SURGASSIST STRAIGHT LINEAR STAPLER AND STRAIGHT LINEAR 4 ROW NO KNIFE DIGITAL LOADING UNITS WITH RELOADS
K030653 · Power Medical Interventions, Inc. · GDW · Mar 31, 2003 · General, Plastic Surgery
Device Facts
Record ID
K030653
Device Name
SURGASSIST STRAIGHT LINEAR STAPLER AND STRAIGHT LINEAR 4 ROW NO KNIFE DIGITAL LOADING UNITS WITH RELOADS
Applicant
Power Medical Interventions, Inc.
Product Code
GDW · General, Plastic Surgery
Decision Date
Mar 31, 2003
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SurgASSIST™ Straight Linear Stapler and Straight Linear 4 Row No Knife Digital Loading Units® with Reloads have applications for general and endoscopic surgery in gastrointestinal, gynecological, general abdominal and thoracic surgical procedures for resection, transection, and creation of anastomoses.
Device Story
SurgASSIST™ Straight Linear Stapler and 4-Row No-Knife Digital Loading Units (DLUs) are single-patient-use surgical staplers. The device consists of a DLU attached to a flexible shaft (FlexShaft) connected to a Power Console (PC). The surgeon operates the device via a handheld electronic Remote Control Unit (RCU). The DLU features an anvil and staple cartridge that compresses tissue, forms staples, and (in stapler models) cuts tissue. The device is used in general and endoscopic surgery to secure tissue layers. Each DLU/Reload contains an electronic memory module that identifies the component type and prevents reuse of fired cartridges. The system provides mechanical power for stapling and cutting, allowing the surgeon to adjust the DLU attitude for optimal access. The output is a stapled (and optionally cut) tissue line, facilitating resection and anastomosis.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Materials: Titanium staples. Energy: Mechanical power via Power Console. Connectivity: Electronic memory module in DLU for identification and reuse prevention. Sterilization: Ethylene Oxide (ETO). Form factor: 15mm diameter DLU, axial to FlexShaft. Software: Embedded firmware for DLU identification and lockout.
Indications for Use
Indicated for general and endoscopic surgery in gastrointestinal, gynecological, general abdominal, and thoracic procedures requiring resection, transection, and creation of anastomoses. Contraindicated for tissue requiring excessive force to compress to specified heights (1.2mm-2.0mm) or tissue that compresses easily below these heights.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
SurgASSIST™ System with Straight Linear Cutter Digital Loading Units® (K020719)
Endopath ETS Flex45 No-Knife Articulating Linear Stapler (K020779)
Submission Summary (Full Text)
{0}------------------------------------------------
K030653 (1 of 7)
MAR 3 1 2003
## Special 510(k) Device Modification PREMARKET NOTIFICATION SAFETY AND EFFECTIVENESS SUMMARY
# SurgASSIST Straight Linear Stapler and Straight Linear 4 Row No Knife Digital Loading Units® with Reloads
In Accordance with 21 CFR section 807.92 Power Medical Interventions, Inc., is submitting the following safety and effectiveness summary.
- 1) Submitter Information:
Power Medical Interventions, Inc. 110 Union Square Drive New Hope, PA 18938 267-775-8151 Ph 267-775-8123 Fax
Applicant:
Barbara J. Whitman
Date of Notification: February 28, 2003
- 2) Name of Device:
| Trade Name: | SurgASSIST™<br>Straight Linear Stapler DLU with Reloads<br>and<br>Straight Linear 4 Row No Knife DLU<br>with Reloads |
|--------------|----------------------------------------------------------------------------------------------------------------------|
| Common Name: | Linear Stapler with Implantable Staples<br>and Reloads |
Classification Name: Staple, Implantable, GDW
- 3) Predicate Devices:
- A. SurgASSIST™ System with Straight Linear Cutter Digital Loading Units®, 55mm and 30mm with Blue and Green Reloads, with Titanium Implantable Staple, Power Medical Interventions, Inc., New Hope, PA. REF SLC55B, SLC55G, SLCR55B, SLCR55G, SLC30B, SLC30G, SLCR30B, SLCR30G. (K020719).
