← Product Code [JOJ](/productcode/JOJ) · K031152

# PROPPER VAPOR LINE PCD (STEAM STERILIZATION PROCESS CHALLENGE TEST) (K031152)

_Propper Mfg. Co., Inc. · JOJ · Aug 17, 2004 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K031152

## Device Facts

- **Applicant:** Propper Mfg. Co., Inc.
- **Product Code:** [JOJ](/productcode/JOJ.md)
- **Decision Date:** Aug 17, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2800
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The Propper Vapor Line PCD is indicated for use in routine challenge testing steam sterilizers in pre-vacuum cycles operating at 273.2°F/134°C.

## Device Story

Propper Vapor Line PCD is a process challenge test device for pre-vacuum steam sterilization at 273.2°F/134°C. Device consists of a metal capsule with a removable cap, sample holder, and a lumen. During sterilization, the lumen allows residual air to exit and steam to enter the capsule. User places a Propper Vapor Line Integrator into the capsule holder, secures the cap, and places the device in the sterilizer chamber. After the cycle, the integrator is retrieved and examined for a 'PASS' or 'FAIL' reading. The device challenges steam penetration via the lumen's length, bore, and configuration. It provides healthcare providers with a reproducible method to verify sterilization conditions, ensuring effective sterilization and patient safety.

## Clinical Evidence

Bench testing only. Performance compared to Propper Bio-Challenge Test-Pak using Duo-Spore and BI-OK biological indicators in a 4-pulse pre-vacuum steam sterilizer at 273.2°F/134°C. Testing evaluated 'PASS/FAIL' readings against spore death at various exposure times (0.25 to 2.00 minutes). Results showed the device is a reproducible challenge that does not indicate 'PASS' until after complete spore death. Multiple device equivalency study confirmed consistent performance across five units.

## Technological Characteristics

Metal capsule with removable cap and sample holder; lumen-based steam penetration challenge (controlled by length, bore, and configuration). Standalone device; no energy source or software. Designed for pre-vacuum steam sterilization at 273.2°F/134°C.

## Regulatory Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

## Predicate Devices

- Propper Bio-Challenge Test-Pak

## Reference Devices

- Duo-Spore indicator
- BI-OK indicator

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

AUG 1 7 2004

## 510(k) SUMMARY

## Propper Vapor Line PCD Steam Sterilization Process Challenge Test 510(k) Number: K031152

| Submitter Information:             | Propper Manufacturing Co., Inc.<br>36-04 Skillman Avenue<br>Long Island City, NY 11101<br>Attn: Frank E. Platko, Ph.D.<br>Telephone: (718) 392-6650 Ext. 264                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name:                       | (Trade:) Propper Vapor Line PCD Steam Sterilization Process Challenge Test<br>(Common:) Steam Chemical Indicator<br>(Classification:) Physical/Chemical Sterilization Process Indicator                                                                                                                                                                                                                                                                                                                                                                                                      |
| Substantial Equivalence<br>Device: | Propper Bio-Challenge Test-Pak                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Device Description:                | Propper Vapor Line PCD is a process challenge test device for use in pre-<br>vacuum steam sterilization processes at 273.2°F/134°C. It consists of a<br>capsule with a removable cap and sample holder to permit the positioning<br>and removal of the Propper Vapor Line Integrator. One end of the capsule<br>is connected to a lumen. During the sterilization cycle, the open end of the<br>lumen permits residual air to exit and steam to enter the capsule The<br>integrator display will indicate whether or not sterilization conditions were<br>attained inside the metal capsule. |
| Intended Use<br>Of Device:         | The Propper Vapor Line PCD is indicated for use in routine challenge<br>testing steam sterilizers in pre-vacuum cycles operating at 273.2°F/134°C.                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Technological<br>Characteristics:  | The Propper Vapor Line PCD and Propper Bio-Challenge Test-Pak are<br>both Steam Process Challenge test devices designed for use in steam<br>sterilization processes. They are similar in function but differ in design<br>and construction.                                                                                                                                                                                                                                                                                                                                                  |
|                                    | The challenge to steam penetration presented by the Propper Vapor Line<br>PCD is controlled by the length, bore and configuration of the lumen<br>which is the only opening into the capsule. Whereas the corresponding                                                                                                                                                                                                                                                                                                                                                                      |

