PROTOCO2L COLON INSUFFLATOR WITH PERFORMANCE IMPROVEMENTS, MODEL 6400
Applicant
E-Z-Em, Inc.
Product Code
FCX · Gastroenterology, Urology
Decision Date
Apr 16, 2003
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.1500
Device Class
Class 2
Indications for Use
The E-Z-EM PROTOCO2L COLON Insufflator administers and regulates carbon dioxide (CO2) as a distention media to the colon during CT Colonography (CTC or Virtual Colonoscopy) and conventional Colonoscopy.
Device Story
PROTOCO2L Colon Insufflator is an electro-mechanical pneumatic system used to regulate CO2 flow, volume, and pressure for colonic distention during CT colonography or conventional colonoscopy. The device consists of an insufflator unit, consumable tubing set, and accessory cart. It uses gas metering technology to provide digital feedback on colon pressure and distention volume. Performance improvements include a staged start-up flow ramp (1.0 L/min to 3.0 L/min based on volume), automated volume delivery increments (4L initial, 2L boluses thereafter), and enhanced pressure relief (electronic relief at 50 mmHg for 5 seconds; independent mechanical relief at 75 mmHg). The device is used in clinical settings, typically adjacent to a CT gantry, and is operated by healthcare professionals. The output (pressure/volume data) allows clinicians to monitor distention in real-time, ensuring safe and effective bowel preparation for imaging or endoscopic procedures.
Clinical Evidence
Bench testing only. No clinical data provided. Performance specifications were validated through comparative bench testing against the predicate device, focusing on flow rate accuracy, pressure relief valve activation, and volume delivery increments.
Technological Characteristics
Electro-mechanical pneumatic system. Materials: same as predicate. Sensing: digital pressure and volume monitoring. Energy: 100/240 VAC 50/60 Hz. Connectivity: standalone. Standards: IEC/EN 60601-1 (Safety), IEC/EN 60601-1-2 (EMC), UL 2601-1/CSA C22.2 No. 601.1. Features: hydrophobic filter, electronic and redundant mechanical pressure relief valves, 7-segment LED displays for pressure and volume.
Indications for Use
Indicated for patients undergoing CT Colonography or conventional colonoscopy requiring colonic distention via CO2 administration.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows a logo with a globe on the left and the text "ECMI" on the right. Below the logo is the text "Visualize a healthier world". The globe is a wireframe design, and the text "ECMI" is in a bold, sans-serif font.
## PROTOCO2L™ Insufflator With Performance Improvements 510(k) Summary
APR 1 6 2003
| Submitter's Information: | E-Z-EM, Inc. |
|--------------------------|---------------------------------------------------------------|
| | 717 Main Street |
| | Westbury, NY 11590 |
| Telephone Number: | (516) 333-8230, ext. 2377 |
| Fax Number: | (516) 997-0282 |
| Contact: | Judy Hauser, Sr. Manager Regulatory Affairs |
| Date Prepared: | March 13, 2003 |
| Trade Name: | PROTOCO₂L™ Colon Insufflator with Performance<br>Improvements |
| Common Name: | Insufflator, Colonic |
| Classification Name: | Endoscope and accessories |
| | 21 CFR 876.1500 |
| | 78 FCX |
| Predicate Device: | PROTOCO₂L™ Colon Insufflator |
### Indications for Use:
The E-Z-EM PROTOCO2L COLON Insufflator administers and regulates carbon dioxide (CO2) as a distention media to the colon during CT Colonography (CTC or Virtual Colonoscopy) and conventional Colonoscopy.
K013219
### Device Description:
The EZEM PROTOCO>L with performance improvements, as with the predicate device is an insufflation system that is used in conjunction with a consumable and is intended to displace and regulate carbon dioxide (CO2) as a distention media to a patient's colon for purposes of CT Colonography and conventional colonoscopy. As with the predicate, the PROTOCO2L system with performance improvements consists of 3 components. These are the insufflator, consumable and accessory cart.