# 000014
{1}------------------------------------------------
030653 (P.2 of 7)
- B. Endopath ETS Flex45 No-Knife Articulating Linear Stapler, Ethicon Endo-Surgery, Inc., Cincinnati, OH. REF NAB, NAG. (K020779).
- 4) Device Description:
The devices described here are reloadable Straight Linear Stapler and Straight Linear 4 Row No Knife Digital Loading Units® (DLUs) with Reloads for single patient use. All have a maximum diameter of .596" and can be used within a 15 mm cannula. The DLUs are supplied pre-sterilized and ready for use upon removal from their packaging. Reloadable stapling and cutting cartridges are referred to as "Reloads".
Each DLU contains a staple-forming anvil. The anvil acts with the staple cartridge to compress and position layers of tissue in readiness for stapling and cutting. At the same time, the anvil provides support and a means for correctly forming staples while they are closed sequentially along the tissue, followed by the cutting blade, when included. The 15 mm diameter DLUs are axial to the FlexShaft, forming a straight extension to the FlexShaft, to which they are connected.
Loaded Straight Linear Stapler and Straight Linear 4 Row No Knife DLUs are used to close otomies and other common and uncommon openings by applying staples through the tissue and forming the staples to a controlled closed condition to secure the layers of tissue together. Straight Linear Stapler and Straight Linear 4 Row No Knife Reloads contain staples and the means to sequentially force staples toward the anvil. The cutting blade, when included, is advanced immediately behind the staple pushers so that tissue is sequentially stapled and then cut, proceeding from one end of the cartridge to the other.
Straight Linear Stapler DLUs, 55 mm with Blue and Green Reloads have 2 staggered rows of staples on one side of the cutter, coded Blue (1.2 mm) for normal tissue and Green (2.0 mm) for thick tissue. DLUs are available with Blue or Green Reloads installed. Individual Blue and Green Reloads are available.
Straight Linear 4 Row No Knife DLUs, 55 mm with Gray Reloads have two, double-staggered rows of staples, coded Gray (1.2 mm) for normal tissue (same as 55 mm Blue Straight Linear Cutter without the blade). DLUs are available with Gray cartridges installed. "No Knife" distinctively labeled. Individual Gray Reloads are available.
The DLUs are attached to the end of the flexible shaft assembly that contains a pair of flexible rotary drive shafts within an overall flexible tube called the FlexShaft, hereafter referred to as FS. The other end of the FS is connected to a Power Console (PC), which applies mechanical power to the drive shafts.
{2}------------------------------------------------
DLUs have all functions powered by the PC. The FS has a short steerable section at the distal end (near the attached DLU) so that the angle of attack (attitude) of the DLU can be adjusted by the surgeon to optimize patient accessibility.
The surgeon operates a DLU via a hand held electronic Remote Control Unit (RCU) that is plugged into the front panel of the PC.
The DLUs have quick attach and release means of coupling to the FS. No tools are required. DLUs are pushed onto the FS end, snapping and locking into place. To remove a DLU from the FS, a sleeve on the DLU at the junction with the FS, is rotated by hand.
DLU designs allow for attachment of Reloads, but will inhibit attachment of incompatible Reloads. Each Reload has an integral electronic memory module. This identifies the type and size of the Reload being used and prevents a used Reload from being refired.
- 5) Indications For Use
The SuraASSIST™ Straight Linear Stapler and Straight Linear 4 Row No Knife Digital Loading Units® with Reloads have applications for general and endoscopic surgery in gastrointestinal, gynecological, general abdominal and thoracic surgical procedures for resection, transection, and creation of anastomoses.
- 6) Comparison to Predicate Devices
The following table compares the subject Straight Linear Stapler and Straight Linear 4 Row No Knife Digital Loading Units® with Reloads to the previously cleared predicate Straight Linear Cutter Digital Loading Units® with Reloads (K020719).
