The intended uses of the EP-1000 Imagers is high resolution hard copy imaging of digital image source material and through the conversion of electronic signals from a wide variety of direct/indirect medical imaging modality outputs. The hardcopy output includes however is not limited to, nuclear medicine, ultrasound, CT (especially 3-D reconstruction), MRI, and Radiation Therapy planning. Images are suitable for medical image diagnosis use and referral. The system is intended for use by medical radiologists, imaging modality specialists, and communications to referring physicians.
Device Story
EP-1000 is a dry, thermal, color printer/imager for medical diagnostic images. Input: digital images via parallel (IEEE 1284) or USB 1.1 interfaces; device accepts proprietary 24-bit RGB format. Operation: uses directly-modulated discrete-element thin-layer linear thermal print head technology to perform dye-diffusion of color onto ChromaVista paper. Output: continuous tone, photographic quality color/BW hardcopy images. Used in clinical settings by radiologists and imaging specialists. Device does not process or alter image data; host applications handle pre-compensation (gamma, contrast, scaling). Hardcopy output supports medical diagnosis and referral. Benefits include high-resolution, dry-process imaging without wet chemistry.
Clinical Evidence
Bench testing only. Performance verified through electrical compliance tests, simulation, printer resolution pattern testing, and clinical studies. Results confirmed fidelity of modulation transfer function, spatial frequency response, color resolution, density response, and image field uniformity. Compliance with SMPTE resolution and contrast patterns verified.
Technological Characteristics
Thermal dye-diffusion printer; 12.4 pixels/mm (314 dpi) resolution; 256 levels of 16.7 million colors. Interfaces: Parallel (IEEE 1284) and USB 1.1. Media: ChromaVista paper. Electrical safety: UL-1950, CAN/CSA-C22.2 No 950-95. EMC: FCC Part-15, Subpart-B, Class-A. No wet chemistry.
Indications for Use
Indicated for high-resolution hard copy imaging of digital medical image data from modalities including nuclear medicine, ultrasound, CT, MRI, and radiation therapy planning for use by radiologists, imaging specialists, and referring physicians for diagnosis and referral.
Regulatory Classification
Identification
A medical image hardcopy device is a device that produces a visible printed record of a medical image and associated identification information. Examples include multiformat cameras and laser printers.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Codonics NP-1600 Series Medical Printers (K962364)
{0}------------------------------------------------
K030690
APR 3 0 2003
Image /page/0/Picture/2 description: The image shows the logo for Codonics. The logo features a stylized hexagon shape on the left, with a small line extending from the top left corner. To the right of the shape is the word "CODONICS" in bold, sans-serif font. Below the company name is the tagline "We bring the future into focus" in a smaller, italicized font.
| 510(k) Summary | | 807.92(c) |
|----------------|-------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | Submitter Information:<br>1.1 Submitter: | 807.92(a)(1) |
| | Codonics, Inc.<br>17991 Englewood Drive<br>Middleburg Heights, Ohio 44130 | |
| | 1.2 Manufacturing Facility: | |
| | Same as above | |
| | 1.3 Representative: | |
| | Not applicable at this time | |
| | 1.4 Contact: | |
| | Roderick Dayton, Phone:(440) 243-1198 / Fax : (440) 243-1334<br>17991 Englewood Drive<br>Middleburg Heights, Ohio 44130 | |
| | 1.5 Date: February 28, 2003 | |
| 2 | Device Name<br>2.1 Camera, Multiformat, Radiological | 807.92(a)(2) & 807.92(a)(3) |
| | 2.2 Classification Name: | Medical Image Hardcopy Devices |
| | Classification Number:<br>2.3 Classification Code: | 892.2040<br>LMC |
| | 2.4 Trade/Proprietary Name: | EP-1000 Medical Color Dry Imagers |
| | 2.5 Predicate Devices: | Codonics NP-1600 Series Medical Printers (Premarket<br>notification K962364) and Seiko ColorPoint 1720 (Premarket<br>notification K991282). |
### Device Description 3
807.92(a)(4)
# 3.1 Function
The EP-1000 Imagers are dry, thermal, color printer/imagers. The devices produce continuous tone, diagnostic quality B/W and color images on reflective, incident light viewed media. The images produced via dye-diffusion technology are photographic and medical color matched quality.
{1}------------------------------------------------
## 3.2 Scientific Concepts:
Digital images are input directly to industry standard parallel (IEEE 1284) or Universal Serial Bus (USB 1.1) interfaces. The EP-1000 recognizes only its own proprietary image format which employs a 24-bit, RGB encoding for each pixel to be printed. Lossy data compression is not employed. Software drivers can be written or adapted for a plurality of host imaging applications and operating systems, which can translate image data of any format to the EP-1000 format.
The EP-1000 does not process or alter the received image data in any way. Host imaging applications or drivers can be designed to pre-compensate image data with gamma, contrast, rotation, scaling, etc; but the EP-1000 can only print the image data as received.
Imaging is accomplished via directly-modulated discrete-element thin-layer linear thermal print head technology. The recording medium is heat activated dye-diffusion of color onto photographic quality ChromaVista® paper. The action of heat on the dye-diffusion media produces a precision mixing of colors, which diffuse the medium top layer. The image formation is accomplished without wet chemistry processing common to many laser film imaging systems in use today.
