SMART NITINOL STENT ENDOSCOPIC BILIARY SYSTEM

K030409 · Cordis Corp. · FGE · Mar 10, 2003 · Gastroenterology, Urology

Device Facts

Record IDK030409
Device NameSMART NITINOL STENT ENDOSCOPIC BILIARY SYSTEM
ApplicantCordis Corp.
Product CodeFGE · Gastroenterology, Urology
Decision DateMar 10, 2003
DecisionSESU
Submission TypeSpecial
Regulation21 CFR 876.5010
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The S.M.A.R.T. TM Nitinol Stent Endoscopic Biliary System is intended for use in the palliation of malignant neoplasms in the biliary tree.

Device Story

Self-expanding biliary stent system; used for palliation of malignant biliary neoplasms. Device consists of nickel-titanium (Nitinol) alloy stent with tantalum radiopaque markers and 7 French delivery system. Deployed endoscopically in biliary tree to maintain patency. Stent diameters 8-10 mm; lengths 40-80 mm; delivery system length 190 cm; compatible with 0.035" guidewires. Operated by clinicians. Provides mechanical support to biliary structures; benefits patient by relieving obstruction caused by malignant neoplasms.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Nickel-titanium (Nitinol) alloy stent; tantalum radiopaque markers. 7 French delivery system profile. 8, 9, 10 mm expanded diameters; 40, 60, 80 mm lengths. 190 cm delivery system length. 0.035" guidewire lumen. Biocompatible materials.

Indications for Use

Indicated for the palliation of malignant neoplasms in the biliary tree.

Regulatory Classification

Identification

A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.

Special Controls

*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ MAR 1 () 2003 K030409 page 1 of 2 ## 510(k) Summary of Safety and Effectiveness | General Provisions | The name of the device is: | |--------------------|----------------------------| |--------------------|----------------------------| | | Proprietary Name | Common or Usual Name | |--|------------------------------------------------------------------|----------------------| | | S.M.A.R.T. <sup>TM</sup> Nitinol Stent Endoscopic Biliary System | Biliary Stent | | Name of Predicate Devices | The device is substantially equivalent to: | |---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | <ul><li>Cordis S.M.A.R.T.<sup>TM</sup> Nitinol Stent Endoscopic Biliary System (510(k) # K003365 – November 22, 2000) - Cordis Corporation.</li><li>Cordis S.M.A.R.T.<sup>TM</sup> Control Nitinol Stent Transhepatic Biliary System (510(k) # K021898 – August 14, 2002) - Cordis Corporation</li><li>SMART Nitinol stent Transhepatic Biliary System (Enhanced Radiopacity) (510(k) # K001843 – July 20, 2000) - Cordis Corporation</li></ul> | | Classification | Class II. | | Performance Standards | Performance standards have not been established by the FDA under section 514 of the Food, Drug and Cosmetic Act. | | Indications for Use | The S.M.A.R.T. <sup>TM</sup> Nitinol Stent Endoscopic Biliary System is intended for use in the palliation of malignant neoplasms in the biliary tree. | | Device Description | The device description of the proposed S.M.A.R.T. <sup>TM</sup> Nitinol Stent Endoscopic Biliary System is as follows. <ul><li>7 French stent delivery system profile;</li><li>Stent material – Nickel Titanium alloy and tantalum micromarkers;</li><li>Expanded stent diameters, 8, 9, and 10 mm;</li><li>Stent lengths 40, 60, and 80 mm;</li><li>Stent delivery system usable length 190 cm; and</li><li>Guidewire lumen 0.035".</li></ul> | | Biocompatibility | All materials used in the S.M.A.R.T. <sup>TM</sup> Nitinol Stent Endoscopic Biliary System are biocompatible. | {1}------------------------------------------------ K030409 page 2 of 2 The S.M.A.R.T.TM Nitinol Stent Endoscopic Biliary System is substantially Summary of equivalent to the predicate devices. The equivalence was confirmed through Substantial Equivalence pre-clinical testing. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three stripes representing the department's mission. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAR 1 0 2003 Mr. Sam Mirza Manager Regulatory Affairs Cordis Corporation 14201 N.W. 60th Avenue MIAMI LAKES FL 33014 Re: K030409 Trade/Device Name: S.M.A.R.T.™ Nitinol Stent Endoscopic Biliary System Regulation Number: 21 CFR §876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: II Product Code: 78 FGE Dated: February 6, 2003 Received: February 7, 2003 Dear Mr. Mirza: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Warnings section of the device's labeling: The safety and effectiveness of this device for use in the vascular system have not been established. Furthermore, the indication for biliary use must be prominently displayed in all labeling, including pouch, box, and carton labels, instructions for use, and other promotional materials, in close proximity to the trade name, of a similar point size, and in bold print. {3}------------------------------------------------ ## Page 2 - Mr. Sam Mirza If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling. Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling. If you desire specific information about the application of other labeling requirements to your device (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4616. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International, and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Daniel G. Schultz, M.D. Daniel G. Schultz, M.D. Director Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page 1 of 1 510(k) Number (if known): _K030409 Device Name: Cordis S.M.A.R.T.101 Nitinol Stent Endoscopic Biliary System FDA's Statement of the Indications For Use for device: The Cordis S.M.A.R.T.™ Nitinol Stent Endoscopic Biliary System is indicated for use in the palliation of malignant neoplasms in the biliary tree. Perscription Use (Per 21 CFR 801.109) OR Over-the-Counter Use --------------------------------------------------------------------------------------------------------------------------------------------------------- Thiel K. Lysem --- (Division Sign-Off) Division in Reproductive, Abdominal, and Radiological Devices 510(k) Number K030409
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...