K030450 · J. Morita USA, Inc. · JAK · May 6, 2003 · Radiology
Device Facts
Record ID
K030450
Device Name
3D ACCU-I-TOMO XYZ SLICE VIEW TOMOGRAPH
Applicant
J. Morita USA, Inc.
Product Code
JAK · Radiology
Decision Date
May 6, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
The 3D Accu-I-tomo is an x-ray imaging device that acquires a 360 degree rotational sequence of the head and neck areas, including the ENT and dentomaxillofacial areas, for use in diagnostic support. The device accomplishes this task by reconstructing a three-dimensional matrix of the examined volume and producing two-dimensional views of this volume, displaying both two- and three-dimensional images. The device can also be used for fluoroscopy during surgery, mostly for ENT and TMJ applications and mostly with a contrast medium. The device is operated and used by physicians, dentists, and x-ray technologists.
Device Story
3D Accu-I-tomo is a cone beam x-ray CT system for head/neck imaging. Input: 360-degree rotational x-ray sequence. Operation: C-arm assembly with x-ray head and image intensifier; system reconstructs 3D matrix from volume data. Output: 2D and 3D images for diagnostic support; fluoroscopic moving/still images for surgery. Used in clinical settings by physicians, dentists, and technologists. System includes Windows-based PC with '3DXD' driver and '3DX Integrated Information System' software. Healthcare providers use images for pre/post-surgical diagnosis and intraoperative guidance. Benefits include high-resolution imaging with lower radiation than conventional CT.
Clinical Evidence
Bench testing only. Evidence includes software validation report (firmware, driver, and application software), hazard analysis, and system/software requirements verification. Device passed image quality testing and complied with electrical, mechanical, and thermal safety standards (UL2601-1, IEC 60601-1). Biocompatibility of patient-contacting materials (head/chin band, cushions, etc.) confirmed via predicate comparison and established use in medical applications.
Technological Characteristics
Cone beam x-ray CT system. Components: C-arm assembly, x-ray head, image intensifier, beam-limiting device. Materials: Biocompatible plastics for patient-contacting surfaces. Connectivity: Windows-based PC (standalone or networked server/client). Software: 3DXD driver and 3DX Integrated Information System. Safety: Complies with UL2601-1, IEC 60601-1, and 21 CFR Subchapter J. Operation: CT mode (3D reconstruction) and Fluoroscopic mode (moving/still images).
Indications for Use
Indicated for diagnostic support via x-ray imaging of head and neck areas, including ENT and dentomaxillofacial regions. Supports 3D volume reconstruction and 2D viewing, plus intraoperative fluoroscopy (often with contrast) for ENT and TMJ applications. For use by physicians, dentists, and x-ray technologists.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
#### 510(k) SUMMARY
# J. Morita Manufacturing Corporation 3D Accu-I-tomo XYZ Slice View Tomograph
#### Name of Device and Name/Address of Sponsor
Trade or Proprietary Name: 3D Accu-I-tomo XYZ Slice View Tomograph Common Name: Cone beam x-ray CT Classification Name: computed tomography x-ray system Product Code: JAK
J. Morita Manufacturing Corporation 680 Higashihama Minami-cho, Fushimi-ku Kyoto 612-8533 Japan Telephone: 011-81-75-605-2305 011-81-75-605-2350 Facsimile: Date Prepared: January 20, 2003
### Intended Use
The 3D Accu-I-tomo is an x-ray imaging device that acquires a 360 degree rotational sequence of the head and neck areas, including the ENT and dentomaxillofacial areas, for use in diagnostic support. The device accomplishes this task by reconstructing a three-dimensional matrix of the examined volume and producing twodimensional views of this volume, displaying both two- and three-dimensional images. The device can also be used for fluoroscopy during surgery, mostly for ENT and TMJ applications and mostly with a contrast medium. The device is operated and used by physicians, dentists, and x-ray technologists.
## Technological Characteristics and Substantial Equivalence
The C-arm assembly is mounted to the support column, which has an x-ray control panel for loading factors, an emergency stop switch, and a remote control for patient positioning. The C-arm assembly includes an image intensifier and an x-ray head with a beam-limiting device. The control box, which includes operational lights and an x-ray emission button, is typically placed in a separate room from the rest of the system. The patient chair includes an armrest and a headrest to fix the patient's head during the x-ray emission. An optional headrest for ENT applications is available which can be replaced with the headrest for dental applications.
