MODIFICATION TO ENDOVIVE INITIAL PLACEMENT PEG KIT, DIRECT PEJ KIT, PEG SAFETY KIT
Applicant
Boston Scientific Corp
Product Code
KNT · Gastroenterology, Urology
Decision Date
Apr 17, 2003
Decision
SESK
Submission Type
Special
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
EndoVive™ Initial Placement PEG Kit is indicated for enteral nutrition directly into the stomach in both pediatric and adult patients who are unable to consume nutrition by conventional means. The EndoVive™ Initial Placement Direct PEJ Kit is indicated for use for enteral nutritional support and decompression directly into the jejunum when feeding via the upper gastrointestinal tract is contraindicated. The EndoVive™ Initial Placement PEG Safety Kit is indicated for providing nutrition directly into the stomach in patients who are unable to consume nutrition by conventional means. The safety components of the kit are designed to reduce the potential for inadvertent sharps injury to medical personnel during and after the procedure.
Device Story
EndoVive kits facilitate initial placement of gastrostomy (PEG) or jejunostomy (PEJ) tubes for enteral feeding or decompression. Devices are used in clinical settings by medical professionals. The primary modification in this submission is the use of a new material for the domed portion of the PEG/PEJ tubes. The kits include various components for tube placement, including sharps for the safety kit version. Healthcare providers use these tubes to provide direct nutritional access to the stomach or jejunum, bypassing the upper GI tract when necessary. The safety kit specifically aims to mitigate sharps injury risks during the procedure. The device functions as a conduit for nutrition and decompression, directly impacting patient nutritional status.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing and biocompatibility evaluation of the new domed material.
Technological Characteristics
Gastrointestinal feeding/decompression tubes. Modification involves a new material for the domed component. Biocompatibility testing conducted. Class II device under 21 CFR 876.5980.
Indications for Use
Indicated for enteral nutrition/decompression in pediatric and adult patients unable to consume nutrition conventionally or where upper GI feeding is contraindicated. Safety kit version includes features to reduce sharps injury risk to medical personnel.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
K030855
1 of 2
Special 510(k) Premarket Notification New Material for Domed PEG/PEJ
Boston Scientific Corporation March 17, 2003
# APR 1 7 2003
## 510(K) SUMMARY
| SPONSOR: | Boston Scientific Corporation (BSC)<br>Microvasive Endoscopy Division<br>One Boston Scientific Place<br>Natick, MA 01760 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT/SUBMITTER: | Paige Sweeney<br>Senior Regulatory Affairs Specialist |
| DATE OF SUBMISSION: | March 17, 2003 |
| DEVICES: | EndoVive <sup>™</sup> Initial Placement PEG Kit EndoVive <sup>™</sup><br>Initial Placement Direct PEJ Kit<br>EndoVive <sup>™</sup> Initial Placement PEG Safety Kit |
| TRADE NAME: | EndoVive <sup>™</sup> Initial Placement PEG Kit |
| COMMON NAME: | Gastrostomy Tube |
| CLASSIFICATION: | Tubes, Gastrointestinal & Accessories<br>Tube, Gastro-Enterostomy<br>Classified Under 21 CFR Part 876, §5980.<br>Classified as a Class II Device |
| TRADE NAME: | EndoVive <sup>™</sup> Initial Placement Direct PEJ Kit |
| COMMON NAME: | Jejunostomy Tube |
| CLASSIFICATION: | Tube, Feeding<br>Classified Under 21 CFR Part 876, §5980.<br>Classified as a Class II Device |
| TRADE NAME: | EndoVive <sup>™</sup> Initial Placement PEG Safety Kit |
| COMMON NAME: | Gastrostomy Tube |
| CLASSIFICATION: | Tubes, Gastrointestinal & Accessories<br>Tube, Gastro-Enterostomy<br>Classified Under 21 CFR Part 876, §5980.<br>Classified as a Class II Device |
| PREDICATE DEVICES: | EndoVive <sup>™</sup> Initial Placement PEG Kit<br>(K014297)<br>EndoVive <sup>™</sup> Initial Placement Direct PEJ Kit<br>(K020120)<br>EndoVive <sup>™</sup> Initial Placement PEG Safety Kit<br>(K014297) |
| DEVICE DESCRIPTIONS: | The proposed EndoVive <sup>™</sup> Initial Placement PEG<br>Kit, the proposed EndoVive <sup>™</sup> Initial Placement<br>Direct PEJ Kit, and the EndoVive <sup>™</sup> Initial Placement |
{1}------------------------------------------------
K030855
2 of 2
ﻳﻨ
#### INTENDED USES:
COMPARISON OF
CHARACTERISTICS:
Boston Scientific Corporation March 17, 2003
PEG Safety Kit are used during initial placement for direct feeding.
