SYRINGEX SAFETY SYRINGE, MODEL 0322G1.5

K031065 · Syringex Medical, Inc. · MEG · Jun 23, 2003 · General Hospital

Device Facts

Record IDK031065
Device NameSYRINGEX SAFETY SYRINGE, MODEL 0322G1.5
ApplicantSyringex Medical, Inc.
Product CodeMEG · General Hospital
Decision DateJun 23, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5860
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Syringex Safety Syringe® is a sterile, non-toxic, non-pyrogenic, retractable needle, single use, disposable syringe for injecting fluids into or withdrawing fluids from the body while helping to provide protection from accidental needlestick injuries. The Syringex Safety Syringe® functions as a conventional hypodermic syringe except for its added ability to retract and contain the contaminated needle safely inside the syringe barrel immediately after the completion of the patient injection.

Device Story

Syringex Safety Syringe is a single-use, disposable, retractable needle syringe. Device functions as a standard piston syringe for fluid injection or withdrawal. Upon completion of injection, user retracts plunger to pull needle into barrel; mechanism secures needle inside barrel. Plunger stem is broken off to prevent needle re-exposure. Designed for one-handed operation to minimize tissue trauma and needlestick risk. Used by physicians, nurses, and veterinarians in clinical settings. Provides immediate, mandatory needle containment post-injection.

Clinical Evidence

Simulated clinical study conducted at 3 hospitals and 1 veterinary center. Over 60 investigators (doctors, nurses, PAs, veterinarians) tested >600 syringes using Injecta-Pads. 100% of investigators reported positive feedback and preference over the predicate. No injuries reported. Bench testing performed to ISO 7886-1 standards.

Technological Characteristics

Materials: plastic, rubber, stainless steel, adhesive, lubricant. Biocompatibility per ISO 10993. Piston syringe design with integrated retractable needle mechanism. Needle gauges: 18-30G; lengths: 5/16" to 1 1/2". Engagement pressure: ~600g. Sterilization: validated process. Non-pyrogenic (LAL test).

Indications for Use

Indicated for use by clinicians for injecting fluids into or withdrawing fluids from the body, while providing a retractable needle mechanism to reduce the risk of accidental needlestick injuries.

