The Ebit Sanità Org@nizer software is a medical image management software device intended to be used for digital acquiring, viewing, processing, archiving, reporting and communicating ultrasound medical studies by qualified medical professionals, after proper installation on an appropriate hardware platform. This medical device is intended for the sole use of digital acquiring, viewing, processing, archiving reporting and communicating digital ultrasound studies. It does not have any influence in the way the medical image is produced, nor it is intended to control ionizing emissions or monitor vital physiological processes. The medical device is able to manage stand-alone or distributed medical image archives. It specifically provides the following added-value: ®Connecting diagnostic modalities for the purpose of centralising the image archiving for later image viewing, processing and reporting; @Establishing distributed points of access on conventional PCs for distributed image and report viewing and access.
Device Story
Org@nizer software functions as medical image management system for ultrasound studies; acquires, views, processes, archives, reports, and communicates digital images. Operates on standard PC hardware; connects diagnostic modalities to centralize archives; provides distributed access points for clinicians. Input consists of digital ultrasound studies; output includes displayed images, reports, and archived data. Clinicians use software to review still frames and cine-loops; perform measurements (distance, area, perimeter, velocity); apply annotations (text, lines); and adjust window-level, zoom, or panning. Device does not influence image acquisition or control ionizing radiation; intended for use by qualified medical professionals in clinical settings to facilitate diagnostic workflow and image access.
Clinical Evidence
No clinical data; bench testing only. Device functionality verified through comparison of technical specifications and intended use with predicate device.
Technological Characteristics
Software-based image management system; runs on Windows-based PC hardware. Supports DICOM connectivity. Features include image database management, still frame/cine-loop review, annotation (text, lines), and measurements (distance, area, perimeter, velocity). Archiving supported on hard disk, CD-R, and DVD-R. No image compression utilized.
Indications for Use
Indicated for qualified medical professionals to manage, view, process, archive, report, and communicate digital ultrasound studies.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
Applicare Radworks Medical Imaging Software W/ Quality Control Module (K982862)
Submission Summary (Full Text)
{0}------------------------------------------------
K030363
510(k) Ebit Sanità Org@nizer Software Ebit Sanità, S.p.A.
MAR 2 7 2003
# 510(k) Summary
The following safety and effectiveness summary has been prepared pursuant to requirement for 510(k) summaries specified in 21CFR¶807.92(a).
807.92(a)(1)
## Submitter Information
Colleen Hittle Densmore, Official Correspondent
The Anson Group, an Aventor company
7992 Castleway Drive
Indianapolis, IN 46250
| Phone: | (317) 849-1916 |
|------------|----------------|
| Facsimile: | (317) 577-9070 |
Contact Person: Colleen Hittle Densmore
| Date: | January 15, 2003 |
|-------|------------------|
|-------|------------------|
807.92(a)(2)
| Trade Name: | Ebit Sanità Org@nizer Software |
|--------------|--------------------------------|
| Common Name: | System, Image Processing |
| Classification: | |
|-------------------|----|
| Regulatory Class: | II |
| | FR Number | Product Code |
|-------------------------|-----------|--------------|
| Image Processing System | 892.2050 | 90-LLZ |
807.92(a)(3)
Predicate Device(s)
| Applicare | Radworks Medical Imaging Software |
|-----------|-----------------------------------|
| | W/ Quality Control Module |
| | K982862 |
Additional Substantial Equivalence Information is provided in the following substantial Equivalence Comparison Table.
{1}------------------------------------------------
510(k) Ebit Sanità Org@nizer Software Ebit Sanità, S.p.A.
807.92(a)(4)
## Device Description
The Ebit Sanità Org@nizer software is a medical Image Management software device used for digital acquisition, viewing, processing, archiving, reporting and communicating ultrasound studies.
807.92(a)(5)
## Intended Use(s)
The Ebit Sanita Org@nizer software is intended to be used for digital acquiring, viewing, processing, archiving, reporting and communicating ultrasound medical studies by qualified medical professionals, after proper installation on an appropriate hardware platform.
