'RAPID ONE' - PROPOXYPHENE TEST

K030835 · American Bio Medica Corp. · JXN · May 22, 2003 · Clinical Toxicology

Device Facts

Record IDK030835
Device Name'RAPID ONE' - PROPOXYPHENE TEST
ApplicantAmerican Bio Medica Corp.
Product CodeJXN · Clinical Toxicology
Decision DateMay 22, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3700
Device ClassClass 2

Indications for Use

'Rapid One'-Propoxyphene Test is a one-step lateral flow immunoassay for the qualitative detection of 300 ng/ml of propoxyphene and norpropoxyphene in human urine. 'Rapid One'-Propoxyphene Test is intended for professional use. It is not intended for i over-the-counter sales to nonprofessionals. The assay to perform, but should not be used without proper supervision. This immunoassay is a simplified, qualitative screening method that provides only a preliminary result for use in determining the need for additional or confirmatory testing, i.e. gas chromatography/mass spectrometry (GC/MS.) 'Rapid One'-Propoxyphene Test provides only a preliminary analytical result. A more specific alternate chemical method must be used in order to obtain a more confirmed result. GC/MS is the preferred confirmatory method. Clinical and professional judgment should be applied to any drug of abuse test result. Particularly when preliminary results are used.

Device Story

Lateral flow immunoassay for qualitative detection of propoxyphene/norpropoxyphene in human urine; utilizes competitive binding principle. Input: urine sample; colloidal gold-labeled antibody complex; immobilized drug conjugate on membrane. Mechanism: in absence of drug, gold-labeled antibodies bind to immobilized conjugate, forming visible test line; in presence of drug, drug competes for antibody sites, preventing test line formation. Control line confirms proper flow. Output: visual color bands (two for negative, one for positive). Used in clinical settings by professionals; provides preliminary screening results requiring GC/MS confirmation. Benefits: rapid, simplified preliminary assessment to guide clinical decision-making regarding further testing.

Clinical Evidence

Bench testing only. Reproducibility evaluated using control urines at concentrations above and below the 300 ng/ml cutoff. All concentrations verified by GC/MS. Testing performed four times daily for five days. Results confirmed reproducibility of performance.

Technological Characteristics

Lateral flow immunoassay; membrane strip with immobilized drug conjugate and dried colloidal gold-multi-antibody complex. Qualitative visual readout. No external energy source required. Standalone test device.

Indications for Use

Indicated for professional use as a qualitative screening tool for the detection of propoxyphene and its metabolite norpropoxyphene in human urine at a 300 ng/ml cutoff. Not for over-the-counter use.

Regulatory Classification

Identification

A propoxyphene test system is a device intended to measure propoxyphene, a pain-relieving drug, in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of propoxyphene use or overdose or in monitoring levels of propoxyphene to ensure appropriate therapy.