The performance improvements presented here do not affect or change the current indications and contraindications of the PROTOCO2L device. A summary of the performance specification inprovements as compared to the predicate performance specifications are stated as follows:
{1}------------------------------------------------
K c3c854 pg 2 of 11
| Proposed Performance Specification | Predicate Performance Specification |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Start-Up Ramp | Start-Up Ramp |
| At the onset of the procedure, flow of CO2 to<br>the patient to be ramped to the patient in<br>accordance with the following increments: | At the onset of the procedure, flow of CO2<br>to the patient initiates itself to the<br>specified flow rate of: |
| 1.0 Liters/Minute for 0 to 0.5 Liter of<br>Delivered Volume | 3 Liters/Minute |
| 2.0 Liters/Minute for greater than 0.5 to 1.0<br>Liter Delivered Volume | |
| 3.0 Liters/Minute for Delivered Volumes<br>Greater than 1.0 Liter | |
| | |
| Pressure Relief | Pressure Relief |
| Electronic Pressure Relief | Electronic Pressure Relief |
| 50 mm Hg for 5 seconds (absolute) | 6 mm Hg-second (relative) |
| Independent Redundant Mechanical Relief | Independent Redundant Mechanical<br>Relief |
| 75 mm Hg | 50 mm Hg |
| Volume Delivery | Volume Delivery |
| Upon Pressing the RUN/STOP Button at the<br>start of the procedure, the unit will remain in<br>run mode until 4 Liters of CO2 have been<br>delivered. Once 4 Liters has been delivered,<br>the unit automatically returns to stop mode. | Upon pressing the RUN/STOP button at<br>the start of the procedure, the unit will<br>remain in run mode indefinitely until the<br>user presses this button a second time to<br>stop CO2 delivery. |
| Thereafter pressing the RUN/STOP button a<br>2nd time will resume the delivery of CO2 for<br>an additional 2 Liters of CO2. Once an<br>additional 2 Liters has been delivered, the unit<br>automatically returns to stop mode. | Thereafter pressing the RUN/STOP<br>button a 3rd time will resume the delivery<br>of CO2 indefinitely until the user presses<br>the button to stop CO2delivery. |
{2}------------------------------------------------
R030854 pg 3 of 11
Subsequent presses of the RUN/STOP button will deliver additional 2 Liter boluses of CO2 as required by the user.
The user can always depress the RUN/STOP button during any of these volume increments to STOP delivery of CO2.
As with the predicate PROTOCO2L design, the PROTOCO2L with performance improvements unit is still based on the currently marketed Northgate 3315/6600 Insufflator product that is designed and manufactured by Northgate Technologies (OEM manufacturer), Elgin, Illinois (Model No. 6400). Other than the performance improvements outlined here, there are no changes to the insufflator cart and tubing set as prescribed by the original 510(k) for this device.
## Comparative Information Regarding Substantial Equivalence:
The EZEM PROTOCO-L Colon Insufflator with performance improvements and the original PROTOCO>L Colon Insufflator are indicated for distention of the Colon during CT Colonography (CTC or Virtual Colonoscopy) and conventional Colonoscopy.
The proposed and predicate PROTOCO-L Colon Insufflator rely on proven and widely accepted gas metering technology to precisely regulate the flow, volume and pressure of CO2 on a continuous basis. With this technology the user has direct digital feedback on the actual colon pressure and distention volume for the duration of the procedure.
In summary both the proposed PROTOCO2L Colon Insufflator with performance improvements and the predicate PROTOCO2L Colon Insufflator:
- Use CO2 as a colon distention media for radiographic exams ●
- Use electronically controlled metering of CO2
- Have the same operating principle and fundamental scientific technology .
- Allow for the adjustment of distention pressure ●
- Have the same maximum delivery pressure .
- Have the same volume delivered display ●
- . Have the same digital display of pressure and distention volume
- Use the same hydrophobic filter technology to prevent contamination of the device .
- Are made from the same materials ●
- Specify use of an enema tip (rectal catheter) for colon access .
Where their significant differences are:
- The Flow Rate for the initial volume of CO2 is gradually increased to 3 Liters/Minute versus initially ● delivering at 3 Liters/Minute.
E-Z-EM PROTOCO2L Insufflator 510(k) Submission Page 3 of 11
{3}------------------------------------------------
Ko 30 854 Jag 4 8 11
- During delivery, pre-defined volume increments at 4 liters of CO2 and thereafter 2 liter volumes are ● administered versus continuous volume delivery until the operator intervenes.
