The HEMOCHRON ® Response is intended for professional use for hemostasis management in a variety of clinical settings for the quantitative determination of an assortment of coagulation test assays including the following HEMOCHRON Whole Blood Coagulation Assays for In vitro diagnostic use only. Activated Clotting Time (ACT) - All HEMOCHRON types Activated Partial Thromboplastin Time (APTT) - All HEMOCHRON types Prothrombin Time (PT) - All HEMOCHRON types Thrombin Time (TT) Heparin Neutralized Thrombin Time (HNTT) High Dose Thrombin Time (HiTT) Fibrinogen (FIB) Protamine Dose Assay (PDA) - All HEMOCHRON types Heparin Response Time (HRT) - All HEMOCHRON types Protamine Response Time (PRT) - All HEMOCHRON types
Device Story
Portable, dual-well, microprocessor-controlled coagulation instrument; performs whole blood coagulation tests. Input: fresh or citrated whole blood in disposable test tube assays containing reagent and precision magnet. Operation: tube rotated at 37°C; fibrin clot formation displaces magnet; magnetic detectors monitor displacement; elapsed time to displacement displayed as coagulation time. Features: barcode reader, printer, laboratory communication interface, data management (patient/QC storage, operator lockout). Software upgrade (V2.0) adds RxDx module for automated heparin/protamine dosing calculations and enhanced data management via PC software. Used in point-of-care clinical settings by healthcare professionals. Output: coagulation time in seconds; dosing calculations. Benefits: rapid hemostasis management and dosing guidance for clinical decision-making.
Clinical Evidence
Bench testing only. Laboratory correlation studies compared V1.52 and V2.00 using split-sample design across ACT, specialty, and RxDx tube groups (n=15-18 per group). Linear correlations were excellent (r=0.954-0.998). Precision testing conducted on two levels of assay-specific controls (n=10 per instrument type) showed comparable mean, SD, and CV% between versions.
Technological Characteristics
Portable, dual-well, electro-mechanical coagulation instrument. Sensing: magnetic displacement detection. Energy: electrical (microprocessor-controlled). Connectivity: barcode reader, printer, PC interface for data download. Software: V2.0 firmware with RxDx dosing module. Sterilization: N/A (instrument).
Indications for Use
Indicated for professional hemostasis management in clinical settings requiring point-of-care testing. Used for quantitative determination of coagulation assays (ACT, APTT, PT, TT, HNTT, HiTT, FIB, PDA, HRT, PRT) using fresh or citrated whole blood.
Regulatory Classification
Identification
A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.
Special Controls
*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the letters "ITC" in a bold, sans-serif font. The letters are black and are slanted to the right. The letters are large and take up most of the image. The image is simple and has a white background.
Image /page/0/Picture/1 description: The image shows a handwritten or stamped date. The top line reads "K0308/8". The second line reads "APR 02 2003". The date appears to be April 2, 2003.
### 510 (k) Summary
#### HEMOCHRON® Response Instrument/System
March 13, 2003 Prepared:
Submitted by: John Clay International Technidyne Corp. 6 Olsen Ave. Edison, NJ 08820 (732-548-5700) Ext. 265 (732-548-2325) Fax
This summary of safety and effectiveness is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
#### Device Name
| Common / Usual Name: | Whole Blood Coagulation Test System |
|----------------------|---------------------------------------|
| Product Name: | HEMOCHRON® Response Instrument/System |
#### Predicate Device
HEMOCHRON® Response (K983475)
#### Device Description
The Hemochron Response instrument/system described herein is a software upgrade (Version 2.0) to the current Hemochron Response Instrument, which has been cleared under K983475 (May 1999).
The HEMOCHRON® Response is a portable, dual-well microprocessor-controlled coagulation instrument with an integral barcode reader, laboratory communication interface and a printer designed to perform whole blood coagulation tests using fresh or citrated whole blood. The system is intended for use in many clinical settings requiring point-of-care testing. The modified Response instrument performs the same assays as the predicate instrument. There are no changes to the clot detection algorithm.