{3}------------------------------------------------
Pow. Medical Interventions, Inc.
SurgASSIST™ Straight Linear Stapler DLUs with Reloads
Special 510(k) Device Modification PreMarket Notification, February 28, 20
Straight Linear Stapler DLU Product Features Comparison Cha
| Features & Description | Name | Predicate | Predicate | Features & Description | Predicate | Predicate | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|
| SurgASSIST™<br>Straight Linear Stapler, 55 mm<br>With Blue and Green Reloads and<br>Straight Linear 4 Row No Knife<br>Digital Loading Units® (DLUs) | SurgASSIST™<br>Straight Linear Stapler DLUs, 55 mm<br>with Blue and Green Reloads and<br>Straight Linear 4 Row No Knife DLUs | SurgASSIST™<br>Straight Linear Cutter DLUs,<br>55 mm and 30 mm with Blue and<br>Green Reloads | Ethicon Endo-Surgery, Inc.<br>Endopath ETS Flex45 No-Knife<br>Articulating Linear Stapler | SurgASSISTTM<br>Straight Linear Stapler, 55 mm<br>With Blue and Green Reloads and<br>Straight Linear 4 Row No Knife<br>Digital Loading Units® (DLUs) | SurgASSISTTM<br>Straight Linear Cutter DLUs,<br>55 mm and 30 mm with Blue and<br>Green Reloads | Ethicon Endo-Surgery, Inc.<br>Endopath ETS Flex45 No-Knife<br>Articulating Linear Stapler | |
| | | SurgASSIST™<br>Straight Linear Cutter DLUs, 55 mm and<br>30 mm with Blue and Green Reloads | Endopath ETS Flex45 No-Knife<br>Articulating Linear Stapler | Do not use the SurgASSISTTM Straight<br>Linear Stapler DLU, 55mm Blue, on any<br>tissue that requires excessive force to<br>compress to 1.5 mm or on any tissue<br>that compresses easily below 1.5 mm.<br><br>Do not use the SurgASSISTTM Straight<br>Linear Stapler DLU, 55 mm Green, on<br>any tissue that requires excessive force<br>to compress to 2.3 mm or on any tissue<br>that compresses easily below 2.3 mm.<br><br>Do not use the SurgASSISTTM Straight<br>Linear 4 Row No Knife DLU, (Gray), on<br>any tissue that requires excessive force<br>to compress to 1.5 mm or on any tissue<br>that compresses easily below 1.5 mm. | Do not use the SurgASSISTTM System<br>Straight Linear Cutter DLU, 55 mm,<br>30 mm where combined tissue<br>thickness exceeds 2.3 mm or is less<br>than 1.0 mm. | Do not use the instrument with blue<br>reload on any tissue that requires<br>excessive force to compress to 1.5 mm<br>or on any tissue that compresses easily<br>to below 1.5 mm.<br><br>Do not use on ischemic or necrotic<br>tissue.<br><br>Do not use the instruments on the aorta.<br>Do not use any linear cutter on major<br>vessels without making provision for<br>proximal and distal control.<br>Do not use the instrument on solidorgans, such as liver or spleen, where<br>attempted compression would be<br>destructive.<br>These instruments are not intended for<br>use when surgical stapling is<br>contraindicated.<br>Do not use the instruments with green<br>reload on any tissue that requires<br>excessive force to compress to 2.0 mm<br>or on any tissue that compresses easily<br>to below 2.0 mm. | |
| Manufacturer of Record | Power Medical Interventions, Inc. | Power Medical Interventions, Inc. | Ethicon Endo-Surgery, Inc | FDA Class (System) | | | |
| Contract Manufacturer | The MedTech Group<br>South Plainfield, NJ | The MedTech Group<br>South Plainfield, NJ | Ethicon Endo-Surgery, Inc | Number of Staples | SLS55B, SLSR55B - 28 staples<br>SLS55G, SLSR55G - 28 staples<br>SLS55B4, SLSR55B4 - 56 staples | SLC55B, SLCR55B - 56 staples<br>SLC55G, SLCR55G - 56 staples<br>SLC30B, SLCR30B - 32 staples<br>SLC30G, SLCR30G - 32 staples | NAB - 44 staples<br>NAG - 44 staples |
