Retrospective clinical serum samples from private pathology laboratories; Retrospective clinical serum samples from a state health laboratory
Retrospective clinical samples were used to evaluate the sensitivity, specificity, and agreement of the PANBIO EBV VCA-p18 IgG ELISA compared to a predicate device and established EBV serological status.
342 prospective sera (seronegative, acute infectious mononucleosis, and past exposure); Sample Size: 342; Number of Sites: 1
DiaSorin EBV VCA IgG ELISA
Sensitivity, specificity, and agreement relative to EBV serological status
Retrospective clinical performance study
148 frozen retrospective sera (seronegative, acute infectious mononucleosis, and past exposure); Sample Size: 148; Number of Sites: 1
DiaSorin EBV VCA IgG ELISA
Sensitivity, specificity, and agreement relative to EBV serological status
Indications for Use
The Epstein Barr Virus Viral Capsid-p18 Antigen (EBV VCA-p18) IgG ELISA Test is for the qualitative detection of IgG antibodies to EBV VCA in serum as an aid in the clinical laboratory diagnosis of EBV infection in patients with clinical symptoms consistent with Infectious Mononucleosis (IM). The PANBIO EBV VCA-p18 IgG ELISA should be used in conjunction with other EBV serologies.
Device Story
The EBV VCA-p18 IgG ELISA is an in vitro diagnostic test used in clinical laboratories. It utilizes a synthetically produced EBV VCA-p18 peptide antigen attached to polystyrene microwells. Patient serum is added; if IgG antibodies to VCA are present, they bind to the antigen. After washing, peroxidase-conjugated anti-human IgG is added, followed by a TMB/H2O2 substrate system. The enzymatic reaction produces a color change (blue, then yellow after acid stop) proportional to the amount of bound IgG. Results are interpreted by healthcare providers to aid in the diagnosis of EBV infection. The device is intended for use in conjunction with other EBV serologies to support clinical decision-making for patients presenting with symptoms of infectious mononucleosis.
Clinical Evidence
Clinical performance evaluated across two study sites (Australia and USA) using 490 total prospective and retrospective serum samples. Compared against the DiaSorin EBV VCA IgG ELISA and established EBV serological status. In Australian study (n=342), relative sensitivity for acute infection was 68.3% and specificity 100%. In US study (n=148), relative sensitivity for acute infection was 73.9% and specificity 96.0%. Reproducibility testing (n=27 per sample) showed total CVs ranging from 4.1% to 13.9%. Analytical specificity evaluated against a panel of 30 specimens (CMV, VZV, HSV, ANA, RF) showed 1/30 cross-reactivity.
Technological Characteristics
Enzyme-Linked Immunosorbent Assay (ELISA) using synthetic EBV VCA-p18 peptide antigen coated on polystyrene microwells. Detection via peroxidase-conjugated anti-human IgG and TMB/H2O2 substrate. Manual or automated plate processing. Results determined by absorbance measurement.
Indications for Use
Indicated for qualitative detection of IgG antibodies to EBV VCA in human serum as an aid in clinical laboratory diagnosis of EBV infection in patients with symptoms consistent with infectious mononucleosis (IM).
Regulatory Classification
Identification
Epstein-Barr virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Epstein-Barr virus in serum. The identification aids in the diagnosis of Epstein-Barr virus infections and provides epidemiological information on diseases caused by these viruses. Epstein-Barr viruses are thought to cause infectious mononucleosis and have been associated with Burkitt's lymphoma (a tumor of the jaw in African children and young adults) and postnasal carcinoma (cancer).
Predicate Devices
DiaSorin EBV VCA IgG ELISA
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for PANBIO. The logo consists of a circular graphic on the left and the word "PANBIO" in bold, sans-serif letters on the right. The graphic appears to be an abstract design, possibly representing biological elements. The overall impression is a clean and professional logo for a company in the biotechnology or related field.
