MONITOR, MODEL HSM-00200; PROBE, MODEL HSP-05065; JACKET, MODEL HSJ-05065
K031153 · Arrow Intl., Inc. · DPW · Nov 25, 2003 · Cardiovascular
Device Facts
Record ID
K031153
Device Name
MONITOR, MODEL HSM-00200; PROBE, MODEL HSP-05065; JACKET, MODEL HSJ-05065
Applicant
Arrow Intl., Inc.
Product Code
DPW · Cardiovascular
Decision Date
Nov 25, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2100
Device Class
Class 2
Indications for Use
The HemoSonic™ 200 is intended for use in patients with surgical, medical, or anesthetic risk to provide a non-invasive, continuous, real-time composite hemodynamic profile that indicates and monitors cardiovascular status by assessing contractility, flow, and resistance parameters.
Device Story
HemoSonic™ 200 is a non-invasive hemodynamic monitor for anesthesia and intensive care settings. System components include a monitor, transesophageal probe assembly, and probe holder. The device uses transesophageal ultrasound to measure real-time descending aortic blood flow. It employs a combination of M-mode and Pulsed Wave Doppler (PWD) to simultaneously determine aortic diameter and blood flow velocity; these modes operate together via synchronized pulses and cannot be used independently. The monitor processes these inputs to calculate and display a composite hemodynamic profile, including contractility, flow, and resistance parameters. Clinicians use this real-time data to assess cardiovascular status and guide clinical decision-making for patients at risk. The device provides continuous monitoring, potentially benefiting patients by allowing for timely hemodynamic interventions.
Clinical Evidence
Bench testing only. Performance demonstrated via comparative round trip insertion loss tests, comparative pulsatile flow loop tests, and compliance with electrical safety and electromagnetic compatibility standards (EN 60601-1, EN 60601-1-2, and IEC 60601-2-37).
Technological Characteristics
Transesophageal ultrasound system; utilizes two transducers (one for M-mode, one for PWD) for simultaneous aortic diameter and flow measurement. Operates via pulse synchronization. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), and IEC 60601-2-37 (ultrasonic diagnostic equipment).
Indications for Use
Indicated for patients with surgical, medical, or anesthetic risk requiring non-invasive, continuous, real-time hemodynamic monitoring of cardiovascular status, including contractility, flow, and resistance parameters via transesophageal ultrasound.
Regulatory Classification
Identification
A cardiovascular blood flowmeter is a device that is connected to a flow transducer that energizes the transducer and processes and displays the blood flow signal.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows a sequence of alphanumeric characters, specifically "KO31153". The characters are written in a bold, sans-serif font, and they appear to be handwritten or stylized. The numbers and letters are evenly spaced and clearly legible against the white background.
# HemoSonic™ 200 Hemodynamic Monitor 510(k) Summary
| Submitter: | Arrow International Inc.<br>2400 Bernville Road<br>Reading, PA 19605 |
|---------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person: | William G. McLain<br>Manager, Regulatory Affairs<br>Phone: 610-378-0131<br>Fax: 610-478-3188<br>E-mail: bill.mclain@arrowintl.com |
| Date summary prepared: | 4/10/03 |
| Device trade name: | HemoSonic™ 200 Hemodynamic Monitor |
| Device common name: | Cardiac output monitor. |
| Device classification name: | FLOWMETER, BLOOD, CARDIOVASCULAR, DPW, at 21 CFR 870.2100 |
| Legally marketed devices to which the device is substantially equivalent: | Sometec - Dynemo 3000 Non-Invasive Hemodynamic Diagnostic Device (K972798) |
| Description of device: | The HemoSonic™ 200 is a monitoring device that non-invasively measures real-time descending aortic blood flow and other hemodynamic parameters during anesthesia and intensive care. The HemoSonic™ 200 system is composed of a monitor, a transesophageal probe assembly and probe holder |
| Intended use of the device: | The HemoSonic™ 200 is intended for use in patients with surgical, medical, or anesthetic risk to provide a non-invasive, continuous, real-time composite hemodynamic profile that indicates and monitors cardiovascular status by assessing contractility, flow, and resistance parameters. |
| Technological characteristics: | The technological characteristics are unchanged from the predicate device. |
| Performance tests: | The following tests were performed to demonstrate substantial equivalence:<br>☐ Comparative round trip insertion loss test,<br>☐ Comparative pulsitile flow loop test,<br>☐ EN 60601-1: Medical electrical equipment – Part 1:<br>General requirements for safety – IEC 601-1,<br>☐ EN 60601-1-2: Medical electrical equipment – Part 1:<br>General requirements for safety. Collateral standard:<br>Electromagnetic compatibility requirements and tests.<br>IEC 601-1-2, and<br>☐ IEC 60601-2-37: Medical electrical equipment – Part 2-<br>37: Particular requirements for the safety of ultrasonic<br>medical diagnostic and monitoring equipment (Clause<br>42.3 tested under condition 2 only.) |
| Conclusions: | The results of the performance tests demonstrate that the<br>device is substantially equivalent to the legally marketed<br>predicate device. |
{1}------------------------------------------------
### HemoSonic™ 200 Hemodynamic Monitor 510(k) Summary
のお気になる。
and the comments of the comments of the comments of
{2}------------------------------------------------
Public Health Service
Image /page/2/Picture/2 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular in shape, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is a stylized image of an eagle with its wings spread.
NOV 2 5 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Arrow International, Inc. c/o Mr. William G. McLain Manager, Regulatory Affairs 2400 Bernville Road Reading, PA 19612
Re: K031153
HemoSonic™ 200 Hemodynamic Monitor Regulation Number: 21CFR 870.2100 and 876.1500 Regulation Name: Cardiovascular Blood Flowmeter and Endoscope Sheath Regulatory Class: Class II (two) Product Code: DPW and FED Dated: September 26, 2003 Received: September 29, 2003
Dear Mr. McLain:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the HemoSonic™ 200 Hemodynamic Monitor, as described in your premarket notification:
Model HSP 05065
{3}------------------------------------------------
### Page 2 – Mr. William G. McLain
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G. (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health · Document Mail Center (HFZz401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801, please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
{4}------------------------------------------------
Page 3 – Mr. William G. McLain
If you have any questions regarding the content of this letter, please contact Kachi Enyinna at (301) 443-8262.
Sincerely yours,
cMepax Myne
Bram D. Zuckerman, M.D. Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(1)
{5}------------------------------------------------
#### Diagnostic Ultrasound Indications for Use Form
## Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | Combined (specify) | Other (specify) |
|-------------------------------|-------------------|---|---|-----|-----|---------------|-------------------|------------------------|--------------------|-----------------|
| | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | | |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | P | P | | | | | P | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: Combined Use - The system uses a combination of M-mode and PWD to determine both the diameter of and the flow within the aorta. The transesophageal probe contains two transducers, one dedicated to each function, and the two modes operate simultaneously (with appropriate pulse synchronization), and cannot be used independently within the normal operation of the equipment.
> (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
CMeleia Maye
(Division Sign-Off)
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number Ko31153
Prescription Use (Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.