{2}------------------------------------------------
| Product | Ebit Sanità Org@nizer<br>Software<br>(This submission) | Applicare Medical Imaging<br>Radworks Medical Imaging<br>Software with Quality Control<br>Module<br>K982862 |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | 892.2050 LLZ<br>Class II | 892.2050 LLZ<br>Class II |
| Intended use | The Ebit Sanità Org@nizer software<br>is a medical image management<br>software device intended to be used<br>for digital acquiring, viewing,<br>processing, archiving, reporting and<br>communicating ultrasound medical<br>studies by qualified medical<br>professionals, after proper<br>installation on an appropriate<br>hardware platform | The RadWorks Quality Control Module is<br>intended to be used by authorized staff to<br>perform various quality control operations on<br>RadWorks imaging studies before they are made<br>available to other locations on the network.<br>These operations include confirming or editing<br>patient characteristics, reviewing the status<br>history of the study, adding or removing images,<br>combining with another study, renumbering<br>images, editing patient orientation information,<br>and setting or editing routing information. |
| Graphic User Interface | Yes | Yes |
| Platform | PC | PC |
| Operating System | Microsoft Windows<br>based | Microsoft Windows<br>based |
| Display Resolution | 1280x1024 | 2048x2560 |
| Image Resolution | From 8 bits grayscale and 24 bits<br>color | From 8 bits, 256 levels to 24 bits color |
| Image Communication | DICOM Compliant | DICOM Compliant |
| Image Compression | No image compression techniques<br>are utilized | Jpeg Loss-less;<br>Jpeg lossy |
| Image Archiving (Hard<br>Disk) | Yes | Yes |
| Image Archiving<br>(Removable Media) | CD-R; DVD-R, other Dicom Entities | CD-R; MOD; DVD-R, other Dicom Entities |
| Image Review | Still frame, Cine-loops, Window-<br>level, Zoom, Panning, Configurable<br>layout | Still frame, Cine-loops, Window-Level, Zoom,<br>Panning, Configurable layout |
| Image Processing | Annotation:<br>text, lines | Annotation:<br>arrows, text, lines, Circles |
| | Measurements: Distance, area,<br>perimeter, velocity | Measurements: distances, angles, pixel values,<br>pixel distribution, grey level statistics |
| Image Database | Yes | Yes |
| Image Management | Uncompressed / JPEG Lossy /<br>JPEG Loss-less | Uncompressed / JPEG Lossy /<br>JPEG Loss-less |
## Comparison Chart for Substantial Equivalence
.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three heads facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 2 7 2003
Ebit Sanita, S.p.A. % Ms. Colleen Densmore The Anson Group 7229 Castleway Drive INDIANAPOLIS IN 46250
#### Re: K030363
Trade/Device Name: Org(@nizer Version 3.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system
Regulatory Class: II Product Code: 90 LLZ Dated: January 31, 2003 Received: February 4, 2003
Dear Ms. Densmore:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely vours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Page 2
{5}------------------------------------------------
## Indications for Use Statement
| Applicant: | Ebit Sanità S.p.A. |
|---------------------------|--------------------|
| 510(k) Number (if known): | K030363 |
Ebit Sanità Org@nizer Software Device Name:
Indication For Use:
The Ebit Sanità Org@nizer software is a medical image management software device intended to be used for digital acquiring, viewing, processing, archiving, reporting and communicating ultrasound medical studies by qualified medical professionals, after proper installation on an appropriate hardware platform.
This medical device is intended for the sole use of digital acquiring, viewing, processing, archiving reporting and communicating digital ultrasound studies. It does not have any influence in the way the medical image is produced, nor it is intended to control ionizing emissions or monitor vital physiological processes. The medical device is able to manage stand-alone or distributed medical image archives. It specifically provides the following added-value:
®Connecting diagnostic modalities for the purpose of centralising the image archiving for later image viewing, processing and reporting;
@Establishing distributed points of access on conventional PCs for distributed image and report viewing and access.
#### (PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use \/ OR Over The Counter ***_***
(PER 21 CFR, 801 109)
(optional Format 1-2-96)
David A. Symm
---
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K030363
510(K) FILE - ORG@NIZER SOFTWARE
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.