IVD CRYPTOSPORIDIUM ANTIGEN DETECTION ASSAY, MODEL CP-96
Applicant
Ivd Research, Inc.
Product Code
MHJ · Microbiology
Decision Date
Jul 10, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3220
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
This microwell enzyme linked immunoabsorbant assay (ELISA) detection kit is an in vitro diagnostic (IVD) immunoassay for the detection of Cryptosporidium antigen in human feces using peroxidase as the enzyme. The assay may be read visually or with an ELISA reader. This IVD assay is intended to be used with stools that are fresh, frozen or preserved in 10% formalin, SAF or Medical Chemical Corporation's (MCC) Universal fixative. Concentrated samples cannot be used with this IVD.
Device Story
Microwell ELISA kit for qualitative detection of Cryptosporidium antigen in human stool. Input: fecal samples (fresh, frozen, or fixed). Principle: rabbit anti-Cryptosporidium antibody captures antigen; goat anti-Cryptosporidium antibody sandwiches captured antigen; peroxidase-conjugated anti-second antibody and TMB chromogen produce blue color indicating presence of antigen. Output: visual color change or optical density reading via ELISA reader. Used in clinical laboratory settings by trained personnel. Results aid healthcare providers in diagnosing cryptosporidiosis. Benefits: provides rapid, sensitive detection of Cryptosporidium species compared to traditional O&P microscopy.
Clinical Evidence
Bench testing and clinical evaluation performed. Study 1 (n=174): 100% sensitivity (24/24), 97% specificity (146/150). Study 2 (n=44): 100% positive agreement (14/14), 100% negative agreement (30/30). Analytical sensitivity: ~50 ng/mL soluble protein or 10 oocysts/well. No cross-reactivity observed with 30+ common enteric parasites, bacteria, or viruses.
Technological Characteristics
Microwell ELISA format. Uses rabbit anti-Cryptosporidium capture antibody and goat anti-Cryptosporidium detection antibody. Peroxidase/TMB chromogen indicator system. Compatible with ELISA reader or visual interpretation. Analytical sensitivity: 10 oocysts/well.
Indications for Use
Indicated for the qualitative detection of Cryptosporidium species antigen in human fecal specimens (fresh, frozen, or preserved in 10% formalin, SAF, or MCC Universal fixative) to aid in the diagnosis of cryptosporidiosis. Not for use with concentrated fecal samples.
Regulatory Classification
Identification
Entamoeba histolytica serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Entamoeba histolytica in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identify Entamoeba histolytica directly from clinical specimens. The identification aids in the diagnosis of amebiasis caused by the microscopic protozoan parasite Entamoeba histolytica and provides epidemiological information on diseases caused by this parasite. The parasite may invade the skin, liver, intestines, lungs, and diaphragm, causing disease conditions such as indolent ulcers, an amebic hepatitis, amebic dysentery, and pulmonary lesions.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.
{0}------------------------------------------------
# JUL 1 0 2003
# 510(k) Summary of Safety and Effectiveness for IVD Research Inc.'s Cryptosporidium Human Fecal Antigen Detection Microwell ELISA Kit
| Submitter: | IVD Research Inc.<br>5909 Sea Lion Place, Suite D<br>Carlsbad, CA 92008 USA<br>Tel: 760-929-7744 Contact: Dave Lambillotte, President |
|------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Prepared: | 12 May 2003 |
Common name: Cryptosporidium Human Fecal Antigen Detection Microwell ELISA Kit
Proprietary name: IVD Research's Cryptosporidium Human Fecal Antigen Detection Microwell ELISA Kit
Classification name: Entamoeba histolytica serological reagents
Predicate Device: LMD Laboratories' Cryptosporidium Antigen Detection ELISA, 510(k) K955345
# DESCRIPTION AND PRINCIPLE:
This microwell enzyme-linked immunoabsorbant assay (ELISA) detection kit (Cryptosporidium ELISA Kit) is an in vitro diagnostic (IVD) immunoassay for the detection of antigen to Cryptosporidium species in human feces using peroxidase as the indicator enzyme. The assay may be read visually or with an ELISA reader. Concentrated fecal samples cannot be used with this immunoassay. Rather, this IVD Cryptosporidium ELISA Kit is intended to be used with human stools that are fresh, frozen or preserved in 10% formalin, SAF or Medical Chemical Corporation's (MCC's) Universal fixative in a clinical laboratory use setting.
## This ELISA corresponds to FDA Classification Name: Entamoeba Histolytica Serological Reagent, a class II (non-exempt) Device, within the Microbiology Classification Panel, having FDA Reg. Citation Number: 21 CFR 866.3220, and FDA Product Codes: MHJ.
This ELISA is an in vitro immunoassay for the qualitative determination of Cryptosporidium species antigen in feces. The ELISA uses a rabbit anti-Cryptosporidium antibody to capture the antigen from the stool supernatant. A second goat anti-Cryptosporidium antibody is then added which sandwiches the captured antigen. This reaction is visualized by the addition of an anti-second antibody conjugated to peroxidase and the chromogen tetramethylbenzidine (TMB). The resulting blue color development indicates the presence of Cryptosporidium species antigens being bound by the anti-Cryptosporidium antibodies.
