DRI COTININE EIA ASSAY

K030649 · Microgenics Corp. · MRS · May 22, 2003 · Clinical Toxicology

Device Facts

Record IDK030649
Device NameDRI COTININE EIA ASSAY
ApplicantMicrogenics Corp.
Product CodeMRS · Clinical Toxicology
Decision DateMay 22, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3220
Device ClassClass 1

Indications for Use

The DRI® Cotinine Enzyme Immunoassay is intended for the Qualitative and Semi-Quantitative determination of Cotinine in human urine at a cutoff level of 500 ng/mL. This assay is intended for in vitro diagnostic use only. The assay is intended as an aid in the detection of cotinine after use or exposure to tobacco products.

Device Story

DRI® Cotinine EIA Assay is a liquid, ready-to-use homogeneous enzyme immunoassay for human urine. It utilizes mouse monoclonal anti-cotinine antibodies to detect cotinine. Principle: competitive binding between sample cotinine and G6PDH-labeled cotinine for fixed antibody sites. In the absence of sample cotinine, antibody binds labeled drug, decreasing enzyme activity. G6PDH activity is measured spectrophotometrically at 340 nm by converting NAD to NADH; absorbance change is proportional to drug concentration. Used in clinical laboratory settings. Provides preliminary analytical results; requires confirmation via GC/MS. Benefits include rapid screening for tobacco use or exposure.

Clinical Evidence

Bench testing only. Evaluated for sensitivity (LOD 34 ng/mL), linearity, specificity, precision, and accuracy. Accuracy compared to GC/MS reference method showed 98% total agreement (110 true positives, 80 true negatives). Precision studies (total and within-run) showed CVs ≤ 9.4%. Specificity testing confirmed no interference from endogenous substances, urinary pH variations, or cross-reacting compounds.

Technological Characteristics

Liquid, ready-to-use homogeneous enzyme immunoassay. Components: Antibody/Substrate Reagent and Enzyme Conjugate Reagent. Sensing principle: spectrophotometric measurement of G6PDH enzyme activity at 340 nm. Analyte: Cotinine. Cutoff: 500 ng/mL. Antibody: Mouse monoclonal.

Indications for Use

Indicated for qualitative and semi-quantitative determination of cotinine in human urine at a 500 ng/mL cutoff to aid in detecting tobacco use or exposure.