#### 10(k) SUMMARY OF SAFETY AND EFFECTIVENESS 1.10
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is:
## Applicant Information:
| Date Prepared: | 12th March, 2003 |
|----------------|---------------------------------------------------------|
| Name: | PANBIO Limited |
| Address: | 116 Lutwyche Road, Windsor<br>Queensland 4030 Australia |
| Contact Person: | Helen Jennings |
|-----------------|--------------------|
| Phone Number: | +61-(0)7-3357-1177 |
| Fax Number: | +61-(0)7-3357-1222 |
### Device Information:
| Trade Name: | EBV VCA-p18 IgG ELISA |
|----------------------|------------------------------------------|
| Common Name: | EBV VCA IgG EIA Test |
| Classification Name: | Epstein-Barr virus serological reagents. |
## Equivalent Device:
DiaSorin EBV VCA IgG ELISA
## Device Description:
The EBV VCA-p18 IgG ELISA is an Enzyme Linked Immunosorbent Assay for the qualitative detection of IgG antibodies in human serum to EBV VCA antigen.
## Intended Use:
The Epstein-Barr Virus Viral Capsid-p18 Antigen (EBV VCA-p18) IgG ELISA is for the qualitative detection of IgG antibodies to EBV VCA in serum as an aid in the clinical laboratory diagnosis of EBV infection in patients with clinical symptoms consistent with infectious mononucleosis (IM). The PANBIO EBV VCA-p18 IgG ELISA should be used in conjunction with other EBV serologies.
#### Principle of Procedure:
Serum containing antibodies to VCA antigen, when present, combine with EBV VCA-p18 antigen attached to the polystyrene surface of the microwells. The antigen is a synthetically produced peptide. Residual serum is removed by washing and peroxidase conjugated antihuman IgG is added. The microwells are washed and a colourless substrate system, tetramethylbenzidine/hydrogen peroxide (TMB/H₂O₂) is added. The substrate is hydrolysed by the enzyme and the chromogen changes to a blue colour. After stopping the reaction with acid, the TMB becomes yellow. Colour development is indicative of the presence of EBV VCA IgG antibodies in the test sample.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "PANBIO" in bold, black letters. To the left of the word is a circular logo with a stylized design. The logo appears to be textured or distressed, giving it a slightly rough appearance. The overall image has a clean and professional look.
### PERFORMANCE CHARACTERISTICS
### Study Site 1:
342 prospective sera of various ages and genders were tested at a private pathology laboratory in Queensland, Australia for EBV testing. The sera include the following groups: 45 seroneqative, 41 with acute infectious mononucleosis and 256 with past exposure to EBV.
These sera were tested on the PANBIO EBV VCA-p18 IgG ELISA and the DiaSorin EBV VCA IgG ELISA. The PANBIO results were compared to the EBV status of the sera to determine the sensitivity, specificity, and agreement of the assay relative to the EBV serological status (Table 1). Additionally, the DiaSorin results were compared to the EBV serological status (Table 2) and PANBIO results (Table 3), as summarised below.
| PANBIO ELISA | | | |
|--------------------------------------------------------------|----------|----------|-------|
| EBV Status | Positive | Negative | Total |
| Seronegative<br>VCA IgG (-)<br>VCA IgM (-)<br>EBNA IgG (-) | 0 | 45 | 45 |
| Acute<br>VCA IgM (+)<br>EBNA IgG (-) | 28 | 13 | 41 |
| Past Infection<br>VCA IgG (+)<br>VCA IgM (-)<br>EBNA IgG (+) | 242 | 14 | 256 |
| Total | 270 | 72 | 342 |
# TABLE 1 EBV STATUS VERSUS PANBIO ELISA
| Relative Sensitivity (Acute) | = | | | 95% Confidence Interval |
|---------------------------------|---------|--------------|----------|-------------------------|
| Relative Sensitivity (Past) | = | 28/41 | = 68.3% | 51.9 – 81.9% |
| Relative Specificity (Negative) | = | 242/256 | = 94.5% | 91.0 – 97.0% |
| Relative Agreement | = | 45/45 | = 100.0% | 92.1 – 100.0% |
| 315/342 | = 92.1% | 88.7 – 94.7% | | |
#### 95% Confidence Interval
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "PANBIO" in all capital letters. To the left of the word is a circular graphic with a design inside. The letters in the word are bold and black. The graphic is also black and white.