## SUBSTANTIAL EQUIVALENCE COMPARISON
A similar ELISA kit was developed and manufactured by LMD Laboratories (San Diego, CA). The FDA cleared this assay under 510(k) K955345 in 7 June 1996. This current kit uses the same methodology (ELISA) and similar reagents as that cleared predicate assay.
Following is the data that shows this firm's ELISA provides substantial equivalence to the gold standard for parasitology (O&P microscopy) as well as the predicate device. The 95% Confidence Interval (CI) was determined using a Confidence Interval Analysis Microcomputer Program published by the British Medical Journal.
All specimens are human samples supplied by clinical laboratories or in-house collections.
{1}------------------------------------------------
### Study #1 (Outside Lab)
A total of 174 formalin, SAF, fresh/frozen and Universal Fixative preserved stools were tested against the ELISA test The results were interpreted visually.
The following results were obtained.
| | Micro + (O&P) | Micro - (O&P) |
|----------------------|---------------|---------------|
| IVD<br>ELISA + | 24 | 4 |
| IVD<br>ELISA - | 0 | 146 |
| 95% CI = 86% to 100% | | |
Sensitivity: 100% (24/24) Specificity: 97% (146/150) 95% CI = 93% to 99%
No cross-reactions were seen with the following organisms:
Entamoeba hartmanni, Endolimax nana, Entamoeba histolytica/dispar, Entamoeba coli, Blastocystis hominis, Dientamoeba fragilis, Chilomastix mesnili, Giardia lamblia, Cyclospora cayetanensis oocysts, Strongyloides stercoralis, Ascaris lumbricoides, Enterobius vermicularis (pinworm), Diphyllobothrium species, Hymenolepis nana, Clonorchis sinensis, Enteromonas hominis, Trichuris trichiura, Iodamoeba buetschlii, Hookworm (Necator americanus), Schistosoma mansoni, Taenia eggs, Isospora belli, Entamoeba polecki, adenovirus, E. coli, Campylobacter, Salmonella, Shigella, Yersinia and 29 other bacterial species (list available on request).
#### Study #2 (Manufacturer Performed)
A total of 44 fresh or fresh/frozen stools examined by O&P microscopy and/or another commercial ELISA were tested against this IVD ELISA. The following results were obtained.
| | ELISA | ELISA |
|----------------|-------|-------|
| IVD<br>ELISA + | 14 | 0 |
| IVD<br>ELISA - | 0 | 30 |
Positive Agreement = 100% (14/14) Negative Agreement = 100% (30/30)
The breakdown for the samples in the studies is as follows:
| Media | Cryptosporidium + | Other Parasites |
|--------------|-------------------|-----------------|
| Fresh/Frozen | 14 | 0 |
| 10% Formalin | 24 | 25 |
| SAF | 1 | 83 |
| MCC | 2* | 31 |
*These two samples were seeded samples.
#### Analytical Sensitivity
This assay can detect approximately 50 nanograms of soluble protein per ml of Cryptosporidium species antigen or 10 oocysts per well. At these detection levels, an OD range of 0.15 to 0.30 can be expected.
### CONCLUSION
IVD Research's Cryptosporidium Human Fecal Antigen Detection Microwell ELISA Kit uses similar methodology to the predicate device. In testing of various fecal specimens, the assay also showed equivalent sensitivity and specificity to the predicate device and O&P microscopy.
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a bird in flight, composed of three curved lines.
JUL 1 0 2003
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
IVD Research, Inc. c/o Alfredo J. Quattrone, Ph.D., D.A.B.T. California Department of Health Food & Drug Branch P.O. Box 942732 (MS-357) Sacramento, CA 94234
Re: k031059
Trade/Device Name: Cryptosporidium Human Fecal Antigen Detection Microwell ELISA Kit
Regulation Number: 21 CFR 866.3220 Regulation Name: Entamoeba histolytica serological reagents Regulatory Class: Class II Product Code: MHJ Dated: March 27, 2003 Received: March 28, 2003
Dear Dr. Quattrone:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{3}------------------------------------------------
# Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
| 510(k) Number: | Pending this Submission K031059 |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Cryptosporidium Human Fecal Antigen Detection<br>Microwell ELISA Kit |
| Indications for Use: | This microwell enzyme linked immunoabsorbant assay (ELISA)<br>detection kit is an in vitro diagnostic (IVD) immunoassay for the<br>detection of Cryptosporidium antigen in human feces using<br>peroxidase as the enzyme. The assay may be read visually or with<br>an ELISA reader. This IVD assay is intended to be used with<br>stools that are fresh, frozen or preserved in 10% formalin, SAF or<br>Medical Chemical Corporation's (MCC) Universal fixative.<br>Concentrated samples cannot be used with this IVD. |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use
(Optional Format 1-2-96)
Freddie Hi. Poole
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K031059
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.