Regulatory Classification

Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K031065 # JUN 2 3 2003 ### Syringex Safety Syringe® # 510(k) Summary 21 CFR 807.92 #### Date of Summary Preparation: 3-29-2003 Submitter: Syringex Medical, Inc. 18 Beatrice Lane Glen Cove, NY 11542 516-676-4008 fax: 516-671-5935 Contact Person: John Jones 2311 Bear Hills Ct. Draper, UT 84020 Tel: 801-553-2150 Fax: 801-553-1545 #### Device Name Proprietary Device (Trade) Name: Common Device Name: Device Classification Name: 510(k) Number Syringex Safety Syringe® Piston Syringe Syringe, Antistick K031062 #### Legally marketed device to which we claim equivalence: SafeSnap Retractable Needle Syringe U.S. Medical Instruments, Inc. San Diego, CA 92127 510(k) Number K925039, K910520 ### Description of device The Syringex Safety Syringe® is a retractable needle safety syringe. The syringe design employs a retractable mechanism that becomes engaged upon injection of the syringes contents. Withdrawal of the plunger retracts the needle into the syringe barrel, holding it securely in place. The plunger stem is broken off to prevent pushing the needle assembly forward. The syringe may be made in all sizes with needle gauges ranging from 18 to 30. The length of the needle is limited only by the length of the barrel in order to contain the needle. Available in lengths from 5/16" to 1 ½". {1}------------------------------------------------ #### Intended Use The Syringex Safety Syringe® is a sterile, non-toxic, non-pyrogenic, retractable needle, single use, disposable syringe for injecting fluids into or withdrawing fluids from the body while helping to provide protection from accidental needlestick injuries. The Syringex Safety Syringe® functions as a conventional hypodermic syringe except for its added ability to retract and contain the contaminated needle safely inside the syringe barrel immediately after the completion of the patient injection. Summary of technological characteristics of device compared to the predicate device. The Syringex Safety Syringe® is substantially equivalent to the previously marketed U.S. Medical SafeSnap syringe. The SafeSnap allows Luer Lock needles to be interchanged. The Syringex Safety Syringe® is currently provided with a safety needle assembly permanently in place. Both syringes are basically standard piston syringes with mechanisms that lock the needle assemblies to the end of the plungers when the plungers are fully depressed. The SafeSnap requires a hinged lock at the end of the barrel to be released after the plunger locks onto the needle assembly. With both syringes, retracting the plunger pulls the needle assembly into the barrel and the plunger stems are broken off. The SafeSnap requires the broken plunger to be pushed into the other end of the syringe barrel to prevent the needle assembly from moving forward, and risk of exposing the needle sharp. The Syringex Safety Syringe® needle assembly is secured inside the barrel with a mechanism that snaps the needle assembly in place as the plunger is withdrawn. Alternately, the Syringex needle assembly may be configured so that the needle is canted to the side, additionally preventing it from sliding forward. With the Syringex Safety Syringe® the needle can be withdrawn directly from the patient into the barrel, and the operation can be accomplished with a one handed technique. With the SafeSnap a hinged lock must be released with a second hand before the needle assembly can be withdrawn, potentially causing tissue trauma if an attempt is made to withdraw the needle directly from the injection site .. Both syringes employ plastic, rubber, stainless steel, adhesive, and a lubricant. All of the materials used by Syringex have passed Tripartite or ISO 10993 specifications. Biocompatibility and pyrogenicity testing and sterilization will be performed on complete Syringex Safety Syringes® after the assembly operation has been completed and before the syringes are released for use. We plan to use the limulus amebocyte lysate (LAL) Test method to determine that each lot is non-pyrogenic. The "predicate" SafeSnap syringe is provided with needle gages from 20 to 29G and in lengths from ½ " to 1 ½ ". Syringex's Safety Syringe can be used with needles from 18 to 30G and in lengths from 5/16" to 1 ½ ". {2}------------------------------------------------ Specific information about the SafeSnap other than what is contained in its shipping carton and on the individual pouches was unavailable, and U.S. Medical refused to supply any information in accordance with FDA requirements and FOI. ### Discussion of Non-clinical (Bench) Tests to support equivalence Both have transparent barrels. The Syringex barrel is 3.28" Long, .395" o.d. and .340" i.d. with a 3cc scale 2" long, marked at .1cc and numbers at ½ cc markings. The SafeSnap barrel is 3.41" long, .467" o.d. and .38" i.d. with a 3cc scale 1.6" long marked at .1 cc and numbers at ½ cc markings. Syringex has a maximum capacity of just over nominal +10%, the minimum required under ISO 7886-1. SafeSnap has about 53% more capacity than nominal, more than 5 times that required by ISO 7886-1. ### Conclusions drawn from Non-clinical (Bench) tests Bench testing to ISO 7886-1 has been conducted to examine compliance of both syringes. In all cases the Syringex Safety Syringe® was either equivalent to or exceeded the performance of the SafeSnap Syringe. In addition, design verification testing was conducted on the Syringex Safety Syringe® to confirm that the device met all functional specifications and applicable standards. Although SafeSnap's barrel is .13" longer, Syringex has a longer scale and the marks are easier to read than Safesnap. Syringex has a nominal engagement pressure just over 600 grams, while SafeSnap's engagement pressure is over 4200 grams. Thus SafeSnap is much more likely to cause tissue trauma when the safety feature is engaged. #### Summary of Simulated Clinical Tests In a simulated use study conducted at 3 hospitals and a veterinary center by a variety of doctors, nurses, physician's assistants and veterinarians, more than 60 investigators tested over 600 Syringex Safety Syringes® to evaluate the safety and effectiveness of the syringe by injecting fluids into an Injecta-Pad. All injections were successfully completed and 100% of the investigators had a positive response to the syringe evaluation and preferred it over the predicate. No injuries of any kind were reported. The complete Statistical Analysis, Protocol and Questionnaires are attached as Exhibit B. ### Conclusions and statistical summary of Simulated Clinical Tests In all cases during the simulated clinical tests performed by healthcare professionals the Syringex Safety Syringe® was determined to be either equivalent to or exceeded the performance of the SafeSnap syringe. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like figure with three curved lines forming its body and wings. The bird is facing to the right. Encircling the bird is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement. Public Health Service JUN 2 3 2003 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. John O. Jones VP Operations, R & D Svringex Medical, Incorporated 2311 Bear Hills Court Draper, Utah 84020 Re: K031065 Trade/Device Name: Syringex Safety Syringe® Regulation Number: 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: MEG, FMF Dated: June 15, 2003 Received: April 15, 2003 Dear Mr. Jones: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Jones Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Susan Punoe Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ K031065 # JUN 2 3 2003 ## Indications for Use Page 1 of 1 510(k) Number (if known): K031065 Device Name: Syringex Safety Syringe® Indications for Use: The primary intended use of the Syringex Safety Syringe® is "a hypodermic needle and syringe intended to inject fluid into or withdraw fluid from the body". The secondary intended use is that the sharps safety feature will help to reduce the number of sharps injuries by providing an immediate and mandatory elimination of the sharp by its safe, simple and easy retraction into the barrel of the syringe. The feature will help prevent sharps injuries when using the device for its primary intended use. ### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurence of CDRH, Office of Device Evaulation (ODE) Patino Cusenite 510(k) Number: K031065
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...