# TABLE 2 EBV STATUS VERSUS DIASORIN ELISA
| | PANBIO ELISA | | |
|--------------------------------------------------------------|--------------|----------|-------|
| EBV Status | Positive | Negative | Total |
| Seronegative<br>VCA IgG (-)<br>VCA IgM (-)<br>EBNA IgG (-) | 0 | 45 | 45 |
| Acute<br>VCA IgM (+)<br>EBNA IgG (-) | 35 | 6 | 41 |
| Past Infection<br>VCA IgG (+)<br>VCA IgM (-)<br>EBNA IgG (+) | 248 | 8 | 256 |
| Total | 283 | 59 | 342 |
Relative Sensitivity (Acute) Relative Sensitivity (Past) Relative Specificity (Negative) Relative Agreement
95% Confidence Interval = 85.4% = 35/41 70.8 – 94.4% = 248/256 = 96.9% 93.9 -- 98.6%
| = 45/45 | = 100.0% | 92.1 - 100.0% |
|-----------|----------|---------------|
| = 328/342 | = 95.6% | 93.2 - 97.7% |
| TABLE 3 | | | | |
|------------------------------|--|--|--|--|
| PANBIO VERSUS DIASORIN ELISA | | | | |
| PANBIO ELISA | | | |
|--------------|----------|----------|-------|
| DiaSorin | Positive | Negative | Total |
| Positive | 269 | 14 | 283 |
| Negative | 1 | 58 | 59 |
| Total | 270 | 72 | 342 |
| | | | 95% Confidence Interval |
|----------------------|-----------|---------|-------------------------|
| Relative Sensitivity | = 269/283 | = 95.1% | 91.8 – 97.3% |
| Relative Specificity | = 58/59 | = 98.3% | 90.9 – 100.0% |
| Relative Agreement | = 327/342 | = 95.6% | 92.9 – 97.5% |
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "PANBIO" in bold, black letters. To the left of the word is a circular graphic with a stylized design. The design appears to be made up of intersecting lines or shapes, creating a textured effect within the circle.
## Study Site 3:
148 frozen retrospective sera of various ages and genders were submitted to a state health laboratory in Maryland USA for EBV testing. The sera include samples from the following groups: 25 seroneqative samples from patients with acute Infectious Mononucleosis, and 100 samples from patients with past exposure to EBV.
These sera were tested on the PANBIO EBV VCA-p18 IgG ELISA and the DiaSorin EBV VCA IgG ELISA. The PANBIO results were compared to the EBV status of the sera to determine the sensitivity, specificity, and agreement of the assay relative to the EBV serological status (Table 4). Additionally, the DiaSorin results were compared to the EBV serological status (Table 5) and PANBIO results (Table 6), as summarised below.
| EBV Status | Positive | Equivocal* | Negative | Total |
|--------------------------------------------------------------|----------|------------|----------|-------|
| Seronegative<br>VCA IgG (-)<br>VCA IgM (-)<br>EBNA IgG (-) | 0 | 1 | 24 | 25 |
| Acute<br>VCA IgM (+)<br>EBNA IgG (-) | 17 | 1 | 5 | 23 |
| Past Infection<br>VCA IgG (+)<br>VCA IgM (-)<br>EBNA IgG (+) | 100 | 0 | 0 | 100 |
| Total | 117 | 2 | 29 | 148 |
# TABLE 4 EBV STATUS VERSUS PANBIO ELISA PANBIO ELISA
#### 95% Confidence Interval
| Relative Sensitivity (Acute) | = 17/23 | = 73.9% | 51.6 - 89.8% |
|---------------------------------|-----------|----------|--------------|
| Relative Sensitivity (Past) | = 100/100 | = 100.0% | 96.4 - 100% |
| Relative Specificity (Negative) | = 24/25 | = 96.0% | 79.6 - 99.9% |
| Relative Agreement | = 141/148 | = 95.3% | 90.5 - 98.1% |
*Retesting of equivocal samples was not conducted, as the samples were unavailable.
