K030523 · Biolase Technology, Inc. · GEX · Jan 30, 2004 · General, Plastic Surgery
Device Facts
Record ID
K030523
Device Name
WATERLASE
Applicant
Biolase Technology, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jan 30, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Use of Waterlase may be indicated for: General Indications* - Class I, II, III, IV, and V cavity preparation . - Caries removal . - Hard tissue surface roughening or etching . - Enameloplasty, excavation of pits and fissures for placement of sealants . - * For use on adult and pediatric patients Root Canal Hard Tissue Indications - Tooth preparation to obtain access to root canal . - Root canal preparation including enlargement • - Root canal debridement and cleaning Endodontic Surgery (Root Amputation) Indications (soft and hard tissue) - Flap preparation incision of soft tissue to prepare a flap and expose the bone. . - Cutting bone to prepare a window access to the apex (apices) of the root(s). 후 - Apicoectomy amputation of the root end. ■ - Root end preparation for retrofill amalgam or composite. . - Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic 8 tissues (i.e., granulation tissue) from around the apex - I Incision and drainage of periapical abscesses NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation. Bone Surgical Indications - Cutting, shaving, contouring and resection of oral osseous tissues (bone) - + Osteotomy Laser Periodontal Procedures - . Full thickness flap - . Partial thickness flap - . Split thickness flap - . Laser soft tissue curettage - * Laser removal of diseased, inflamed and necrosed soft tissue within the periodontal pocket - . Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium - . Removal of granulation tissue from bony defects - ↓ Sulcular debridement (removal of diseased, inflamed or necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility) - . Osteoplasty and osseous recontouring (removal of bone to correct osseous defects and create physiologic osseous contours) - . Ostectomy (resection of bone to restore bony architecture, resection of bone for grafting, etc.) - t Osseous crown lengthening Soft Tissue Indications Including Pulpal Tissues* Incision, excision, vaporization, ablation and coagulation of oral soft tissues, includina: - Excisional and incisional biopsies . - . Exposure of unerupted teeth - . Fibroma removal - . Flap preparation - incision of soft tissue to prepare a flap and expose the bone - . Flap preparation - incision of soft tissue to prepare a flap and expose unerupted teeth (hard and soft tissue impactions) - Frenectomy and frenotomy . - Gingival troughing for crown impressions . - Gingivectomy * - . Gingivoplasty - Gingival incision and excision . - . Hemostasis - . Implant recovery - . Incision and drainage of abscesses - . Laser soft tissue curettage of the post-extraction tooth so area during apical surgery - Leukoplakia . - Operculectomy - . Oral papillectomies - . Pulpotomy - . Pulp extirpation - Pulpotomy as an adjunct to root canal therapy - Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic . tissues (i.e., granulation tissue) from around the apex - . Root canal debridement and cleaning - # Reduction of gingival hypertrophy - Soft tissue crown lengthening . - Treatment of canker sores, herpetic and aphthous uicers of the oral mucosa • - . Vestibuloplasty * For use on adult and pediatric patients
Device Story
Waterlase is an Er,Cr:YSGG laser system for dental hard and soft tissue procedures. Device uses laser energy combined with water atomization for hard tissue (cutting, etching, removing) and direct laser energy for soft tissue (incising, ablating, coagulating). Flexible fiberoptic handpiece delivers energy; visible light pinpoints treatment area. Clinician adjusts power output, pulse energy, repetition rate, and air/water flow. Used in dental clinics by dentists. Output allows precise tissue removal, debridement, and hemostasis, potentially reducing need for traditional mechanical tools and improving clinical indices in periodontal/endodontic applications.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and previously cleared indications.
Indicated for adult and pediatric patients requiring oral hard and soft tissue procedures, including cavity preparation, caries removal, root canal therapy, endodontic surgery, bone surgery (osteotomy/osteoplasty), periodontal procedures (flap, curettage, debridement), and soft tissue management (excision, ablation, coagulation, biopsy, frenectomy, gingivectomy).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
030523 1/4
| | 510(k) Summary of Safety<br>and Effectiveness Information | JAN 30 2004 |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Regulatory<br>Authority: | Safe Medical Devices Act of 1990,<br>21 CFR 807.92 | |
| Company: | BioLase Technology, Inc.<br>981 Calle Amanecer<br>San Clemente, CA 92673 | |
| Contact: | Ms. Ioana M. Rizoiu<br>BioLase Technology, Inc.<br>981 Calle Amanecer<br>San Clemente, CA 92673<br>(949) 361-1200 (949) 361-0204 Fax | |
| Trade Name: | Waterlase® | |
| Common Name: | Er,Cr:YSGG laser | |
| Classification Name: | Surgical laser instrument | |
| Classification Code: | 79 GEX, MXF, DZI | |
| Equivalent Devices: | | |
| BioLase Technology, Inc. | Waterlase® | |
### Device Description:
The Waterlase hydrokinetic dental laser system is a diverse device utilized to perform a variety of dental applications. For hard tissue procedures the Waterlase uses the Erbium, Chromium: Yttrium, Scandium, Gallium Garnet (Er,Cr.YSGG) laser in combination with advanced water atomization technology to cut, remove, roughen and etch tissues. Soft tissue procedures are performed using a different mode of operation where direct Er,Cr.YSGG laser energy is applied to incise, excise or ablate these tissues. In soft tissue procedures the water spray is applied for hydration, cooling or to keep tissues clean. For hard tissue applications the spray is part of the tissue removing process as well as hydration, cooling and keeping the tissues clean.
