← Product Code [DQY](/productcode/DQY) · K030671

# MEDTRONIOC 7F LAUNCHER BIOPSY GUIDE CATHETER, MODEL LA7RVBB, MEDTRONIC 8F LAUNCHER BIOPSY CATHETER, MODEL LA8RVBB (K030671)

_Medtronic Vascular · DQY · Jul 31, 2003 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K030671

## Device Facts

- **Applicant:** Medtronic Vascular
- **Product Code:** [DQY](/productcode/DQY.md)
- **Decision Date:** Jul 31, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Medtronic 7F and 8F Launcher Biopsy guide Catheters provide a pathway through which therapeuti devices are introduced. The guiding catheter is intended to be used to for the percutaneous introduction of biopsy devices.

## Device Story

The Medtronic 7F and 8F Launcher Biopsy Guide Catheter is a percutaneous intravascular catheter used to provide a pathway for the introduction of therapeutic devices, specifically biopsy forceps, into the coronary or peripheral system, including the right ventricle. The device consists of a braided proximal shaft, an inner liner, a secondary segment (8F only), a primary segment, a sleeve, and a soft distal tip with a lubricious inner coating. It is operated by a physician in a clinical setting to facilitate access for biopsy procedures. The device is provided in 7F and 8F sizes with specific curve styles to assist in navigation. By providing a stable pathway, it enables the physician to deliver biopsy instruments to the target site, aiding in diagnostic tissue sampling.

## Clinical Evidence

Bench testing only. Biocompatibility testing was successfully performed on all materials. Performance testing verified that the catheters meet all applicable standards and specifications for their intended use.

## Technological Characteristics

Braided proximal shaft, inner liner, secondary segment (8F), primary segment, sleeve, and soft distal tip. Inner lumen features a thin lubricious coating. Available in 7F and 8F sizes. Sterilization method not specified. No software or electronic components.

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- 7F Launcher Guide Catheter ([K022764](/device/K022764.md))
- 8F Launcher Guide Catheter ([K023402](/device/K023402.md))
- SHERPA, SHERPA II (commercialized as Vector) and ASCENT biopsy Guide Catheters ([K944668](/device/K944668.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## JUL 3 1 2003

| VI. | 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS<br>(Pursuant to Section 12, Safe Medical Devices Act of 1990) |                                                                                                                                                                                                                                            |
|-----|----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| A.  | Submitter Information:                                                                                   |                                                                                                                                                                                                                                            |
|     | Name:                                                                                                    | Medtronic, Inc.                                                                                                                                                                                                                            |
|     | Address:                                                                                                 | 37A Cherry Hill Drive, Danvers, MA 01923                                                                                                                                                                                                   |
|     | Phone:                                                                                                   | 978-777-0042                                                                                                                                                                                                                               |
|     | Fax:                                                                                                     | 978-777-0390                                                                                                                                                                                                                               |
|     | Contact Person:                                                                                          | Fred Boucher, Regulatory Affairs Manager                                                                                                                                                                                                   |
|     | Date of Preparation:                                                                                     | February 28, 2003                                                                                                                                                                                                                          |
| B.  | Device Name:                                                                                             |                                                                                                                                                                                                                                            |
|     | Trade Name:                                                                                              | Medtronic® 7F and 8F Launcher Biopsy Guide<br>Catheter                                                                                                                                                                                     |
|     | Common Name:                                                                                             | Guide Catheter                                                                                                                                                                                                                             |
|     | Classification Name:                                                                                     | Intravascular Catheter / Percutaneous Catheter                                                                                                                                                                                             |
| C.  | Predicate Device Names                                                                                   | Medtronic® 7F and 8F Launcher Guide Catheters<br>Medtronic SHERPA II (Vector)Guide Catheters                                                                                                                                               |
| D.  | Device Description:                                                                                      | Medtronic® 7F and 8F Launcher Biopsy Guide<br>Catheter                                                                                                                                                                                     |
| E.  | Intended Use                                                                                             | The Medtronic 7F and 8F Launcher Biopsy guide<br>Catheters provide a pathway through which therapeuti<br>devices are introduced. The guiding catheter is<br>intended to be used to for the percutaneous<br>introduction of biopsy devices. |