| 510(k) Clearance Numbers | Subject of this Notification | K020719 | K020779 | | | | |
| Product Codes | SLS55B, SLSR55B, SLS55G,<br>SLSR55G, SLS55B4, SLSR55B4 | SLC55B, SLCR55B, SLC55G,<br>SLCR55G, SLC30B, SLCR30B,<br>SLC30G, SLCR30G | NAB45 and NAG45 | | | | |
| Intended use | Have applications for general and<br>endoscopic surgery in gastrointestinal,<br>gynecological, general abdominal and<br>thoracic surgical procedures for<br>resection, transection, and creation of<br>anastomoses. | Have applications for general and<br>endoscopic surgery in gastrointestinal,<br>gynecological, general abdominal and<br>thoracic surgical procedures for<br>resection, transection, and creation of<br>anastomoses. | Intended for transection, resection,<br>and/or creation of anastomoses. The<br>instruments have application in multiple<br>open or minimally invasive general,<br>gynecologic, urologic (including radical<br>prostatectomy), thoracic, and pediatric<br>surgical procedures. They can be used<br>with staple line or tissue buttressing<br>materials such as bovine pericardium. | | | | |
K 030653 (P.4 of 7)
000017
{4}------------------------------------------------
Pow. Medical Interventions, Inc.
SurgASSIST™ Straight Linear Stapler DLUs with Reloads
Special 510(k) Device Modification PreMarket Notification, February 28, 2
# Straight Linear Stapler DLU Product Features Comparison Cham
K030653 (P.Sofm)
{5}------------------------------------------------
Pow.. Medical Interventions, Inc.
SurgASSIST™ Straight Linear Stapler DLUs with Reloads
Special 510(k) Device Modification PreMarket Notification, February 28, 20
# raight Linear Stapler DLU Product Features Comparison Cham
| Features & Description | SurgASSIST™<br>Straight Linear Stapler, 55 mm<br>With Blue and Green Reloads and<br>Straight Linear 4 Row No Knife<br>Digital Loading Units® (DLUs) | Predicate<br>SurgASSIST™<br>Straight Linear Cutter DLUs,<br>55 mm and 30 mm with Blue and<br>Green Reloads | Predicate<br>Ethicon Endo-Surgery, Inc.<br>Endopath ETS Flex45 No-Knife<br>Articulating Linear Stapler |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| Rows of Staples | SLS55B, SLSR55B - 2 rows<br>SLS55G, SLSR55G - 2 rows<br>SLS55B4, SLSR55B4 - 4 rows | SLC55B, SLCR55B - 4 rows<br>SLC55G, SLCR55G - 4 rows<br>SLC30B, SLCR30B - 4 rows<br>SLC30G, SLCR30G - 4 rows | NAB - 4 rows<br>NAG - 4 rows |
| Knife/Scalpel | SLS55B, SLSR55B - scalpel included<br>SLS55G, SLSR55G - scalpel included<br>SLS55B4, SLSR55B4 - NO KNIFE | SLC55B, SLCR55B - scalpel included<br>SLC55G, SLCR55G - scalpel included<br>SLC30B, SLCR30B - scalpel included<br>SLC30G, SLCR30G - scalpel included | NAB - NO KNIFE<br>NAG - NO KNIFE |
| Staple Diameter | SLS55B, SLSR55B - .21 mm<br>SLS55G, SLSR55G - .23 mm<br>SLS55B4 - .21 mm | SLC55B, SLCR55B - .21 mm<br>SLC55G, SLCR55G - .23 mm<br>SLC30B, SLCR30B - .23 mm<br>SLC30G, SLCR30G - .23 mm | NAB - .23 mm<br>NAG - .23 mm |
| Staple Height | SLS55B, SLSR55B - 1.2 mm<br>SLS55G, SLSR55G - 2.0 mm<br>SLS55B4 - 1.2 mm | SLC55B, SLCR55B - 1.2 mm<br>SLC55G, SLCR55G - 2.0 mm<br>SLC30B, SLCR30B - 1.2 mm<br>SLC30G, SLCR30G - 2.0 mm | NAB - 1.5 mm<br>NAG - 2.0 mm |
| DLU Internal Power | None | None | None |
| Digital Information | Memory module containing digital data<br>for identification, etc. | Memory module containing digital data<br>for identification, etc. | None |