A flexible fiberoptic handpiece delivers the Waterlase laser e nergy. A visible light emitted from the handpiece distal end pinpoints the area of treatment. In both hard and soft tissue applications the power output, pulse energy, repetition rate and air and water flow rates are adjustable to specific user requirements.
{1}------------------------------------------------
K030523
## Indications for Use:
## Hard Tissue
.
## (General Indications*
- Class I, II, III, IV and V cavity preparation
2/4
- . Caries removal
- Hard tissue surface roughening or etching .
- Enameloplasty, excavation of pits and fissures for placement of sealants .
- * For use on adult and pediatric patients
## Root Canal Hard Tissue Indications
- Tooth preparation to obtain access to root canal
- Root canal preparation including enlargement
- Root canal debridement and cleaning .
### Bone Surgical Indications
.
- Cutting, shaving, contouring and resection of oral osseous tissucs (bone)
- . Osteotomy
### Endodontic Surgery (Root Amputation) Indications
- Flap preparation incision of soft tissue to prepare a flap and expose the bone .
- Cutting bone to prepare a window access to the apex (apices) of the root(s). .
- . Apicoectomy - amputation of the root end.
- Root end preparation for retrofill amalgam or composite .
- Removal of pathological tissues (i.e. cysts, neoplasm or abscess) and . hyperplastic tissues (i.e., granulation tissue) from around the apcx. Note: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation.
## Laser Periodontal Procedures
- Full thickness flap .
- . Partial thickness flap
- Split thickness flap .
- Laser soft tissue curettage .
- Laser removal of discased, infected, inflamed and necrosed soft tissue t the provinsity of the first of the first the first the first the first the first the first the first the first the finitial was and the finalisting the finalism of the fini within the periodontal pocket
- Removal of highly inflamed edematous tissuc affected by bacteria . penetration of the pocket lining and junctional epithelium
- Removal of granulation tissue from bony defects .
- Sulcular debridement (removal of diseased, inflamcd or necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility)
{2}------------------------------------------------
030523 3/4
WaterLase™ Expanded Indications for Use BioLase Technology, Inc. October 29, 2003
- Osteoplasty and osseous recontouring (removal of bone to correct osseous defects and create physiologic osseous contours)
- Ostectomy (resection of bone to restore bony architecture, resection of bone for grafting, etc.)
- Osseous crown lengthening
## Soft Tissue Indications Including Pulpal Tissues*
Incision, excision, vaporization, ablation and coagulation of oral soft tissues, including:
- Excisional and incisional biopsies .
- . Exposure of unerupted teeth
- . Fibroma removal
- Flap preparation incision of soft tissue to prepare a flap and expose the bone .
- Flap preparation incision of soft tissue to prepare a flap and expose the 이 unerupted teeth (hard and soft tissue impactions)
- Frenectomy and frenotomy .
- Gingival troughing for crown impressions *
- Gingivectomy *
- Gingivoplasty ◆
- Gingival incision and excision ◆
- Hemostasis& coagulation �
- s Implant recovery
- . Incision and drainage of abscesses
- * Incision and drainage of periapical abscesses.
- Laser soft tissue curettage of the post-extraction tooth sockets and the . periapical area during apical surgery
- . Leukoplakia
- . Operculectomy
- Oral papillectomies .
- Pulpotomy .
- . Pulp extirpation
- . Pulpotomy as an adjunct to root canal therapy
- ◆ Reduction of gingival hypertrophy
- Removal of pathological tissues (i.e. cysts, neoplasm or abscess) and . hyperplastic tissues (i.e. granulation tissue). Note: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation.
- Root canal debridement and cleaning .
- Soft tissue crown lengthening .
- Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa ◆
- . Vestibuloplasty
* For use on adult and pediatric patients
## Cautions and Contraindications:
All clinical procedures performed with Waterlase must be subjected to the same clinical judgment and care as with traditional techniques. Patient risk must always
{3}------------------------------------------------
030513 4/4
WaterLase™ Expanded Indications for Use BioLase Technology, Inc. October 29, 2003
be considered and fully understood before clinical treatment. The clinician must completely understand the patient's medical history prior to treatment. Exercisc caution for gencral medical conditions that might contraindicate a local procedure. Such conditions may include allergy to local or topical anesthetics, heart discase, lung disease, bleeding disorders, sleep apnea or an immune system deficiency. Medical clearance from patient's physician is advisable when doubt exists regarding treatment.
#### Substantial Equivalence:
There are no unique applications, indications, materials or specifications presented herein. All the presented indications for use retain the same meaning as their equivalent indications cleared by the FDA in K011041 (Waterlase", expanded soft tissue indications including Sulcular Debridement), K012511 (Waterlase " indications for Root Canal), K013908 (Waterlase , indications for oral osseous procedures) and K022803 (Waterlase , indications for Endodontic Surgery).
#### Conclusion:
The indications included herein are the same as those included in previous clearances. Waterlase is therefore substantially equivalent in relation to the previous clearances.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus, featuring three parallel lines that curve and converge.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 3 0 2004
Ms. Ioana M. Rizoiu Vice President, Clinical Research and Development Biolase Technology, Inc. 981 Calle Amanecer San Clemente, California 92673
Re: K030523 Trade/Device Name: Waterlase® Regulation Number: 21 CFR 878.4810 Regulation Name: Lascr surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Codc: GEX Dated: October 31, 2003 Received: November 3, 2003
Dear Ms. Rizoiu:
We have reviewed your Scction 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adultcration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Codc of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# 510(k) Number (if known): K030523
Device Name: Waterlase®
Indications for Use:
Use of Waterlase may be indicated for:
## General Indications*
- Class I, II, III, IV, and V cavity preparation .
- Caries removal .
- Hard tissue surface roughening or etching .
- Enameloplasty, excavation of pits and fissures for placement of sealants .
- * For use on adult and pediatric patients
## Root Canal Hard Tissue Indications
- Tooth preparation to obtain access to root canal .
- Root canal preparation including enlargement •
- Root canal debridement and cleaning
## Endodontic Surgery (Root Amputation) Indications (soft and hard tissue)
- Flap preparation incision of soft tissue to prepare a flap and expose the bone. .
- Cutting bone to prepare a window access to the apex (apices) of the root(s). 후
- Apicoectomy amputation of the root end. ■
- Root end preparation for retrofill amalgam or composite. .
- Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic 8 tissues (i.e., granulation tissue) from around the apex
- I Incision and drainage of periapical abscesses
- NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation.
## Bone Surgical Indications
- Cutting, shaving, contouring and resection of oral osseous tissues (bone)
- + Osteotomy
## Laser Periodontal Procedures
- . Full thickness flap
- . Partial thickness flap
- . Split thickness flap
- . Laser soft tissue curettage
Mark A. Milken
Neurological Devices
K030523
{7}------------------------------------------------
- * Laser removal of diseased, inflamed and necrosed soft tissue within the periodontal pocket
- . Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium
- . Removal of granulation tissue from bony defects
- ↓ Sulcular debridement (removal of diseased, inflamed or necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility)
- . Osteoplasty and osseous recontouring (removal of bone to correct osseous defects and create physiologic osseous contours)
- . Ostectomy (resection of bone to restore bony architecture, resection of bone for grafting, etc.)
- t Osseous crown lengthening
## Soft Tissue Indications Including Pulpal Tissues*
Incision, excision, vaporization, ablation and coagulation of oral soft tissues, includina:
- Excisional and incisional biopsies .
- . Exposure of unerupted teeth
- . Fibroma removal
- . Flap preparation - incision of soft tissue to prepare a flap and expose the bone
- . Flap preparation - incision of soft tissue to prepare a flap and expose unerupted teeth (hard and soft tissue impactions)
- Frenectomy and frenotomy .
- Gingival troughing for crown impressions .
- Gingivectomy *
- . Gingivoplasty
- Gingival incision and excision .
- . Hemostasis
- . Implant recovery
- . Incision and drainage of abscesses
- . Laser soft tissue curettage of the post-extraction tooth so area during apical surgery
- Leukoplakia .
- Operculectomy
- . Oral papillectomies
- . Pulpotomy
- . Pulp extirpation
- Pulpotomy as an adjunct to root canal therapy
- Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic . tissues (i.e., granulation tissue) from around the apex
- . Root canal debridement and cleaning
- # Reduction of gingival hypertrophy
"K030523
{8}------------------------------------------------
- Soft tissue crown lengthening .
- Treatment of canker sores, herpetic and aphthous uicers of the oral mucosa •
- . Vestibuloplasty
- * For use on adult and pediatric patients
for Mark A. Mellers
K030523
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
or
Over-The-Counter-Use _________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.