VITAL SIGNS MONITOR W/ARRHYTHMIA DETECTION AND ST ANALYSIS, MODEL 8100
Applicant
Criticare Systems, Inc.
Product Code
MHX · Cardiovascular
Decision Date
Apr 17, 2003
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1025
Device Class
Class 2
Indications for Use
This system is intended to monitor physiological parameters of patients within any healthcare environment. The user, responsible to interpret the monitored data made available, will be a professional health care provider. Physiological data, system alarms and patient data analysis will be available to the care provider from the monitor.
Device Story
8100 Vital Signs Monitor displays real-time physiological data (ECG, NIBP, IBP, SpO2, Temp, Respiration, CO2, N2O, O2, anesthetic agents); integrates Brentwood Medical arrhythmia detection and ST analysis software. Device uses existing ECG hardware and microprocessor to process signals; provides limit alarms, strip chart recordings, and data trends. Used in clinical settings by healthcare professionals; output displayed on flat-panel TFT LCD. Enables clinicians to monitor patient status, detect arrhythmias, and analyze ST segments to inform clinical decision-making and patient care.
Clinical Evidence
No clinical data. Bench testing only. Performance of integrated Brentwood algorithms verified using AAMI EC-57 testing and validation with a representative group of prerecorded ECG waveforms containing abnormal beats to confirm alarm timeliness.
Technological Characteristics
Molded plastic enclosure; Active TFT LCD color display; membrane key panels and rotary push-button navigation. Monitors ECG, NIBP, IBP, SpO2, Temp, Respiration, CO2, N2O, O2, and anesthetic agents. Connectivity to VitalView 24 central monitor. Software-based arrhythmia detection and ST analysis algorithms.
Indications for Use
Indicated for monitoring physiological parameters of patients in any healthcare environment by professional health care providers.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Brentwood Medical arrhythmia/ST algorithms (K013717)
Submission Summary (Full Text)
{0}------------------------------------------------
K030613
page 1 of 2
17 February 2003
APR 1 7 2003
## 510(k) Summary
12
#### 8100 Vital Signs Monitor w/ Arrhythmia Detection and ST Analysis
| Contact: | Alex Kaplan |
|-----------------------------------------|---------------------------------------------------|
| | Director of QA & RA |
| | Criticare Systems, Inc. |
| | 20925 Crossroads Circle |
| | Waukesha, WI 53186 USA |
| | 262-798-8282 Voice |
| | 262-798-8290 FAX |
| Trade Name: | 8100 Vital Signs Monitor w/Arrhythmia and ST |
| Common Name: | Vital Signs Monitor |
| Classification Name: | Monitor, Physiological, Patient (74 MWI) |
| Substantial Equivalence is claimed to : | CSI Model 8100/8500 Vital Signs Monitor (K012059) |
#### Device Description:
The 8100 monitor interprets and displays real time physiological data of the patient including waveforms and numerical data. The 8100 can be custom configured to monitor ECG, Noninvasive BP (NIBP), Invasive BP (IBP), S.O2, Temperature, Respiration, CO2, N.O. O2 and Halogenated Anesthetic Agents. Arrhythmia Detection and ST Analysis software from Brentwood Medical has been incorporated into the 8100 system and uses the extant ECG hardware inputs and microprocessor to enable these capabilities. All user interface elements are provided by the existing 8100 monitor. For each patient vital parameter, the 8100 will be capable of providing limit alarms and alerts, printing of strip chart recordings and storing data trends for retrospective review.
#### Intended Use:
This system is intended to monitor physiological parameters of patients within any healthcare environment. The user, responsible to interpret the monitored data made available, will be a professional health care provider. Physiological data, system alarms and patient data analysis will be available to the care provider from the monitor.
{1}------------------------------------------------
#### Comparison with predicate device:
Criticare Systems Inc. has developed and distributed physiological monitoring devices worldwide since its inception in 1984. The 8100 Vital Signs Monitoring System w/ Arrhythmia Detection and ST Analysis (8100) has the same functions and specifications as the predicate device, the 8100 / 8500 Vital Signs Monitoring System (8100 / 8500), with the addition of Arrhythmia Detection and ST Analysis software algorithms from Brentwood Medical (Brentwood algorithms). The 8100 / 8500 Vital Signs Monitoring System was approved for distribution under the FDA 510(k) file number K012059 and Brentwood Medical's algorithms were previously listed under FDA 510(k) file number K013717. The 8100 / 8500 predicate system is capable of monitoring a number of physiological parameters including ECG, NIBP, IBP, ETCO3, N,O. O., Halogenated Anesthetic Agents, Respiration, Temperature and SpO2. The patient data collected by the 8100 monitor is displayed for the user on a flat panel display as on the predicate device. The 8100 monitor utilizes Active TFT LCD color display technology. Membrane key panels and rotary push button navigation provides a user interface equivalent to the predicate device. The packaging design of the 8100 monitor is molded plastic, as was the predicate 8100/8500 Vital Signs Monitor. It incorporates audible and visual alarm functions that are activated when set parameter limits are exceeded. The monitoring system is also capable of communicating patient data to the VitalView 24 central monitor for display and alarms as appropriate.
#### Determination of Substantial Equivalence:
The 8100 monitor has been confirmed to have equivalent performance to the predicate devices for each patient monitoring modality. Additionally, the 8100 complies with safety and performance standards (detailed below) for each monitoring modality and verification of compliance has been completed. AAMI EC-57 testing was performed by Brentwood Medical in their previous submittal (K013717). To document the performance of the Brentwood algorithms when integrated into the 8100 and verify the timeliness of ECG morphology-related alarms, the 8100 was tested with a representative group of prerecorded ECG waveforms with abnormal beats. The patient monitoring technologies present in the 8100 monitor have been in clinical use for at least six years in the predicate 8100 / 8500 (and it's predicate devices, the Scholar 2200 and POET IQ monitors). CSI's field experience with these modalities in the predicate devices has been satisfactory. This combination of equivalence testing, applicable objective standards compliance and field experience substantiates a high level of confidence in the safety and efficacy of the 8100 monitor.
Therefore, the 8100 monitor is substantially equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with its wings spread, with the profiles of two human faces incorporated into the design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 1 7 2003
Criticare Systems, Inc. c/o Mr. Alex Kaplan Director of QA & RA 20925 Crossroads Circle Suite 100 Waukesha, WI 53186-4054
Re: K030613
Trade Name: Vital Signs Monitor Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm Regulatory Class: Class III (three) Product Code: MHX Dated: April 3, 2003 Received: April 7, 2003
Dear Mr. Kaplan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Alex Kaplan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
K. Walter Tull
Bi am D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# 510(k) NUMBER (IF KNOWN) : _ KQ 306/3
DEVICE NAME: Vital Signs Monitor
#### Indications for Use
This system is intended to monitor physiological parameters of patients within any healthcare environment. The user, responsible to interpret the monitored data made available, will be a professional health care provider. Physiological data, system alarns and patient data analysis will be available to the care provider from the monitor.
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED.)
Prescription Use Qver - the - Counter - Use_ (Per 21 CFR 801.109) (Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.