The patented clot detection mechanism is an electro-mechanical system consisting of two test wells into which disposable test tube assays are inserted. The test tube assays contain specific reagent for the test performed and a precision magnet. Immediately after adding a blood sample to the test tube and pressing the start button, the test tube is placed in the test well and is automatically rotated at a slow controlled speed and incubated at 37°C. When a fibrin clot begins to form, it causes the magnet in the test tube to be displaced. Two magnetic detectors located in the test well continuously monitor the precise position of the magnet. When a pre-determined displacement occurs, the elapsed time form the start of the test and the clot endpoint is displayed as the coagulation time in seconds.
{1}------------------------------------------------
The HEMOCHRON® Response contains extensive data management capabilities including the following:
- Patient and quality control result storage
- . Input of operator and patient identification
- Designation of quality control level ●
- Date and time stamp for all test results .
- Printing of results
- . Tagging of test results with date and time
- Operator Lockout ●
- Storage of operator identification numbers .
- . Downloading of stored patient and OC data to a PC
- Configurable date and time formats
### New Features of the Hemochron Response Instrument
The modifications to the Hemochron Response instrument include a software revision to the instrument. The major components of the software modification include the following:
- . Activation of the automated heparin and protamine dosing calculations for patient management. This system is the RxDx module and is described in a previous submission (K010193) for the Hemochron Kaolin-Activated heparin and protamine dosing assays.
- Enhanced programming capability and flexibility for the user through the use of the . instrument keypad and the Hemochron Response Data Management Software (HRDM), which is provided for use in a personal computer (PC). The use of the Hemochron Response Data Management Software has been described in the predicate device submission and operator's manual (K983475).
The user interface for the clinician remains consistent with the previously cleared system with the exception of the enhanced features described in the revised operators manual, the RxDx module and the features associated with the (HRDM).
### Statement of Intended Use
The HEMOCHRON ® Response is intended for professional use for hemostasis management in a variety of clinical settings for the quantitative determination of an assortment of coagulation test assays including the following HEMOCHRON Whole Blood Coagulation Assays for In vitro diagnostic use only.
Activated Clotting Time (ACT) - All HEMOCHRON types Activated Partial Thromboplastin Time (APTT) - All HEMOCHRON types Prothrombin Time (PT) - All HEMOCHRON types Thrombin Time (TT) Heparin Neutralized Thrombin Time (HNTT) High Dose Thrombin Time (HiTT) Fibrinogen (FIB) Protamine Dose Assay (PDA) - All HEMOCHRON types Heparin Response Time (HRT) - All HEMOCHRON types Protamine Response Time (PRT) - All HEMOCHRON types
{2}------------------------------------------------
#### Summary of Performance Data
#### Laboratory Correlations
Laboratory correlation studies were conducted between the Hemochron Response V1.52 (Current Version) and Hemochron Response V2.00 (Modified Version) instruments using a split sample design. Heparin sensitivity testing was conducted on the test tubes from each of three groups of tests, namely ACT (including ACT, KACT and P214), specialty tubes (including APTT FWB, APTT Citrate, PT FWB, PT Citrate, TT, HNTT, Fibrinogen diluted and undiluted) and RxDx tubes (including HRTB, HRTP, PDAO, PRT200, each in Celite or kaolin versions). Blood samples were drawn from four different donors and heparinized in vitro. Six HR V2.00 and six HR V1.52 were used for these laboratory studies.
> The linear correlations between the instruments were excellent. (Y=HR V2.00, X=HR V1.52) HR=Hemochron Response
| Test Assay Group | Correlation Equation | r value | n |
|------------------|----------------------|---------|----|
| ACT Tubes | $y=0.98x-9.36$ | 0.995 | 18 |
| Specialty Tubes | $y=1.04x-0.45$ | 0.954 | 15 |
| RxDx Tubes | $y=1.00x-16.32$ | 0.998 | 17 |
#### Precision Testing
Precision testing was conducted using two levels of assay-specific controls on two Response V 1.52 and two Response V2.00 (n=10 per instrument type) instruments. Representative assays from each of three groups of tests including ACT (ACT test tubes), APTT Fresh Whole Blood, HiTT (Specialty test tubes) and PRT (RxDx test tubes) were used in the test matrix.