#### F. Technological Characteristics Summary

- 1 . The Medtronic 7F and 8F Launcher Biopsy Guide Catheters are identical to the Medtronic 7F and 8F Launcher Guide Catheters regarding materials, components, design, packaging and sterilization.
- The indications for use are identical to the Medtronic SHERPA II 2. (commercialized as Vector) regarding the percutaneous introduction of biopsy forceps and the Medtronic Launcher Guide Catheters regarding coronary and peripheral access. They are all indicated to provide a pathway through which therapeutic devices are introduced. The Medtronic Launcher Biopsy Guide Catheter indications for use are as follows:

{1}------------------------------------------------

Medtronic Launcher Biopsy Guide Catheter is designed to provide a pathway through which therapeutic devices are introduced. This includes the introduction of biopsy forcep devices. The guiding catheter is intended to be used in accessing the coronary or peripheral system.

- 3. Similar to the Medtronic Launcher Guide Catheters the Medtronic Launcher Biopsy Catheters will be available in 7F and 8F sizes. Similar to the Medtronic SHERPA II (Vector) Biopsy Guide Catheters, the Medtronic Launcher Biopsy Catheters will be available in one specific curve style.
- 4. The Medtronic® Launcher Biopsy Guide Catheters are constructed with a braided proximal shaft with an inner liner a secondary segment (8F only), primary segment, a sleeve and a soft distal tip. The inner lumen of the catheter has a thin lubricious coating.
- న. All appropriate biocompatibility tests were successfully performed on the materials used for the Medtronic Launcher Biopsy Guide Catheter.
- 6. Results of testing verified that the Medtronic Launcher Biopsy Guide Catheters meet all applicable standards and specifications and are deemed adequate for the intended use. The guide catheters are considered to be substantially equivalent to the following device:
	- . 7F Launcher Guide Catheter (K022764)
	- 8F Launcher Guide Catheter (K023402) .
	- SHERPA, SHERPA II (commercialized as Vector) and ASCENT ● biopsy Guide Catheters (K944668)

{2}------------------------------------------------

Public Health Service

Image /page/2/Picture/2 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized symbol that resembles three overlapping human figures or abstract shapes, possibly representing people or services provided by the department.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 3 1 2003

Medtronic, Inc. c/o Mr. Fred Boucher Regulatory Affairs Manager 37A Cherry Hill Drive Danvers, MA 01923

Re: K030671

> Medtronic® 7F and 8F Launcher Biopsy Guide Catheter Regulation Number: 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: June 17, 2003 Received: June 18, 2003

Dear Mr. Boucher:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

### Page 2 -- Mr. Fred Boucher

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

D. Zuker
MP

ram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

# Attachment 2 长 63067 |

### D. Indications For Use

- Medtronic 7F and 8F Launcher Right Ventricular Biopsy Device Name: Guide Catheters
- The Medtronic 7F and 8F Launcher Right Ventricular Indications for Use: Biopsy Guiding Catheters are intended to be used for the percutaneous introduction of biopsy forceps into the right ventricle of the heart.

None Contraindications:

Concurrence of CDRH, Office of Device Evaluation (ODE)

| (Division Sign-Off)                |         |
|------------------------------------|---------|
| Division of Cardiovascular Devices |         |
| 510(k) Number                      | K030671 |

| Prescription Use<br>(Per 21 CFR 801.109) |  |
|------------------------------------------|--|
|------------------------------------------|--|

OR

| Over-the-Counter Use |  |
|----------------------|--|
|----------------------|--|

---

**Source:** [https://fda-staging.innolitics.com/device/K030671](https://fda-staging.innolitics.com/device/K030671)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com