# RITA CLIPS FOR INFUSION PUMPS (K030936)

_Rita Medical Systems, Inc. · MRZ · Apr 1, 2003 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K030936

## Device Facts

- **Applicant:** Rita Medical Systems, Inc.
- **Product Code:** [MRZ](/productcode/MRZ.md)
- **Decision Date:** Apr 1, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5725
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

The RITA® Clips are used with Harvard 2 Syringe Pump to increase the syringe holding capacity of the Pump from two to five syringes to allow one Harvard 2 Syringe Pump to supply local delivery of saline through multiple irrigation ports of one RITA® Electrosurgical Device of the RITA® System. The RITA® System is intended for use in percutaneous, laparoscopic, or intraoperative coagulation and ablation of soft tissue, including the partial or complete ablation of non-resectable liver lesions.

## Device Story

RITA Clips are a mechanical syringe holder accessory for the Harvard 2 Syringe Pump; increases capacity from two to five 20-cc syringes. Used in clinical settings (percutaneous, laparoscopic, or intraoperative) by physicians to facilitate saline irrigation during soft tissue coagulation and ablation procedures using the RITA Electrosurgical Device. The device does not contact the patient. It enables the delivery of saline through multiple irrigation ports of the electrosurgical device, supporting the ablation of soft tissue lesions.

## Clinical Evidence

Bench testing only. Testing verified flow rate and alarm function of the Harvard 2 Syringe Pump when equipped with RITA Clips. No patient contact occurs; no biocompatibility testing required.

## Technological Characteristics

Mechanical syringe holder accessory; accommodates five 20-cc syringes. Designed for use with the Harvard 2 Syringe Pump. No electrical components or software. Non-patient contacting.

## Regulatory Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

## Predicate Devices

- Harvard Clinical Technology - Harvard 2 Syringe Pump

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS Y 030 936 9.0

| Contact<br>Information      | RITA Medical Systems                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                             | APR 01 2003                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                             | 967 N. Shoreline Blvd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                             | Mountain View, CA 94043                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                             | Phone: 650 314-3400 Fax: 650 890-3905                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                             | Contact Person: Vicki Hacker, VP of Clinical Affairs                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| General<br>Provisions       | Trade Name: RITA® Clips                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                             | Common/Classification Name: Infusion Pump Accessory                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Name of<br>Predicate Device | Harvard Clinical Technology - Harvard 2 Syringe Pump                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Classification              | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Performance<br>Standards    | Performance standards have not been established for infusion pump accessory<br>(syringe holder) under section 514 of the Food, Drug and Cosmetic Act.                                                                                                                                                                                                                                                                                                                                                                              |
| Intended Use                | The RITA® Clips are used with Harvard 2 Syringe Pump to increase the syringe<br>holding capacity of the Pump from two to five syringes to allow one Harvard 2<br>Syringe Pump to supply local delivery of saline through multiple irrigation ports of<br>one RITA® Electrosurgical Device of the RITA® System. The RITA® System is<br>intended for use in percutaneous, laparoscopic, or intraoperative coagulation and<br>ablation of soft tissue, including the partial or complete ablation of non-resectable<br>liver lesions. |
| Device<br>Description       | The RITA® Clips are a syringe holder accessory to the Harvard 2 Syringe Pump to<br>accommodate five (5) 20-cc syringes.                                                                                                                                                                                                                                                                                                                                                                                                            |
| Performance<br>Data         | The RITA® Clips and Harvard 2 Syringe Pump underwent testing to verify flow<br>rate and alarm function. The devices passed the test criteria. The RITA® Clips do<br>not directly contact the patient therefore no biocompatibility testing was required.                                                                                                                                                                                                                                                                           |

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image contains the words "Public Health Service". The text is in a simple, sans-serif font. The words are stacked on top of each other, with "Public Health" on the top line and "Service" on the bottom line.

Image /page/1/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of three human profiles facing to the right. Above the profiles are three curved lines that resemble feathers or wings.

APR 0 1 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

RITA Medical Systems Incorporated C/O Mr. Charles Mack Responsible Third Party Official Underwriters Laboratories Incorporated 2600 N.W. Lake Road Camas, Washington 98601-8542

Re: K030936

Trade/Device Name: RITA ® Clips for Infusion Pumps Regulation Number: 880.5725 Regulation Name: Infusion Pump Accessory Regulatory Class: II Product Code: MRZ Dated: March 19, 2003 Received: March 25, 2003

Dear Mr. Mack:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Page 2 - Mr. Mack

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Susan Raanes

Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## INTENDED USE 3.0

## Indications for Use Statement

510(K) Number (if known)

RITA® Clips for Infusion Pumps Device Name

> The RITA® Clips are used with Harvard 2 Syringe Pump to increase the syringe holding capacity of the Pump from two to five syringes to allow one Harvard 2 Syringe Pump to supply local delivery of saline through multiple irrigation ports of one RITA® Electrosurgical Device of the RITA® System. The RITA System is intended for use in percutaneous. laparoscopic, or intraoperative coagulation and ablation of soft tissue, including the partial or complete ablation of non-resectable liver lesions.

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use v (per 21 CFR 801).

OR

Over the Counter Use _________________________________________________________________________________________________________________________________________________________

*Talcus Cicerite*

Division of Anesthesiology, General Hospital, Infection Control, Dental Devices

510(k) Number:

March 6, 2003

---

**Source:** [https://fda-staging.innolitics.com/device/K030936](https://fda-staging.innolitics.com/device/K030936)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).