- Electro-pneumatic pressure relief valve is at 50 mmHg for 5 seconds versus a relative 6 mm Hg -. second pressure relief relative to set pressure.
- Independent redundant mechanical pressure relief at 75 mm Hg. .
A comparison between the EZEM PROTOCO2L with software modifications and the predicate PROTOCO2L device is presented in the following tables.
{4}------------------------------------------------
K 030 854 Pg. 5 of 4
# Comparative Summary Tables
| | Proposed Device | Predicate Device |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Insufflator Unit, General | E-Z-EM PROTOCO2L WITH<br>PERFORMANCE<br>IMPROVEMENTS | E-Z-EM PROTOCO2L |
| Indication for Use | Administer and regulate CO2 as a<br>distention media for the colon during<br>CT Colonography (CTC or Virtual<br>Colonoscopy)<br>and<br>conventional<br>colonoscopy. | Same as proposed |
| Design | Electro-Mechanical Pneumatic<br>System to regulate both Flow and<br>Pressure of CO2 | Same as proposed |
| Anatomical Sites | Rectal administration of CO2 via<br>enema tip or catheter lumen | Same as proposed |
| Radiation | No ionizing radiation emitted | Same as proposed |
| Thermal | No thermal energy introduced into<br>patient | Same as proposed |
| | Proposed Device | Predicate Device |
| Insufflator Unit,<br>General | E-Z-EM PROTOCO₂L WITH<br>PERFORMANCE<br>IMPROVEMENTS | E-Z-EM PROTOCO₂L |
| Insufflation Unit | Located Adjacent to CT Gantry | Same as Proposed |
| Power Supply | Switching Power Supply in<br>Dedicated Enclosure | Same as Proposed |
| Flow Rate | 1 to 3 Liters/Minute based on the<br>volume delivered; Flow is ramped<br>up during first Liter:<br><br>Delivery volume from 0 to 0.5<br>Liter @ 1 Liter/Minute maximum<br><br>Delivery volume greater than 0.5<br>to 1 Liter @ 2 Liters/Minute<br>maximum<br><br>Delivery volume greater than 1<br>Liter @ 3 Liters/Minute<br>maximum<br><br>Accuracy: Not to exceed + 20%<br>@ 3 Liters/ minute | 1 to 3 Liters/Minute, Fixed<br>Accuracy: Not to exceed + 20%<br>@ 3 Liters/ minute |
| Delivery Pressure | 0 to 25 mm Hg in increments of 1<br>mm Hg<br>Accuracy: +/- 10% | Same as proposed |
| | Proposed Device | Predicate Device |
| Insufflator Unit, General | E-Z-EM PROTOCO₂L WITH<br>PERFORMANCE IMPROVEMENTS | E-Z-EM PROTOCO₂L |
| Delivery Volume | 1 to 999 Liters in 0.1 Liter increments<br>Accuracy: +/- 20 % | Same as proposed |
| Pressure Limiting | Yes | Yes |
| Over-Pressure Relief | Yes | Yes |
| | Electrical Relief Valve opens when colon<br>pressure goes above 50 mmHg for<br>sustained 5 seconds<br><br>Independent redundant Mechanical Relief<br>Valve opens when colon pressure is at 75<br>mmHg | Electrical Relief Valve opens<br>when colon pressure exceeds set<br>pressure by 5 mmHg for 2 seconds<br><br>Independent redundant<br>Mechanical Relief Valve opens<br>when colon pressure is at 50<br>mmHg |
{5}------------------------------------------------
K 0.30 854 Pg 6 of "
{6}------------------------------------------------
K. 030 854 Pg 7 of ""
{7}------------------------------------------------
K 0308.54
Pg 8 of 11
| | Proposed Device | Predicate Device |
|--------------------------------------------|---------------------------------------------------|------------------|
| Insufflator Unit,<br>displays and Controls | E-Z-EM PROTOCO₂L WITH<br>PERFORMANCE IMPROVEMENTS | E-Z-EM PROTOCO₂L |
| Display | Insufflation Pressure: 2-Seven Segment<br>LEDs | Same as Proposed |
| | Volume: 3-Seven Segment LEDs | |
| | Supply Volume: Graduated LED Array | |
| Pressure | Rotary Adjustment Knob | Same as Proposed |
| Flow Rate | Fixed, No Adjustment | Same as Proposed |