## 3.3 Physical And Performance Characteristics:
In the case of medical image hard copy devices, important performance characteristics, which affect the effectiveness and safety, relate to the fidelity of the modulation transfer function. Spatial frequency response, color resolution, density response and full image field uniformity combine to affect the final image. Characteristic response of thermal print head and film response must be mapped and compensated for to achieve suitable performance.
Pixel size (81µ for the EP-1000 Imagers) produces a pixel resolution of 12.4 pixels/mm or 314 dpi. The color palate produces 256 levels of 16.7 millions colors. The SMPTE resolution and contrast pattern and uniform density response function confirms quality suitable for the intended medical imaging use.
#### ব Device Intended Use: 807.92(a)(5)
- 4.1 The intended uses of the EP-1000 Imagers is high resolution hard copy imaging of digital image source material and through the conversion of electronic signals from a wide variety of direct/indirect medical imaging modality outputs. The hardcopy output includes however is not limited to, nuclear medicine, ultrasound, CT (especially 3-D reconstruction), MRI, and Radiation Therapy planning. Images are suitable for medical image diagnosis use and referral. The system is intended for use by medical radiologists, imaging modality specialists, and communications to referring physicians.
### 5 Device Technological Characteristics: 807.92(a)(6)
- 5.1 The characteristics of the EP-1000 Imagers compare substantially to the Codonics NP-1600 Series Medical Printers (Premarket notification K962364), the Codonics Horizon® Ci Medical Imager (Premarket notification K021054) with respect to color, reflective media printing, and Seiko ColorPoint 1720 (Premarket notification K991282) in system function and intended uses. The technology and applications are substantially equivalent to models of printers already cleared to market by the FDA. Differences of note do not affect safety and effectiveness of the device, intended use, or application methods. The device operates in a manner substantially equivalent to other cleared devices in this category.
{2}------------------------------------------------
### 6 Testing and Equivalence:
## 807.92(b)(1), 807.92(b)(2) & 807.92(b)(3)
6.1 In the code implementation, electrical compliance tests, simulation, printer resolution pattern testing, and clinical studies, results and outcomes have been thoroughly reviewed with proper operation and intended functions verified. The device passed a series of electrical safety tests including UL-1950 and CAN/CSA-C22.2 No 950-95. The devices comply with electromagnetic standards of FCC Part-15, Subpart-B, Class-A. Laboratory tests have documented expected results consistent with predicate devices currently in commercial distribution, particularly with regard to the predicate NP-1600.
Codonics believes the EP-1000 Imagers to be substantially equivalent to Medical Image Hardcopy Devices currently in commercial distribution in the U.S. We have selected the Codonics NP-1600 Series Medical Printers (Premarket notification K962364) and the Seiko ColorPoint 1720 (Premarket notification K991282) as the predicate devices for our claim of substantial equivalence, Attachment 6 contains information describing these predicate devices and provides a comparison of the EP-1000 Imagers to the predicate device(s) and describes how any differences of note are substantially equivalent.
### Hazard Analysis and Safety Concerns 7
7.1Hazard analysis on this product has been performed throughout the product concept and testing phases of the product development and implementation. This process has emphasized:
- Identification of potential hazards, their causes and their effects ●
- . Development of methodologies to control the occurrence of hazards and to constrain their effects;
- . Determine any effect on patient safety and system effectiveness
The potential hazards associated with this product are not different than those of other hardcopy image components. These are primarily related to the failure of computer system components, and may be variously obviated by decisions taken by the end users of the product. None of the failures are expected to materially contribute to patient death or injury.
It is our conclusion that no hardware or software component, operating in a properly configured environment, whose latent design defect would be expected to result in death or injury of the patient. Thus the "level of Concern" is "Minor".
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three heads facing to the right. The eagle is surrounded by a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 3 0 2003
Mr. Rick Dayton Director of Quality Codonics, Inc. 17991 Englewood Drive MIDDLEBURG HEIGHTS OH 44130 Re: K030690 Trade/Device Name: EP-1000 Medical Color Dry Imager Regulation Number: 21 CFR 892.2040 Regulation Name: Medical image hardcopy device Regulatory Class: II Product Code: 90 LMC Dated: March 4, 2003 Received: March 5, 2003
Dear Mr. Dayton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________
EP-1000 Medical Color Dry Imagers Device Name:
EP-1000 Models:
Indications For Use:
The intended uses of the EP-1000 Imagers is high resolution hard copy imaging of digital image source material and through the conversion of electronic signals from a wide variety of direct/indirect medical imaging modality outputs. The hardcopy output includes however is not limited to, nuclear medicine, ultrasound, CT (especially 3-D reconstruction), MRI, and Radiation Therapy planning. Images are suitable for medical image diagnosis use and referral. The system is intended for use by medical radiologists, imaging modality specialists, and communications to referring physicians.
The intended uses are identical to the Codonics NP-1600 Series Medical Printers (Premarket notification K962364) as a color, reflective media imager. The intended uses are identical to the Seiko ColorPoint 1720 (Premarket notification K991282) in terms of color, reflective media output.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use _V OR Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109)
(Optional Format 1-2-96)
David A. Legron
(Division Sian-Off) Division of Reproductive. A and Radiological Devices 510ki Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.