K030636 · Clinisurg Vascular Products, LLC · LJT · Jun 5, 2003 · General Hospital
Device Facts
Record ID
K030636
Device Name
C-PORT
Applicant
Clinisurg Vascular Products, LLC
Product Code
LJT · General Hospital
Decision Date
Jun 5, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5965
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The C-Port family of products will be used as a subcutaneously implanted device where repeated access to the vascular system is the therapy of choice for delivery of medications, fluids, nutritional solutions and blood products or withdrawal of blood.
Device Story
C-Port is a subcutaneously implantable vascular access device consisting of a titanium chamber with a silicone septum and a silicone or polyurethane catheter. The device is implanted by a physician in the soft tissue near the clavicle; the catheter is tunneled to a large vessel and positioned in the superior vena cava. The system provides a port for repeated needle access via Huber needles for fluid infusion or blood withdrawal. The kit includes standard surgical accessories (syringes, trocars, guide wires, sheaths). The device benefits patients by providing a reliable, long-term route for therapy, reducing the need for repeated venipuncture. Healthcare providers access the port through the skin; output is the delivery of fluids or withdrawal of blood. Clinical decision-making is supported by the ability to administer necessary therapies or obtain samples as required by the patient's condition.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is based on bench testing, including catheter-to-port connection, septum puncture, port leak, fluid dynamics, biological evaluation (toxicity), and catheter tensile/elongation strength testing.
Technological Characteristics
Titanium chamber/septum retainer, silicone septum, silicone rubber or polyurethane catheter. Energy source: None (mechanical). Connectivity: None. Sterilization: Ethylene oxide gas. Form factor: Subcutaneous port with catheter. Algorithm: None.
Indications for Use
Indicated for patients requiring repeated vascular access for delivery of medications, fluids, nutritional solutions, blood products, or blood withdrawal. Suitable for adults and pediatric patients.
Regulatory Classification
Identification
A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.
{0}------------------------------------------------
C-Port and Catheter Kits
K030636
Section 10
510(k) Summary
## CliniSurg Vascular Products 5454 Amesbury Dr. #2107 Dallas, TX 75206
## CliniSurg Vascular Products Premarket Notification 510(k) Summary for the C-Port Family of Implanted Vascular Access Devices
## Part l General Information
| 1. | Submitter name: | CliniSurg Vascular Products, LLC |
|-----|-----------------------------------|----------------------------------------------------------------------|
| 2. | FDA Establishment Number: | (number to be determined) |
| 3. | Address: | 5454 Amesbury Dr. #2107<br>Dallas, TX 75206 |
| 4. | Phone: | 214-691-5935<br>(cell) 214-549-9208 |
| 5. | Fax: | Not available at this filing |
| 6. | Contact person: | Donald F. Hults |
| 7. | Date prepared: | February 14, 2003 |
| 8. | Device trade or proprietary name: | C-Port (family of Vascular Access Devices) |
| 9. | Device common name or usual name: | Implantable Infusion Ports |
| 10. | Device classification name: | Port and catheter, implanted, subcutaneous, intravascular ( 85-LJT ) |
{1}------------------------------------------------
#### Device classification: 11.
The device is a Class II Device
#### 12. Substantial equivalency:
Claimed against the following device(s):
- Titan Port or Vortex Port, from Norfolk Medical Products, now . Horizon Medical Products since November 1998 (#K830000 -1983).
#### 13. Type of medical device:
The CliniSurg family of Vascular Access devices is a group of subcutaneously implantable ports with a catheter either pre-attached or attachable for application by physicians in indicated therapies.
#### 14. Description of Device:
#### CAUTION: CAUTION: The device is to be used by or on the order of a physician.
The device has a port that is a titanium chamber with a silicone membrane designed for repeated needle insertion. The port comes in various sizes to enable use with larger adults or smaller pediatric patients or the more emaciated adult patient. A higher or lower profile device is chosen by patient evaluation. The port is connected to a catheter that is long enough to insert into the vena cava for positioning to enable fluid infusion into larger vessels. The catheter is fixed to the port with a hub either during a procedure to implant the device or it comes pre-assembled (physician choice for use). The catheter has a series of radiopaque marks to enable depth determination when the catheter is inserted into the vena cava. A kit is provided to aid in catheter placement, insertion, and port implantation. Items such as syringes, vein introducers, trocars, guide wires, sheaths, fixation hubs, Huber needles, and infusion sets may be part of a total kit.
#### 15. Indications for use statement and product function:
The C-Port family of Vascular Access Devices will be used as a subcutaneously implanted device where repeated access to the vascular system is the therapy of choice for delivery of medications, fluids, nutritional solutions, and blood products or withdrawal of blood.
The C-Port family of ports and catheters provides a simple method for delivery of volumes of medications and fluids via a chamber leading to a catheter and opening into a large vessel in the body. The catheter is inserted into a large vessel and terminated in the superior vena cava. The port is implanted subcutaneously in the soft tissue near the clavicle on the patient's right upper chest wall. Medications or fluids can then be provided as necessary.
