K030724 · Zimmer, Inc. · LZO · Jun 3, 2003 · Orthopedic
Device Facts
Record ID
K030724
Device Name
ALUMINA CERAMIC FEMORAL HEADS, 28 AND 32 MM
Applicant
Zimmer, Inc.
Product Code
LZO · Orthopedic
Decision Date
Jun 3, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Alumina Ceramic Femoral Heads are modular components designed for use in total hip arthroplasty and indicated for the following: Patients suffering from severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head, and nonunion of previous fractures of the femur; patients with congenital hip dysplasia, protrusio acetabuli, or slipped capital femoral epiphysis; patients suffering from disability due to previous fusion; patients with previously failed endoprostheses and/or total hip components in the operative extremity; and patients with acute neck fractures.
Device Story
Alumina ceramic femoral heads (28mm and 32mm) serve as modular components in total hip arthroplasty. Devices feature 12/14 bore designed to mate with titanium or cobalt-chromium alloy femoral stems via 12/14 neck taper. Intended for single-use; implanted by surgeons during hip replacement procedures. Ceramic material provides articulation surface for hip joint. Benefit includes restoration of hip function and pain relief for patients with degenerative or traumatic hip conditions.
Clinical Evidence
Bench testing only. Mechanical testing performed per FDA Draft Guidance for the Preparation of Premarket Notifications for Ceramic Ball Hip Systems. Results demonstrate equivalence to currently marketed devices and capacity to withstand in vivo loading.
Technological Characteristics
Material: Alumina (Al2O3) ceramic. Form factor: 28 mm and 32 mm diameter femoral heads with 12/14 bore. Connectivity: None (mechanical interface). Sterilization: Not specified. Software: None.
Indications for Use
Indicated for patients requiring total hip arthroplasty due to severe hip pain/disability from rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis, femoral head fracture nonunion, congenital hip dysplasia, protrusio acetabuli, slipped capital femoral epiphysis, previous fusion, failed prior endoprostheses/total hip components, or acute neck fractures.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Zimmer. The logo consists of a large, stylized "Z" inside of a circle on the left, and the word "zimmer" in bold, sans-serif font on the right. The logo is black and white.
## JUN - 3 2003
## Summary of Safety and Effectiveness
030724
page 1 of 2
| Submitter: | Zimmer, Inc.<br>P.O. Box 708<br>Warsaw, IN 46581-0708 |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Karen Cain<br>Manager, Regulatory Affairs<br>Telephone: (574) 372-4219<br>Fax: (574) 372-4605 |
| Date: | March 3, 2003 |
| Trade Name: | Alumina Ceramic Femoral Heads, 28 and 32 mm |
| Common Name: | Ceramic Femoral Head Prosthesis |
| Classification Name<br>and Reference: | Hip joint metal/ceramic/polymer semiconstrained<br>cemented or nonporous uncemented prosthesis<br>21 CFR § 888.3353 |
| Predicate Devices: | Zimmer Ceramic Femoral Heads, K905739 and<br>K914676, cleared March 7, 1991, and January 13,<br>1992, respectively. |
| Device Description: | The Alumina Ceramic Femoral Heads are single-<br>use devices manufactured from Al <sub>2</sub> O <sub>3</sub> . Both 28 and<br>32 mm diameter heads are available in three neck<br>lengths each. The 12/14 bore of the ceramic ball<br>mates with Zimmer titanium and cobalt-chromium<br>alloy femoral stems having a corresponding 12/14<br>neck taper. |
| Indications for Use: | The Alumina Ceramic Femoral Heads are modular<br>components designed for use in total hip<br>arthroplasty and indicated for the following:<br>Patients suffering from severe hip pain and<br>disability due to rheumatoid arthritis, osteoarthritis,<br>traumatic arthritis, polyarthritis, collagen disorders,<br>avascular necrosis of the femoral head, and<br>nonunion of previous fractures of the femur;<br>patients with congenital hip dysplasia, protrusio |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Zimmer Biomet. The logo consists of a stylized letter "Z" inside a circle, followed by the word "zimmer" in a bold, sans-serif font. The logo is black and white and appears to be a vector graphic.
Image /page/1/Picture/2 description: The image shows handwritten text that appears to be a combination of an identifier and page information. The text "KO 30724" is written on the top line, likely serving as a document or project identifier. Below this, the text "page 2 of 2" indicates that this is the second page of a two-page document.
Comparison to Predicate Device:
Performance Data:
acetabuli, or slipped capital femoral epiphysis; patients suffering from disability due to previous fusion; patients with previously failed endoprostheses and/or total hip components in the operative extremity; and patients with acute neck fractures.
The Alumina Ceramic Femoral Heads are substantially equivalent to the femoral heads listed above as predicate devices. Both designs are intended to function as a modular femoral head component in hip arthroplasty and are manufactured from AbO3 ceramic.
Mechanical testing was performed per the FDA Draft Guidance for the Preparation of Premarket Notifications for Ceramic Ball Hip Systems. Test results indicate that the Alumina Ceramic Femoral Heads are equivalent to devices currently on the market and capable of withstanding in vivo loading.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird with three stylized wings. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN - 3 2003
Ms. Karen Cain Manager, Regulatory Affairs Zimmer, Inc. P.O. Box 708 Warsaw, Indiana 46581-0708
Re: K030724
Trade/Device Name: Alumina Ceramic Femoral Heads, 28 and 32 mm Regulation Numbers: 21 CFR 888.3353 Regulation Names: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: II Product Codes: LZO Dated: March 3, 2003 Received: March 7, 2003
Dear Ms. Cain:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic
{3}------------------------------------------------
Page 2- Ms. Karen Cain
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
R. Mark N. Wilkerson
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
KO30724
510(k) Number (if known):
Device Name:
Alumina Ceramic Femoral Heads, 28 and 32 mm
Indications for Use:
The Alumina Ceramic Femoral Heads are modular components designed for use in total hip arthroplasty and indicated for the following:
Patients suffering from severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head, and nonunion of previous fractures of the femur; patients with congenital hip dysplasia, protrusio acetabuli, or slipped capital femoral epiphysis; patients suffering from disability due to previous fusion; patients with previously failed endoprostheses and/or total hip components in the operative extremity; and patients with acute neck fractures.
(Please do not write below this line - Continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation

(Division Sign-Off)
Division of General. Restorative
and Neurological Devices
| 510(k) Number | K030724 |
|---------------|---------|
|---------------|---------|
| Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-The-Counter Use<br>(Optional Format 1-2-96) |
|------------------------------------------|----|--------------------------------------------------|
|------------------------------------------|----|--------------------------------------------------|
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.