| How Supplied | Sterile - Single Patient Use | Sterile - Single Patient Use | Sterile - Single Patient Use |
# 000019
03d6S3(f.6of7)
{6}------------------------------------------------
Pow
# traight Linear Stapler DLU Product Features Comparison Cha
(continued from continued from previous page)
| 1<br>1 | | |
|--------|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | |
| t<br>1 | | |
| 1 | | |
| | | |
| | | |
| | | |
| | | |
| 1 | | |
| | | |
| 1 | | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
| | | |
| ) | | |
| | | |
| | | |
| | | |
| 1 | | |
| | | |
| | | |
| | | |
| | | |
| . | | |
| | | |
| | | |
| | | |
| | | |
| Features & Description | Method of Sterilization | Digital Loading Unit® | Reloads |
|------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|--------------------------------|-----------------------------|
| SurgASSIST™™<br>Straight Linear Stapler, 55 mm<br>with Blue and Green Reloads and<br>Straight Linear 4 Row No Knife<br>Digital Loading Units® (DLUS) | Ethylene Oxide Gas (ETO) | Blister Tray<br>with Tyvek Lid | Tyvek Pouch |
| Predicate<br>SurgASSIST™™<br>Straight Linear Cutter DLUs,<br>55 mm and 30 mm with Blue and<br>Green Reloads | Ethylene Oxide Gas (ETO) | Blister Tray with<br>Tyvek Lid | Tyvek Pouch |
| Predicate<br>Ethicon Endo-Surgery, Inc.<br>Endopath ETS Flex45 No-Knife<br>Articulating Linear Stapler | Irradiation | Blister Tray with Tyvek Lid | Blister Tray with Tyvek Lid |
Image /page/6/Picture/5 description: The image contains handwritten text that appears to be an alphanumeric code or identifier. The text reads "KU30653(F.70P7)". The characters are written in a cursive style, and the text is slightly slanted.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# MAR 3 1 2003
Ms. Barbara J. Whitman Regulatory Affairs Manager Power Medical Interventions, Inc. 110 Union Square Drive New Hope, Pennsylvania 18938
Re: K030653
Trade/Device Name: SurgASSIST™ Straight Linear Stapler Digital Loading Unit® with Reloads and Straight Linear 4 Row No Knife Digital Loading Unit® with Reloads
Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: II Product Code: GDW Dated: February 28, 2003 Received: March 3, 2003
Dear Ms. Whitman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic' Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{8}------------------------------------------------
Page 2 - Ms. Barbara J. Whitman
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
(Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
Power Medical Interventions, Inc. New Hope, PA 18938
510(k) No. K K934653
Device Name:
SurgASSIST™ Straight Linear Stapler Digital Loading Unit® with Reloads And Straight Linear 4 Row No Knife Digital Loading Unit® with Reloads
### INDICATIONS FOR USE:
The SurgASSIST™ Straight Linear Stapler and Straight Linear 4 Row No Knife Digital Loading Units® with Reloads have applications for general and endoscopic surgery in gastrointestinal, gynecological, general abdominal and thoracic surgical procedures for resection, transection, and creation of anastomoses.
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Prescription Use Per 21CFR §801.109
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
Miriam C. Phrost
(Division Sign-Off) Division of General, Restorative and Neurological Devices
OR
510(k) Number_________________________________________________________________________________________________________________________________________________________________
000012
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.