Special Controls

*Classification.* Class II (special controls). A propoxyphene test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ MAY 22 2003 K030835 ### 510 (k) Summary #### Submitter's Name/Address: American Bio Medica Corportion 122 Smith Road Kinderhook, NY 12106 Date of Preparation of this Summary: Device Trade or Proprietary Name: ## Contact Person: Henry Wells VP Product Development Phone: 410 992-4734 Fax: 410 992-0328 March 12, 2003 'Rapid One'-Propoxyphene Test Propoxyphene Test System Classification Number/Class: Classification Name: Device Common/Usual Name or {no classification regulation]/Class II This 510(k) Summary is being submitted in accordance with the requirement of 21 CFR 807.92. The assigned 510(k) number is: K030835 Predicate Device: MedTox Diagnostics, Inc. 2-Panel Propoxyphene/MAMP-MDMA Test. (510 (k) No. K002141). ## Test Description: The assay employed in the 'Rapid One'-Propoxyphene Test is based on the same principle of highly specific reactions between antigens and antibodies. This assay is a one-step, competitive, immunoassay for the detection of propoxyphene and its metabolite norpropoxyphene in human urine. The test device consists of a membrane strip onto which a drug conjugate has been immobilized and a colloidal goldmulti-antibody complex is dried at one end of the membrane. In the absence of any drug in the urine sample, the colloidal gold-antibody complex moves with the urine by capillary action to contact the immobilized drug conjugates. Antibody-antigen reactions occur forming visible lines in the 'test' area. When drug is present in the urine sample, the drug or metabolite will compete with its corresponding drug conjugate in the test area for the limited antibody sites on the colloidal gold-labeled antibody complex. If sufficient amount of drug is present, it will fill all of the available antibody binding sites, thus preventing attachment of the labeled antibody to the drug conjugate. An absence of a color band (line) in the 'test' area is indicative of a positive result. {1}------------------------------------------------ A control band (line), comprised of a different antibody/antigen reaction, is present on the membrane strip. The control line is not influenced by the presence or absence of drug in the urine, and therefore, should be present on all reactions. A negative urine will produce two colored bands, and a positive sample will produce only one band. ## Intended use: 'Rapid One'-Propoxyphene Test is used for the qualitative detection of propoxyphene and norpropoxyphene in human urine. This immunoassay is a simplified screening method that provides only a preliminary result for use in determining the need for additional or confirmatory testing. i.e. gas chromatography/mass spectrometry (GC/MS.) #### Performance Characteristics: 'Rapid One'-Propoxyphene Test will detect propoxyphene or norpropoxyphene at 300 ng/ml. Reproducibility was evaluated using control urines containing concentrations above and below the stated cut-off. Negative controls were also used. All concentrations were verified by GC/MS. Each sample was tested four times, twice daily, for five days. The results confirmed the reproducibility of the 'Rapid One'-Propoxyphene Test performance. ### Conclusion: 'Rapid One'-Propoxyphene Test is substantially equivalent to the previously cleared propoxyphene section of MedTox 2 Panel PPX/MAMP-MDMA Test (510(k) No. K002141). {2}------------------------------------------------ Image /page/2/Picture/0 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The symbol is enclosed within a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is written around the border of the circle. # DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Dr. Henry Wells Vice President Product Development American Bio Medica Corporation 9110 Red Branch Road Suite B Columbia, Maryland 21045 MAY 22 2003 Re: k030835 > Trade/Device Name: Rapid One - Propoxyphene Test Regulation Number: 21 CFR § 862.3700 Regulation Name: Propoxyphene Test System Regulatory Class: II Product Code: JXN Dated: March 14, 2003 Received: March 17, 2003 Dear Dr. Wells: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. {3}------------------------------------------------ Page 2 - If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Steven Sutman Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page of K030835 510(k) Number (if known):__ Device Name: _________________________________________________________________________________________________________________________________________________________________ Indications For Use: 'Rapid One'-Propoxyphene Test 'Rapid One'-Propoxyphene Test is a one-step lateral flow immunoassay for the qualitative detection of 300 ng/ml of propoxyphene and norpropoxyphene in human urine. 'Rapid One'-Propoxyphene Test is intended for professional use. It is not intended for i over-the-counter sales to nonprofessionals. The assay to perform, but should not be used without proper supervision. This immunoassay is a simplified, qualitative screening method that provides only a preliminary result for use in determining the need for additional or confirmatory testing, i.e. gas chromatography/mass spectrometry (GC/MS.) 'Rapid One'-Propoxyphene Test provides only a preliminary analytical result. A more specific alternate chemical method must be used in order to obtain a more confirmed result. GC/MS is the preferred confirmatory method. Clinical and professional judgment should be applied to any drug of abuse test result. Particularly when preliminary results are used. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety | Prescription Use | | |----------------------|--------| | (Per 21 CFR 801.109) | 510(k) | | | K030835 | |--|---------| | | OR | | | Over-The-Counter Use | |--|----------------------| |--|----------------------| (Optional Format 1-2-96)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...