| Level 1 | HR V2.00 | HR V1.52 | Level 2 | HR V2.00 | HR V1.52 |
|------------|----------|----------|------------|----------|----------|
| total mean | 127.5 | 130.8 | total mean | 269.3 | 261.8 |
| SD | 7.6 | 8.1 | SD | 13.8 | 9.0 |
| CV% | 6.0% | 6.2% | CV% | 5.1% | 3.4% |
| mean diff | 2.52% | | mean diff | 2.86% | |
d. CDI Na ారా (Control
### Assay: APTT FWB (Control used: O101 Normal and Abnormal)
| Level : | HR V2.00 | HR V 1.52 | Level 2 | HR V2.00 | HR V1.52 | |
|------------|----------|-----------|------------|----------|----------|--|
| total mean | 103.5 | 106.6 | total mean | 335.6 | 321.9 | |
| SD | 7.8 | 9.1 | SD | 17.9 | 16.9 | |
| CV% | 7.5% | 8.5% | CV% | 5.3% | 5.2% | |
| mean diff | 2.91% | | mean diff | | 4.26% | |
{3}------------------------------------------------
| Assay: PRT | (Control used: RxDX WBC Normal and PRT Level II) | | | | |
|------------|--------------------------------------------------|----------|------------|----------|----------|
| Level 1 | HR V2.00 | HR V1.52 | Level 2 | HR V2.00 | HR V1.52 |
| total mean | 151.8 | 156.5 | total mean | 359.5 | 376.4 |
| SD | 7.7 | 8.8 | SD | 14.9 | 10.3 |
| CV% | 5.1% | 5.6% | CV% | 4.2% | 2.7% |
| mean diff | 3.51% | | mean diff | 4.49% | |
#### Assav: PRT (Control used: PxDx WRC Normal and PPT Lavel II)
#### Assay: HiTT
(Control used: RxDx WBC Normal and HiTT Level II)
| Level 1 | HR V2.00 | HR V1.52 | Level 2 | HR V2.00 | HR V1.52 | |
|------------|----------|---------------------|------------|----------|----------|--|
| total mean | 57.1 | રેતે રહ્યું રેતા વિ | total mean | 168.6 | 183.9 | |
| SD | 8.2 | 8.0 | SD | 26.4 | 24.0 | |
| CV% | 14.4% | 13.5% | CV% | 15.7% | 13.0% | |
| mean diff | 4.19% | | mean diff | 8.32% | | |
| | | | | | | |
## Conclusion of Risk analysis and Design Controls
Based on the extensive verification and validation testing, it can be concluded that the potential risks associated with the modified system have been resolved and that safety and effectiveness of the HEMOCHRON Response Instrument/System are not compromised.
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features a stylized caduceus-like symbol with three intertwined lines, representing health and medicine. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. John Clay Director, Regulatory Affairs International Technidyne Corporation 8 Olsen Avenue Edison, New Jersey 08820
Re: k030818
> Trade/Device Name: HEMOCHRON® Response Instrument/System Regulation Number: 21 CFR & 864.5425 Regulation Name: Multipurpose System for In Vitro Coagulation Instruments Regulatory Class: II Product Code: JPA Dated: March 13. 2003 Received: March 14, 2003
APR 0 2 2003
Dear Mr. Clay:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{5}------------------------------------------------
Page 2 -
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# 510(k) Number (If Known):
Device name: HEMOCHRON® Response
#### Indications for Use:
The HEMOCHRON ® Response is intended for professional use for hemostasis management in a variety of clinical settings for the quantitative determination of an assortment of coagulation test assays including the following HEMOCHRON Whole Blood Coagulation Assays
Activated Clotting Time (ACT) - All HEMOCHRON types Activated Partial Thromboplastin Time (APTT) - All HEMOCHRON types Prothrombin Time (PT) - All HEMOCHRON types Thrombin Time (TT) Heparin Neutralized Thrombin Time (HNTT) High Dose Thrombin Time (HiTT) Fibrinogen (FIB) Protamine Dose Assay (PDA) - All HEMOCHRON types Heparin Response Time (HRT) - All HEMOCHRON types Protamine Response Time (PRT) - All HEMOCHRON types
For In Vitro Diagnostic Use Only
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Josephine Bautista
(Division Sign Off)
Division of Clinical Laboratory Devices
510(k) Number K030818
Prescription Use or Over-the- Counter Use
Per 21 CFR 801.109
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.