{1}------------------------------------------------

510(k) Summary---continued

:

|                       | challenge by the Propper Bio-Challenge Test-Pak is determined by a             steam-resistant paper load and a rate controlling barrier film.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Nonclinical<br>Tests: | The safety and effectiveness of the Propper Vapor Line PCD Steam             Sterilization Process Challenge Test were determined by comparing the             performance characteristics of this device directly to that of the Propper             Bio-Challenge Test-Pak in side-by-side testing in a steam sterilizer.<br>Before each test cycle, a Propper Vapor Line Integrator was placed into             the slot of the PCD sample holder which was then inserted into the body             of the metal capsule and secured in position with the screw cap. The PCD             device was placed inside the sterilizer chamber on the lower rack over the             drain, side-by-side with two Propper Bio-Challenge Test-Paks. One pack             contained a Duo-Spore indicator and the other a BI-OK indicator. Then,             a conventional 4-pulse pre-vacuum steam sterilization cycle was run.             Cycle tests were run at 273.2°F/134°C for different exposure times.<br>After each sterilization cycle, the Vapor Line integrator was retrieved and             examined to determine the “PASS” or “FAIL” reading. The Duo-Spore             strip was aseptically removed from its glassine envelope, transferred to             trypticase soy broth and incubated at 56°C. The spore strip was examined             for growth on a daily basis for seven days. The BI-OK vial was activated,             incubated at 56°C for 48 hours and then examined for growth. A second             sterilizer study was conducted to establish multiple device equivalency.             Complete test results appear in tabular form at the end of this summary |
| Test<br>Conclusions:  | The Propper Vapor Line PCD is a highly reproducible challenge test. It             will not indicate acceptable “PASS” results until after complete spore             death has occurred. Based on the performance data, the test is safe and             effective for its intended use. It is substantially equivalent in performance             to commercial biological challenge tests used in pre-vacuum steam             sterilizer cycle testing.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |

{2}------------------------------------------------

#### VAPOR LINE PCD vs BIO-CHALLENGE TEST-PAK Exposure Time vs % Fail/Survival

| 273.2F(134C)                    |      |      |      |      |      |      |
|---------------------------------|------|------|------|------|------|------|
| Exposure Time (min.)            | 0.25 | 0.60 | 1.00 | 1.50 | 1.75 | 2.00 |
| % Survived (Duo-Spore)*         | 0    | 0    | 0    | 0    | 0    | 0    |
| % Survived (BI-O.K.)            | 80   | 40   | 0    | 0    | 0    | 0    |
| % Fail (Propper Vapor Line PCD) | 100  | 100  | 100  | 100  | 15   | 0    |
| Total No. of Tests              | 5    | 20   | 20   | 5    | 20   | 20   |

* Complete spore death occurred prior to 0.25 minutes exposure

{3}------------------------------------------------

# PROPPER VAPOR LINE PCD MULTIPLE DEVICE EQUIVALENCY STUDY

### Temperature @ 273.2F/134C

| Exposure Time (min.)    | 1.50       | 1.75       | 2.00       |
|-------------------------|------------|------------|------------|
|                         | Fail/Total | Fail/Total | Fail/Total |
| PCD 1*                  | 1/1        | 3/6        | 0/6        |
| PCD 2                   | 1/1        | 4/6        | 0/6        |
| PCD 3                   | 1/1        | 3/6        | 0/6        |
| PCD 4                   | 1/1        | 2/6        | 0/6        |
| PCD 5                   | 1/1        | 2/6        | 0/6        |
| % FAIL (Vapor Line PCD) | 100        | 47         | 0          |
| Total Number of Tests   | 5          | 30         | 30         |

* PCD # I also used in comparative testing vs Bio-Challenge Test Pak

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 1 7 2004

Dr. John D. Dyckman Vice President, Research and Development Propper Manufacturing Company, Incorporated 36-04 Skillman Avenue Long Island City, New York 11101

Re: K031152

Trade/Device Name: Propper Vapor Line PCD Steam Sterilization Process Challenge Test Regulation Number: 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: JOJ Dated: May 28, 2004 Received: June 2, 2004

Dear Dr. Dyckman:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your end have determined the device is substantially equivalent (for the recroniced above and in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device mendments, or to devices that have been reclassified in accordance with the provisions of Amendinons, or to arretic Act (Act) that do not require approval of a premarket the Federal F 60a; Drag). You may, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include controls providents of a registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 Nr.), it may of such to back of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.

{5}------------------------------------------------

Page 2 - Dr. Dyckman

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice and ifements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Ques

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

K031152 510(k) Number (if known):_

Device Name: Propper Vapor Line PCD Steam Sterilization Process Challenge Test

Indications For Use:

The PROPPER VAPOR LINE PCD is indicated for use as a routine process challenge test for steam sterilizers in pre-vacuum cycles at 273.2°F/134°C.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Kai Murray

(Division Sign-Off) (Division of Anesthesiology, General Hospital, Infection Control, Dental

510(k) Number: K031152

Prescription Use (Pcr 21 CFR 801.109) OR

Over-The-Counter Use X

(Optional Formal 1-2-96)

---

**Source:** [https://fda-staging.innolitics.com/device/K031152](https://fda-staging.innolitics.com/device/K031152)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