EndoVive™ Initial Placement PEG Kit is indicated for enteral nutrition directly into the stomach in both pediatric and adult patients who are unable to consume nutrition by conventional means.
The EndoVive™ Initial Placement Direct PEJ Kit is indicated for use for enteral nutritional support and decompression directly into the jejunum when feeding via the upper gastrointestinal tract is contraindicated.
The EndoVive™ Initial Placement PEG Safety Kit is indicated for providing nutrition directly into the stomach in patients who are unable to consume nutrition by conventional means. The safety components of the kit are designed to reduce the potential for inadvertent sharps injury to medical personnel during and after the procedure.
The proposed devices are substantially equivalent to currently marketed devices, as they are identical with the exception of the proposed domed PEG/PEJ material.
The proposed devices are substantially equivalent to PERFORMANCE DATA: currently marketed devices in terms of performance characteristics and were tested for biocompatibility.
{2}------------------------------------------------
Image /page/2/Picture/12 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized symbol that resembles a human figure or a bird in flight, composed of several curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 1 7 2003
Ms. Paige Sweeney Senior Regulatory Affairs Specialist Boston Scientific Corporation Microvasive Endoscopy One Boston Scientific Place NATICK MA 01760
Re: K030855
Trade/Device Name: EndoVive™ Initial Placement PEG Kit EndoVive™ Initial Placement Direct PEJ Kit EndoVive™ Initial Placement PEG Safety Kit Regulation Number: 21 CFR §876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: 78 KNT Dated: March 17, 2003 Received: March 18, 2003
Dear Ms. Sweeney:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act). You may, therefore, market the device, subject to the general controls provisions of Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any
{3}------------------------------------------------
Page 2 - Ms. Paige Sweeney
Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
In addition, we have determined that your device kit contains Iodine swabs, lubricating jelly, antibiotic ointment, and 1% Xylocaine, which are subject to regulation as drugs.
Our substantially equivalent determination does not apply to the drug component[s] of your device. We recommend you first contact the Center for Drug Evaluation and Research before marketing your device with the drug components. For information on applicable Agency requirements for marketing these drugs, we suggest you contact:
Director, Division of Drug Labeling Compliance (HFD-310) Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville. Maryland 20857 (301) 594-0101
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation, please contact the Office of Compliance at (301) 594- . Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address http://www.fda.gov/dsma/dsmamain.html.
Sincerely yours.
David A. Segner
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
KO 30855
Special 510(k) Premarket Notification New Material for Domed PEG/PEJ
Boston Scientific Corporation March 17, 2003
### INDICATIONS FOR USE STATEMENTS
| 510(k) Number<br>(if known) | To be determined | Page 1 of 1 |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Device Names | EndoVive™ Initial Placement PEG Kit<br>EndoVive™ Initial Placement Direct PEJ Kit<br>EndoVive™ Initial Placement PEG Safety Kit | |
| Indications for Use | EndoVive™ Initial Placement PEG Kit is indicated for enteral<br>nutrition directly into the stomach in both pediatric and adult patients<br>who are unable to consume nutrition by conventional means. | |
| | The EndoVive™ Initial Placement Direct PEJ Kit is indicated for<br>nutritional support and decompression directly into the jejunum when<br>feeding via the upper gastrointestinal tract is contraindicated. | |
| | The EndoVive™ Initial Placement PEG Safety Kit is indicated for<br>providing nutrition directly into the stomach in patients who are<br>unable to consume nutrition by conventional means. The safety<br>components of the kit are designed to reduce the potential for<br>inadvertent sharps injury to medical personnel during and after the<br>procedure. | |
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------~ = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = =
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109) OR
Over the Counter Use _________________________________________________________________________________________________________________________________________________________
David R. Lehman
(Division Sign-Off) Division of Reproductive, Ab and Radiological Devices 510(k) Number
000101
Proprietary and Confidential Information of Boston Scientific Corporation
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.