K031132 · Eastman Kodak Company · LLZ · Nov 26, 2003 · Radiology
Device Facts
Record ID
K031132
Device Name
CONSULTIVA CASE INPUT STATION, MODEL CIS-1
Applicant
Eastman Kodak Company
Product Code
LLZ · Radiology
Decision Date
Nov 26, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Consultiva™ Case Input Station is a software application that facilitates the digitization of medical images from a commercial digitizer and archives the digital images. The Consultiva™ Case Input Station is intended for use by a technician under the supervision of a medical professional.
Device Story
Consultiva™ Case Input Station (CIS) is a Windows-based software application; functions as a standalone system or as an initial component in a larger medical imaging workflow. Input: medical images from a connected commercial digitizer. Operation: user initiates scans via a graphical user interface (UI); software interfaces with digitizer via high-level programming interfaces to manage scan initiation and status messaging. Output: digitized medical images saved to local hard disk; status information (e.g., Pass/Fail) displayed to user. Quality control: includes procedures to verify digitizer performance using test films and automated quality checks on each image to ensure sufficient quality. Used in clinical settings by technicians; output used by medical professionals for diagnostic imaging workflows.
Clinical Evidence
Bench testing only. Verification procedures with pass/fail criteria were performed to ensure the software met all specified requirements.
Technological Characteristics
Windows-based software application; interfaces with external commercial digitizers via high-level programming interfaces. Features include image digitization, archiving, and quality control procedures. Connectivity: standalone or networked for downstream integration. Software-based image quality checks and digitizer performance verification.
Indications for Use
Indicated for use by technicians under medical professional supervision to facilitate digitization and archiving of medical images from commercial digitizers.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for MiraMedica. The logo consists of a stylized graphic above the company name. The graphic appears to be a curved shape with vertical lines, possibly representing hair or a flame. The company name, "MiraMedica," is written in a simple, sans-serif font below the graphic.
Image /page/0/Picture/2 description: The image shows handwritten text that appears to be a document identifier and page number. The text "K031132" is written on the top line. Below that, the text "page 1 of 2" indicates that this is page 1 of a 2-page document.
Page E-1
## 510(k) Summary
| Submitter's Name: | MiraMedica, Inc. | |
|----------------------------|----------------------------------------------------------------|---------------|
| Submitter's Address: | 15466 Los Gatos Blvd., Suite 109-171, Los Gatos, CA, 95032 | |
| Submitter's Telephone: | (408) 858-0718 | |
| Contact Name: | Wido Menhardt | |
| Date Summary was Prepared: | April 04, 2003 | |
| Trade or Proprietary Name: | Consultiva™ Case Input Station | |
| Common or Usual Name: | System, Image Processing, Radiological | |
| Classification Name: | Picture Archiving and Communications System (21 CFR 892. 2050) | |
| Predicate Devices: | Device Name | 510(k) Number |
| | ExpressSuite | K011049 |
| | ImagClear | K992467 |
| | IMPAX Workstations | K022292 |
#### Description of the Device and Summary of the Technological Characteristics:
The Consultiva™ Case Input Station (CIS) is a Windows-based program that facilitates the digitization of medical images and saves these digital images to a specified location on hard disk. Although the CIS is a stand-alone system, it requires a digitizer to get data. The CIS interfaces with the digitizer through high level programming interfaces that allow the CIS to initiate scans and receive messages from the digitizer. The CIS provides a User Interface (UI) that allows the user to initiate scans and verify the completion of the scans (visually) as well as view error information.
The CIS can operate as a stand-alone system to collect medical images. It can also be the first step in a larger medical system. It interfaces with downstream products by sending and receiving messages. If configured in this mode, the CIS can display status information (e.g. Pass of Fail) to the user through the UI.
The CIS also provides some Quality Control (OC) procedures to verify digitizer performance. These procedures are initiated from the CIS UI and require specific test films to be digitized and processed. In the regular workflow, the CIS also implements some quality checks on each image to ensure that the image is of sufficient quality.
{1}------------------------------------------------
Page E-2
Image /page/1/Picture/1 description: The image shows the text "K031132 page 2 of 2" in a handwritten font. Below this text is a stylized graphic and the text "Arra Medica" in a smaller font. The text and graphic elements are arranged in a visually distinct manner.
#### Indications for Use:
The Consultiva™ Case Input Station is a software application that facilitates the digitization of medical images from a commercial digitizer and archives the digital images. The Consultiva™ Case Input Station is intended for use by a technician under the supervision of a medical professional.
#### Substantial Equivalence:
The Consultiva™ Case Input Station is similar to predicate devices such as the VIDAR ExpressSuite, Agfa IMPAX Workstation, and the Titan (formerly DBA) ImagClear. The intended use for the CIS and the predicate devices is for the digital acquisition and archiving of medical images. All of the devices operate on commercially available computer systems such as a Windows-based system and all interface with commercially available digitizers.
#### Testing:
Various tests of the software to verify system specifications are being performed. Verification procedures with pass/fail criteria were developed to ensure that the product met all the specified requirements.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three lines forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
### NOV 26 2003
Wido Menhardt, Ph.D. General Manager, CAD Business Segment Eastman Kodak Company 100 Centry Center Court, Suite 600 SAN JOSE CA 95112
Re: K031132 Trade/Device Name: Consultiva™ Case Input Station Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communication system Regulatory Class: II Product Code: 90 LLZ Dated: October 27, 2003 Received: October 30, 2003
Dear Dr. Menhardt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Page 2
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo for a company called "MiraMedica". The logo consists of a stylized, abstract symbol above the company name. The symbol appears to be a curved shape, possibly representing a human form or a stylized letter. The text "MiraMedica" is written in a serif font, with the first letter of each word capitalized.
# D Indications for Use Statement
| Ver/ 3 - 4/24/96 | |
|---------------------------------------------|---------|
| Applicant: MiraMedica, Inc. | |
| 510(k) Number (if known): | K031132 |
| Device Name: Consultiva™ Case Input Station | |
| Indications For Use: | |
The Consultiva™ Case Input Station is a software application that facilitates the digitization of
medical images from a commercial digitizer and archives the digital images. The Consultiva™
Case Input Station is intended for use by a technician under the supervision of a medical
professional.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF
NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
| 510(k) Number | K031132 |
|---------------|---------|
|---------------|---------|
| Prescription Use per 21 CFR 801.109 | Over the Counter Use |
|-------------------------------------------------|----------------------|
|-------------------------------------------------|----------------------|
Consultiva™ Case Input Station 510(k) Application - CONFIDENTIAL
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.