion Set is intended for intravenous administration of fluids and/or withdrawal of blood specimens using a syringe, luer adapter, or other compatible/appropriate devices. The winged infusion set may be used for any patient population with consideration given to patient size. Additionally, after withdrawal of the needle from the patient's vein, the shield cover can be manually activated to cover the needle to minimize risk of accidental needlestick.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use
Pattuce Cucenit
(Optional Format 1-2-96)
n of Anesthesiology. General Hospital
510(k) Number: K031266
003
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.