ENVAN HEART RATE MONITOR

K030656 · Envan, LLC · DRT · Apr 8, 2003 · Cardiovascular

Device Facts

Record IDK030656
Device NameENVAN HEART RATE MONITOR
ApplicantEnvan, LLC
Product CodeDRT · Cardiovascular
Decision DateApr 8, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2300
Device ClassClass 2

Indications for Use

The EnVan Heart Rate Monitor is indicated for use to measure the heart rate of patients who require routine periodic vital signs monitoring. It is intended to provide heart rate information on a wristwatch display for interpretation by a healthcare professional in a hospital, or ambulatory care setting. It is neither designed nor intended for use in the place of cardiac (electrocardiograph) monitoring. The EnVan Heart Rate Monitor is not suitable for use in emergency rooms, intensive care or cardiac care units, or operating room or post anesthesia units where cardiac monitors, pacemakers and other types of electrical or electronic medical devices are in use. These devices may interfere with the operation of the En Van Heart Rate Monitor.

Device Story

EnVän Heart Rate Monitor measures patient heart rate for routine vital signs monitoring. System consists of conductive synthetic rubber pads (IndePad™) for signal acquisition and a digital wristwatch for processing and display. Healthcare professional wears watch to view heart rate, which updates every few seconds. Used in hospital, outpatient, or ambulatory settings. Not intended to replace ECG monitoring. Contraindicated in environments with other electronic medical devices (ER, ICU, OR) due to potential signal interference. Provides periodic heart rate data to assist clinicians in patient assessment.

Clinical Evidence

No clinical data provided; substantial equivalence based on technological and performance characteristics equivalent to predicate devices.

Technological Characteristics

Conductive synthetic rubber pads (IndePad™) for signal acquisition; digital wristwatch display; periodic heart rate update mechanism. Class II device (21 CFR 870.2300).

Indications for Use

Indicated for patients requiring routine, periodic vital signs monitoring. Not for use in emergency rooms, intensive care, cardiac care, operating rooms, or post-anesthesia units due to potential interference with other electronic medical devices.

Regulatory Classification

Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ EnVän Health Care p / /2 ## 7.0 PREMARKET NOTIFICATION 510(K) SUMMARY | Sponsor: | EnVän LLC | APR 0 8 2003 | |-----------------------|-------------------------------------------------------------|--------------| | | P.O. Box 660827 | | | | Birmingham, Alabama 35266 | | | | Telephone: (205) 824-2280 | | | | Fax: (205) 824-2278 | | | | Contact: Swaid N. Swaid, MD, FACS | | | Manufacturer: | EnVän LLC | | | | P.O. Box 660827 | | | | Birmingham, Alabama 35266 | | | | Telephone: (205) 824-2280 | | | | Fax: (205) 824-2278 | | | | Contact: Swaid N. Swaid, MD, FACS | | | Registration: | To be assigned | | | Contact Person: | Marie Marlow | | | | M Squared Associates, Inc. | | | | 719 A Street, NE | | | | Washington DC 20002 | | | | Telephone: 202-546-1262 | | | | Fax: 202-546-3848 | | | | E-mail: mmarlow@msquaredassociates.com | | | Trade Name of Device: | EnVän Heart Rate Monitor | | | Common Name: | Heart Rate Monitor | | | Classification name: | Cardiac Monitor (including cardiotachometer and rate alarm) | | | Product Code: | DRT | | | Regulation Class: | II | | | Regulation Number: | \$870.2300 | | Device Description: The EnVän Heart Rate Monitor is currently marketed as a consumer product for use as a sports training aid, and intended for use in conjunction with either exercise devices or {1}------------------------------------------------ exercise routines with no medical indications. Clearance of this 510(k) premarket notification will allow the device to be labeled for use to measure heart rate in patients who require routine, periodic vital signs monitoring. Indications for Use: The EnVan Heart Rate Monitor is indicated for use to measure the heart rate of patients who require routine periodic vital signs monitoring. It is intended to provide heart rate information on a wristwatch display for interpretation by a healthcare professional in a hospital, outpatient or ambulatory care setting. It is neither designed nor intended for use in the place of cardiac (electrocardiograph) monitoring. The EnVan Heart Rate Monitor is not suitable for use within emergency rooms, intensive care or cardiac care units, or operating room or post anesthesia units where cardiac monitors, pacemakers and other types of electrical or electronic medical devices are in use. These devices may interfere with the operation of the EnVan Heart Rate Monitor. Major Components: Heart rate signals are acquired and transmitted by conductive synthetic rubber pads (IndePad™) then processed and displayed on a digital wristwatch worn by the healthcare professional. The heart rate display on the screen is updated every few seconds, to assure an accurate measurement of heart rate. ## Basis for Substantial Equivalence Predicate Devices: Criticare Systems Vital Signs Monitor (K022435) and Masimo SET® Radical Pulse Oximeter (K992340) - The EnVan Heart Rate Monitor has indications for use equivalent to those for the heart rate . monitoring capabilities of the predicate devices. - The EnVan Heart Rate Monitor has technological characteristics, performance characteristics, ● and instructions for use equivalent to those for the heart rate monitoring capabilities of the predicate devices. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" written around it. APR 0 8 2003 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 EnVän, LLC c/o Ms. Marie Marlow President M Squared Associates, Inc. 719 A Street, NE Washington, DC 20002 Re: K030656 Trade Name: EnVän® Heart Rate Monitor Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II (two) Product Code: DRT Dated: February 28, 2003 Received: March 3, 2003 Dear Ms. Marlow: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {3}------------------------------------------------ Page 2 - Ms. Marie Marlow or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Sponsor: EnVän K030656 not yet assigned 510(k) number: Device Name: EnVän® Heart Rate Monitor ## Indications for Use: The EnVan Heart Rate Monitor is indicated for use to measure the heart rate of patients who require routine periodic vital signs monitoring. It is intended to provide heart rate information on a wristwatch display for interpretation by a healthcare professional in a hospital, or ambulatory care setting. It is neither designed nor intended for use in the place of cardiac (electrocardiograph) monitoring. The EnVan Heart Rate Monitor is not suitable for use in emergency rooms, intensive care or cardiac care units, or operating room or post anesthesia units where cardiac monitors, pacemakers and other types of electrical or electronic medical devices are in use. These devices may interfere with the operation of the En Van Heart Rate Monitor. (Division Sign-O Division of Cardiovas 510(k) Number
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...