{2}------------------------------------------------
#### 16. Contraindications or cautions for use:
A complete listing of the possible complications in the implantation and use of the C-Port family of ports is listed in the USER'S MANUAL which is supplied with the product. Possible complications in the use of the C-Port include, but are not limited to, infection, erosion, extrusion of the device, hematomas, clot formation, thrombosis, catheter fragmentation, and embolization, and occlusion. Improper placement of the catheter in the body has been shown to cause the catheter to be severed from a "pinching" effect by the clavicle and the first rib. In the placement of the catheter into the vein and on into the superior vena cava, caution should be exercised to be sure that the catheter does not pass between the clavicle and the first rib.
#### 17. Methods (ways) of application:
The method of application is to prepare and insert the catheter into the vein (by "cut down" method or "percutaneous" method) and on into the vena cava. Then the proximal end of the catheter is tunneled subcutaneously to an area of cut down where the port is to be placed beneath the skin and secured to the fascial lavers of tissue. The catheter is ioined to the port (in attachable models) and the port is secured to the tissue. All wounds are then closed normally.
Intravenous fluids and / or nutritional fluids may then be administered by needle puncture of the septum in the port or periodic blood samples may be acquired if appropriate flushing techniques are followed (instructions for this are found in the USER'S MANUAL).
#### 18. Special precautions for disposal of the device or emptied packages (container):
The sharps used in the procedure should be disposed of according to institution policy. Any remaining parts or empty packages do not require special handling when disposing.
#### 19. Information on sterilization method(s):
#### a) For sterile products:
All C-Port family products that are manufactured by CliniSurg Vascular Products are sterilized by ethylene oxide gas. Sterilization details for validation. verification, bioburden, etc. can be found in the section on Risk Assessment (7.0) and Sterilization (8.0) in the CliniSurg Vascular Products 510(k).
#### For non-sterile products: b)
Not applicable. The C-Port family of products (ports and catheter kits) are packaged sterile. Any recognized open or damaged packaging should be noted, instructions in the User's Manual followed, and the product returned for evaluation and potential repackaging and re-sterilization or disposal.
{3}------------------------------------------------
#### Validity period: 20.
As long as the package/product is unopened and undamaged the product is valid. The materials that make up the components of the C-Port family of products do not deteriorate over time.
#### Special precautions for handling and transportation: 21.
Keep the product dry and do not open the package.
#### 22. Description of package:
A C-Port product (port and catheter) is a basic cut-down set and is packaged in a plastic tray with:
- 1-Huber Point Needle 1)
- 2) 1-Huber Needle Infusion Set
- 1-Vein pick (retracting device) 3)
- 1-Tunneling Trocar 4)
- 5) 1-Blunt Needle
The plastic tray is heat sealed in a polyethylene/nylon Tyvek header bag (pouch) and sterilized. If requested, a percutaneous introducer kit is packaged with the C-Port product. This Port/Introducer Set is listed with its own catalog numbers and must be ordered as such. The percutaneous introducer kit contains:
- 1-Sheath 1)
- 2) 1 - Dilator
- 3) 1-Guide Wire
- 4) 1-Svringe
- 5) 1-Needle
In addition, the C-Port is packaged with a User's Manual and Patient Implant stickers.
Each kit is packaged in an external fiberboard box. Each box is labeled appropriately to match the item inside the box.
#### 23. Labeling:
Each product is labeled on its plastic container with a stick-on label containing the following information:
Company Name, Location, Contact Number Product Name by Brand Name and Common Name Model No. Size of catheter Lot Number registration Units/package indicator Restricted Device: Federal (U.S.A.) law restricts this device to sale by or on the order of a physician Sterile in unopened undamaged package For sinale use only Package contains information as noted in item 22 above
{4}------------------------------------------------
#### 24. Instructions:
Detailed instructions for use and care of the C-Port are in the USER'S MANUAL. A simple listing of implant instructions is noted here:
- Before implanting inspect the port. Do not use if holes, cracks, or surface 1) contaminants are visible.
- Flush all air from the port prior to placement using the 20ga. Huber point 2) needle and syringe with heparinized saline.
- The selected site for the reservoir body should be over a bony structure 3) and in a location both convenient and comfortable for the patient.
- Place the catheter into the vein using the "cut-down" technique or by 4) using a percutaneous introducer.
- Place the tip of the catheter in an area of high blood flow when placing it 5) in the venous system. Fluoroscopy is recommended to verfiv proper placement of the catheter tip in the superior vena cava.
- ର) Take care not to serrate the catheter tip or occlude it during the catheter placement process. Leave sufficient slack upon placement so patient movement does not stress the catheter.
- Position the pocket for the reservoir so that the suture line is not directly 7) over the port. Do not place the port too deep or too shallow. A depth of approximately 5mm under the skin surface is recommended as the optimal placement depth.
- Cut the catheter to the proper length and moisten all components with 8) saline.
