HandyDop™ is a device designed to transmit and receive ultrasonic energy into and from the pregnant woman, by means of continuous wave (doppler) echoscopy. The device is indicated for use to detect the flow of blood within the body. When using it with the 2 MHz probe, HandyDop™ is indicated for use for the early detection of fetal life, detection of multiple pregnancies, fetal screening from early gestation through delivery, and general indication of fetal well being. When using it with the 4 MHz or 8 MHz probes, HandyDop™ is indicated for use for detection of blood flow in the peripheral vascular system for assisting in the detection of peripheral vascular disease.
Device Story
Handydop is a handheld, battery-operated, continuous wave (CW) Doppler ultrasound device. It utilizes three interchangeable probes (2 MHz, 4 MHz, 8 MHz) to transmit and receive ultrasonic energy. The device detects blood flow, converting signals into audible modulated noise via an integrated speaker. Operated by clinicians in clinical settings, the device features a simple control interface consisting of an ON/OFF switch and a loudness wheel. The output provides real-time auditory feedback of blood flow, assisting healthcare providers in fetal monitoring and peripheral vascular assessment. The device is cordless, powered by two 9V batteries or accumulators.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological characteristics and intended use comparisons to the predicate device. Bench testing of acoustic output measurements is required post-clearance.
Technological Characteristics
Handheld CW Doppler ultrasound system. Materials: ABS-plastic housing, Araldit-D (duroplastic) encapsulated quartz crystals. Transducers: 2 MHz (fetal), 4 MHz and 8 MHz (peripheral vascular). Power: 2x 9V batteries or accumulators. Interface: ON/OFF switch, volume control wheel. Output: Audible speaker. No software or digital processing described.
Indications for Use
Indicated for pregnant women (2 MHz probe) for fetal life detection, multiple pregnancy detection, and fetal screening/well-being assessment. Indicated for patients (4 MHz/8 MHz probes) for peripheral vascular blood flow detection to assist in diagnosing peripheral vascular disease.
Regulatory Classification
Identification
A fetal ultrasonic monitor is a device designed to transmit and receive ultrasonic energy into and from the pregnant woman, usually by means of continuous wave (doppler) echoscopy. The device is used to represent some physiological condition or characteristic in a measured value over a period of time (e.g., perinatal monitoring during labor) or in an immediately perceptible form (e.g., use of the ultrasonic stethoscope). This generic type of device may include the following accessories: signal analysis and display equipment, electronic interfaces for other equipment, patient and equipment supports, and component parts. This generic type of device does not include devices used to image some relatively unchanging physiological structure or interpret a physiological condition, but does include devices which may be set to alarm automatically at a predetermined threshold value.
Predicate Devices
Medasonics' Cadence Doppler Ultrasound System (K991441)
Submission Summary (Full Text)
{0}------------------------------------------------
SEP 1 2 2003
# 510(k) Summary As Required by 21 section 807.92 ( c )
1-Submitter Name: Elcat GmbH
| 1-Submitter Name: | Eleat GmbH |
|---------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2-Address: | Schießstättstraße 29<br>D-82515 Wolfratshausen<br>Germany |
| 3-Phone: | 011 49 8171 4214-0 |
| 4-Fax: | 011 49 8171 4214-47 |
| 5-Contact Person: | Mr Bernd Marquardt (General Manager) |
| 6-Date summary prepared: | January 30th, 2003 |
| 7- Official Correspondent: | Mansour Consulting LLC |
| 8- Address: | 1308 Morningside Park Dr<br>Alpharetta, GA 30022 USA |
| 9- Phone: | (770) 908-8180 |
| 10- Fax: | (425) 795-9341 |
| 11- Contact person: | Jay Mansour, president |
| 12-Device Trade or Proprietary Name: | HANDYDOP |
| 13-Device Common or usual name: | Doppler |
| 14-Device Classification Name: | MONITOR, ULTRASONIC, FETAL |
| 15-Substantial Equivalency is claimed against the following device: | Medasonics' Cadence Doppler Ultrasound System, 510k #k991441 (refer<br>to Appendix 2 for FDA website printout)<br>This notification for HANDYDOP is of the ABBREVIATED type as per<br>the declaration of conformity included in this summary |
### 16-Description of the Device:
The Handydop is a handheld, unidirectional, continuous wave (CW) Doppler with three changeable probes. 2MHz CW probe for foetal heartbeat detection, 4MHz CW probe and 8 MHz CW probe for peripheral vascular applications. You can hear the blood flow from a vein as a modulated noise on the speaker of the handydop. The Handydop works cordless with two 9V Battery's or Accumulators. The housings of the Handydop and the probes are made of ABS-plastics. The quartz crystals where encapsulated with Araldit-D (a Duroplastic). As control panel there are only the ON/OFF switch and the loudness wheel.
### 17-Intended use of the device: (Indications for use typed on a separate FDA form)
HandyDop™ is a device designed to transmit and receive ultrasonic energy into and from the pregnant woman, by means of continuous wave (doppler) echoscopy. The device is indicated for use to detect the flow of blood within the body.