A Windows-based personal computer is connected to the 3D Accu-I-tomo, on which is installed "3DXD" driver software and "3DX Integrated Information System" application software. For the network environment, "3DXD" driver software and "3DX Integrated Information System" are installed in the server computer and "3DX Integrated Information System" is installed in the client computer. The Windows-based personal computer is not manufactured by J. Morita Manufacturing Corp.
{1}------------------------------------------------
The 3D Accu-I-tomo operates in either CT mode or Fluoroscopic mode. In CT mode, it generates three-dimensional images that make possible precise diagnosis of the head and neck areas. It produces high resolution images and uses less x-radiation in comparison with conventional computed tomography x-ray systems. Fluoroscopy is available using the same x-ray head and image intensifier that are used for the CT mode. Fluoroscopy is mainly used with a contrast medium for ENT applications and surgery of the TM Joint. The same device can provide three-dimensional images and fluoroscopy. Both two- and three-dimensional images are taken for diagnosis before and after surgery. The fluoroscopic setting has three modes: a fluoroscopic moving image mode, a still image mode, and a testing mode. In the testing mode, it can take but not save a moving image.
The 3D Accu-I-tomo is substantially equivalent to the (i) Siemens Medical Systems' Siremobil C02 (K#973598), (ii), Siemens Siremobil Iso-C 3D Imaging Option (K#003266), and (iii) NIM s.r.l.'s NewTom Model QR-DVT 9000 (K#003787). The 3D Accu-I-tomo has similar general intended uses, principles of operation, and technological characteristics as the previously cleared predicate devices. Although there are minor difference in the characteristics of the 3D Accu-I-tomo and the predicate devices, these differences do not raise new questions of safety or efficacy.
The software used in the 3D Accu-I-tomo has been successfully validated by J. Morita Manufacturing Corp. The software validation report describes: the development process for the device's firmware; and driver software and application software installed in Windows-based personal computer; the software change control and code revision procedures; the system and software requirements; the software handling and storage procedures; a hazard analysis; and a software/firmware certification that the company followed the above-described procedures and policies.
The 3D Accu-I-tomo was tested to ensure compliance with UL2601-1 and IEC 60601-1, and its collateral standards, and it complied with the applicable requirements. The 3D Accu-I-tomo will be tested and will comply with the applicable requirements of 21 CFR Subchapter J prior to marketing. The 3D Accu-I-tomo also passed the image quality testing.
The 3D Accu-I-tomo complies with the applicable thermal, mechanical, and electrical safety requirements of UL2601-1, IEC 60601-1, and its collateral standards and will comply with the applicable requirements of 21 CFR Subchapter J prior to marketing.
The 3D Accu-I-tomo uses biocompatible plastics on all body contacting surfaces. such as the head/chin band, headrest cushion, headrest support, side head support, subnasal point rest and ear piece, etc. Such plastics have been used in predicate devices or widely used in other medical applications in which the plastic is in contact with the patient's body.
40139551.doc
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
MAY 0 6 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
J. Morita Manufacturing Corporation % Mr. Keith A. Barritt Fish & Richardson, P.C. 1425 K Street, N. W. 11" Floor WASHINGTON DC 20005
Re: K030450 Trade/Device Name: 3D Accu-I-tomo XYZ Slice View Tomograph Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: 90 JAK Dated: February 10, 2003 Received: February 11, 2003
Dear Mr. Barritt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing. (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): Device Name: 3D Accu-I-tomo XYZ Slice View Tomograph Indications For Use: Κοζο 450
> The 3D Accu-I-tomo is an x-ray imaging device that acquires a 360 degree rotational sequence of the head and neck areas, including the ENT and dentomaxillofacial areas, for use in diagnostic support. The device accomplishes this task by reconstructing a three-dimensional matrix of the examined volume and producing two-dimensional views of this volume, displaying both two- and threedimensional images. The device can also be used for fluoroscopy during surgery, mostly for ENT and TMJ applications and mostly with a contrast medium. The device is operated and used by physicians, dentists, and x-ray technologists.
(DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use OR
Over-The-Counter Use
(Per 21 CFR 801.109)
40139586.doc
David A. Larson
Division Sign-Off Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.