PROXIMATE HCS HEMORRHOIDAL CIRCULAR STAPLER AND ACCESSORIES, MODEL PPH01

K030925 · Ethicon Endo-Surgery, Inc. · GDW · May 15, 2003 · General, Plastic Surgery

Device Facts

Record IDK030925
Device NamePROXIMATE HCS HEMORRHOIDAL CIRCULAR STAPLER AND ACCESSORIES, MODEL PPH01
ApplicantEthicon Endo-Surgery, Inc.
Product CodeGDW · General, Plastic Surgery
Decision DateMay 15, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4750
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The PROXIMATE® HCS Hemorrhoidal Circular Stapler and Accessories have application for general surgical treatment of anorectal wall defects by means of transanal stapling and resection of mucosal and musculomucosal tissue.

Device Story

PROXIMATE HCS Hemorrhoidal Circular Stapler and Accessories; mechanical surgical instrument set. Facilitates delivery of circumferential, staggered, double-row of staples; simultaneously resects segment of compressed soft tissue. Used in Procedure for Prolapse and Hemorrhoids (PPH) and other anorectal tissue stapling/resection procedures. Operated by surgeons in clinical/surgical settings. Provides mechanical tissue fixation and resection; assists in repair of anorectal wall defects.

Clinical Evidence

Published clinical data and information demonstrate the general utility of the HCS device to staple and resect anorectal tissue in procedures similar to the previously cleared indications, supporting substantial equivalence.

Technological Characteristics

Mechanical intraluminal stapler; delivers circumferential, staggered, double-row of staples; performs simultaneous tissue resection. Identical to predicate device. No software, electronics, or energy sources.

Indications for Use

Indicated for general surgical treatment of anorectal wall defects via transanal stapling and resection of mucosal and musculomucosal tissue.

Regulatory Classification

Identification

An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ **MAY 15 2003** K0309225 # PROXIMATE® HCS Hemorrhoidal Circular Stapler and Accessories (PPH01) # 510(k) Summary of Safety and Effectiveness #### Company: Date Prepared: March 21, 2003 Ethicon Endo-Surgery, Inc. 4545 Creek Rd. Cincinnati, OH 45242 #### Contact: Name: Carol Sprinkle Title: Regulatory Affairs Specialist #### Name of Device: Trade Name: PROXIMATE® HCS Hemorrhoidal Circular Stapler and Accessories Common Name: Intraluminal Stapler (ILS) Classification Name: Implantable Staple, General & Plastic Surgery Surgical Device #### Predicate Device: PROXIMATE® HCS Hemorrhoidal Circular Stapler and Accessories (K991030). #### Device Description: The PROXIMATE HCS Hemorrhoidal Circular Stapler and Accessories are a set of instruments that facilitate delivery of a circumferential, staggered, double-row of staples while simultaneously resecting a segment of compressed soft tissue. The set is commonly used in the Procedure for Prolapse and Hemorrhoids (PPH). The set is also used for other applications where circular or semicircular stapling of anorectal tissue is desired. #### Indications for Use: The PROXIMATE® HCS Hemorrhoidal Circular Stapler and Accessories have application for general surgical treatment of anorectal wall defects by means of transanal stapling and resection of mucosal and musculomucosal tissue. ### Technological Characteristics: The PROXIMATE HCS Hemorrhoidal Circular Stapler is identical to the predicate device. The only changes are in the Indications for Use statement and Instructions for Use. ### Clinical Evidence: Published clinical data and other information demonstrate the general utility of the HCS device to staple and resect anorectal tissue in various procedures similar to the previously cleared Indications for Use for surgical treatment of hemorrhoidal disease, thus meeting all the requirements of substantial equivalence. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image is the seal of the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle with three human profiles embedded within the eagle's body. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # MAY 1 5 2003 Ms. Carol Sprinkle Regulatory Affairs Specialist Ethicon Endo-Surgery, Inc. 4545 Creek Road Cincinnati, Ohio 45242 Re: K030925 Trade/Device Name: PROXIMATE® HCS Hemorrhoidal Circular Stapler and Accessories Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: II Product Code: GDW Dated: March 21, 2003 Received: March 24, 2003 Dear Ms. Sprinkle : We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ Page 2 - Ms. Carol Sprinkle This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Miriam C. Provost (Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Page | of _ of _( 510(k) Number (if known): _ K D30925 Device Name: PROXIMATE® HCS Hemorrhoidal Circular Stapler and Accessories Indications for Use: The PROXIMATE® HCS Hemorrhoidal Circular Stapler and Accessories have application for general surgical treatment of anorectal wall defects by means of transanal stapling and resection of mucosal and musculomucosal tissue. Miriam C. Provost (Division Sign-Off) Division of General, Restorative and Neurological Devices 510(k) Number K030925 PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) > Concurrence of CDRH, Office of Device Evaluation (ODE) (Optional Format 3-10-98)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...