WAKO AUTOKIT MICRO ALBUMIN, WAKO AUTOKIT MICRO CALIBRATOR SET, MICRO ALBUMIN CONTROL SET
K030320 · Wako Chemicals USA, Inc. · JIS · Apr 11, 2003 · Clinical Chemistry
Device Facts
Record ID
K030320
Device Name
WAKO AUTOKIT MICRO ALBUMIN, WAKO AUTOKIT MICRO CALIBRATOR SET, MICRO ALBUMIN CONTROL SET
Applicant
Wako Chemicals USA, Inc.
Product Code
JIS · Clinical Chemistry
Decision Date
Apr 11, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1150
Device Class
Class 2
Indications for Use
A urinary protein or albumin (nonquantitative) test system is a device intended to identify proteins or albumin in urine. Idetnification of urinary protein or albumin (nonquantitative) is used in the diagnosis and treatment of disease conditions such as renal or heart diseases or thyroid disorders, which are characterized by proteinuria or albuminuria.
Device Story
Wako Autokit Micro Albumin is an in vitro diagnostic assay for measuring albumin in urine. The device utilizes an immunoturbidimetric principle: patient urine samples are mixed with a buffer and anti-human albumin antibody (goat); albumin in the sample binds to the antibody to form insoluble aggregates, increasing solution turbidity. This turbidity is measured optically, with the signal intensity proportional to the albumin concentration. The assay is intended for clinical laboratory use to detect microalbuminuria, a marker for early diabetic nephropathy. Healthcare providers use the quantitative results to monitor diabetes control and assess the risk of progressive kidney damage. The device provides a minimum detectable level of 0.33 ug/dL.
Clinical Evidence
Bench testing only. Precision studies confirmed acceptable day-to-day performance. Comparison studies against the predicate device (Wako Micro Albumin B assay) yielded a correlation coefficient of 0.9984 and a regression equation of y = 1.0179x - 0.9619. The minimum detectable level is 0.33 ug/dL.
Technological Characteristics
Immunoturbidimetric assay. Reagents include buffer and goat anti-human albumin antibody. Optical measurement of turbidity proportional to albumin concentration. Standalone diagnostic reagent kit for use with clinical chemistry analyzers.
Indications for Use
Indicated for the identification of urinary protein or albumin in patients to assist in the diagnosis and treatment of disease conditions characterized by proteinuria or albuminuria, such as renal disease, heart disease, or thyroid disorders.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
Wako Micro Albumin B assay
Submission Summary (Full Text)
{0}------------------------------------------------
APR 1 1 2003
# Nako
## Wako Diagnostics
1600 Bellwood Road, Richmond, VA 23237 U.S.A.
K030320
### 510(k) Summary of Safety and Effectiveness
Diabetic nephropathy, which is accompanied by irreversible kidney damage and persistent proteinuria, is a major cause of death in persons with insulin-dependent diabctes mellitus and a main reason to initiate hemodialysis. Therefore, detection of kidney (glomerular) damage that is minimal and reversible is important. Microalbuminuria is a condition characterized by increased urinary excretion of albumin in the absence of overt nephropathy. It has been reported in several sludies to predict development of diabetic nephropathy and its mortality risk in both diabetes mellitus of insulin-dependent and non-insulin-dependent.
Because micro-albuminuria may be reversible if diabctes is well controlled, its early detection may be very beneficial in treatment programs for diabetes.
#### Principle of the method
When a sample is mixed with Buffer and Antibody, albumin in the sample combines specificaaly with anti-human albumin antibody (goat) in the Antibody to yeild an insoluble aggregate that causes increases turbidity in the solution. The degree of the turbidity of solution can be measured optically and is proportional to the concentration of albumin in the patient's sample.
Precision studies indicate acceptable values can be obtained on a day to day basis. The minimum detectable level of this method is estamiated to be 0.33 ug/dL. In comparison studies against the prodicate, Wako Micro Albumin B assay, a correlation cocfficient of 0.9984 and a regression cquation of y = 1.0179x - 0.9619 was obtained.
TELEPHONE (877) 714-1924 FACSIMILE (804) 271-7791 E-MAIL diagnostics@wakousa.com WEBSITE www.wakousa.com
A division of Wako Chernicals USA, Inc
{1}------------------------------------------------
#### References:
Burtis, C.A. and Ashwood, E.R., Ed.: Tietz Textbook of Clinical Chemistry, 200 Ed., Sanuders, Philadelphia, 1994.
Lothar Thomas, M.D., Ed.: Clinical Laboratory Diagnostics DG Klinische Chemie Mitteilungen 26 (1995) page 5 Mount.J.N. : Journal of Clinical Pathology, 22, 12 (1986) Rosenstock, j. and Raskin, P .: Diabetes Care, 9, 529 (1986) Mogensen, Christensen, C.K. and Vittinghus, E .: Diabetes, 32 (Suppl.2), 64 (1983) Viberti, G.C. et al : Kidney International, 21, 714 (1982) Schmitz, A. and Vaeth, M. : Diabetic Medicine, 5, 126 (1988) Mogensen, C.E. et al : N.. Engl. J. Med., 311, 89 (1984) Mogensen, C.E.:N. Engl. J. Med., 310, 356 (1984)
Kuzellallory
---
Tonya Mallory, Executive Man
April 17, 2003
April 7, 2003 Wako Diagnostics Wako Chemicals USA, Inc. . 1600 Bellwood Road Richmond, VA 23237
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure, represented by three curved lines that resemble a person's head and torso. The figure is positioned to the right of the department's name, which is arranged in a circular pattern around the logo. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
APR 1 1 2003
Ms. Lori Creasy Regulatory Affairs Specialist Wako Diagnostics 1600 Bellwood Road Richmond, VA 23237
Re: k030320
> Trade/Device Name: Wako Autokit Micro Albumin Regulation Number: 21 CFR 862.1645 Regulation Name: Urinary protein or albumin (nonquantitative) test system Regulatory Class: Class II Product Code: JIS; JIQ; JJX Dated: January 29, 2003 Received: January 30, 2003
Dear Ms. Creasy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809): and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{3}------------------------------------------------
Page 2 –
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K630320
## Indications for Use:
A urinary protein or albumin (nonquantitative) test system is a device intended to identify proteins or albumin in urine. Idetnification of urinary protein or albumin (nonquantitative) is used in the diagnosis and treatment of disease conditions such as renal or heart diseases or thyroid disorders, which are characterized by proteinuria or albuminuria.
Proprietary Name: Wako Autokit Micro Albumin
Established Registration Number: 1627434
Premarket Notification 510 (k) Number:
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use (Optional Format 1-2-96)
Sean Cooper
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K030320
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.