- Slide the compression boot over the catheter. 9)
- 10) Slide the catheter over the barbed outlet tube of the reservoir.
- Slide the compression boot forward until the catheter and the outlet tube 11) are completely covered.
- 12) Test by gently tugging on the catheter.
- 13) Secure the port to the underlying fascia with at least three nonabsorbable sutures.
- 14) After suturing has been satisfactorily completed. flush the incision with an appropriate antibiotic to ensure a sterile pocket.
- 15) Before closure, check patency and flow through the C-Port by x-ray. fluoroscopy, or by an imaging technique of choice.
- 16) After each use, always leave the C-Port filled with a heparinized saline solution in a concentration recommended by your institution.
{5}------------------------------------------------
# Part II
# Detailed product characteristics
#### 1. Structure (elements) of medical device material:
| Component | Material |
|--------------------------|------------------------------------|
| C-Port Base Plate (Body) | Titanium |
| C-Port Septum Retainer | Titanium |
| Port Septum | Silicone |
| Catheter | Silicone Rubber or<br>Polyurethane |
| Connector Pin | Titanium |
| Catheter Boot | Silicone Rubber |
All the materials in these components have been used in the market in predicate devices and other medical devices for an extended period of time. These materials do not present any new or modified mechanical or chemical specifications and are substantially equivalent to devices currently cleared for marketing in the United States.
### lf product is used with another medical material, present the name and safety conditions for their use:
The CliniSurg Vascular Products C-Port line of drug delivery depots should be accessed with a Huber point needle. It should be used according to standard hospital protocol for needle use.
#### 2. Testing of components
All the components of the C-Port have been tested and meet the required specifications or items called for in the Guidance documents on Implanted ports, Long-term catheters, Sterility, and Biocompatibility. Types of testing that have been documented are as follows:
- Catheter to Port connection test a)
- Septum puncture test b)
- Port leak test C)
- d) Fluid dynamics test
- e) Biological evaluation (toxicity)
- Catheter body tensile strength f)
- Catheter elongation strength g)
{6}------------------------------------------------
#### 3. Kit certification
The C-Port is supplied with a kit for convenience in implanting by the surgeon and the institution. The items supplied in the kit for introducing the catheter to the venous system and for assisting in placement of the port are standard products commonly used in the marketplace for many applications. No new materials or product configurations have been introduced.
## Part III
## Risk Assessment
The C-Port is an implantable device requiring surgery for implantation. The risks associated with surgery in general are well documented and understood. There are no risks involved with the implantation of this device over the general risks associated with any surgical procedure.
Possible complications are presented in the literature that pertain to this class of devices. Possible notations are: infection, device erosion, hematoma, necrosis, device extrusion, clot formation, catheter fragmentation, embolization, catheter occlusion, A-V fistula, cardiac arrhythmia, cardiac puncture, cardiac tamponade, drug extravasation, fibrin sheath, endocarditis, implant reactions, migration of the port/catheter, perforation, or laceration of the vessels, thoracic duct injury, thromboembolism, thrombosis, and superior vena cava syndrome. Improper placement of the catheter with the body has been shown to cause the catheter to be severed due to a "pinching" effect by the clavicle and the first rib. This device presents no known additional risks to the patient that are not well documented and for which there is already a prescribed therapy.
This device is similar to other devices currently on the market for which there is an abundance of medical information.
All of the materials used in this product are purchased to specifications that specify their use in a medical device and pose no additional or unknown risk to the patient.
All these materials have been successfully used in other Vascular Access products and present no unusual or unacceptable risks to the patient.
{7}------------------------------------------------
# Part IV
## Information on clinical studies performed
No clinical studies were performed as part of this 510(K) process as the product is identical to products currently cleared for marketing and can be applied for under the substantial equivalence claim.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference of the circle. Inside the circle is a stylized image of an eagle, with three curved lines representing the eagle's wings and head.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN - 5 2003
Mr. Donald F. Hults Manager CliniSurg Vascular Products, LLC 5454 Amesbury Drive, # 2107 Dallas, Texas 75206
Re: K030636
Trade/Device Name: C-Port Regulation Number: 880.5965 Regulation Name: Subcutaneous, Intravascular, Implanted, Intravascular Infusion Port and Catheter Regulatory Class: II Product Code: LJT Dated: May 19, 2003 Received: May 20, 2003
Dear Mr. Hults:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{9}------------------------------------------------
### Page 2 - Mr. Hults
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Susan Runner, DDS, MA
Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
510(k) Number (if known):
Device Name: C-Port (family of Implanted Vascular Access Devices) Port and catheter, Implanted, Subcutaneous, Intravascular - LJT
Indications for Use:
The C-Port family of products will be used as a subcutaneously implanted device where repeated access to the vascular system is the therapy of choice for delivery of medications, fluids, nutritional solutions and blood products or withdrawal of blood.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurence of CDRH, Office of Device Evaluation (ODE)
Patricia Cucenit
(Division Sign-Off) (Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
510(k) Number: K030636
(Optional Format 3-10-98)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.