Note: "Serological" sensitivity and specificity refers to the comparison of the PANBIO assay results to that of other assays normally used to diagnose EBV associated IM. There was not an attempt to correlate the assay's results with disease presence or absence. No judgement can be made on the comparison's accuracy to predict disease. Since the above studies were performed on a pre-selected, retrospective, populations for the assay's positive and negative predictive value may be done or inferred.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "PANBIO" in bold, black letters. To the left of the word is a circular graphic with a black, abstract design. The design appears to be two intertwined shapes, possibly letters, with a textured, almost distressed look. The overall impression is a logo or branding element for a company or product named Panbio.
| | TABLE 5 | | |
|--|---------|----------------------------------|--|
| | | EBV STATUS VERSUS DIASORIN ELISA | |
| EBV Status | Positive | Negative | Total |
|--------------------------------------------------------------|----------|----------|-------|
| Seronegative<br>VCA IgG (-)<br>VCA IgM (-)<br>EBNA IgG (-) | 2 | 23 | 25 |
| Acute<br>VCA IgM (+)<br>EBNA IgG (-) | 23 | 0 | 23 |
| Past Infection<br>VCA IgG (+)<br>VCA IgM (-)<br>EBNA IgG (+) | 100 | 0 | 100 |
| Total | 125 | 23 | 148 |
| | | 95% Confidence Interval |
|---------------------------------|--------------------|-------------------------|
| Relative Sensitivity (Acute) | = 23/23 = 100.0% | 85.2 – 100% |
| Relative Sensitivity (Past) | = 100/100 = 100.0% | 96.4 – 100% |
| Relative Specificity (Negative) | = 23/25 = 92.0% | 74.0 - 99.0% |
| Relative Agreement | = 146/148 = 98.6% | 95.2 - 99.8% |
# TABLE 6 PANBIO VERSUS DIASORIN ELISA
| DiaSorin | Positive | Equivocal* | Negative | Total |
|----------|----------|------------|----------|-------|
| Positive | 117 | 1 | 7 | 125 |
| Negative | 0 | 1 | 22 | 23 |
| Total | 117 | 2 | 29 | 148 |
| | | | 95% Confidence Interval |
|----------------------|-----------|---------|-------------------------|
| Relative Sensitivity | = 117/125 | = 93.6% | 87.8 – 97.2% |
| Relative Specificity | = 22/23 | = 95.7% | 78.0 – 99.9% |
| Relative Agreement | = 139/148 | = 93.9% | 88.8 - 97.2% |
*Retesting of equivocal samples was not conducted, as the samples were unavailable.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "PANBIO" in bold, black letters. To the left of the word is a circular graphic that appears to be a stylized representation of organic matter. The overall impression is that this is a logo or branding element for a company named Panbio, possibly related to biology or organic products.
### REPRODUCIBILITY
### Study Sites 1, 4 & 5:
The reproducibility of the PANBIO EBV VCA-p18 IgG ELISA was determined by testing 8 sera 3 times each on three different days at three Australian study sites. Two sites were private pathology laboratories and the third site was PANBIO Limited. Within-run, between day, between site and total precision were estimated by analysis of variance (ANOVA Type II). The results are presented in table 7 below.
