← Product Code [FZP](/productcode/FZP) · K030926

# FAST THROW STAINLESS STEEL SURGICAL SUTURE (K030926)

_Advanced Surgical Products, Inc. · FZP · Jul 21, 2003 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K030926

## Device Facts

- **Applicant:** Advanced Surgical Products, Inc.
- **Product Code:** [FZP](/productcode/FZP.md)
- **Decision Date:** Jul 21, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The "Fast Throw" Stainless Steel Surgical Suture is indicated for use in superficial skin closure.

## Device Story

The "Fast Throw" Stainless Steel Surgical Suture is a sterile, single-use, knot-less device designed for superficial skin wound closure. It features barbs along its length and a T-shaped backstop at the proximal end. During use, the surgeon passes the suture through the wound and cinches the tissue against the T-shaped backstop to secure the closure without requiring traditional knots. The device is intended for use by clinicians in a surgical setting.

## Clinical Evidence

No clinical data. Evidence consists of bench testing (needle attachment pull strength, tensile testing per USP standards) and in vivo testing in a controlled environment.

## Technological Characteristics

Material: Stainless steel. Design: Monofilament suture with integrated barbs and a proximal T-shaped backstop. Sterilization: Sterile, single-use. Testing: Performed per USP standards for needle attachment and tensile strength.

## Regulatory Identification

An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.

## Predicate Devices

- LOOK Inc. 316L STAINLESS STEEL SURGICAL SUTURE ([K933686](/device/K933686.md))
- ETHI-PACK SURGICAL STAINLESS STEEL SUTURE ([K931271](/device/K931271.md))
- STAINLESS STEEL SUTURE ([K931915](/device/K931915.md))
- CLIP, IMPLANTABLE UNITED STATES SURGICAL ([K970793](/device/K970793.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K030926/P1/2

# JUL 2 1 2003

#### IV. Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92

510(k) Number: TBD

#### Applicant Information:

| Date Prepared:                   | March 10, 2003                                                          |
|----------------------------------|-------------------------------------------------------------------------|
| Name:<br>Address:                | Advanced Surgical Products<br>681 Manzanita Ave<br>Sunnyvale, CA. 94085 |
| Contact Person:<br>Phone Number: | Jeffrey S. Jones<br>650-207-6987                                        |

408-734-5316

#### Device Information:

Facsimile Number:

| Classification: | Class II                                             |
|-----------------|------------------------------------------------------|
| Trade Name:     | Advanced Surgical Products, Inc. "Fast Throw" Suture |
| Common name:    | STAINLESS STEEL SURGICAL SUTURE                      |

#### Predicate Device:

The Advanced Surgical Products "Fast Throw" Stainless Steel Surgical Suture is substantially equivalent in intended use and method of operation to the following predicate devices:

| Name:        | LOOK Inc. 316L STAINLESS STEEL SURGICAL SUTURE |
|--------------|------------------------------------------------|
| Manufacture: | LOOK, Inc.                                     |
| 510(k) #:    | K933686                                        |
| Name:        | ETHI-PACK SURGICAL STAINLESS STEEL SUTURE      |
| Manufacture: | ETHICON, Inc.                                  |
| 510(k) #:    | K931271                                        |
| Name:        | STAINLESS STEEL SUTURE                         |
| Manufacture: | SURGIMETRICS INTERNATIONAL, LTD.               |
| 510(k) #:    | K931915                                        |

{1}------------------------------------------------

K030926/P2/2

Name: Manufacture: 510(K) #:

CLIP, IMPLANTABLE UNITED STATES SURGICAL K970793

#### Device Description:

The Advanced Surgical Products "Fast Throw" Stainless Steel Surgical Suture is a sterile single use knot-less suture that is primarily designed for skin wound closure The device incorporates barbs along the length of the suture and works in conjunction with a Tshaped backstop incorporated at the proximal end of the device. As the suture is passed through the wound, the wound is cinched up to the T-shaped backstop.

#### Intended Use:

The "Fast Throw" Stainless Steel Surgical Suture is indicated for use in superficial skin closure.

#### Comparison to Predicate Devices:

The "Fast Throw" Stainless Steel Surgical Suture is substantially equivalent in intended use and method of operation to 316L monofilament Stainless Steel sutures and Implantable Clips currently in use.

#### Pre-Clinical Data:

Per USP Standards, needle attachment pull strength was performed as well as tensile testing on sterilized sutures.

In Vivo testing was performed with sterilized, packaged sutures in a controlled environment.

#### Summary:

Based on the product technical information, intended use, performance and biocompatibility information provided in this pre-market notification, Advanced Surgical Products, Inc. "Fast Throw" Stainless Steel Surgical Suture has been shown to be substantially equivalent to the currently marketed predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name encircling a symbol. The symbol consists of three stylized, overlapping shapes that resemble abstract human figures or birds in flight. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

### JUL 2 1 2003

Mr. Jeffrey S. Jones President Advanced Surgical Products 681 Manzanita Avenue Sunnyvale, California 94085

Re: K030926

Trade/Device Name: "Fast Throw" Stainless Steel Surgical Suture Regulation Number: 21 CFR 878.4300, 878.4495 Regulation Name: Implantable clip, Stainless steel suture Regulatory Class: II Product Code: FZP, GAQ Dated: May 22, 2003 Received: May 27, 2003

Dear Mr. Jones:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. Jeffrey S. Jones

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Miriam C. Provost

Celia M. Witten, Ph.D., M.D. Director Division of General. Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Ko30926

## II. Indications for Use Form

510(k) number (if known):

Device Name:

Advanced Surgical Products, Inc. "Fast Throw" Stainless Steel Surgical Suture

#### Indications for Use:

The "Fast Throw" Stainless Steel Surgical Suture is indicated for use in superficial skin closure.

Muriàm C. Provost

(Division Sign-Off) Division of General, Restorative and Neurological Devices

510(k) Number K030926

---

**Source:** [https://fda-staging.innolitics.com/device/K030926](https://fda-staging.innolitics.com/device/K030926)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).