Regulatory Classification

Identification

A carbon monoxide test system is a device intended to measure carbon monoxide or carboxyhemoglobin (carbon monoxide bound to the hemoglobin in the blood) in blood. Measurements obtained by this device are used in the diagnosis and treatment of or confirmation of carbon monoxide poisoning.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K030649 # MAY 2 2 2003 # 510(K) Summary This 510(k) summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(K) number is: Unknown | Submitter Information (21 CFR 807.92(a)(1)) | | |---------------------------------------------|----------------------------------------------------------------------------------------------------| | Submitter: | Microgenics Corporation | | | 46360 Fremont Boulevard | | | Fremont, CA 94538 | | | phone: (510) 979-5023 | | | fax: (510) 979-5223 | | Contact: | Name: Lakshmi Anne | | | Title: Director of Product Development | | Summary date: | February 27, 2003 | | | Name of Device and Classification (21 CFR 807.92(a)(2)) | | Name (trade): | DRI® Cotinine EIA Assay | | Name (usual): | Cotinine Assay | | Classification: | Cotinine Enzyme Immunoassay has been placed in Class II (MKU) by the<br>Bureau of Medical Devices. | Identification of Legally Marketed Predicate Device(s) (21 CFR 807.92 (a)(3)) DRI® Cotinine EIA Assay is substantially equivalent to OTI AUTO-LYTE® Cotinine EIA Assay (OraSure Technologies, Inc, Bethlehem, PA), cleared under premarket notification K072481. DRI® Cotinine EIA Assay is identical or similar to its predicate in terms of intended use, method principle, risk to the patient, and clinical performance. Description of Device (21 CFR 807.92 (a)(4)) The DRI® Cotinine EIA Assay is a liquid ready-to-use homogeneous enzyme immunoassay. The assay uses specific antibodies, which can detect cotinine drugs in urine. The assay is based on competition between a drug labeled with glucose-6-phosphate dehydrogenase (G6PDH) enzyme, and free drug from the urine sample for a fixed amount of specific antibody binding sites. In the absence of free drug from the sample. the specific antibody binds the drug labeled with G6PDH causing a decrease in enzyme activity. This phenomenon creates a direct relationship between drug concentration in urine and enzyme activity. The enzyme G6PDH activity is determined spectrophotometrically at 340 nm by measuring its ability to convert nicotinamide adenine dinucleotide (NAD) to NADH. {1}------------------------------------------------ # Intended Use (21 CFR 807.92 (a)(5)) «The DRI® Cotinine Enzyme Immunoassay is intended for the Qualitative and Semi-Quantitative determination of Cotinine in human urine at a cutoff level of 500 ng/mL. This assay is intended for in vitro diagnostic use only. The assay is intended as an aid in the detection of cotinine after use or exposure to tobacco products." ### Similarities to the Predicate(s) (21 CFR 807.92 (a)(6)) A summary table of the similarities and differences between DRI® Cotinine EIA Assay and the predicate device follows. | Device Name | AUTO-LYTE® Cotinine EIA | DRI® Cotinine EIA Assay | |---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | (K072481) | | | Indications for<br>Use | "The OraSure Technologies, Inc. (OTI)<br>AUTO-LYTE® Cotinine EIA is a<br>homogeneous immunoassay intended for<br>the Qualitative and Semi-Quantitative<br>analysis of cotinine in human urine.<br>Cotinine is a metabolite of nicotine, and<br>the OTI kit is used as an aid in the<br>detection of cotinine after use of tobacco<br>products or other products containing<br>nicotine. FOR IN VITRO DIAGNOSTIC<br>USE." | "The DRI® Cotinine Enzyme<br>Immunoassay is intended for the<br>Qualitative and Semi-Quantitative<br>determination of Cotinine in human urine<br>at a cutoff level of 500 ng/mL. This<br>assay is intended for in vitro diagnostic<br>use only. The assay is intended as an aid in<br>the detection of cotinine after use or<br>exposure to tobacco products." | | Method<br>Principle | The OTI AUTO-LYTE® Cotinine EIA is a<br>homogeneous enzyme immunoassay<br>used for the analysis of cotinine in urine.<br>The assay is based on competition<br>between drug in the sample and drug<br>labeled with the enzyme glucose-6-<br>phosphate dehydrogenase (G6PDH) for<br>antibody binding sites. Enzyme activity<br>decreases upon binding to the antibody;<br>free drug in the sample is proportional to<br>enzyme activity. Active enzyme converts<br>oxidized nicotinamide adenine<br>dinucleotide (NAD) to NADH, resulting<br>in an absorbance change that is measured<br>spectrophotometrically at 340 nm. | The DRI® Cotinine Assay is a liquid,<br>ready-to-use homogeneous enzyme<br>immunoassay. The assay is based on<br>competition between cotinine labeled<br>with glucose-6-phosphate dehydrogenase<br>(G6PDH) enzyme and free cotinine from<br>in the sample for a fixed amount of<br>cotinine-specific antibody binding sites.<br>The enzyme G6PDH activity is<br>determined spectrophotometrically at 340<br>nm by measuring its ability to convert<br>nicotinamide adenine dinucleotide (NAD)<br>to NADH. | | Components | - Reagent A<br>- Reagent B | - Antibody/Substrate Reagent<br>- Enzyme Conjugate Reagent | | Anti-Cotinine<br>Antibody | Sheep Polyclonal | Mouse Monoclonal | | Risk to patient | Not included in package insert. | This assay provides only a preliminary<br>analytical result. A more specific<br>alternate chemical method must be used<br>in order to obtain a confirmed analytical | # Comparison Table: AUTO-LYTE® Cotinine EIA (K072481) and DRI® Cotinine EIA Assay {2}------------------------------------------------ | Device Name | AUTO-LYTE® Cotinine EIA (K072481) | DRI® Cotinine EIA Assay | |-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------| | | | result. Gas chromatography/mass<br>spectrometry (GC/MS) is the preferred<br>confirmatory method. | | Clinical<br>Performance | Accuracy: Accuracy against a Carbon<br>Monoxide Monitor reference method<br>indicated a relative sensitivity of 93.6%<br>and a relative specificity of 74.0% | Accuracy: Accuracy against a GC/MS<br>reference method was 98% (110 true<br>positives, —80 true negatives) | | | Total Imprecision: Percent rate CVs<br>across 4 levels of analyte concentration<br>(300 ng/mL-5000 ng/mL) were ≤ 1.09%. | Total Imprecision: Percent rate CVs<br>across 3 levels of analyte concentration<br>(300 ng/mL, 500 ng/mL, and 700 ng/mL<br>were ≤ 0.8%.) | ### Brief Discussion of Nonclinical/Clinical Data (21 CFR 807.92(b)(1, 2)) The DRI® Cotinine EIA Assay was evaluated via a series of traditional laboratory studies. These studies included the performance characteristics of sensitivity, linearity, specificity, precision, and accuracy. The assay showed good sensitivity with an LOD of 34 ng/mL. Precision studies indicated good reproducibility of results at the critical points of the measurement range (distinguishing positive from negative interpretations), as dose %CVs for both total and within-run testing were ... 9.4%. Accuracy studies showed good performance of the DRI® Cotinine EIA Assay as compared to the GC/MS reference method. The % Total Agreement is 98%. Specificity testing demonstrated that the DRI® Cotinine EIA Assay is not affected by common endogenous substances, variations in urinary pH levels, structurally unrelated pharmaceutical compounds, or potentially cross-reacting compounds. #### Performance Data - Conclusions (21 CFR 807.92 (b)(3)) The DRI® Cotinine EIA Assay has been shown to be substantially equivalent to the predicate device. and safe and effective for its intended use. {3}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image is a logo for the Department of Health & Human Services. The logo features a stylized image of an eagle with three lines representing its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. Public Health Service Food and Drug Administratio 2098 Gaither Road Rockville MD 20850 Ms. Lakshmi Anne Director of Product Development Microgenics Corporation 46360 Fremont Boulevard Fremont, CA 94538 MAY 2 2 2003 Re: k030649 Trade/Device Name: DRI® Cotinine EIA Assay Regulation Number: 21 CFR 862.1150 Regulation Name: Calibrator Regulatory Class: Class II Product Code: JIT, MKU Dated: February 27, 2003 Received: February 28, 2003 Dear Ms. Anne: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {4}------------------------------------------------ #### Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Steven Sutman Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ 510(K) Number (if known): DRI® Cotinine Enzyme Immunoassay Device Name: Indications For Use: "The DRI® Cotinine Enzyme Immunoassay is intended for the Qualitative and Semi-Quantitative determination of Cotinine in human urine at a cutoff level of 500 ng/mL. This assay is intended for in vitro diagnostic use only. The assay is intended as an aid in the detection of cotinine after use or exposure to tobacco products." Sean Cooper Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K030649 (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...