# TABLE 7 PANBIO EBV VCA-p18 IgG Study
| | | Within | | Between Day | | Between Site | | Total | | |
|----------|----|--------|------|-------------|------|--------------|------|-------|------|-------|
| Sample | n | *Mean | *S.D | CV | *S.D | CV | *S.D | CV | *S.D | CV |
| Positive | 27 | 2.98 | 0.17 | 5.7% | 0.06 | 2.0% | 0.11 | 3.8% | 0.20 | 6.7% |
| Cut-off | 27 | 1.00 | 0.05 | 4.5% | 0.00 | 0.0% | 0.00 | 0.0% | 0.04 | 4.1% |
| Negative | 27 | 0.18 | 0.02 | 8.9% | 0.00 | 2.4% | 0.02 | 12.6% | 0.02 | 13.9% |
| #1 | 27 | 5.11 | 0.37 | 7.2% | 0.07 | 1.3% | 0.00 | 0.0% | 0.37 | 7.2% |
| #2 | 27 | 5.06 | 0.34 | 6.7% | 0.14 | 2.8% | 0.00 | 0.0% | 0.35 | 7.0% |
| #3 | 27 | 5.35 | 0.32 | 6.0% | 0.07 | 1.3% | 0.50 | 9.4% | 0.53 | 9.9% |
| #4 | 27 | 1.30 | 0.06 | 4.9% | 0.00 | 0.0% | 0.02 | 1.3% | 0.06 | 4.8% |
| #5 | 27 | 1.93 | 0.11 | 5.6% | 0.04 | 2.0% | 0.05 | 2.8% | 0.12 | 6.3% |
| #6 | 27 | 1.81 | 0.11 | 6.1% | 0.03 | 1.6% | 0.12 | 6.5% | 0.15 | 8.3% |
| #7 | 27 | 0.97 | 0.05 | 5.0% | 0.02 | 1.9% | 0.04 | 4.4% | 0.06 | 6.4% |
| #8 | 27 | 0.90 | 0.08 | 9.0% | 0.00 | 0.0% | 0.07 | 7.3% | 0.10 | 10.8% |
### Precision Measures (Using Cut-Off Ratio*)
All values are calculated from Ratios (Cut-off using O.D) SD = Standard Deviation; CV = Coefficient of Variation
Note: Standard Deviation results have been rounded to two decimal places for tabulation purposes.
*Cut-off Ratio is calculated as the Absorbance of the Sample divided by the Mean Absorbance of the Cut-off.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "PANBIO" in bold, black letters. To the left of the word is a circular graphic that appears to be an abstract design. The graphic is also in black and white and has a textured appearance. The overall image is simple and clean, with a focus on the company name and logo.
### POTENTIAL CROSS-REACTIVITY
### Study Site 5:
This study consisted of a panel of 30 specimens screened for IgG antibodies detectable by ELISA to disease types other than Epstein-Barr virus. The purpose of this study was to establish the analytical specificity of the EBV VCA-p18 IgG ELISA, through the analysis of specimens from patients with diseases that have the potential for cross-reactivity. Each of the specimens included in the study was characterised with respect to disease diagnosis and analysed with the EBV VCA-p18 IgG ELISA. Table 9 on the following page lists the crossreactivity results for each type of specimen included in the disease panel. Table 8 below provides a summary of the data presented in Table 9 (see next page).
| Disease<br>(IgG Antibodies) | Total Specimens | Positive Result |
|-----------------------------|-----------------|-----------------|
| Cytomegalovirus | 5 | (0/5) |
| Varicella zoster | 10 | (0/10) |
| Herpes simplex virus 1 | 8 | (1/8) |
| Herpes simplex virus 2 | 1 | (0/1) |
| Anti-Nuclear Antibody | 3 | (0/3) |
| Rheumatoid Factor | 3 | (0/3) |
| Total Antibody | 30 | (1/30) |
### TABLE 8 - PANBIO EBV VCA-p18 IgG CROSS-REACTIVITY SPECIMEN PANEL SUMMARY
Results indicate that one specimen (1/30) was positive when analysed with the EBV VCA-p18 IgG ELISA. The overall result of the above disease panel is consistent with good analytical specificity for the EBV VCA-p18 IgG ELISA.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "PANBIO" in bold, black letters. To the left of the word is a circular logo with a stylized design. The logo appears to have a textured or distressed effect, giving it a slightly rough or vintage look. The overall impression is of a company or brand with a strong, established presence.
