SYNCHRON SYSTEMS OPIATE REAGENT

K031002 · Beckman Coulter, Inc. · DJG · Apr 29, 2003 · Clinical Toxicology

Device Facts

Record IDK031002
Device NameSYNCHRON SYSTEMS OPIATE REAGENT
ApplicantBeckman Coulter, Inc.
Product CodeDJG · Clinical Toxicology
Decision DateApr 29, 2003
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.3650
Device ClassClass 2

Indications for Use

Opiate 300 nq (OP) Reagent, in conjunction with the SYNCHRON® System OP 300 Urine Calibrators, is intended for the qualitative determination of opiates in human urine, at a cutoff value of 300 ng/mL (morphine), on SYNCHRON Systems. The Opiate assay provides a rapid screening procedure for determining the presence of opiates in urine. This test provides only a preliminary analytical result: a positive result by this assay should be confirmed by another generally accepted non-immunological method, such as thin layer chromatography (TLC), gas chromatography (GC), or gas chromatography/mass spectrometry (GC/MS). GC/MS is the preferred confirmatory method.12 Clinical consideration and professional judgement should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Device Story

In vitro diagnostic reagent kit for qualitative opiate detection in human urine; used on Beckman Coulter SYNCHRON CX and LX clinical chemistry systems. Reagent utilizes modified antibodies to detect opiates at 300 ng/mL cutoff. System performs automated analysis; provides preliminary screening results for clinical laboratory professionals. Positive results require confirmation via non-immunological methods (e.g., GC/MS). Assists clinicians in identifying potential opiate presence; supports clinical decision-making regarding drug use.

Clinical Evidence

Bench testing only. Validation testing performed to support equivalency of the modified reagent antibody compared to the predicate device.

Technological Characteristics

In vitro diagnostic reagent kit; antibody-based immunoassay; designed for use on SYNCHRON CX and LX automated clinical chemistry analyzers. Reagent packaged in 250-test cartridges.

Indications for Use

Indicated for the qualitative screening of opiates in human urine at a 300 ng/mL cutoff for clinical laboratory use.

Regulatory Classification

Identification

An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.

Special Controls

*Classification.* Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K031002 APR 2 9 2003 # 510(k) SUMMARY # 1.0 Submitted Bv: Kim Walker Senior Requlatory Affairs Specialist Beckman Coulter, Inc. 200 S. Kraemer Blvd. W-104 Brea, CA 92822-8000 Telephone: (714) 961-4912 FAX: (714) 961-4123 ## Date Submitted 2.0 March 25, 2003 #### 3.0 Device Name(s): - 3.1 Proprietary Names SYNCHRON Systems Opiate Reagent - 3.2 Classification Names Opiate test system. [862.3650] #### 4.0 Legally Marketed Device The SYNCHRON Systems Opiate Reagent claims substantial equivalence to the SYNCHRON Systems Opiate Reagent currently in commercial distribution. (FDA 510(k) Number K944606) #### 5.0 Device Description The SYNCHRON Systems Opiate (OP) reagent is designed for optimal SYNCHRON performance on on the CX (CX4/4CE/4A/4PRO, CX5/5CE/5A/5PRO. CX7/7RTS/7Δ/7PRO. CX9ALX/9PRO) and LX (LX20/PRO/LXi) Systems. The reagent kit contains one 250-test cartridge that is packaged separately from the associated calibrators. {1}------------------------------------------------ # 6.0 Intended Use Opiate 300 nq (OP) Reagent, in conjunction with the SYNCHRON® System OP 300 Urine Calibrators, is intended for the qualitative determination of opiates in human urine, at a cutoff value of 300 ng/mL (morphine), on SYNCHRON Systems. The Opiate assay provides a rapid screening procedure for determining the presence of opiates in urine. This test provides only a preliminary analytical result: a positive result by this assay should be confirmed by another generally accepted non-immunological method, such as thin layer chromatography (TLC), gas chromatography (GC), or gas chromatography/mass spectrometry (GC/MS). GC/MS is the preferred confirmatory method.12 Clinical consideration and professional judgement should be applied to any drug of abuse test result, particularly when preliminary positive results are used. # 7.0 Comparison to the Predicate (Description of the Modification to the Legally Marketed Device) The SYNCHRON Systems Opiate reagent antibody has been modified for drug cross-reactivity. ## 8.0 Summary of Performance Data Performance data from validation testing supports equivalency. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines above a wavy line. APR 2 9 2003 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Ms. Kim Walker Senior Regulatory Affairs Specialist Beckman Coulter, Inc. 200 S. Kraemer Boulevard M/S W-104, Box 8000 Brea, CA 92822-8000 Re: k031002 Trade/Device Name: SYNCHRON® Systems Opiate Reagent Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate Test System Regulatory Class: Class II Product Code: DJG Dated: March 25, 2003 Received: March 31, 2003 Dear Ms. Walker: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {3}------------------------------------------------ # Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Steven Sutman Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ page _1_ of _1_ 510(k) Number (if known): SYNCHRON® Systems Opiate Reagent Device Name: Indications for Use: Opiate 300 ng (OP) Reagent, in conjunction with the SYNCHRON® System OP 300 Urine Calibrators, is intended for the qualitative determination of opiates in human urine, at a cutoff value of 300 ng/mL (morphine), on SYNCHRON Systems. The Opiate assay provides a rapid screening procedure for determining the presence of opiates in urine. This test provides only a preliminary analytical result: a positive result by this assay should be confirmed by another generally accepted non-immunological method, such as thin layer chromatography (TLC), gas chromatography (GC), or gas chromatography/mass spectrometry (GC/MS). GC/MS is the preferred confirmatory method.12 Clinical consideration and professional judgement should be applied to any drug of abuse test result, particularly when preliminary positive results are used. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) i OR Prescription Use (per 21 CFR 801.109) Over-the-Counter Use Optional Format 1-2-96 Dean Cooper (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number K031002 Concurrence of CDRH, Office of Device Evaluation (ODE) Beckman Coulter, Inc., Special 510(k) Device Modification SYNCHRON® Systems Opiate Reagent File: OP_s510(k).doc 03/03
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...