When using it with the 2 MHz probe, HandyDop™ is indicated for use for the early detection of fetal life, detection of multiple pregnancies, fetal screening from early gestation through delivery, and general indication of fetal well being.
When using it with the 4 MHz or 8 MHz probes, HandyDop™ is indicated for use for detection of blood flow in the peripheral vascular system for assisting in the detection of peripheral vascular disease.
PAGE 2
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features a stylized eagle with three lines representing its wings, symbolizing health, services, and people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Public Health Service
SEP 1 2 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
ELCAT Gmbh % Mr. Jay Mansour President Mansour Consulting, LLC 1308 Morningside Park Dr. ALPHARETTA GA 30022
Re: K030466
Trade Name: HandyDop™ Regulation Number: 21 CFR 884.2660 Regulation Name: Fetal ultrasonic monitor and accessories Regulatory Class: II Product Code: 85 KNG Dated: July 23, 2003 Received: July 28, 2003
Dear Mr. Mansour:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the HandyDop™, as described in your premarket notification:
### Transducer Model Number
| 2MHz |
|------|
| 4MHz |
| 8MHz |
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA
{2}------------------------------------------------
### Page 2 - Mr. Mansour
may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G. (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801, please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
{3}------------------------------------------------
Page: 3 – Mr. Mansour
If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.
117
, i
Sincerely yours,
Yamila Brogdon
Nancy C. Brogdon
Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
# Enclosure ﻧ
ﺎ
.
{4}------------------------------------------------
# Diagnostic Ultrasound Indications for Use Form
# Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | N | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | N | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
| N= new indication; P= previously cleared by FDA; E= added under Appendix E | | | | | | | | | | |
| Additional Comments: HANDY DOT IS DESIGNED TO TRANSMIT AND RECEIVE ULTRASOUND<br>ENERGY INTO & THRU THE PREGNANT WOMAN BY MEANS OF CONTINUOUS WAVE (DOPPLER)<br>(IDENTIFY TO DETECT FLOW OF BLOOD WITHIN THE BODY<br>2 MHZ PROBE IS FOR USE FOR EARLY DETECTION OF FETAL LIFE, DETECTION OF<br>MULTIPLE PREGNANCIES, FETAL SCREENING FROM EARLY GESTATION THRU | | | | | | | | | | |
Prescription Use (Per 21 CFR 801.109)
Nancy Higdon
(Division Sign-Off) / Division of Reproductive, Abdor and Radiological Devices 540(k) Number _
{5}------------------------------------------------
| SUBJECT | Siok # K030466 |
|---------|----------------|
| DATE | 07/23/03 |
CONTINUATION OF COMMENTS
DELIVERY, AND GENERAL INDICATION OF FETAL WELL BEING.
4 MHZ & 8 MHZ PROBES ARE FOR USE FOR DETECTION
of BLOOD FLOW IN THE PERIPHERAL VASCULAR SYSTEM
OF THE BODY FOR ASSISTING IN THE DETECTION OF
PERIPHERAL VASCULAR DISEASE. <END>
PAGE 55
.
t
{6}------------------------------------------------
#### Diagnostic Ultrasound Indications for Use Form
#### Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | N | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
| N= new indication; P= previously cleared by FDA; E= added under Appendix E | | | | | | | | | | |
| Additional Comments: 2 MHz Probe is indicated for use for<br>The early detection of fetal life, detection of<br>multiple pregnancies, fetal screening from early<br>gestation through delivery, and general indication<br>of fetal well being. | | | | | | | | | | |
Concurrence of CDRH, Office of Device Evaluation (ODE)
David R. Legum
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number -
Prescription Use (Per 21 CFR 801.109)
F-3
{7}------------------------------------------------
#### Diagnostic Ultrasound Indications for Use Form
# Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | N | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
| N= new indication; P= previously cleared by FDA; E= added under Appendix E | | | | | | | | | | |
| Additional Comments: 4 MAZ PROBE IS INDICATED FOR USE FOR DETECTION OF BLOOD FLOW IN THE PERIPHERAL VASCULAR SYSTEM OF THE BODY FOR ASSISTING IN THE DETECTION OF PERIPHERAL VASCULAR DISEASE | | | | | | | | | | |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
ca Evaluation (ODE)
David G. Lynn
(Division Sign Off)
(Division Sign-Off) Division of Reproductive, Abdomin and Radiological Devices 510(k) Number _
Prescription Use (Per 21 CFR 801.109)
{8}------------------------------------------------
#### Diagnostic Ultrasound Indications for Use Form
### Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic utrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | N | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: 8 MHZ PROBE IS INDICATED FOR USE FOR DETECTION OF BLOOD FLOW IN THE PERIPHERAL VASCULAR SYSTEM OF THE BODY FOR ASSISTING IN THE DETECTION of PERIPHERAL VASCULAR DISEASE
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David R. Lipton
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number _
Prescription Use (Per 21 CFR 801.109)
F-3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.