| | Gull IFA<br>Merifluor EBV IgG<br>Batch No. EB100.091 | PANBIO<br>EBV VCA-p18 IgG ELISA<br>Batch No. 02282 | Sample<br>IgG Antibodies | |
|----|------------------------------------------------------|----------------------------------------------------|--------------------------|-----------|
| 1 | N | 1.35 | N | CMV IgG |
| 2 | N | 1.82 | N | CMV IgG |
| 3 | N | 1.72 | N | CMV IgG |
| 4 | N | 3.56 | N | CMV IgG |
| 5 | N | 5.92 | N | CMV IgG |
| 6 | N | 3.56 | N | VZV IgG |
| 7 | N | 1.82 | N | VZV IgG |
| 8 | N | 1.18 | N | VZV IgG |
| 9 | N | 3.44 | N | VZV IgG |
| 10 | N | 1.35 | N | VZV IgG |
| 11 | N | 1.28 | N | VZV IgG |
| 12 | N | 5.92 | N | VZV IgG |
| 13 | N | 1.82 | N | VZV IgG |
| 14 | N | 1.60 | N | VZV IgG |
| 15 | N | 1.47 | N | VZV IgG |
| 16 | N | 3.56 | N | HSV 1 IgG |
| 17 | N | 1.97 | N | HSV 1 IgG |
| 18 | N | 5.92 | N | HSV 1 IgG |
| 19 | N | 1.47 | N | HSV 1 IgG |
| 20 | N | 1.28 | N | HSV 1 IgG |
| 21 | N | 1.60 | N | HSV 1 IgG |
| 22 | N | 2.04 | N | HSV 1 IgG |
| 23 | N | 13.73 | P | HSV 1 IgG |
| 24 | N | 5.70 | N | HSV 2 IgG |
| 25 | N | 0.75 | N | ANA |
| 26 | N | 0.73 | N | ANA |
| 27 | N | 0.84 | N | ANA |
| 28 | N | 0.58 | N | RF |
| 29 | N | 0.58 | N | RF |
| 30 | N | 0.67 | N | RF |
# TABLE 9 -- PANBIO EBV VCA-p18 IgG CROSS-REACTIVITY SPECIMEN PANEL
# INTERPRETATION
| | | -------<br>aunivoca | - |
|--------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|----------------|
| DANDIC | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | A CHARROOM CO | -------------- |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "PANBIO" in bold, black letters. To the left of the word is a circular logo with a design that appears to be abstract. The logo is also in black and has a distressed or textured appearance. The overall image has a clean, minimalist aesthetic.
#### METHODS OF DATA ANALYSIS 1.11
# Relative Sensitivity:
True Positives / (True Positives + False Negatives)
# Relative Specificity:
True Negatives / (True Negatives + False Positives)
# Relative Agreement:
(True Positives + True Negatives) / Total Samples
# 95% Confidence Interval:
CIA 'Confidence Interval Analysis' Software Program' from "Statistics with Confidence" by Prof. M. J. Gardner and British Medical Journal (1991). Version 1.1.
# ANOVA Analysis of Variance Type II:
Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline. NCCLS (1999), EP5-A Vol. 19 No. 2.
{9}------------------------------------------------
Public Health Service
Image /page/9/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES" are arranged in a circular pattern around the caduceus. The logo is black and white.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
JUN 2 7 2003
Ms. Kate Wersin Regulatory Affairs Officer PANBIO Limited 116 Lutwyche Road, Windsor Brisbane, Queensland, 4030 Australia
Re: k030863
> Trade/Device Name: EBV VCA-p18 IgG ELISA Regulation Number: 21 CFR 866.3235 Regulation Name: Epstein-Barr Virus Serological Reagents Regulatory Class: Class I Product Code: LSE Dated: May 13, 2003 Received: May 16, 2003
Dear Ms. Wersin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{10}------------------------------------------------
Page 2 –
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the word "PANBIO" in bold, black letters. To the left of the word is a circular graphic that is also in black. The graphic appears to be a stylized representation of some sort of organic material, possibly related to biology or nature, given the company name.
March 12, 2003
510(k) Number: K030863
# Device Name: EBV VCA-p18 IgG ELISA
The Epstein Barr Virus Viral Capsid-p18 Antigen (EBV VCA-p18) IgG ELISA Test is for the qualitative detection of IgG antibodies to EBV VCA in serum as an aid in the clinical laboratory diagnosis of EBV infection in patients with clinical symptoms consistent with Infectious Mononucleosis (IM). The PANBIO EBV VCA-p18 IgG ELISA should be used in conjunction with other EBV serologies.
# PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-the Counter Use (Optional Format 1-2-96)
Freddie Tu-Poole
(Division Sign-Off) Division of Clinical La 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.