| Power | Single Acting Button | Same as Proposed |
| Stop/Run | Single Acting Button | Same as Proposed |
| Volume Reset | Single Acting Button | Same as Proposed |
| Over Pressure Alarm Reset | Single Acting Button | Same as Proposed |
| Consumable Connection | Medical, CPC locking Connection | Same as Proposed |
{8}------------------------------------------------
K 030854 py. 7 9/1
| | Proposed Device | Predicate Device |
|---------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| Insufflator Unit,<br>Environmental<br>Requirements | E-Z-EM PROTOCO2L WITH<br>PERFORMANCE IMPROVEMENTS | E-Z-EM PROTOCO2L |
| Operating Temperature<br>Operating Humidity,<br>Operating Altitude &<br>Storage Temperature | Meets Requirements set forth in<br>IEC/EN60601-1, Medical Electrical<br>Equipment Part 1: General Requirements<br>for Safety | Same as proposed |
| Electromagnetic Compatibility<br>(EMC) | Meets Requirements set forth in IEC/EN<br>60601-1-2, Medical Electrical Equipment<br>Part 1: Collateral Standard,<br>Electromagnetic Compatibility<br>CISPR 11<br>IEC 801-2<br>IEC 801-3<br>IEC 801-4<br>IEC 801-5 | Same as proposed |
| UL/CSA | Medical Device Listing to UL 2601-1 / CSA C22.2 No. 601.1, Electrical Class I,<br>Type B Isolation rating | Same as proposed |
{9}------------------------------------------------
K 030 854 Pg 10 of 11
| Insufflator Unit,<br>Environmental<br>Requirements | Proposed Device<br>E-Z-EM PROTOCO2L WITH<br>PERFORMANCE IMPROVEMENTS | Predicate Device<br>E-Z-EM PROTOCO2L |
|----------------------------------------------------|----------------------------------------------------------------------|--------------------------------------|
| Energy Requirements | 100/240 VAC 50/60 Hz<br>Auto Seeking | Same as proposed |
| Shock and Vibration | International Safe Transit Authority<br>(ISTA), Project 3C | Same as proposed |
:
.
{10}------------------------------------------------
\$\qquad K_{0}30.854\$
| Insufflator Consumable | Proposed Device<br>E-Z-EM PROTOCO₂L WITH<br>PERFORMANCE IMPROVEMENTS | Predicate Device<br>E-Z-EM<br>PROTOCO₂L |
|------------------------|------------------------------------------------------------------------------------------------------------|-----------------------------------------|
| Consumable Tubing Set | E-Z-EM Designed and Manufactured non-<br>sterile tubing set using existing enema tip<br>for rectal access. | Same as<br>proposed |
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or movement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 1 6 2003
Ms. Judy Hauser Sr. Manager Regulatory Affairs E-Z-EM, Inc. 717 Main Street WESTBURY NY 11590
Re: K030854
Trade/Device Name: E-Z-EM PROTOCO,L Colon Insufflator Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: 78 FCX Dated: March 17, 2003 Received: March 18, 2003
Dear Ms. Hauser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{12}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{13}------------------------------------------------
Confidential
KO30854
# INDICATIONS FOR USE
510(k) Application: Special 510(k)
E-Z-EM PROTOCO₂L COLON INSUFFLATOR Device Name:
Indications for Use:
The E-Z-EM PROTOCO2L COLON Insufflator administers and regulates carbon dioxide (CO2) as a distention media to the colon during CT Colonography (CTC or Virtual Colonoscopy) and conventional Colonoscopy.
Concurrence of CDRH, Office of Device Evaluation (ODE)
**Prescription Use**
(Per 21 CFR 801.109)
OR
Over-the-Counter Use
Nancy C. Brogdon
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices
510(k) Number
E-Z-EM PROTOCO3L Insufflator 510(k) Submission Page 6 of 26
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.